VDRIVE SYSTEM/ VDRIVE DUO/ VDRIVE WITH V-CAS

K141530 · Stereotaxis, Inc. · DXX · Dec 18, 2014 · Cardiovascular

Device Facts

Record IDK141530
Device NameVDRIVE SYSTEM/ VDRIVE DUO/ VDRIVE WITH V-CAS
ApplicantStereotaxis, Inc.
Product CodeDXX · Cardiovascular
Decision DateDec 18, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1290
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Vdrive® system is intended to stabilize, navigate and remotely control: · Compatible Intracardiac Echocardiography (ICE) catheters to facilitate visualization of cardiac structure during the performance of cardiac procedure when used in conjunction with the V-Sono™ disposable sets in the Vdrive® system, · Compatible loop (circular) mapping catheters to facilitate movement of the catheter during the performance of electrophysiological procedures when used in conjunction with the V-Loop™ disposable sets in the Vdrive® system, and · Compatible fixed curve transseptal sheaths and catheters to facilitate movement of the sheath and catheter when used in conjunction with the V-CAS™ disposable sets in the Vdrive® system and with the Niobe® Magnetic Navigation System (MNS). The Vdrive® with V-Sono™ disposable is indicated for remotely controlling the advancement, retraction, rotation and anterior-posterior deflection of compatible ultrasound catheters inserted into the right atrium. Compatible catheters at this time include Biosense Webster, Inc. Soundstar™ 3D Ultrasound Catheters and AcuNav™ Ultrasound Catheters. Other models of ICE catheters have not been tested with the Vdrive™ system. The Vdrive® with V-Loop™ disposable is indicated to remotely control the advancement, retraction, tip deflection and loop size of compatible loop catheters inserted across the septum into the left atrium using conventional procedures. Compatible catheters at this time include Biosense Webster Lasso 2515 NAV Circular Mapping Catheters. Other models of loop catheters have not been tested with the Vdrive® system. The Vdrive® with V-CAS™ disposable is indicated for remotely controlling the advancement, retraction, and rotation of compatible fixed curve transseptal sheaths, and the advancement and retraction of compatible magnetic electrophysiology (EP) mapping and ablation catheters inside the patient's heart when used in conjunction with a Stereotaxis Magnetic Navigation System. Compatible fixed curve sheaths at this time include the St. Jude Medical® Transseptal Sheath and Swartz™ Braided Transseptal Sheath. Other models of transseptal sheaths and mapping/ablation catheters have not been tested with the Vdrive® system. Vdrive® with V-CAS™ is contraindicated for vascular access sites other than the groin. It is not intended to advance the EP mapping and ablation catheters through the coronary vasculature nor the coronary sinus. The transseptal sheath is not to be moved while the EP catheter is actively delivering therapy. The Vdrive Duo™ is an optional accessory intended for remotely controlling the Vdrive® system when one arm of the device is equipped with one disposable set (V-Sono™, V-Loop™ or V-CAS™) and the other arm is equipped with a different available disposable set. During the procedure, the Vdrive Duo™ allows selection between the disposable sets.

Device Story

Robotic catheter control system; stabilizes/navigates ICE catheters, loop mapping catheters, and transseptal sheaths. Hardware includes control box, adjustable arms, drive units, and user interface (tableside/dedicated controllers). V-CAS™ disposable kit (handle clamps, support tube, drape) enables remote advancement, retraction, and rotation of sheaths/catheters. Used in cardiac labs by electrophysiologists; operates in conjunction with Niobe® Magnetic Navigation System for magnetic steering. Physician uses interface to manipulate catheters/sheaths remotely, reducing manual handling. Benefits include precise navigation and stabilization during cardiac procedures. Vdrive Duo™ allows simultaneous use of two different disposable sets.

Clinical Evidence

Prospective, randomized multi-center clinical trial (VERSATILE Study, NCT01656772) compared Vdrive® with V-Loop™ to manual navigation in 120 subjects. Vdrive Duo™ used in 33 subjects (V-Loop™ + V-Sono™ n=18; V-Loop™ + V-CAS™ n=15). No device-related adverse events reported. Electrical PV isolation achieved in 96.6% (57/59) of targeted PVs using V-CAS™ controlled ablation catheters. Porcine animal study confirmed safety/effectiveness of V-CAS™ movements.

