← Product Code [DYB](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DYB) · K162427

# Vado Bi-Directional Steerable Sheath (K162427)

_Kalila Medical, Inc. · DYB · Jan 27, 2017 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DYB/K162427

## Device Facts

- **Applicant:** Kalila Medical, Inc.
- **Product Code:** [DYB](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DYB.md)
- **Decision Date:** Jan 27, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1340
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The Vado® Bi-Directional Steerable Sheath is introducing various cardiovascular catheters into the vasculature and into the chambers of the heart including the heart through the interatrial septum.

## Device Story

Vado® Bi-Directional Steerable Sheath is a catheter introducer system comprising a dilator and a steerable sheath. Used in cardiac procedures to provide intracardiac access and stable catheter positioning. Operator (physician) uses a handle-mounted deflection knob to steer the sheath tip (clockwise ≥ 180°, counterclockwise ≥ 90°). Includes a hemostasis valve to minimize blood loss and a side port with a 3-way stopcock for aspiration, fluid infusion, blood sampling, and pressure monitoring. Radiopaque marker allows visualization under fluoroscopy. Distal tip vent holes facilitate aspiration. Device is used in clinical settings to facilitate the placement of cardiovascular catheters.

## Clinical Evidence

No clinical data provided. Substantial equivalence supported by bench performance testing, including biocompatibility (cytotoxicity, sensitization, irritation, systemic toxicity, pyrogenicity, hemolysis, complement activation, thrombosis), radiopacity, corrosion resistance, dimensional inspection, deflection cycling, leakage resistance, torque strength, kink resistance, and tensile strength. Packaging integrity also verified.

## Technological Characteristics

Catheter introducer system; 8.8 Fr sheath; 91 cm length. Features rotary-type deflection mechanism. Materials include radiopaque markers. Sterilized via EO gas. Connectivity: None (mechanical device).

## Regulatory Identification

A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.

## Predicate Devices

- Vado Steerable Sheath ([K140420](/device/K140420.md))
- Agilis NxT Steerable Introducer ([K061363](/device/K061363.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus, which is a symbol often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

January 27, 2017

Kalila Medical, Inc. Carrie Neuberger Regulatory Affairs 745 Camden Ave, Suite A Campbell, California 95008-4146

Re: K162427

Trade/Device Name: Vado® Bi-Directional Steerable Sheath Regulation Number: 21 CFR 870.1340 Regulation Name: Catheter Introducer Regulatory Class: Class II Product Code: DYB Dated: December 20, 2016 Received: December 28, 2016

Dear Carrie Neuberger:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Fernando
Fernando Aguel-S
Aguel

for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K162427

## Device Name

Vado® Bi-Directional Steerable Sheath

Indications for Use (Describe)

The Vado® Bi-Directional Steerable Sheath is introducing various cardiovascular catheters into the vasculature and into the chambers of the heart including the heart through the interatrial septum.

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
|----------------------------------------------|---------------------------------------------|
|----------------------------------------------|---------------------------------------------|

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image contains the logo for Kalila Medical. The logo consists of an orange and gray abstract shape resembling two intertwined curves, followed by the text "Kalila Medical" in a combination of orange and gray. The word "Kalila" is in orange, while "Medical" is in gray.

