← Product Code [DYB](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DYB) · K173287

# Elite HV Radial (K173287)

_Galt Medical Corp. · DYB · Dec 21, 2017 · Cardiovascular · SESK_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DYB/K173287

## Device Facts

- **Applicant:** Galt Medical Corp.
- **Product Code:** [DYB](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DYB.md)
- **Decision Date:** Dec 21, 2017
- **Decision:** SESK
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1340
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The Galt Medical Corp Elite HV Radial is indicated to facilitate placing a catheter through the skin into a vein or artery including but not limited to the radial artery.

## Device Story

The Elite HV Radial is a catheter introducer system used in percutaneous procedures to facilitate vascular access. The device consists of a valved sheath (with sidearm and stopcock) and a dilator, often packaged with a guidewire, entry needle, and flushing syringe. The sheath features a hydrophilic coating to reduce frictional resistance during insertion and removal. Both the sheath and dilator contain barium sulfate for fluoroscopic visualization. The device is operated by clinicians in a clinical setting. The system functions by providing a conduit for catheter placement; the integrated valve minimizes blood loss during the procedure. No design changes were made to the existing Elite HV platform; this submission expands the indication to include radial artery access, utilizing existing size configurations appropriate for this anatomy.

## Clinical Evidence

No clinical data provided. Substantial equivalence is supported by bench testing and biocompatibility evaluation. Bench testing included particulate, mechanical properties, lubricity, coating durability, and dimensional comparison tests. Accelerated aging (4-year equivalent) was performed. Biocompatibility testing followed ISO 10993-1, including cytotoxicity, sensitization, irritation, systemic toxicity, hemolysis, hemocompatibility, and EO residuals.

## Technological Characteristics

Introducer system comprising a valved sheath (silicone or TPE valve) and dilator. Materials include barium sulfate for radiopacity and a hydrophilic coating. Sizes range from 4F-6F with lengths 5cm-25cm. Sterilization via ethylene oxide (EO). No software or electronic components.

## Regulatory Identification

A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.

## Predicate Devices

- Glidesheath ([K152173](/device/K152173.md))
- Elite HV ([K043525](/device/K043525.md))
- Gateway Hemostasis Valve ([K152528](/device/K152528.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 21, 2017

Galt Medical Corp. David Derrick Director of Quality and Regulatory Affairs 2220 Merritt Drive Garland, Texas 75087

# Re: K173287

Trade/Device Name: Elite HV Radial Regulation Number: 21 CFR 870.1340 Regulation Name: Catheter Introducer Regulatory Class: Class II Product Code: DYB Dated: October 13, 2017 Received: October 16, 2017

# Dear David Derrick:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. However, you are responsible to determine that the medical devices you use as components in the kit have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were legally on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit/tray. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Kenneth J. Cavanaugh -S

for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below

510(k) Number (if known)

K173287

Device Name ELITE HV RADIAL

Indications for Use (Describe)

The Galt Medical Corp Elite HV Radial is indicated to facilitate placing a catheter through the skin into a vein or artery including but not limited to the radial artery.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

__ Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## K173287

### Section 5 – 510(k) Summary

| Application Date:      | October 13, 2017                                                                                                                                                                                         |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Application Type:      | Traditional 510(k)                                                                                                                                                                                       |
| Applicant Information  | Galt Medical Corporation<br>2220 Merritt Dr.<br>Garland, TX 75041<br>Phone: 214-778-1306<br>Fax: 972-271-4706                                                                                            |
| Official Contact:      | David Derrick<br>Director of Quality and Regulatory Affairs<br>Galt Medical Corporation<br>2220 Merritt Dr.<br>Garland, TX 75041<br>Phone: 214-778-1306<br>Fax: 972-271-4706<br>dderrick@galtmedical.com |
| Device Name:           | Elite HV Radial                                                                                                                                                                                          |
| Device Model Number:   | TBD                                                                                                                                                                                                      |
| Classification Name:   | Catheter Introducer (DYB),<br>21 CFR 870.1340                                                                                                                                                            |
| Device Classification: | Class II (Cardiovascular)                                                                                                                                                                                |
| Predicate Device:      | Glidesheath (K152173)<br>Elite HV (K043525<br>Gateway Hemostasis Valve (152528)                                                                                                                          |
| Manufacturer:          | Galt Medical<br>2220 Merritt Drive<br>Garland, TX 75041<br>Phone: 214-778-5177<br>Fax: 972-271-4706                                                                                                      |

Establishment Registration Number: 1649395

Intended Use: The Galt Medical Corp Elite HV Radial is indicated to facilitate placing a catheter through the skin into a vein or artery including but not limited to the radial artery.