Technological Characteristics

Robotic catheter control system; electromechanical drive units; sterile single-use disposable kits (handle clamps, support tubes, drapes). Connectivity: interfaces with Niobe® Magnetic Navigation System. Software-driven user interface (tableside/dedicated controllers). Sterilization: sterile disposables.

Indications for Use

Indicated for patients undergoing diagnostic or therapeutic cardiac electrophysiology procedures or cardiac imaging. V-Sono™ used for ICE catheter control in right atrium. V-Loop™ used for circular mapping catheter control in left atrium. V-CAS™ used for fixed curve transseptal sheath and magnetic EP catheter control in heart. Contraindicated for vascular access sites other than groin; not for coronary vasculature or coronary sinus access.

Regulatory Classification

Identification

A steerable catheter control system is a device that is connected to the proximal end of a steerable guide wire that controls the motion of the steerable catheter.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 December 18, 2014 Stereotaxis. Inc. % Diana Horwitz, Ph.D., Rac Regulatory Consultant 2995 Steven Martin Dr, Fairfax, Virginia 22031 Re: K141530 Trade/Device Name: Vdrive, Vdrive with V-CAS, Vdrive Duo Regulation Number: 21 CFR 870.1290 Regulation Name: Steerable Catheter Control System Regulatory Class: Class II Product Code: DXX. DOX Dated: November 12, 2014 Received: November 12, 2014 Dear Diana Horwitz, Ph.D., Rac, We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR {1}------------------------------------------------ Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, # Melissa A. Torres -S For Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ### Indications for Use 510(k) Number (if known) K141530 Device Name Vdrive® Vdrive® with V-CASTM Vdrive Duo™ Indications for Use (Describe) The Vdrive® system is intended to stabilize, navigate and remotely control: · Compatible Intracardiac Echocardiography (ICE) catheters to facilitate visualization of cardiac structure during the performance of cardiac procedure when used in conjunction with the V-Sono™ disposable sets in the Vdrive® system, · Compatible loop (circular) mapping catheters to facilitate movement of the catheter during the performance of electrophysiological procedures when used in conjunction with the V-Loop™ disposable sets in the Vdrive® system, and · Compatible fixed curve transseptal sheaths and catheters to facilitate movement of the sheath and catheter when used in conjunction with the V-CAS™ disposable sets in the Vdrive® system and with the Niobe® Magnetic Navigation System (MNS). The Vdrive® with V-Sono™ disposable is indicated for remotely controlling the advancement, retraction, rotation and anterior-posterior deflection of compatible ultrasound catheters inserted into the right atrium. Compatible catheters at this time include Biosense Webster, Inc. Soundstar™ 3D Ultrasound Catheters and AcuNav™ Ultrasound Catheters. Other models of ICE catheters have not been tested with the Vdrive™ system. The Vdrive® with V-Loop™ disposable is indicated to remotely control the advancement, retraction, tip deflection and loop size of compatible loop catheters inserted across the septum into the left atrium using conventional procedures. Compatible catheters at this time include Biosense Webster Lasso 2515 NAV Circular Mapping Catheters. Other models of loop catheters have not been tested with the Vdrive® system. The Vdrive® with V-CAS™ disposable is indicated for remotely controlling the advancement, retraction, and rotation of compatible fixed curve transseptal sheaths, and the advancement and retraction of compatible magnetic electrophysiology (EP) mapping and ablation catheters inside the patient's heart when used in conjunction with a Stereotaxis Magnetic Navigation System. Compatible fixed curve sheaths at this time include the St. Jude Medical® Transseptal Sheath and Swartz™ Braided Transseptal Sheath. Other models of transseptal sheaths and mapping/ablation catheters have not been tested with the Vdrive® system. Vdrive® with V-CAS™ is contraindicated for vascular access sites other than the groin. It is not intended to advance the EP mapping and ablation catheters through the coronary vasculature nor the coronary sinus. The transseptal sheath is