← Product Code [KRA](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/KRA) · K163701

# Direxion and Direxion HI-FLO Torqueable Microcatheters (K163701)

_Boston Scientific Corporation · KRA · May 16, 2017 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/KRA/K163701

## Device Facts

- **Applicant:** Boston Scientific Corporation
- **Product Code:** [KRA](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/KRA.md)
- **Decision Date:** May 16, 2017
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 870.1210
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

The Direxion and Direxion HI-FLO Torqueable Microcatheters are intended for peripheral vascular use. The pre-loaded Fathom and Transend Guidewires can be used to selectively introduce and position the microcatheter in the peripheral vasculature. The microcatheter can be used for controlled and selective infusion of diagnostic, embolic, or therapeutic materials into the vessel.

## Device Story

Torqueable microcatheters used for peripheral vascular access; facilitates selective infusion of diagnostic, embolic, or therapeutic materials. Device consists of microcatheter with hydrophilic coating and radiopaque distal marker; available in small (Direxion) and large (Direxion HI-FLO) lumen configurations. Operated by physicians in clinical settings under fluoroscopic guidance. Pre-loaded with Fathom or Transend steerable guidewires to aid navigation through tortuous anatomy. Distal tip is steam-shapeable; proximal end features standard luer with rotating hemostatic valve or Y-adapter. Provides controlled delivery of therapeutic agents; enables access to challenging peripheral vessels; improves procedural efficiency through pre-loaded guidewire systems.

## Clinical Evidence

Bench testing only. Performance testing included dimensional requirements (OD/ID), guide catheter compatibility, surface defect analysis, coating length, distal joint integrity, radiopaque marker tensile strength, embolic coil compatibility, tip shape, packaging mandrel removal force, sterile barrier integrity, and system removal from packaging. All tests met predetermined acceptance criteria.

## Technological Characteristics

Peripheral vascular microcatheter; small (2.5F-2.7F) and large (2.9F-3F) lumen variants. Features hydrophilic coating, radiopaque distal markers, and steam-shapeable tips. Includes pre-loaded steerable guidewires (Fathom-16, Transend 14/18). Proximal interface uses standard luer with RHV or Y-adapter. Mechanical device; no software or energy source.

## Regulatory Identification

A continuous flush catheter is an attachment to a catheter-transducer system that permits continuous intravascular flushing at a slow infusion rate for the purpose of eliminating clotting, back-leakage, and waveform damping.

## Predicate Devices

- Boston Scientific Direxion Microcatheters ([K142259](/device/K142259.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the circle is an abstract symbol consisting of three stylized human profiles facing to the right. The profiles are stacked on top of each other, with the top profile being the largest and the bottom profile being the smallest.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 16, 2017

Boston Scientific Corporation Ms. Liz Johnston Regulatory Affairs Specialist Three Scimed Place Maple Grove, MN 55311

Re: K163701

Trade/Device Name: Direxion and Direxion HI-FLO Torqueable Microcatheters Regulation Number: 21 CFR 870.1210 Regulation Name: Continuous Flush Catheter Regulatory Class: Class II Product Code: KRA Dated: April 17, 2017 Received: April 18, 2017

Dear Ms. Johnston:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Sincerely,

Fernand Aguel - S

for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K163701

Device Name

Direxion and Direxion HI-FLO Torqueable Microcatheters

Indications for Use (Describe)

The Direxion and Direxion HI-FLO Torqueable Microcatheters are intended for peripheral vascular use. The pre-loaded Fathom and Transend Guidewires can be used to selectively introduce and position the microcatheter in the peripheral vasculature. The microcatheter can be used for controlled and selective infusion of diagnostic, embolic, or therapeutic materials into the vessel.

Type of Use (Select one or both, as applicable)

| <span style="font-family: DejaVu Sans, sans-serif">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|----------------------------------------------------------------------------------------------------------|
| <span style="font-family: DejaVu Sans, sans-serif">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)  |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# K163701 510(k) Summary

