Bullfrog Micro-Infusion Device

K153501 · Mercator Medsystems, Inc. · KRA · Apr 15, 2016 · Cardiovascular

Device Facts

Record IDK153501
Device NameBullfrog Micro-Infusion Device
ApplicantMercator Medsystems, Inc.
Product CodeKRA · Cardiovascular
Decision DateApr 15, 2016
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1210
Device ClassClass 2
AttributesTherapeutic

Indications for Use

In selective areas of peripheral and coronary vessels, the Bullfrog Micro-Infusion Device is intended for the infusion of diagnostic and therapeutic agents that are indicated for delivery into the vessel wall or perivascular area. The Bullfrog Micro-Infusion Device is also intended for the infusion of diagnostic and therapeutic agents intraluminally.

Device Story

Wire-guided, single-operator, endovascular catheter; features perpendicular microneedle contained within semi-rigid polymer actuator balloon. Device advanced to target peripheral or coronary vasculature; hydraulic actuation moves microneedle through external elastic lamina for delivery of diagnostic/therapeutic agents to adventitial and perivascular tissues. Compliant stabilizing balloon inflates with actuator to provide counter-force for needle seating; needle retracted via vacuum. Used by physicians in clinical settings for targeted drug delivery; benefits include localized treatment of vessel wall or perivascular areas, potentially improving therapeutic efficacy while minimizing systemic exposure.

Clinical Evidence

Bench testing included mechanical and fluid delivery performance, biocompatibility (ISO 10993), and sterilization validation. In-vivo safety and effectiveness studies were conducted. All tests met pre-determined specifications and acceptance criteria.

Technological Characteristics

Wire-guided endovascular catheter; semi-rigid polymer actuator balloon; perpendicular microneedle; compliant stabilizing balloon. Materials are biocompatible per ISO 10993. Available in three model sizes (2-4 mm, 3-6 mm, 4-8 mm). Hydraulic actuation mechanism.

Regulatory Classification

Identification

A continuous flush catheter is an attachment to a catheter-transducer system that permits continuous intravascular flushing at a slow infusion rate for the purpose of eliminating clotting, back-leakage, and waveform damping.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of a human figure with three faces in profile, one behind the other, with flowing lines representing the body. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 April 15, 2016 Mercator Medsystems, Inc. % Adam Harris Senior Regulatory Manager Target Health, Inc. 261 Madison Avenue 24th Floor New York, New York 10016 Re: K153501 Trade/Device Name: Bullfrog Micro-infusion Device Regulation Number: 21 CFR 870.1210 Regulation Name: Continuous Flush Catheter Regulatory Class: Class II Product Code: KRA Dated: March 16, 2016 Received: March 17, 2016 Dear Adam Harris: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in {1}------------------------------------------------ the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Kenneth J. Cavanaugh -S for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # 4. INDICATIONS FOR USE STATEMENT 510(k) Number (if known): K153501 Device Name: Mercator MedSystems Bullfrog® Micro-Infusion Device Indications For Use: In selective areas of peripheral and coronary vessels, the Bullfrog Micro-Infusion Device is intended for the infusion of diagnostic and therapeutic agents that are indicated for delivery into the vessel wall or perivascular area. The Bullfrog Micro-Infusion Device is also intended for the infusion of diagnostic and therapeutic agents intraluminally. Prescription Use X AND/OR (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) {3}------------------------------------------------ ## 5. 510(K) SUMMARY 510(k) Applicant Mercator MedSystems, Inc. 2200 Powell Street, Suite 530 Emeryville, CA 94608 Telephone: (510) 614-4550 Facsimile: (510) 614-4560 Contact Person: Adam Harris, Target Health Phone: (646) 218-2009 Email: aharris@targethealth.com Date of Summary: 30 November 2015 Device Overview Bullfrog® Micro-Infusion Device Trade Name: Common Name: Continuous Flush Infusion Catheter (per 21 CFR 870.1210) Classification: Continuous Flush Infusion Catheter 21 CFR 870.1210 Product Code KRA Panel: Cardiovascular Predicate Devices | 510(k)<br>Number | Trade or Proprietary or Model Name | Manufacturer | |------------------|--------------------------------------------|---------------------| | K062752 | Mercator MicroSyringe II Infusion Catheter | Mercator MedSystems | ### Device Description The Mercator MedSystems Bullfrog Micro-Infusion Device is a wire-guided, singleoperator, endovascular catheter that consists of a perpendicular microneedle, which is sheathed by and contained within a semi-rigid polymer actuator balloon. The device is designed to be advanced to target vasculature and hydraulically actuated to move the microneedle through the external elastic lamina to deliver substances to adventitial and perivascular tissues. A compliant stabilizing balloon inflates with the actuator to provide a force opposite the needle tip for proper seating of the needle is retracted within the sheathing structure by vacuuming the hydraulic actuator. ### Intended Use In selective areas of peripheral and coronary vessels, the Bullfrog Micro-Infusion Device is intended for the infusion of diagnostic and therapeutic agents that are indicated for {4}------------------------------------------------ delivery into the vessel wall or perivascular area. The Bullfrog Micro-Infusion Device is also intended for the infusion of diagnostic and therapeutic agents intraluminally. # Indications for Use N/A ## Technological Characteristics All materials used in the manufacture of the Bullfrog Micro-Infusion Device are suitable for this use and have been used in several previously cleared products. ### Performance Data Performance testing of the Bullfrog Micro-Infusion Device included mechanical and fluid delivery performance, biocompatibility, sterilization validation, in-vivo safety and effectiveness studies. All tests met the pre-determined specifications and acceptance criteria. ### Safety and Effectiveness The Bullfrog Micro-Infusion Device labeling contains instructions for use and any necessary cautions and warnings to assure safe and effective use of the device. The biocompatibility assessment was conducted in accordance with ISO 10993, Biological Evaluation of Medical Devices. ### Comparison to Predicate Devices The Bullfrog Micro-Infusion Device is substantially equivalent to the predicate. The Bullfrog Micro-Infusion Device has the same intended use, design and materials, mechanical safety, methods of introduction and methods of operation as the MicroSyringe II/Bullfrog Micro-Infusion Device from K062752. The Bullfrog Micro-Infusion Device has new technological characteristics related to currently available model sizes, in comparison to the predicate. The Bullfrog Micro-Infusion Device is available in three model sizes (2-4 mm, 3-6 mm and 4-8 mm), while the MicroSyringe II/Bullfrog Micro-Infusion Device from K062752 was cleared with data from only one model size (3-6 mm). Risk analysis has shown that the Bullfrog Micro-Infusion Device has the same risk profile as the predicate.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...