← Product Code [DYB](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DYB) · K123475

# SHOOCIN INTRODUCER KIT (K123475)

_Lepu Medical Technology (Beijing) Co., Ltd. · DYB · Feb 3, 2014 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DYB/K123475

## Device Facts

- **Applicant:** Lepu Medical Technology (Beijing) Co., Ltd.
- **Product Code:** [DYB](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DYB.md)
- **Decision Date:** Feb 3, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1340
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The ShoocinTM Introducer Kits are intended for use to facilitate the introduction of guide wires, catheters and other accessory medical devices through the skin into a vein or artery and minimize blood loss associated with such introduction.

## Device Story

Shoocin Introducer Kit is a vascular access device consisting of a puncture needle, guide wire with collimator, sheath introducer, and dilator. Puncture needle creates initial access; guide wire acts as a track for sheath insertion. Sheath introducer provides a conduit for interventional devices; includes hemostasis valve and side port with 3-way stopcock to minimize blood loss. Dilator provides stability during sheath deployment. Components are radiopaque for fluoroscopic visualization. Used in clinical settings by physicians to gain vascular access for interventional procedures. Benefits include facilitated device delivery and reduced hemorrhage during access.

## Clinical Evidence

Bench testing only. No clinical data provided. Performance testing included biocompatibility (ISO 10993-1, -4, -5, -10, -11), fracture strength, leakage performance, radiodetectability, and guide wire/dilator accessibility. All tests met specified requirements.

## Technological Characteristics

Components: Stainless steel needle, stainless steel guide wire, polycarbonate/acrylobutyistyrene/polyethylene/polyurethane/silicone plastics. Sheath/dilator contain bismuth for radiopacity. Sterilization: Ethylene oxide per AAMI/ANSI/ISO 11135:2007. Shelf life: 2 years per ISO 11070. Connectivity: None. Software: None.

## Regulatory Identification

A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.

## Predicate Devices

- Pinnacle Precision Access System ([K111606](/device/K111606.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K.123475
FEB - 3 2014

## Section 5: 510k) Summary

ﻪ ﺍ

The Summary of Safety and Ifferintion on the Shocial" in the submitted hat or revience with the requirences in the requirences in the requirences in the regirents in the rece

| Applicant:           | Lepu Medical Technology (Beijing) Co., Ltd.<br>No. 37 Chaoqian Road<br>Changping District, Beijing 102200<br>P.R. China                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
|----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Telephone:           | +86-10-80120641                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Contact:             | Shan Zhao                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Date:                | November 6, 2012                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Name:                | ShoocinTM Introducer Kit                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Classification Name: | Catheter Introducer, 870.1340                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Product Code:        | DYB                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Predicate:           | Pinnacle Precision Access System, Terumo Corporation, K111606 with market<br>clearance dates of October 3, 2011.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Description:         | The ShoocinTM Introducer Kit consists of a puncture needle, a guide wire with<br>a guide wire collimator, a sheath introducer, and a dilator. The puncture<br>needle incorporates a lumen, which provides a conduit for the insertion of the<br>guide wire into the vascular system. The flexible stainless steel guide wire is<br>utilized as a guiding mechanism for the insertion of the introduction sheath<br>into the vascular system. The guide wire contains a wire collimator, which<br>assists in funneling the wire through the lumen of the puncture needle and<br>contains marking that are visible under fluoroscopy, which can determine in<br>length of the guide wire within the vascular system. The sheath introducer<br>provides a conduit for introducing other interventional devices, including<br>guide wires and interventional catheters, into the vascular system. The main<br>components of this assembly are a non-hydrophilic coated sheath introducer,<br>hemostasis valve housing, and a side port with tubing connected to a 3-way<br>stopcock/valve. The dilator is used to provide support and stability to the<br>sheath introducer during deployment into the vascular system. The proximal<br>end of the dilator includes a luer port and has a tapered, atraumatic distal tip.<br>Both the sheath and dilator contain bismuth, making these devices visible<br>under fluoroscopy. |
| Intended Use:        | The ShoocinTM Introducer Kits are intended for use to facilitate the<br>introduction of guide wires, catheters and other accessory medical devices<br>through the skin into a vein or artery and minimize blood loss associated with<br>such introduction.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |

