← Product Code [KRA](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/KRA) · K161962

# VAMP Venous/Arterial Blood Management Protection System (K161962)

_Edwards Lifesciences, LLC · KRA · Nov 28, 2016 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/KRA/K161962

## Device Facts

- **Applicant:** Edwards Lifesciences, LLC
- **Product Code:** [KRA](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/KRA.md)
- **Decision Date:** Nov 28, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1210
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Pediatric

## Indications for Use

VAMP Adult and VAMP Plus: To be used only for blood withdrawal. The blood sampling system is indicated for use on patients requiring periodic withdrawal of blood samples from arterial and central line catheters, including peripherally inserted central venous catheters, which are attached to pressure monitoring lines. VAMP Jr.: To be used only for blood withdrawal. The blood sampling system is indicated for use on pediatric patients (including neonates) requiring periodic withdrawal of blood samples from umbilical, arterial and central line catheters, including peripherally inserted central catheters and central venous catheters, which are attached to pressure monitoring lines.

## Device Story

Sterile, single-use, needleless closed blood sampling system; connects to pressure monitoring lines and catheters (arterial, central, umbilical). Facilitates withdrawal of heparinized/diluted blood (clearing volume) from catheter; allows collection of undiluted blood samples via VAMP Blood Transfer Unit (BTU) or Direct-Draw unit; reinfuses clearing volume to patient post-sampling. Reduces infection risk, needle sticks, and blood waste. Used in clinical settings; operated by healthcare professionals. Output is blood sample in vacuum tube; enables diagnostic testing. Benefits patient by minimizing blood loss and reducing exposure to needle-related hazards.

## Clinical Evidence

Bench testing only. Device successfully passed MRI safety testing.

## Technological Characteristics

Sterile, single-use, needleless closed blood sampling system. Components include split septum membrane (BTU) or direct-draw vacuum tube connection. Compatible with disposable/reusable pressure transducers. MRI safe.

## Regulatory Identification

A continuous flush catheter is an attachment to a catheter-transducer system that permits continuous intravascular flushing at a slow infusion rate for the purpose of eliminating clotting, back-leakage, and waveform damping.

## Predicate Devices

- VAMP Venous/Arterial Blood Management Protection System ([K885281](/device/K885281.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling three human profiles facing to the right, with flowing lines suggesting movement or transformation.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 28, 2016

Edwards Lifesciences LLC Jennifer Wilbur Manager, Critical Care Regulatory Affairs, Program Management One Edwards Way Irvine, California 92614

Re: K161962

Trade/Device Name: VAMP Venous/Arterial Blood Management Protection System Regulation Number: 21 CFR 870.1210 Regulation Name: Continuous Flush Catheter Regulatory Class: Class II Product Code: KRA Dated: October 26, 2016 Received: October 27, 2016

Dear Jennifer Wilbur:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known)

### K161962

Device Name

VAMP Venous/Arterial Blood Management Protection System

Indications for Use (Describe) VAMP Adult and VAMP Plus:

To be used only for blood withdrawal.

The blood sampling system is indicated for use on patients requiring periodic withdrawal of blood samples from arterial and central line catheters, including peripherally inserted central venous catheters, which are attached to pressure monitoring lines.

VAMP Jr.:

To be used only for blood withdrawal.

The blood sampling system is indicated for use on pediatric patients (including neonates) requiring periodic withdrawal of blood samples from umbilical, arterial and central line catheters, including peripherally inserted central catheters and central venous catheters, which are attached to pressure monitoring lines.

Type of Use (Select one or both, as applicable)

| <span></span> Prescription Use (Part 21 CFR 801 Subpart D) | <span></span> |
|------------------------------------------------------------|---------------|
|------------------------------------------------------------|---------------|

| <span></span> Over-The-Counter Use (21 CFR 801 Subpart C) | <span></span> |
|-----------------------------------------------------------|---------------|
|-----------------------------------------------------------|---------------|

