OptiOne Single-Head Contrast Delivery System with Pedestal; with Ceiling Suspension;Injector, OptiOne Base System

K152361 · Liebel-Flarsheim Company, LLC · IZQ · Jan 14, 2016 · Cardiovascular

Device Facts

Record IDK152361
Device NameOptiOne Single-Head Contrast Delivery System with Pedestal; with Ceiling Suspension;Injector, OptiOne Base System
ApplicantLiebel-Flarsheim Company, LLC
Product CodeIZQ · Cardiovascular
Decision DateJan 14, 2016
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1650
Device ClassClass 2

Indications for Use

The OptiOne Single-Head Contrast Delivery System is a contrast delivery system and is designed to inject radiopaque contrast media into a patient's vascular system to obtain diagnostic images with computed tomography (i.e. "CT") equipment.

Device Story

OptiOne is a motor-driven, microprocessor-controlled contrast delivery system for CT imaging; used in healthcare facilities by clinicians/technologists. System components include powerhead, console, and power supply. Operator programs flow rate, volume, pressure, and timing via touch screen console. Powerhead utilizes motor control circuits, ball screw, and syringe push ram to deliver contrast. Features include OptiBolus for optimized flow, remote start for radiation safety, and electrical safety stops. Output is controlled contrast delivery to patient vascular system; enables diagnostic image acquisition. Benefits include precise, programmable contrast administration for improved image quality.

Clinical Evidence

Bench testing only. Verification and validation activities confirmed device performance against design inputs and specifications. Compliance with recognized standards including AAMI/ANSI ES60601-1, IEC 60601-1-2, IEC 60601-1-6, and IEC 62304 was demonstrated.

Technological Characteristics

Electromechanical syringe drive system; plastic and metal construction. Features: 4-phase protocol capability, 40-protocol storage, 98°F syringe heater, 0.1-10 mL/s flow rate, 50-325 psi pressure limit. Connectivity via relays, optical couplings, and CAN serial interface. Software lifecycle per IEC 62304. Usability per IEC 62366.

Indications for Use

Indicated for patients requiring radiopaque contrast media injection into the vascular system for diagnostic CT imaging.