{1}------------------------------------------------

## Section 5: 510k) Summary

ﺰ

:

·

| Specification<br>Comparison: |                           |                                                       | The Shoocin™ Introducer Kit specifications are similar to the Pinnacle Access<br>System (K111606) and the differences do not raise any new issues of safety or<br>effectiveness.                                                                                                                                                               |
|------------------------------|---------------------------|-------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                              |                           | Pinnacle                                              | Shoocin                                                                                                                                                                                                                                                                                                                                        |
|                              | Sheath Diameter:          | 4F through 11F                                        | 4F through 11F                                                                                                                                                                                                                                                                                                                                 |
|                              | Sheath Length:            | 10cm - 25cm                                           | 7cm - 23cm                                                                                                                                                                                                                                                                                                                                     |
|                              | Guide Wire Length:        | 45cm                                                  | 45cm and 80 cm                                                                                                                                                                                                                                                                                                                                 |
|                              | Guide Wire OD:            | .035" - .038"                                         | .018" - .038"                                                                                                                                                                                                                                                                                                                                  |
|                              | Introducer Needle:        | 19 – 21G                                              | 18 – 20G                                                                                                                                                                                                                                                                                                                                       |
|                              | Dilator:                  | 15.5cm or 16.0cm                                      | 7cm, 11cm, 16cm, 23cm                                                                                                                                                                                                                                                                                                                          |
|                              |                           |                                                       |                                                                                                                                                                                                                                                                                                                                                |
| Materials<br>Comparison:     |                           |                                                       | The Shoocin ™ Introducer Kit materials have been subjected to<br>biocompatibility tests and the differences between the Pinnacle Access System<br>(K111606) do not raise any new issues of safety or effectiveness.                                                                                                                            |
|                              |                           | Pinnacle                                              | Shoocin                                                                                                                                                                                                                                                                                                                                        |
|                              | Needle:                   | Stainless Steel                                       | Stainless Steel                                                                                                                                                                                                                                                                                                                                |
|                              | Needle Luer:              | Styrene-Butadiene                                     | Polycarbonate                                                                                                                                                                                                                                                                                                                                  |
|                              | Guide Wire:               | Stainless Steel or Nitinol<br>w/Palladium distal coil | Stainless Steel                                                                                                                                                                                                                                                                                                                                |
|                              | Guide Wire<br>Collimator: | Polypropylene                                         | Polycarbonate                                                                                                                                                                                                                                                                                                                                  |
|                              | Dilator Tube:             | Polypropylene w/Bismuth<br>subcarbonate               | Polypropylene                                                                                                                                                                                                                                                                                                                                  |
|                              | Dilator Luer:             | Polypropylene                                         | Acrylobutyistyrene                                                                                                                                                                                                                                                                                                                             |
|                              | Outer Sheath:             | ETFE w15% BiO3, w/2%<br>TiO2                          | Fluorinated Ethylene<br>Propylene                                                                                                                                                                                                                                                                                                              |
|                              | Sheath base:              | Styrene elastomer                                     | Acrylobutyistyrene                                                                                                                                                                                                                                                                                                                             |
|                              | Side Port Tubing          | Polybutadiene                                         | Polyurethane                                                                                                                                                                                                                                                                                                                                   |
|                              | Hemostasis Valve:         | Silicone Rubber                                       | Silicone                                                                                                                                                                                                                                                                                                                                       |
|                              | 3-Way Stop Cock:          | Polypropylene                                         | Acrylobutyistyrene                                                                                                                                                                                                                                                                                                                             |
|                              | Cap:                      | Polypropylene                                         | Polyethylene                                                                                                                                                                                                                                                                                                                                   |
|                              | Valve:                    | Polypropylene                                         | Polyethylene                                                                                                                                                                                                                                                                                                                                   |
| Sterilization:               |                           |                                                       | The method used is based on practices recommended by AAMI / ANSI / ISO<br>11135:2007 and provides a Sterility Assurance Level (SAL) of 106                                                                                                                                                                                                     |
| Shelf Life:                  |                           |                                                       | In accordance with ISO 11070 the real time aging of Shoocin ™ Introducer Kit<br>demonstrated that the performance of the specific components met the<br>standard requirements without any significant difference to product<br>performance requirements before aging. So the product is stable and reliable<br>within the two-year useful life |