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# SECTION 5 – 510(k) SUMMARY

| 510(k) Submitter                    | Edwards Lifesciences, LLC                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                             |
|-------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person                      | Primary Contact<br>Jennifer Wilbur<br>Manager, Regulatory Affairs, PM<br>Jennifer_Wilbur@edwards.com<br>Office: 949-756-4436<br>Cell: 508-776-4525<br>Fax: 949-809-2984<br>Anticipated to be on maternity<br>leave from November 17, 2016 –<br>February 8, 2017                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Backup Contact<br>Deana Boushell<br>Senior Manager, Regulatory Affairs, PM<br>Deana_Boushell@edwards.com<br>Office: 949-756-4472<br>Cell: 508-254-3289<br>Fax: 949-809-2967 |
| Date Prepared                       | November 23, 2016                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                             |
| Trade Name                          | VAMP Venous/Arterial Blood Management Protection System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |                                                                                                                                                                             |
| Common Name                         | Closed Blood Sampling System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |                                                                                                                                                                             |
| Classification Name                 | Catheter, Continuous Flush (CFR 21 870.1210)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |                                                                                                                                                                             |
| Regulation<br>Class/Product Code    | Class II<br>KRA                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                                             |
| Predicate Device(s)                 | K885281: VAMP Venous/Arterial Blood Management Protection System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                                                                             |
| Device Description                  | The subject Edwards Venous/Arterial Blood Management Protection (VAMP)<br>System is a sterile, single use device that provides a safe and convenient<br>method for the withdrawal of blood samples when attached to pressure<br>monitoring lines. The subject device is a needleless closed blood sampling<br>system designed to reduce infection, needle sticks, and blood waste<br>associated with blood sampling.<br>The VAMP blood sampling system is designed for use with disposable and<br>reusable pressure transducers and for connection to central line catheters<br>(inclusive of peripherally inserted central catheters and central venous<br>catheters), arterial catheters, and umbilical catheters where the system can be<br>flushed clear after sampling. The VAMP blood sampling system is used for the<br>drawing and retention of heparinized/diluted blood (or clearing volume) from<br>the catheter or cannula within the line, allowing undiluted blood samples to be<br>drawn from an in-line sampling site. At the completion of sample draw, the<br>mixed heparin and blood solution (clearing volume) is reinfused into the<br>patient to reduce fluid loss to the patient.<br>The collected blood can be transferred to a vacuum tube via a VAMP Blood<br>Transfer Unit (BTU) or a VAMP Direct-Draw unit. The main purpose of the<br>BTU is to provide a conduit (split septum membrane) in which a blood-filled<br>syringe and blunt cannula assembly can transfer the blood sample to a<br>vacuum tube without using a needle. The Direct-Draw Unit performs in the |                                                                                                                                                                             |
|                                     | same fashion, except it allows for direct connection of the vacuum tube and<br>doesn't require use of separate sampling syringe.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                                                                             |
|                                     | VAMP Adult and VAMP Plus                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |                                                                                                                                                                             |
| Indications for<br>Use/Intended Use | To be used only for blood withdrawal.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                             |
|                                     | The blood sampling system is indicated for use on patients requiring periodic<br>withdrawal of blood samples from arterial and central line catheters, including<br>peripherally inserted central catheters and central venous catheters, which are<br>attached to pressure monitoring lines.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                             |
|                                     | VAMP Jr.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |                                                                                                                                                                             |
|                                     | To be used only for blood withdrawal.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                             |
|                                     | The blood sampling system is indicated for use on pediatric patients (including<br>neonates) requiring periodic withdrawal of blood samples from umbilical,<br>arterial and central line catheters, including peripherally inserted central<br>catheters and central venous catheters, which are attached to pressure<br>monitoring lines.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |                                                                                                                                                                             |
| Comparative<br>Analysis             | There are no physical differences between the subject and predicate devices.<br>The proposed changes to the predicate device include an update to the<br>indications of VAMP Adult, VAMP Plus and VAMP Jr. to clarify that central line<br>catheters are inclusive of central venous catheters and peripherally inserted<br>central catheters. This clarification has no impact to the intended use and is<br>only serving to clarify what is meant by central line catheter.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                             |
| Functional/ Safety<br>Testing       | The VAMP blood sampling system has successfully passed MRI safety<br>testing.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                             |
| Conclusion                          | The Edwards Venous/Arterial Blood Management Protection System and its<br>accessories have been demonstrated to be substantially equivalent to the<br>predicate Venous/Arterial Blood Management Protection System and its<br>accessories (K885281).                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                             |

{4}------------------------------------------------

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/KRA/K161962](https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/KRA/K161962)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