not to be moved while the EP catheter is actively delivering therapy. The Vdrive Duo™ is an optional accessory intended for remotely controlling the Vdrive® system when one arm of the device is equipped with one disposable set (V-Sono™, V-Loop™ or V-CAS™) and the other arm is equipped with a different available disposable set. During the procedure, the Vdrive Duo™ allows selection between the disposable sets. Type of Use (Select one or both, as applicable) ال Prescription Use (Part 21 CFR 801 Subpart D) _ Over-The-Counter Use (21 CFR 801 Subpart C) {3}------------------------------------------------ PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. #### FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ ## 510(k) Summary per 21CFR §807.92 | Submitter's<br>information | Stereotaxis, Inc.<br>4320 Forest Park Ave, Suite 100<br>St. Louis, MO 63108<br>Contact: John Nadelin, VP Regulatory & Quality<br>Telephone: 314-678-6130 | |-----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Device/<br>classification<br>name | Device Name: Vdrive®, Vdrive® with V-CAST™, Vdrive Duo™<br>Classification/Common name: System, Catheter Control, Steerable<br>Classification Number: 870.1290<br>Product Code: DXX, DQX<br>Classification Panel: Cardiovascular<br>Predicate Devices:<br>• Cardiodrive® Catheter Advancement System, K071029 (Stereotaxis)<br>• Vdrive® with V-Loop™, K140804 (Stereotaxis)<br>• Vdrive® with V-Sono™, K122659 (Stereotaxis)<br>• Vdrive Duo™, K133396 (Stereotaxis)<br>• Artisan S Control Catheter, K090365 (Hansen Medical) | | Device<br>description | Vdrive® with V-CAST™ is intended to control a compatible fixed curve transseptal<br>sheath and catheter during diagnostic and therapeutic cardiac procedures and is<br>comprised of four major components:<br>1. Vdrive® Hardware - control box, adjustable arm, drive unit and support<br>structure or Vdrive Duo™ (K133396) - with two adjustable arms<br>2. Vdrive® User Interface – combination of software-driven<br>(a) Tableside Controller and (b) dedicated Vdrive® Controller<br>3. V-CAST™ Disposable Kit – Handle Clamps, Catheter Support Tube and<br>Drape. These components are disposable, sterile, single use devices.<br>4. V-Loop™ Disposable Kit (K140804)<br>5. V-Sono™ Disposable Kit (K122659) | | Intended use | The Vdrive® system is intended to stabilize, navigate and remotely control:<br>• Compatible Intracardiac Echocardiography (ICE) catheters to facilitate<br>visualization of cardiac structure during the performance of cardiac<br>procedure when used in conjunction with the V-Sono™ disposable sets in<br>the Vdrive® system,<br>• Compatible loop (circular) mapping catheters to facilitate movement of the<br>catheter during the performance of electrophysiological procedures when<br>used in conjunction with the V-Loop™ disposable sets in the Vdrive® | {5}------------------------------------------------ Image /page/5/Picture/1 description: The image is a logo for Stereotaxis. The logo features an orange abstract design on the left, followed by the word "STEREOTAXIS" in a bold, sans-serif font. Below the word, there is a tagline that reads "Improving Life with Robotic Precision and Safety" in a smaller font. system, and - Compatible fixed curve transseptal sheaths and catheters to facilitate movement of the sheath and catheter when used in conjunction with the V-CASTM disposable sets in the Vdrive® system and with the Niobe® Magnetic Navigation System (MNS). The Vdrive® with V-Sono™ disposable is indicated for remotely controlling the advancement, retraction, rotation and anterior-posterior deflection of compatible ultrasound catheters inserted into the right atrium. Compatible catheters at this time include Biosense Webster. Inc. Soundstar™ 3D Ultrasound Catheters and Acuson AcuNav™ Ultrasound Catheters. Other models of ICE catheters have not been tested with the Vdrive® system. The Vdrive® with V-Loop™ disposable is indicated to remotely control the advancement, retraction, rotation, tip deflection and loop size of compatible loop catheters inserted across the septum into the left atrium using conventional procedures. Compatible catheters at this time include Biosense Webster Lasso 2515 and Lasso 2515 NAV Circular Mapping Catheters. Other models of loop catheters have not been tested with the Vdrive® system. The Vdrive® with