Regulatory Classification

Special Controls

*Classification.* Class II (special controls). The device, when it is a non-patient contacting balloon inflation syringe intended only to inflate/deflate balloon catheters and monitor pressure within the balloon, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the United States Department of Health and Human Services. The logo features a stylized caduceus, which is a symbol of medicine and healing. The caduceus is composed of three human profiles facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circle around the caduceus. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 January 14, 2016 Liebel-Flarsheim Company, LLC Craig Buehler Sr. Regulatory Affairs Specialist 2111 East Galbraith Rd Cincinnati, Ohio 45237 Re: K152361 Trade/Device Name: OptiOne Single-Head Contrast Delivery System Regulation Number: 21 CFR 870.1650 Regulation Name: Angiographic Injector And Syringe Regulatory Class: Class II Product Code: IZQ Dated: December 14, 2015 Received: December 15, 2015 Dear Craig Buehler: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Kenneth J. Cavanaugh -S for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K152361 Device Name OptiOne Single-Head Contrast Delivery System Indications for Use (Describe) The OptiOne Single-Head Contrast Delivery System is a contrast delivery system and is designed to inject radiopaque contrast media into a patient's vascular system to obtain diagnostic images with computed tomography (i.e. "CT") equipment. | Type of Use (Select one or both, as applicable) | | |----------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------| | <span style="font-family: sans-serif;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: sans-serif;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for Mallinckrodt Pharmaceuticals. The logo consists of a geometric shape on the left, made up of overlapping cubes in shades of green, blue, and purple. To the right of the shape is the company name, "Mallinckrodt" in a bold, dark blue font, with the word "Pharmaceuticals" underneath in a lighter blue, smaller font. # Section 5: 510(k) Summary ### I. SUBMITTER Liebel-Flarsheim Company LLC 2111 East Galbraith Rd. Cincinnati. OH 45237 Establishment Number: 1518293 | Contact: | Mr. Craig Buehler | |----------------|-----------------------------------| | Title: | Sr. Regulatory Affairs Specialist | | Telephone: | (314) 654-3188 | | Fax: | (314) 654-3344 | | Date Prepared: | 19 August 2015 | ### II. DEVICE Name of Device: OptiOne Single-Head Contrast Delivery System Common or Usual Name: Power Injector Classification Name: Injector, Contrast Medium, Automatic (21 CFR 870.1650) Regulatory Class: II Product Code: IZQ | Model Number | Description | |--------------|-------------------------------------------------------------------------| | 847002 | Injector, OptiOne Base System | | S8472 | OptiOne Single-Head Contrast Delivery System with Pedestal | | S8473 | OptiOne Single-Head Contrast Delivery System with Ceiling<br>Suspension | ### III. PREDICATE DEVICE OptiVantage DH Injector System with Enhanced Communication (K063503) ## IV. DEVICE DESCRIPTION: The OptiOne Single-Head Contrast Delivery System (OptiOne) is intended to inject radiopaque contrast media into a patient's vascular system to obtain diagnostic images when used in conjunction with computed tomography equipment. Each injection is accomplished with a motor-driven syringe mechanism with microprocessor control of the flow rate, volume, pressure and timing. The OptiOne consists of the following key components: a Powerhead, a Console, and a Power Supply. The Power {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the Mallinckrodt Pharmaceuticals logo. The logo consists of a geometric shape on the left and the company name on the right. The geometric shape is made up of several overlapping cubes in shades of green, blue, and purple. The company name is written in a dark blue font, with the word "Mallinckrodt" on top and the word "Pharmaceuticals" below. Supply delivers power to the Powerhead and Console. The Console provides remote control and programmability of the injection through a touch screen display system. The Powerhead contains the components to perform the actual injection such as motor control circuits, ball screw and syringe push ram. The OptiOne is intended for use by doctors, radiology technologists and other licensed medical practitioners in a healthcare facility or hospital with computed tomography equipment. The OptiOne incorporates the same materials as its predicate device and will not contact the patient under normal operation conditions. #### V. INDICATIONS FOR USE The OptiOne Single-Head Contrast Delivery System is a contrast delivery system and is designed to inject radiopaque contrast media into a patient's vascular system to obtain diagnostic images when used with computed tomography (i.e. "CT") equipment. The OptiOne and the OptiVantage predicate device both have the same indication for use. ### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE The OptiOne Single-Head Contrast Delivery System (OptiOne) was developed using the predicate device, OptiVantage DH Injector System with Enhanced Communication (OptiVantage), as the basis for the injector design. Many of the same components and accessories found in and used with the OptiVantage are also found and used with the OptiOne. The main difference between the OptiOne and its predicate is the OptiVantage is a dual head injector and the OptiOne is a single head injector. The OptiOne has the same technological characteristics as the predicate OptiVantage. These similarities and the differences reducing the powerhead from the predicate device's dual head to a single head are described in the table beginning on the next page. {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the logo for Mallinckrodt Pharmaceuticals. The logo consists of a geometric design on the left, made up of overlapping cubes in shades of green, blue, and purple. To the right of the geometric design is the company name, "Mallinckrodt," in a bold, dark blue font, with the word "Pharmaceuticals" underneath in a smaller, lighter blue font. | Feature | OptiOne Single-Head Contrast<br>Delivery System | OptiVantage DH Injector System<br>with Enhanced Communication<br>(K063503) | |------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------| | Indication for Use | is a contrast delivery system and is<br>designed to inject radiopaque<br>contrast media into a patient's<br>vascular system to obtain<br>diagnostic images when used with<br>computed tomography (i.e. "CT")<br>equipment. | Same | | Number of Phases | 4 phases per protocol maximum | 6 phases per protocol maximum | | Protocol Storage | 40 protocols | Same | | Syringe Sizes | 125 mL, 100 mL, 75 mL, 50 mL<br>pre-filled<br>200 mL empty | Same | | Single or Dual Syringe<br>Modality | Single Syringe Capability | Single and Dual Syringe Capability | | Syringe Drive System | Electromechanical | Same | | Syringe Heater | 98° ± 6° F (37° ± 3° C) nominal.