← Product Code [DYB](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DYB) · K092575

# ENGAGE/ENGAGE TR (K092575)

_St Jude Medical · DYB · Oct 8, 2009 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DYB/K092575

## Device Facts

- **Applicant:** St Jude Medical
- **Product Code:** [DYB](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DYB.md)
- **Decision Date:** Oct 8, 2009
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1340
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** 3rd-Party Reviewed

## Indications for Use

The introduction of angiographic catheters, closed end catheters, balloon catheters, and electrodes into a blood vessel (including but not limited to femoral, radial, and brachial access) where minimizing blood loss is essential.

## Device Story

Engage Introducer facilitates vascular access for catheters and electrodes; minimizes blood loss during procedures. Device comprises hemostasis sheath assembly and dilator. Sheath includes snap-lock hub, hemostasis valve, and side port with 3-way stopcock. Dilator provides support and guidewire compatibility (0.035"-0.038"). Used by clinicians in clinical settings for femoral, radial, or brachial access. Operates as mechanical conduit for interventional devices. No electronic or software components.

## Clinical Evidence

No clinical or animal data provided. Substantial equivalence established via bench testing, including functional, packaging, sterilization, biocompatibility, and shelf-life verification.

## Technological Characteristics

Hemostasis sheath assembly and dilator; 4-9F sizes; 12cm-25cm lengths; 0.035"-0.038" guidewire compatibility. Features snap-lock hub, hemostasis valve, and side port with 3-way stopcock. Mechanical device; no energy source, software, or connectivity.

## Regulatory Identification

A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.

## Predicate Devices

- Engage/Engage TR Introducer ([K091137](/device/K091137.md))
- Ultimum Hemostasis Introducer ([K001346](/device/K001346.md))
- Fast Cath Hemostasis Introducer ([K914090](/device/K914090.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K092575

# Section 7. 510(k) Summary

#### 7.1 Applicant Information

OCT - 8 2009

| Submitted by:   | St. Jude Medical<br>6500 Wedgwood Road<br>Maple Grove, MN 5531 |
|-----------------|----------------------------------------------------------------|
| Contact Person: | Linh Pham                                                      |
| Telephone:      | 763-383-2586                                                   |
| Fax:            | 763-383-2559                                                   |
| Date Prepared:  | 14 August, 2009                                                |

#### Device Information 7.2

| Classification Name: | Introducer, Catheter         |
|----------------------|------------------------------|
| Common Name:         | Hemostasis Introducer        |
| Trade Name:          | Engage Introducer            |
| Classification:      | Class II per 21 CFR 870.1340 |
| Product Code:        | DYB                          |

#### 7.3 Device Description

The modified Engage Introducer is essentially a modification of the previously cleared Engage/Engage TR/Ultimum introducer (K091137). The modified Engage Introducer is intended to provide easy access to the vascular system, while providing convenient temporary closure of the access port during catheter exchanges. The Engage Introducer sheaths range in effective length from a nominal 12cm to a nominal 25cm. The Engage Introducer sheaths included in the scope of this submission range in sizes from 4-9F ACT (Active Clotting Time). Guidewire compatibility ranges from 0.035" to 0.038".

The Engage device consists of two primary components: the Hemostasis sheath assembly and the dilator. The Hemostasis sheath assembly is the vessel access device and the dilator fits inside the sheath providing support. The dilator lumen is designed to provide a 

{1}------------------------------------------------

close fit to appropriately sized guidewire. At the proximal end of the Hemostasis sheath is a snap-lock hub which is equipped with a Hemostasis valve and side port with approximately 8 cm of tubing attached and ending with a 3-way stopcock. Some of the Engage devices are packaged with FDA cleared devices such as guidewires and needles.

#### 7.4 Intended Use

There is no change to the intended use of the modified Engage Introducer as it is identical to the predicate Engage/Engage TR Introducer, K091137-April 22, 2009. The introduction of angiographic catheters, closed end catheters, balloon catheters, and electrodes into a blood vessel (including but not limited to femoral, radial, and brachial access) where minimizing blood loss is essential.

## 7.5 Predicate Device Comparison/Technological Characteristics

The modified Engage Introducer included in this Special 510(k) submission shares the same intended use as the predicate Engage/Engage TR Introducer (K091137. Avril. 22. 2009), which is indicated for the introduction of angiographic catheters, closed end catheters, balloon catheters, and electrodes into a blood vessel (including but not limited to femoral, radial, and brachial access) where minimizing blood loss is essential. The modified Engage, covered by this submission, is substantially equivalent to the St. Jude Medical Engage/Engage TR Introducer (K091137, April, 22, 2009), Ultimum Hemostasis Introducer (K001346. May 24, 2000), and Fast Cath Hemostasis Introducer (K914090, October 28, 1991).

The modifications to the Engage device do not affect the intended use of the system and there is no alteration in the fundamental scientific technology of the device. The Engage Introducer covered by this Special 510(k) submission is similar in function and technological characteristics, mechanism of action and intended use as the market cleared predicated devices, Engage/Engage TR Introducer and Ultimum Hemostasis Introducer, and Fast-Cath Introducer (K091137, K001346, & K914090).

{2}------------------------------------------------

#### 7.6 Test Summary

The Engage Introducer product family is required to pass predetermined design performance criteria. The summary of Engage test performance data is provided in this 510k submission. Based on passing verification criteria for functional, packaging, sterilization, biocompatibility, and shelf life tests, the Engage Introducer performs substantially equivalent to predicate devices. Given the scope of the modifications incorporated to create the proposed Engage Introducer, no additional animal or clinical data was deemed necessary.

## 7.7 Substantial Equivalence

The Engage Introducer covered by this submission is substantially equivalent to the previously cleared Engage/Engage TR Introducer (K091137, April, 22, 2009), Ultimum Hemostasis Introducer (K001346, May 24, 2000), and Fast Cath Hemostasis Introducer (K914090, October 28, 1991), given equivalent intended use, principles of operation and similar technological characteristics.

#### 7.8 Conclusion

In conclusion, the modified Engage Introducer is substantially equivalent to the market cleared Engage/Engage TR Introducer, Ultimum Hemostasis Introducers, and Fast Cath Hemostasis Introducer (K091137, K001346, K914090 respectively).

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services USA. The seal is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. In the center of the seal is an abstract image of an eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002

St. Jude Medical c/o Mr. Mark Job 1394 25th St. NW Buffalo, MN 55313

OCT - 8 2009

Re: K092575

Engage™ Introducer Regulation Number: 21 CFR 870.1340 Regulation Name: Catheter Introducer Regulatory Class: Class II Product Code: DYB Dated: September 19, 2009 Received: September 23, 2009

11:12

Dear Mr. Job:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. 1

{4}------------------------------------------------

Page 2 - Mr. Mark Job

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

una R. vri Mner

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

# Section 6. Indication For Use

510(k) Number: K092575 Device Name: Engage™ Introducer Indication for Use:

The introduction of angiographic catheters, closed end catheters, balloon catheters, and electrodes into a blood vessel (including but not limited to femoral, radial, and brachial access) where minimizing blood loss is essential.

Prescription Use X

(Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use -------

(21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

sima R. Lames

(Division Sign-Off) Division of Cardiovascular Devices

510(k) Number k092575

St. Jude Medical Engage Introducer 510(k)

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DYB/K092575](https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DYB/K092575)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