# 510(k) Summary¨¨¨¨

| 510(k) Number                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | K162427                                                                                                                                  |               |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------|---------------|
| Submitter Name and Address                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                          |               |
| Name:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Kalila Medical, Inc.                                                                                                                     |               |
| Address:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | 745 Camden Avenue, Suite A<br>Campbell, CA 95008                                                                                         |               |
| Telephone:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | 408-637-2150                                                                                                                             |               |
| Fax:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | 408-370-6325                                                                                                                             |               |
| Contact:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Carrie Neuberger                                                                                                                         |               |
| Date Prepared:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | December 20, 2016                                                                                                                        |               |
| General Device Information                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                          |               |
| Device Name:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Vado® Bi-Directional Steerable Sheath                                                                                                    |               |
| Common Name:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Catheter introducer                                                                                                                      |               |
| Classification:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | 21 CFR 870.1340<br>A catheter introducer is a sheath used to<br>facilitate placing a catheter through the skin<br>into a vein or artery. |               |
| Device Class:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Class II                                                                                                                                 |               |
| Product Code:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | DYB                                                                                                                                      |               |
| Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |                                                                                                                                          |               |
| Manufacturer                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Device Name                                                                                                                              | 510(k) Number |
| Kalila Medical                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Vado Steerable Sheath                                                                                                                    | K140420       |
| St. Jude Medical                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Agilis NxT Steerable<br>Introducer                                                                                                       | K061363       |
| Device Description                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                          |               |
| The Vado® Bi-Directional Steerable Sheath consists of a dilator and steerable<br>sheath which are designed to provide intracardiac access and flexible, stable<br>catheter positioning in the cardiac anatomy. A hemostasis valve in the handle<br>minimizes blood loss during catheter introduction and/or exchange. A side port<br>with 3-way stopcock is provided for air or blood aspiration, fluid infusion,<br>blood sampling and pressure monitoring. A deflection knob on the handle<br>deflects the tip of the sheath clockwise ≥ 180° and counterclockwise ≥ 90°.<br>Distal tip vent holes facilitate aspiration and a radiopaque marker permits<br>visualization of the sheath tip under fluoroscopy. |                                                                                                                                          |               |
| Indications for Use                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |                                                                                                                                          |               |
| The Vado Bi-Directional Steerable Sheath is indicated for introducing various<br>cardiovascular catheters into the vasculature and into the chambers of the<br>heart including the left side of the heart through the interatrial septum.                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |                                                                                                                                          |               |
| Comparison to the Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                          |               |
| The Vado Bi-Directional Steerable Sheath has the same indications for use and<br>intended use as the predicate devices. The Vado Bi-Directional Steerable Sheath                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |                                                                                                                                          |               |

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Kalila Medical. The logo consists of an orange and gray abstract shape on the left, followed by the words "Kalila Medical" in orange and gray. The word "Kalila" is in orange, while the word "Medical" is in gray.

utilizes the same or similar materials, design principals and fundamental scientific technology as the predicate device(s).

| Device<br>Characteristic          | Subject Device:<br>Vado Bi-Directional<br>Steerable Sheath                                                                                                                                                                                                        | Predicate Device:<br>Vado Steerable Sheath<br>(K140420)                                                                                                                                                                                            | Predicate Device:<br>Agilis NxT Introducer<br>(K061363)                                                                                                                                                         |
|-----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Description                       | Dilator/introducer with<br>hemostatic hub<br>designed to introduce<br>sheath into the vessel;<br>steerable catheter<br>introducer                                                                                                                                 | Dilator/introducer<br>with hemostatic hub<br>designed to introduce<br>sheath into the vessel;<br>steerable catheter<br>introducer                                                                                                                  | Dilator/introducer with<br>hemostatic hub<br>designed to introduce<br>sheath into the vessel;<br>steerable catheter<br>introducer                                                                               |
| Indications<br>for Use            | The Vado Bi-Directional<br>Steerable Sheath is<br>indicated for<br>introducing various<br>cardiovascular<br>catheters into the<br>vasculature and into the<br>chambers of the heart<br>including the left side of<br>the heart through the<br>interatrial septum. | The Vado Steerable<br>Sheath is indicated for<br>introducing various<br>cardiovascular<br>catheters into the<br>vasculature and into<br>the chambers of the heart<br>including the left<br>side of the heart<br>through the interatrial<br>septum. | The Agilis NxT<br>Steerable Introducer is<br>indicated when<br>introducing various<br>cardiovascular<br>catheters into the heart,<br>including the left side of<br>the heart through the<br>interatrial septum. |
| Intended Use                      | Catheter Introducer                                                                                                                                                                                                                                               | Catheter Introducer                                                                                                                                                                                                                                | Catheter Introducer                                                                                                                                                                                             |
| Handle<br>Deflection<br>Mechanism | Rotary type deflection                                                                                                                                                                                                                                            | Rotary type deflection                                                                                                                                                                                                                             | Rotary type deflection                                                                                                                                                                                          |
| Sheath<br>French Size             | 8.8 Fr                                                                                                                                                                                                                                                            | 8.8 Fr                                                                                                                                                                                                                                             | 8.5 Fr                                                                                                                                                                                                          |
| Sheath<br>Length<br>(Total)       | 91 cm                                                                                                                                                                                                                                                             | 83 cm                                                                                                                                                                                                                                              | 91 cm (and 81cm)                                                                                                                                                                                                |
| Dilator<br>Length<br>(Usable)     | 93 cm                                                                                                                                                                                                                                                             | 85 cm                                                                                                                                                                                                                                              | 94 cm (and 85 cm)                                                                                                                                                                                               |
| Guide wire<br>compatibility       | 0.032" max                                                                                                                                                                                                                                                        | 0.032" max                                                                                                                                                                                                                                         | 0.032" max                                                                                                                                                                                                      |
| Tip<br>Deflection                 | Bi-Directional<br>(180°/90°)                                                                                                                                                                                                                                      | Uni-Directional<br>(140°)                                                                                                                                                                                                                          | Bi-Directional<br>(180°/90°)                                                                                                                                                                                    |
| Packaging<br>Configuration        | Tray inside pouch<br>inside shelf carton                                                                                                                                                                                                                          | Tray inside pouch<br>inside shelf carton                                                                                                                                                                                                           | Tray inside pouch<br>inside shelf carton                                                                                                                                                                        |
| Sterilization                     | EO gas                                                                                                                                                                                                                                                            | EO gas                                                                                                                                                                                                                                             | EO gas                                                                                                                                                                                                          |

