← Product Code [KRA](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/KRA) · K050630

# PIVOT STEERABLE MICROCATHETER (K050630)

_Boston Scientific - Precision Vascular · KRA · Mar 22, 2005 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/KRA/K050630

## Device Facts

- **Applicant:** Boston Scientific - Precision Vascular
- **Product Code:** [KRA](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/KRA.md)
- **Decision Date:** Mar 22, 2005
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 870.1210
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

The "PivotTM" Steerable Microcatheter is intended to be used to access tortuous vasculature for sub-selective controlled infusion or delivery of diagnostic, embolic, and therapeutic agents into the distal, peripheral, coronary, and neurovasculature, and for guide wire exchange/support during diagnostic or interventional procedures.

## Device Story

Pivot Steerable Microcatheter: 1.9F/2.4F tubular device (~150 cm); lumen for delivery of contrast, drugs, or embolics; lubricious coating on distal segment; two radiopaque markers at distal tip for fluoroscopic visualization; standard luer adapter at proximal end. Used by physicians to access distal, tortuous vasculature over guide wire; steerable like guide wire. Enables sub-selective infusion/delivery of agents; provides guide wire support. Benefits patient by facilitating access to complex vascular anatomy for diagnostic or interventional procedures.

## Clinical Evidence

Bench testing only. Verification included dimensional inspection, hub integrity, flow rate, burst strength, tensile strength, guidewire compatibility, torsional strength, torqueability, and corrosion resistance per ISO 10555-1. All results acceptable.

## Technological Characteristics

1.9F/2.4F tubular catheter; polymeric construction; hydrophilic coating; radiopaque markers; standard luer adapter. Tested per ISO 10555-1. Mechanical device; no software or energy source.

## Regulatory Identification

A continuous flush catheter is an attachment to a catheter-transducer system that permits continuous intravascular flushing at a slow infusion rate for the purpose of eliminating clotting, back-leakage, and waveform damping.

## Predicate Devices

- SDS 0.018" ([K020733](/device/K020733.md))
- SDS 0.010" ([K032494](/device/K032494.md))
- Pivot "SDS" ([K033726](/device/K033726.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K050630

## Summary of Safety and Effectiveness

Prepared 8 March 2005

| General<br>Provisions    | Submitter of 510(k)<br>Premarket Notification:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Boston Scientific - Precision Vascular<br>2405 West Orton Circle<br>West Valley City, UT 84119<br>Phone: 801.974.1700<br>Fax: 801.974.1740 |                                |
|--------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------|
|                          | Contact Person:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Rick Gaykowski<br>Vice President, Regulatory/Clinical<br>Affairs & Quality Systems                                                         |                                |
|                          | Device Trade Name:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | "PivotTM" Steerable Microcatheter                                                                                                          |                                |
|                          | Device Generic Name:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Infusion Catheter                                                                                                                          |                                |
|                          | The predicate devices are listed below.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                            |                                |
| Predicate<br>Devices     | Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Manufacturer                                                                                                                               | 510(k) Number/Date<br>Pro Code |
|                          | <i>SDS</i> 0.018"                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | PVS, Inc.                                                                                                                                  | K020733, 21 March 2002<br>KRA  |
|                          | <i>SDS</i> 0.010"                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | PVS, Inc.                                                                                                                                  | K032494, 05 Sep 2003<br>KRA    |
|                          | <i>Pivot</i> "SDS"                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | PVS, Inc.                                                                                                                                  | K033726, 16 Jan 2004<br>KRA    |
| Classification           | Class II, 21 CFR 870.1210, Continuous Flush Catheter, KRA                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                            |                                |
| Performance<br>Standards | Performance standards have not been established by FDA under section 514 of the<br>Federal Food, Drug and Cosmetic Act.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                            |                                |
| Intended Use             | The "PivotTM" Steerable Microcatheter is intended to be used to access tortuous<br>vasculature for sub-selective controlled infusion or delivery of diagnostic, embolic, and<br>therapeutic agents into the distal, peripheral, coronary, and neurovasculature, and for<br>guide wire exchange/support during diagnostic or interventional procedures.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                                                                                                                            |                                |
| Device<br>Description    | The "PivotTM" Steerable Microcatheter is a 1.9F/2.4F (nominal distal/proximal) tubular<br>device, ~150 cm in length, with a lumen to be used for delivery of contrast, drugs, or<br>embolics. The lumen is constructed from a polymeric material and has an inside<br>diameter of 0.017". The device is coated on the outer diameter with a lubricious<br>coating over the distal segment of the device. Two radiopaque markers are positioned<br>at the distal tip of the device to aid visualization under fluoroscopy. The proximal end of<br>the device has a standard luer adapter for attachment of accessories and can be used<br>to flush the lumen. The subject device has the ability to access distal, tortuous<br>vasculature over a guide wire, deliver embolics and agents, and has the ability to be<br>steered like a guide wire as needed. |                                                                                                                                            |                                |