…… ち "

, . .. !

· 《中语·

*

. '.

ﺒﻴ

. '

{2}------------------------------------------------

| <b>Section 5: 510k) Summary</b>                     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |                                              |                                                                                                                          |
|-----------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|
| Biocompatibility:                                   | The ShoocinTM Introducer Kit produced by Lepu Medical was assessed<br>against the International Standard ISO 10993-1, "Biological evaluation of<br>medical devices. Part 1. Guidance on selection of tests." The ShoocinTM<br>Introducer Kit would be classified as an External Communicating Device in<br>contact with the Circulating Blood for a Limited Duration (<24 hours). The<br>following test would be required for any patient / user contacting material:                                                                                                                                      |                                              |                                                                                                                          |
|                                                     | Test                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Standard                                     | Results                                                                                                                  |
|                                                     | Haemolytsis                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | ISO 10993-4                                  | Both the test article in direct<br>contact with blood and the test<br>article extract were non-<br>hemolytic.            |
|                                                     | Cytotoxicity                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | ISO 10993-5                                  | The test article extract showed<br>no evidence of causing cell<br>lysis or toxicity                                      |
|                                                     | Maximum<br>Sensitization                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | ISO 10993-10                                 | The test article extracts showed<br>no evidence of causing delayed<br>dermal contact sensitization in<br>the guinea pig. |
|                                                     | Intracutaneous<br>Irritation                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | ISO 10993-10                                 | The test article met the<br>requirements for the SC and SO<br>test extracts.                                             |
|                                                     | Systemic Toxicity                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | ISO 10993-11                                 | There was no mortality or<br>evidence of systemic toxicity<br>from the extracts injected into<br>mice.                   |
|                                                     | USP Pyrogen Study                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | ISO 10993-11                                 | The test article was judged as<br>nonpyrogenic.                                                                          |
|                                                     | <i>In Vivo</i><br>Thromboresistance                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | ISO 10993-4                                  | The test article and control<br>were thromboresistant and<br>comparable.                                                 |
|                                                     | Partial Thromboplastin<br>Time                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | ISO 10993-4                                  | The test article would be<br>considered a minimal activator<br>and met the requirements of the<br>test.                  |
| Performance Testing:                                | The ShoocinTM Introducer Kit successfully passed all of the following<br>performance tests:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |                                              |                                                                                                                          |
|                                                     | ShoocinTM                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Introducer Kit Performance Testing           |                                                                                                                          |
|                                                     | Sheath Introducer                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Sheath Introducer should be free of defects  |                                                                                                                          |
|                                                     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Sheath Introducer ID                         |                                                                                                                          |
|                                                     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Sheath Introducer Effective Length           |                                                                                                                          |
|                                                     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Radiodetectability                           |                                                                                                                          |
|                                                     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Dilator accessibility                        |                                                                                                                          |
|                                                     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Fracture Strength Test                       |                                                                                                                          |
|                                                     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Leaking Performance Test                     |                                                                                                                          |
| <b>Section 5: 510k) Summary</b>                     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |                                              |                                                                                                                          |
| Performance Testing:                                | The ShoocinTM Introducer Kit successfully passed all of the following performance tests:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |                                              |                                                                                                                          |
|                                                     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | ShoocinTM Introducer Kit Performance Testing |                                                                                                                          |
|                                                     | Dilator                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Dilator should be free of defects            |                                                                                                                          |
|                                                     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Dilator OD                                   |                                                                                                                          |
|                                                     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Dilator ID                                   |                                                                                                                          |
|                                                     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Dilator Effective Length                     |                                                                                                                          |
|                                                     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Radiodetectability                           |                                                                                                                          |
|                                                     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Guide wire accessibility                     |                                                                                                                          |
|                                                     | Guide wire                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Fracture Strength Test                       |                                                                                                                          |
|                                                     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Surface free from impurity                   |                                                                                                                          |
|                                                     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Guide wire length                            |                                                                                                                          |
|                                                     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Guide wire OD                                |                                                                                                                          |
|                                                     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Radiodetectability                           |                                                                                                                          |
|                                                     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Corrosion Resistance                         |                                                                                                                          |
|                                                     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Fracture Strength Test                       |                                                                                                                          |
|                                                     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Flexing Resistance Test                      |                                                                                                                          |
|                                                     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Associative Guide Wire Strength Test         |                                                                                                                          |
|                                                     | Puncture Needle                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Surface free from defects                    |                                                                                                                          |
|                                                     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Needle OD                                    |                                                                                                                          |
|                                                     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Needle ID                                    |                                                                                                                          |
|                                                     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Needle Effective Length                      |                                                                                                                          |
|                                                     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Accessibility - Guide wire                   |                                                                                                                          |
|                                                     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Corrosion Resistance                         |                                                                                                                          |
|                                                     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Joint Strength - Needle to hub               |                                                                                                                          |
| <b>Substantial<br/>Equivalency<br/>Information:</b> | The information provided in this submission, comparing intended use, principle of operation and performance, the ShoocinTM Introducer Kit device is substantially equivalent to existing legally marketed device.                                                                                                                                                                                                                                                                                                                                                                                          |                                              |                                                                                                                          |
| <b>Conclusion:</b>                                  | The information provided in this submission and comparing intended use, principle of operation and overall technological characteristics (i.e. puncture needle, guide wire, dilator, and sheath introducer to obtain access to the vascular system), the ShoocinTM Introducer Kit supports a determination of substantially equivalent to existing legally marketed predicate device Pinnacle Precision Access System. Any technological differences between the ShoocinTM Introducer Kit and the predicate Pinnacle Precision Access System device do not raise new questions of safety or effectiveness. |                                              |                                                                                                                          |

