The CXI™ TriForce is intended to be percutaneously introduced into blood vessels and support a wire guide while performing percutaneous peripheral intervention. The product is also intended for injection of radiopaque contrast media for the purpose of angiography.
Device Story
CXI™ TriForce Peripheral Crossing Set facilitates wire guide exchange, infusion, and wire guide support during percutaneous peripheral intervention. Set includes 5.0 French Flexor® introducer, 4.0 French CXI™ Support Catheter, and Peel-Away® sheath. Introducer provides support; catheter allows 0.035-inch wire guide passage. Used in clinical settings by physicians. Catheter features hydrophilic coating and multi-layer shaft (nylon, stainless steel braid, polyimide, polyurethane, PTFE). Device enables contrast media injection for angiography. Benefits include improved vessel access and procedural support.
Clinical Evidence
Bench testing only. Testing included tensile strength, liquid leakage, static burst, dynamic burst, flow rate, and biocompatibility (ISO 10993-1). All tests met predetermined acceptance criteria in conformance with BS EN ISO 11070 and ISO 10555-1 standards.
Technological Characteristics
Set includes 5.0F Flexor® introducer (stainless steel coil reinforced nylon), 4.0F CXI™ Support Catheter (nylon, stainless steel braid, polyimide, polyurethane, PTFE shaft with hydrophilic coating), and Peel-Away® sheath. Manual operation. No electronic components or software. Sterilized.
Indications for Use
Indicated for patients undergoing percutaneous peripheral intervention requiring wire guide support and/or angiography via radiopaque contrast media injection.
Regulatory Classification
Identification
A continuous flush catheter is an attachment to a catheter-transducer system that permits continuous intravascular flushing at a slow infusion rate for the purpose of eliminating clotting, back-leakage, and waveform damping.
K212725 — ViperCross Support Catheters · Cardiovascular Systems, Inc. · Sep 24, 2021
K210322 — XO Cross Support Catheter · Transit Scientific, LLC · Oct 21, 2021
K093052 — CROSSCATH SUPPORT CATHTER · Cook, Inc. · Dec 8, 2009
Submission Summary (Full Text)
{0}------------------------------------------------
K111263 page 1 of 3
1-2
Cook Incorporated CXI™ TriForce Peripheral Crossing Set Traditional 510(k) 13 July 2011
## 5. 510(k) Summary
JUL 2 5 2011
# CXI™ TriForce Peripheral Crossing Set 510(k) Summary 21 CFR 807.92
## 1. Submitter Information:
| Applicant: | Cook Incorporated |
|-----------------------|--------------------------|
| Address: | 750 Daniels Way |
| | Bloomington, IN 47404 |
| Phone Number: | (800) 468-1379 |
| Fax Number: | (812) 332-0281 |
| Contact: | Molly Busenbark, MA, RAC |
| Contact Address: | Cook Incorporated |
| | 750 Daniels Way |
| | Bloomington, IN 47404 |
| Contact Phone Number: | 812-339-2235 Ext. 2162 |
| Contact Fax Number: | 812-332-0281 |
### 2. Device Information:
| Trade name: | CXITM TriForce Peripheral Crossing Set |
|-----------------|----------------------------------------|
| Common name: | Continuous flush catheter |
| Classification: | Class II |
| Regulation: | 21 CFR 870.1210 |
| Product Code: | KRA |
# 3. Predicate Device:
The CXI™ TriForce Peripheral Crossing Set (hereafter referred to as the CXI™ TriForce) is substantially equivalent to the CXI™ Support Catheter, manufactured by Cook Incorporated, which is cleared under 510(k) number K072724. The CXI™ TriForce is also substantially equivalent to the FineCross™ MG Coronary Micro-Guide Catheter and the Glidesheath™ manufactured by Terumo Medical Corporation, which are cleared under 510(k) numbers K082519 and K102008, respectively.
# 4. Comparison to Predicate Device:
It has been demonstrated that the CXI™ TriForce is comparable to the predicate devices in terms of design, intended use, materials, fundamental technology, and principal of operation.
# 5. Device Description:
The CXI™ TriForce is designed to facilitate wire guide exchange, infusion, and wire guide support. Each set contains the following sterile components:
{1}------------------------------------------------
K111263 page 2 of 3
1-3
Cook Incorporated CXI™ TriForce Peripheral Crossing Set Traditional 510(k) 13 July 2011
- Outer 5.0 French Flexor® introducer .
- Inner 4.0 French CXI™ Support Catheter .
- Peel-Away® sheath .