Per 21 CFR 807.92

| Submitter's Name and Address | Boston Scientific Corporation<br>Three Scimed Place<br>Maple Grove, MN 55311<br>USA                                                                                                                                                                                                                                                                                                                                                                                          |
|------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contract Information         | Liz Johnston<br>Regulatory Affairs Specialist<br>Phone: 763-494-1676<br>Fax: 763-494-2222<br>Email: Liz.Johnston@bsci.com                                                                                                                                                                                                                                                                                                                                                    |
| Date Prepared                | 22 December 2016                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Proprietary Name             | Single Product Configurations:<br>• Direxion™ Torqueable Microcatheter<br>• Direxion™ HI-FLO™ Torqueable Microcatheter<br>Preloaded Guidewire System Configurations:<br>• Direxion™ Fathom™-16 System Pre-Loaded Torqueable Microcatheter<br>• Direxion™ Transend™-14 System Pre-Loaded Torqueable Microcatheter<br>• Direxion™ HI-FLO™ Fathom™-16 System Pre-Loaded Torqueable Microcatheter<br>• Direxion™ HI-FLO™ Transend™-18 System Pre-Loaded Torqueable Microcatheter |
| Common Name                  | Continuous Flush Catheters                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Classification               | Class II per CFR 870.1210                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Predicate Device             | Boston Scientific Direxion Microcatheters:<br>• K142259, KRA, September 10, 2014                                                                                                                                                                                                                                                                                                                                                                                             |

{4}------------------------------------------------

# Device Description

The Direxion Microcatheter is available in small and large lumens. The Direxion Torqueable Microcatheter (Direxion) is a small lumen microcatheter with a distal outside diameter of 2.5F (0.85 mm), and a maximum outside diameter of 2.7F (0.95 mm). It has an inside diameter of 0.021 in (0.5 mm) minimally in the proximal and distal regions. The microcatheter lumen is able to accommodate steerable guidewires with diameters ≤ 0.018 in (0.47 mm).

The Direxion HI-FLO Torqueable Microcatheter (Direxion HI-FLO) is a large lumen microcatheter with a distal outside diameter of 2.9F (1.00 mm), and a maximum outside diameter of 3F (1.05 mm). It has an inside diameter of 0.027 in (0.6 mm) minimally in the proximal and distal regions. The microcatheter lumen is able to accommodate steerable guidewires with diameters ≤ 0.021 in (0.53 mm).

The Direxion and Direxion HI-FLO Microcatheters are available in a variety of tip shapes (Straight, Bern, Swan, and J) to aid with accessing challenging anatomy. The distal outer surface of the microcatheter is coated with a hydrophilic coating. A radiopaque marker is located at the distal tip to facilitate fluoroscopic visualization. Some Direxion Microcatheters have a second marker 3 cm proximal to the first marker. The distal tip of the microcatheter is steam shapeable. The proximal end incorporates a standard luer with rotating hemostatic valve (RHV) or Y-adapter.

The Direxion and Direxion HI-FLO Microcatheters are available with the following preloaded guidewires:

# Fathom-16 Steerable Guidewire

- 0.016 in (0.41 mm) diameter; 140 or 180 cm lengths
# Transend 14/18 Steerable Guidewires

- 0.014 in (0.37 mm) or 0.018 in (0.47 mm) diameters; 135, 165 or 190 cm lengths
The guidewires have a hydrophilic coating to provide lubricity, which aids in the navigation of distal, tortuous vasculature. The guidewires are radiopaque to allow for visualization under fluoroscopy and the tips are shapeable.

Accessories may include a RHV or Y-adapter, steam shaping mandrel, microcatheter introducer, quidewire introducer and torque device.

{5}------------------------------------------------

# Intended Use / Indications for Use

The Direxion and Direxion HI-FLO Torqueable Microcatheters are intended for peripheral vascular use. The pre-loaded Fathom and Transend Guidewires can be used to selectively introduce and position the microcatheter in the peripheral vasculature. The microcatheter can be used for controlled and selective infusion of diagnostic, embolic, or therapeutic materials into the vessel.

# Comparison of Technological Characteristics

The Direxion and Direxion HI-FLO Torqueable Microcatheters are similar in fundamental design, function, materials, packaging, sterilization, operating principle, intended use / indication for use and fundamental technology as the predicate device. Modifications were made to the coating length, pre-shaped tips (Bern, J. Swan), and packaging of the Direxion Microcatheter.

# Performance Data

The following testing was conducted for design elements and performance characteristics deemed appropriate to demonstrate equivalence to the predicate device. The Direxion Microcatheters met the predetermined acceptance criteria ensuring substantial equivalence to the predicate device. No new safety or performance issues were raised during testing.

- OD/ID Dimensional Requirements
- Guide Catheter Compatibility ●
- Microcatheter Surface Defects ●
- Microcatheter Coating Length ●
- Distal Joint Integrity ●
- RO Marker band Tensile ●
- Embolic Coil Compatibility
- Initial Tip Shape
- Packaging Mandrel Removal Force ●
- Sterile Barrier Integrity ●
- Device Containment in Packaging ●
- Microcatheter System Removal from Packaging ●

# Conclusion

Boston Scientific has demonstrated that the modifications to the Direxion Microcatheters are substantially equivalent in fundamental design, technology, function, materials, packaging, operating principle, and intended use / indication for use as the predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/KRA/K163701](https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/KRA/K163701)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