<br>Maintains the temperature of pre-<br>heated contrast (not intended to<br>heat cold/room temperature<br>contrast). | Same | | Auto Syringe Fill Rate | 7.0 mL/s<br>Tolerance: +/-5% | Same | | Manual Fill Rate | 2.0 - 15 mL/s<br>Tolerance: +/- 1 mL/s | Same | | Flow Rate | 0.1 - 10 mL/s adjustable in<br>increments of 0.1 mL/s<br>Tolerance: +/-0.05 mL/s for Flow<br>Rates less than 1 mL/s<br>+/-5 % for Flow Rates 1 mL/s to<br>10 mL/s | Same | | Peak Pressure Limit | Pounds per Square Inch (psi):<br>50 - 325 adjustable in 5 psi<br>increments<br>kPa: 345 – 2240 adjustable in<br>34 kPa increments | Same | | Feature | OptiOne Single-Head Contrast<br>Delivery System | OptiVantage DH Injector System<br>with Enhanced Communication<br>(K063503) | | Remote Start | Use of the remote hand switch<br>allows the operator to perform<br>injections from outside the area of<br>direct radiation. | Same | | Scan Delay | 0 - 600 seconds adjustable in<br>increments of 1 second | Same | | Phase Delay | 0 – 600 seconds adjustable in<br>increments of 1 second or<br>permanent pause | Same | | Inject Delay | 0 - 600 seconds adjustable in<br>increments of 1 second | Same | | OptiBolus | Yes - The OptiBolus feature is<br>used to deliver an exponentially<br>decaying flow rate injection that<br>optimizes the contrast usage and<br>provides an extended period of<br>uniform enhancement of the area<br>of interest. | Same | | Universal Console | 12.25 W x 2.5 H x 8.5 D inches<br>(311 W x 64 H x 216 D mm)<br>5.8 lbs (2.6 kg) | Same | | Powerhead | 7 W x 6 H x 12.25 D inches<br>(178 W x 152 H x 311 D mm)<br>10.4 lbs (4.7 kg) | 12.5 W x 6 H x 8 D inches<br>(318 W x 152 H x 203 D mm)<br>14.5 lbs (6.57 kg) | | Power Supply | 10 W x 4 H x 9 D inches<br>(267 W x 102 H x 228 D mm)<br>6 lbs (2.7 kg) | Same | | Post Injection<br>Readout | Displayed on Powerhead and<br>Console | Same | | OEM Interface | Available Via Relays and Optical<br>Couplings / Serial (CAN) | Same | | Safety Stop<br>Mechanism | Electrical Stop When Injection<br>Parameters are Out of<br>Specification | Same | | Volume Remaining<br>Display | Displayed on Powerhead and<br>Console | Same | | Materials | Plastic and Metal | Same Plastic and Metal | | Target Population | Humans | Same | | Feature | OptiOne Single-Head Contrast<br>Delivery System | OptiVantage DH Injector System<br>with Enhanced Communication<br>(K063503) | | Sterility (Syringe) | Injectors are not sterile products,<br>but syringes and disposables are<br>provided sterile. | Same | {6}------------------------------------------------ Image /page/6/Picture/0 description: The image shows the logo for Mallinckrodt Pharmaceuticals. The logo consists of a geometric shape made up of overlapping cubes in shades of green, blue, and purple. To the right of the geometric shape is the company name, "Mallinckrodt" in a bold, dark blue font, with the word "Pharmaceuticals" underneath in a lighter, thinner font. {7}------------------------------------------------ Image /page/7/Picture/0 description: The image shows the logo for Mallinckrodt Pharmaceuticals. On the left is a geometric design of overlapping cubes in shades of green, blue, and purple. To the right of the design is the company name, "Mallinckrodt" in large, bold, navy blue font, with the word "Pharmaceuticals" underneath in a smaller, lighter blue font. #### VII. PERFORMANCE DATA Extensive verification and validation activities were performed to ensure the OptiOne Single-Head Contrast Delivery System design outputs met all design inputs and the customer requirements were successfully met. The testing confirms the OptiOne Single-Head Contrast Delivery System meets the required specifications. The OptiOne Single-Head Contrast Delivery System was also tested in conformance to the following recognized standards: | Standard | |-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | IEC ISO 14971 Second edition 2007-03-01 Application of Risk Management to<br>Medical Devices | | ISO 15223-1 Second Edition 2012-07-01 Medical Devices - Symbols to be<br>Used with Medical Device Labels, Labelling, and Information to be Supplied -<br>Part 1 | | AAMI / ANSI ES60601-1:2005/(R)2012 and C1:2009/(R)2012 and,<br>a2:2010/(r)2012 Medical Electrical Equipment -- Part 1: General Requirements<br>for Basic Safety and Essential Performance | | IEC 60601-1-2 Edition 4.0 2014-02 Medical Electrical Equipment - Part 1-2:<br>General Requirements for Basic Safety and Essential Performance - Collateral<br>Standard: Electromagnetic Disturbances - Requirements and Tests | | IEC 60601-1-6 Edition 3.1 2013-10 Medical Electrical Equipment - Part 1-6:<br>General Requirements for Basic Safety and Essential Performance - Collateral<br>Standard: Usability | | IEC 62304 First Edition 2006-05 Medical Device Software - Software Life<br>Cycle Processes. (Software/Informatics) | | IEC 62366 Edition 1.1 2014-01 Medical Devices - Application of Usability<br>Engineering to Medical Devices. (General I) | {8}------------------------------------------------ Image /page/8/Picture/0 description: The image shows the Mallinckrodt Pharmaceuticals logo. The logo consists of a geometric design on the left and the company name on the right. The geometric design is made up of overlapping cubes in shades of green, blue, and purple. The company name is written in a dark blue font, with "Mallinckrodt" on the top line and "Pharmaceuticals" on the bottom line. ## VIII. CONCLUSIONS The OptiOne Single-Head Contrast Delivery System was developed under the Liebel-Flarsheim Company LLC Design Control processes. The verification and validation activities ensure the OptiOne meets all customer requirements and product specifications. The OptiOne Single-Head Contrast Delivery System has the same intended use as its predicate. The powerhead and software was modified from the predicate device to reflect the clinical use of a single head contrast delivery system from a dual head injector. Based on the data contained in this submission, Liebel-Flarsheim Company LLC believes the OptiOne Single-Head Contrast Delivery System is substantially equivalent to its predicate.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...