The outer Flexor® introducer is manufactured with a stainless steel coil reinforced nylon construction which provides both flexibility and support to the sheath during use. The outer Flexor introducer is used to support the inner CXI™ Support Catheter during a procedure. The outer Flexor introducers are available in length measurements of 55 or 90 centimeters and can be manufactured with an angled or a straight tip. The inner CXI™ Support Catheter is manufactured with 4.0 French braided flexible kink resistant shaft materials with hydrophilic coating. The flexible shaft material is a laminated material consisting of a nylon layer, stainless steel braiding and an inner polyimide, polyurethane layer, and an innermost polytetrafluoroethylene layer. The inner CXI™ Support Catheters allow acceptance of a 0.035 inch diameter (0.89 millimeters) wire guide. The inner CXI™ Support Catheter will support a wire guide while performing percutaneous peripheral intervention. The CXI™ Support Catheters are available in length measurements of 65 or 100 centimeters and can be manufactured with an angled or a straight tip.
The tetrafluoroethylene Peel-Away® sheath is included in the set for assistance when inserting the inner CXI™ Support Catheter into the outer Flexor® introducer.
## 6. Intended Use:
The CXI™ TriForce is intended to be percutaneously introduced into blood vessels and support a wire guide while performing percutaneous peripheral intervention. The product is also intended for injection of radiopaque contrast media for the purpose of angiography.
### 7. Technological Characteristics:
The proposed CXI™ TriForce consists of a Flexor® introducer, a CXI™ Support Catheter, and a Peel-Away® sheath. The CXI™ TriForce is substantially equivalent to the CXI™ Support Catheter (K072724), the FineCross™ MG Coronary Micro-Guide Catheter (K082519), and the Glidesheath™ (K102008). No new technological aspects are being introduced with the proposed device.
The proposed CXI™ TriForce was subjected to applicable testing to assure reliable design and performance under the testing parameters.
# 8. Test Data:
The following tests were performed to demonstrate that the CXI™ TriForce meets applicable design and performance requirements and supports a determination of substantial equivalence. Additionally, appropriate engineering tests were also performed on aged product to ensure that the CXI™ TriForce meets the performance requirements throughout the duration of shelf life.
- Tensile Strength Testing shows the tensile strength during proper clinical use . should not fracture or rupture the introducer or catheter. In conformance with the
{2}------------------------------------------------
K111263 page 3 of 3
applicable sections of BS EN ISO 11070 and ISO 10555-1, the predetermined acceptance criteria were met.
- Liquid Leakage Testing shows there would be no liquid leakage from the . introducer, introducer valve, or catheter during proper clinical use. In conformance with the applicable sections of ISO 11070 and ISO 10555-1, the predetermined acceptance criteria were met.
- . Static Burst - Testing shows the pressures reached during proper clinical use (maximum pressure at maximum flow rate) are less than the static burst pressure of the catheter, and should not fracture or rupture the catheter. The predetermined acceptance criteria were met.
- . Dynamic Burst - Testing shows the pressures reached during proper clinical use (maximum pressure at maximum flow rate) are less than the dynamic burst pressure of the catheter, and should not fracture or rupture the catheter. The predetermined acceptance criteria were met.
- Flow Rate Testing shows the pressure exerted at the maximum flow rate during . proper clinical use should not fracture or rupture the catheter. The predetermined acceptance criteria were met.
- Biocompatibility Testing shows the device is biocompatible. In conformance with . the applicable sections of ISO 10993-1, the predetermined acceptance criteria were met.
In conclusion, the results of these tests provide reasonable assurance that the device is as safe and as effective as the predicate devices and supports a determination of substantial equivalence.
{3}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its body and wing. The eagle is positioned within a circle, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged around the circle's perimeter.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Cook Incorporated c/o Ms. Molly Busenbark 750 Daniels Way Bloomington, IN 47404
JUL 25 2011
Re: K111263
Trade/Device Name: CXI TriForce Peripheral Crossing Set Regulation Number: 21 CFR 870.1210 Regulation Name: Continuous Flush Catheter Regulatory Class: Class II (two) Product Code: KRA Dated: July 13, 2011 Received: July 14, 2011
Dear Ms. Busenbark:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Br
Di
Bram D. Zuckerman. M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
K111263 page 161
### 4. Indications for Use Statement
510(k) Number (if known): K | | 26 3
CXI™ TriForce Peripheral Crossing Set Device Name:
Indications for Use:
The CXIM TriForce Peripheral Crossing Set is intended to be percutaneously introduced into blood vessels and support a wire guide while performing percutaneous peripheral intervention. The product is also intended for injection of radiopaque contrast media for the purpose of angiography.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE) diovascular Devices
510(k) Number K111263
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.