Where design differences exist between the subject device and the predicate device, performance testing demonstrates that these differences do not adversely affect safety and effectiveness.

This submission supports the position the Vado® Bi-Directional Steerable Sheath, FG1_008, is substantially equivalent to the Vado Steerable Sheath (K140420) and the Agilis NxT Steerable Introducer (K061363).

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Kalila Medical. The logo consists of an orange square with a stylized "S" inside, followed by the words "Kalila Medical" in orange and gray. The word "Kalila" is in orange, and the word "Medical" is in gray.

Summary of Non-Clinical and Clinical Testing

The 510(k) notice contains summaries of device and packaging performance tests, shelf life tests, biocompatibility, and sterilization studies conducted to evaluate the performance characteristics of the Vado® Bi-Directional Steerable Sheath, FG1 008. The data presented demonstrates that the Vado® Bi-Directional Steerable Sheath met its functional and performance characteristics in accordance with applicable industry standards and is equivalent to the predicate devices.

To verify the Vado® Bi-Directional Steerable Sheath met its functional and performance requirements, representative sterilized samples of the device underwent sterilization, biocompatibility, bench testing, packaging integrity, and shelf life testing.

Biocompatibility Tests:

- 1. Cytotoxicity (MEM Elution)
- 2. Sensitization (Guinea Pig Maximization Sensitization)
- 3. Irritation (Intracutaneous Reactivity Irritation in Rabbits)
- 4. Acute Systemic Toxicity
- 5. Pyrogen (Materials Mediated in Rabbits)
- 6. Hemolysis (Direct Contact)
- 7. Hemolysis (Indirect Contact)
- 8. Complement Activation
- 9. Thrombosis

### Bench Performance Tests:

- 1. Radiopacity
- 2. Corrosion Resistance
- 3. Sheath and Dilator Visual Inspection
- 4. Sheath Dimension Inspection
- 5. Dilator Dimension Inspection
- 6. Sheath Deflection
- 7. Sheath Curvature Dimensions
- 8. Insertion and Deflection Cycling
- 9. Aspiration Air Ingress
- 10. Valve Leakage Resistance
- 11. Sheath Leakage Resistance
- 12. Device Preparation
- 13. Valve Integrity and Functionality
- 14. Shaft Torque Strength
- 15. Kink Resistance
- 16. Marker Band Location
- 17. Tensile Strength

Packaging Integrity Tests:

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image contains the logo for Kalila Medical. The logo consists of an orange and gray abstract shape on the left, followed by the words "Kalila Medical" in orange and gray. The word "Kalila" is in orange, while the word "Medical" is in gray.

- 1. Pouch Seal Strength
- 2. Gross Leak Detection

Clinical testing is not provided in this submission.

Statement of Equivalence

The Vado Bi-Directional Steerable Sheath has the same indications for use and the same or similar technological characteristics as the predicate devices. Testing has demonstrated that the Vado Bi-Directional Steerable Sheath does not raise any new questions of safety and effectiveness. Based on this and the data provided in this pre-market notification, the subject device and the predicate devices have been shown to be substantially equivalent.

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DYB/K162427](https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DYB/K162427)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