Device Description: The Galt Medical Corp Elite HV product has been on the market since February of 2005, and was cleared under 510(k) K043525. The clearance included sizes from 4-9FR, and lengths from 5-110 cm with the following indication for use:

{4}------------------------------------------------

The Sheath Introducer system is indicated for use in percutaneous procedures to introduce catheters and other intravascular devices into the vasculature.

The purpose of this 510(k) is to expand that indication for use for the Elite HV products to include radial placement. There were no design changes required as the current Galt offering included sizes which are appropriate for radial placement. The only limitation of the Radial placement claim is that the range of sizes is more limited due to the target anatomy.

The Elite HV and Elite HV Radial consists of an introducer (valved sheath with sidearm and stopcock and dilator), which are packaged together with a guide wire. The Elite HV product is configured with a silicone valve or a TPE valve cleared under 510(k) K152528. Some product configurations also include an entry needle, guide wire inserter and a flushing syringe. The Elite HV Radial is used to facilitate placing a catheter through the skin into a vein or artery including but not limited to the radial artery. The sheath is coated with hydrophilic coating to minimize frictional resistance when inserting or removing the sheath from the patient's blood vessel. In addition, the sheath and dilator contain barium sulfate, making these devices visible under fluoroscopy.

Comparison of Technological Characteristics: The subject device Elite HV Radial and the predicate device Terumo Glidesheath have identical indication statements and are of similar design. The subject and predicate device are available in similar configurations.

|                               | Subject Device                                                                                                                                                                                     | Predicate Device                                                                                                                                                                    | Predicate Device                                                                                                                                                           | Predicate Device                                                                                                                                                                                        |
|-------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Mfr. /<br>Product             | Galt Medical / Elite<br>HV Radial                                                                                                                                                                  | Galt Medical / Elite<br>HV                                                                                                                                                          | Terumo /<br>Glidesheath™                                                                                                                                                   | Galt Medical /<br>Gateway Hemostasis<br>Valve                                                                                                                                                           |
| 510(k)<br>Number              |                                                                                                                                                                                                    | K043525                                                                                                                                                                             | K152173                                                                                                                                                                    | K152528                                                                                                                                                                                                 |
| Device<br>Classification<br>n | 870.1340                                                                                                                                                                                           | 870.1340                                                                                                                                                                            | 870.1340                                                                                                                                                                   | 870.4290                                                                                                                                                                                                |
| Product<br>Code               | DYB                                                                                                                                                                                                | DYB                                                                                                                                                                                 | DYB                                                                                                                                                                        | DTL                                                                                                                                                                                                     |
| Intended<br>use               | The Galt Medical<br>Corp Elite HV Radial<br>is indicated to<br>facilitate placing a<br>catheter through the<br>skin into a vein or<br>artery including but<br>not limited to the<br>radial artery. | The Galt Medical Corp<br>Elite HV is indicated<br>for use in<br>percutaneous<br>procedures to<br>introduce catheters<br>and other<br>intravascular devices<br>into the vasculature. | The Glidesheath™ is<br>indicated to facilitate<br>placing a catheter<br>through the skin into<br>a vein or artery<br>including but not<br>limited to the radial<br>artery. | The attachable Cath<br>Lab Hemostasis Valve<br>is indicated to<br>minimize blood loss<br>during introduction<br>of catheters,<br>guidewires and other<br>intravascular devices<br>into the vasculature. |
| Design                        | Introducer with valved<br>sheath and dilator in a<br>range of sizes and<br>lengths                                                                                                                 | Introducer with valved<br>sheath and dilator in a<br>range of sizes and<br>lengths                                                                                                  | Introducer with valved<br>sheath and dilator in a<br>range of sizes and<br>lengths                                                                                         | Introducer with valved<br>hub and luer lock<br>connector                                                                                                                                                |
| Color                         | Sheath Hub – White                                                                                                                                                                                 | Sheath Hub - White                                                                                                                                                                  | Sheath Hub - Various                                                                                                                                                       | Sheath Hub - White                                                                                                                                                                                      |