{1}------------------------------------------------

| Technological<br>Characteristics         | Technological similarities between the "Pivot™" Steerable Microcatheter and             predicate devices include the basal tubular design and dimensions, polymeric             materials and construction, and hydrophilic coating. In instances where the             technological characteristics may differ slightly, it has been demonstrated that there             are no new questions raised regarding safety and efficacy of the "Pivot™" Steerable             Microcatheter.                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
|------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Safety and<br>Performance<br>Tests       | Performance testing of the "Pivot™" Steerable Microcatheter was conducted in             accordance with ISO 10555-1, Sterile, Single-Use Intravascular Catheters -- Part 1.             Verification testing for the subject device included dimensional inspection, hub             integrity, flow rate measurements, burst strength, tensile strength, guidewire             compatibility testing and performance under simulated conditions. Subject product             testing has yielded acceptable results.<br>In addition, torsional strength, torqueability, and corrosion resistance tests have also             yielded acceptable results. The results of these tests, in conjunction with the             substantial equivalence claims as outlined in the premarket notification, effectively             demonstrate the "Pivot™" Steerable Microcatheter is substantial equivalent to the             cited predicate devices. |
| Summary of<br>Substantial<br>Equivalence | Based on the indications for use, technological characteristics, and safety and             performance testing, the subject "Pivot™" Steerable Microcatheter meets the             minimum requirements that are considered adequate for its intended use and is             substantially equivalent in design, materials, sterilization, principles of operation and             indications for use to current commercially available catheters/cited predicates.                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol of three horizontal lines curving upwards, resembling a human figure with outstretched arms.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAR 2 2 2005

Mr. Rick Gaykowski Mice President, Regulatory/Clinical Affairs & Quality Systems Boston Scientific - Precision Vascular 2405 West Orton Circle West Valley City, UT 84119

K050630 Re:

Trade/Device Name: Pivor™ Steerable Microcatheter Regulation Number: 21 CFR 870.1210 Regulation Name: Continuous flush catheter Regulatory Class: II Product Code: KRA Dated: March 10, 2005 Received: March 11, 2005

Dear Mr. Gaykowski:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for assessment to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set

{3}------------------------------------------------

Page 2 – Mr. Rick Gaykowski

Fage 2 - "htt read as a proposion (21 CFR Part 820); and if applicable, the electronic
forth in the quality systems in and inco (Sections 531-542 of the Act); 21 CFR 1000-105 forth in the quality systems (QS) regulation (2) CFR Pat 620); and 11 CFF
product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. 101 - 12 - 12 - 12 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -This letter will allow you to begin marketing your device of your device to a legally
premarket notification. The FDA finding of subscification for your device to a legally remarket notification. The FDA finding of substantial on your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please
s and the results of the support (240) 276.0120 . Also, please note the r If you desire specific advice for your device on our labeling nease not the regulation entitled,
contact the Office of Compliance at (240) 276-0120. Also, please not 97). contact the Office of Compliance at (240) 270-01-20. Proxet 807.97). You may obtain
"Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obta "Misbranding by reference to premarket notification "(21 cf 1 cr from the Division of Small
other general information on your responsibilities under the Act from the Division other general information on your responsibilities under (800) 638-204) or 2001)
Manufacturers, International and Consumer Assistance at its todalfred wordindex.html. Manufacturers, International and Consumer Assistance at les to restor in over index.html.
(301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/i

Sincerely yours,

Durna R. Kirchner

Bram D. Zuckerman, M.D.

Director

Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Page 1 of 1

510(k) Number (if known): ¥650630_

Device Name: "Pivot™" Steerable Microcatheter

Indications for Use:

The "Pivot™" Steerable Microcatheter is intended to be used to access tortuous vasculature for sub-The "Pivot™" Steerable Microcather is intended to activer and therapeutic agents into the distal,
selective controlled infusion of diagnosic, emblic, and therapeupport during selective controlled infusion or delivery of diagnostic, endonic, and therebels agence of the counter of the county, or interventional procedures.

:

. . .

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY

Concurrence of CDRH, Office of Device Evaluation (ODE)

| Prescription Use     | <span></span> ✓ | OR | Over-the-Counter Use |  |
|----------------------|-----------------|----|----------------------|--|
| (Per 21 CFR 801.109) |                 |    |                      |  |

|               | <div style="text-align:left;">Donna P. Vechinel</div> <div style="text-align:left;">(Division Sign-Off)</div> <div style="text-align:left;">Division of Cardiovascular Devices</div> |
|---------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | K050630                                                                                                                                                                              |

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/KRA/K050630](https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/KRA/K050630)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