:

{3}------------------------------------------------

4

{4}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three wavy lines underneath, representing the department's mission to protect the health of all Americans. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle symbol. The logo is printed in black and white.

## 11 February 3, 2014

Lepu Medical Technology (Beijing) Co., Ltd. c/o Arthur Goddard FDA Regulatory and Quality Systems Consultant 1531 Felton Road South Euclid, Ohio 44121-2722

Re: K123475

Trade Name: Shoocin™ Introducer Kit Dated: January 21, 2014 Received: January 27, 2014

Dear Mr. Goddard:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

## Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

{5}------------------------------------------------

Page 2 - Mr. Goddard

" If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDeyices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

M. Zillberman

for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Section 4: Indication for Use Summary

| 510(k) Number (if known): | K123475                              |
|---------------------------|--------------------------------------|
| Device Name:              | Shoocin <sup>TM</sup> Introducer Kit |

Indications For Use:

The Shoocin<sup>TM</sup> Introducer Kits are intended for use to facilitate the introduction of guide wires, catheters and other accessory medical devices through the skin into a vein or artery and minimize blood loss associated with such introduction.

| Prescription Use     | X | OR | Over-The-Counter-Use |
|----------------------|---|----|----------------------|
| (Per 21 CFR 801.109) |   |    |                      |

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

M. Z. Hillman

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

1

------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

.

.

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DYB/K123475](https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DYB/K123475)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