{5}------------------------------------------------

|              | Subject Device                                                                                                                                                                                                                | Predicate Device                                                                                                                                                                           | Predicate Device                                                                                                                                                                                  | Predicate Device                                                                                              |
|--------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|
|              | Sheath cannula-<br>White<br>Sheath cap- Various<br>indicating Fr size<br>Dilator hub- White<br>Dilator cannula- Blue<br>Sidearm with<br>stopcock - Clear<br>Valve silicone - clear<br>natural<br>Valve TPE - clear<br>natural | Sheath cannula- White<br>Sheath cap- Various<br>indicating Fr size<br>Dilator hub- White<br>Dilator cannula- Blue<br>Sidearm with stopcock<br>- Clear<br>Valve silicone - clear<br>natural | Sheath cannula-<br>White<br>Sheath cap- Various<br>indicating Fr size<br>Dilator hub- Various<br>indicating Fr size<br>Dilator cannula- Blue<br>Sidearm with<br>stopcock - Clear<br>Valve - clear | Sheath cap- Various<br>indicating Fr size<br>Sidearm with<br>stopcock - Clear<br>Valve TPE - clear<br>natural |
| Sizes        | 4F - 6F<br>5cm to 25cm lengths                                                                                                                                                                                                | 4F - 9F<br>5cm to 110cm lengths                                                                                                                                                            | 4F - 6F<br>10cm to 25cm<br>lengths                                                                                                                                                                | Maximum 8FR                                                                                                   |
| Dilator Lock | Dilator is retained in<br>sheath cap                                                                                                                                                                                          | Dilator is retained in<br>sheath cap                                                                                                                                                       | Dilator is retained on<br>sheath cap                                                                                                                                                              | Not applicable                                                                                                |

{6}------------------------------------------------

Substantial Equivalence and Summary of Bench Testing: The technological differences between the subject device and predicate device have been evaluated through bench and biocompatibility tests to provide evidence of substantial equivalence. The Elite HV Radial is substantially equivalent to the specified predicate device based on comparisons of the devices functionality, compatibility, technological characteristics, and indications for use.

Testing was conducted according to protocols based on international standards and Galt Medical requirements. Functional Testing included the following:

- . Particulate Test
- Mechanical Properties Test
- Lubricity Test
- . Durability of coating
- . Dimensional comparison test

Additionally, accelerated aging was performed on the subject devices to an equivalent age of 4 years, and the functional testing was repeated.

Biocompatibility testing was performed on the subject device in accordance with ISO 10993-1. This was done in order to update the biocompatibility testing to current standards, and was not done as a result of any changes to the product. Biocompatibility testing included the following:

- Cytotoxicity (MEM Elution)
- Sensitization (Magnusson - Kligman)
- . Irritation (Intracutaneous injection)
- Systemic Toxicity (Systemic injection)
- Hemolysis (Direct Contact)
- Hemocompatibility (Complement Activation)
- In-Vitro Hemocompatibility (Dog Thrombogenicity)
- Systemic Toxicity (Materials Mediated Pyrogen)
- EO Residuals (EO, ECH, & EG)

Additionally the subject device was adopted into the existing ethylene oxide sterilization cycle for the Galt product cleared under K043525 & K152528.

Packaging of the subject device will remain unchanged from the Galt Elite HV products. Current packaging shelf life testing was provided in the predicate submission.

Conclusion: It will be shown in this 510(k) submission that the differences between the Galt Elite HV Radial and the predicate devices do not raise any new questions regarding safety and effectiveness. The Galt Elite HV Radial as designed and manufactured is determined to be substantially equivalent to the current marketed predicate device.

End of Section

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DYB/K173287](https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DYB/K173287)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