V-CAS™ disposable is indicated for remotely controlling the advancement, retraction, and rotation of compatible fixed curve transseptal sheaths, and the advancement and retraction of compatible magnetic electrophysiology (EP) mapping and ablation catheters inside the patient's heart when used in conjunction with a Stereotaxis Magnetic Navigation System. Compatible fixed curve sheaths at this time include the St. Jude Medical® Transseptal Sheath and Swartz™ Braided Transseptal Sheath. Other models of transseptal sheaths and mapping/ablation catheters have not been tested with the Vdrive® system. Vdrive® with V-CASTM is contraindicated for vascular access sites other than the groin. It is not intended to advance the EP mapping and ablation catheters through the coronary vasculature nor the coronary sinus. The transseptal sheath is not to be moved while the EP catheter is actively delivering therapy. The Vdrive Duo™ is an optional accessory intended for remotely controlling the Vdrive® system when one arm of the device is equipped with one disposable set (V-Sono™, V-Loop™ or V-CAS™) and the other arm is equipped with a different available disposable set. During the procedure, the Vdrive Duo™ allows selection between the disposable sets. {6}------------------------------------------------ Image /page/6/Picture/1 description: The image shows the Stereotaxis logo. The logo consists of an orange abstract shape on the left, followed by the word "STEREOTAXIS" in a bold, sans-serif font. Below the word "STEREOTAXIS" is the tagline "Improving Life with Robotic Precision and Safety" in a smaller font. #### Technological Characteristics | Device<br>Character-<br>istic | Subject<br>Device<br>Vdrive® with<br>V-CASTM.<br>Vdrive Duo™M | Predicate 1<br>Cardiodrive®<br>(K071029) | Predicate 2<br>Vdrive® with<br>V-LoopTM<br>(K140804) | Predicate 3 & 4<br>Vdrive® with<br>V-Sono™<br>(K122659.<br>K133396) | Predicate 5<br>Artisan S<br>Control<br>Catheter<br>(K090365) | |---------------------------------------|------------------------------------------------------------------------|----------------------------------------------------------|---------------------------------------------------------------|---------------------------------------------------------------------|--------------------------------------------------------------| | Location of<br>Catheter tip | Right atrium<br>or left atrium | Same | Same | Right atrium | Same | | Initial<br>Placement of<br>catheter | Manual<br>placement by<br>electrophysiol<br>ogist under<br>fluoroscopy | Same | Same | Same | Same | | Type of<br>Procedure | Diagnostic or<br>therapeutic EP<br>procedures | Same | Diagnostic EP<br>procedures | Cardiac imaging | Diagnostic<br>EP<br>procedures | | Sheath<br>Movements | Remote<br>retraction and<br>advancement.<br>rotation | Does not<br>control sheath<br>(this is done<br>manually) | N/A<br>(no sheath) | N/A<br>(no sheath) | Remote<br>retraction,<br>advance-<br>ment, and<br>deflection | | Catheter<br>Movements | Advance-<br>retract<br>Magnetic<br>steering | Same | Advance/<br>retract,<br>rotation,<br>deflection,<br>loop size | Advance/ retract,<br>rotation,<br>deflection | None.<br>Catheter is<br>captured by<br>sheath | | Visualiza-tion<br>During<br>Procedure | Fluoroscopy,<br>standard of<br>care | Same | Same | Same | Same | | Catheter Tip<br>Movement | Niobe<br>(magnetic<br>navigation) | Same | Vdrive<br>manipulation | Vdrive<br>manipulation | Articulating<br>sheath<br>moves<br>catheter tip | | Single Use<br>Disposable | Sterile handle<br>clamp<br>Catheter<br>Support<br>Drape | Sterile Drive<br>unit<br>Bracket | Same | Same | Sheath<br>cassette<br>Irrigation<br>manifold<br>Drape | #### Performance data Performance data establish the substantial equivalence of the Vdrive® with V-CAS™ compared to the predicate devices. Performance data included software verification and validation data, bench performance testing, animal testing, and reference to a prospective randomized multi-center clinical trial of the Vdrive® with V-Loop™ system in navigation of circular mapping catheters compared to conventional manual methods of navigation. Performance testing was conducted for electrical safety, EMC compatibility, sterilization and shelf life and packaging. {7}------------------------------------------------ Image /page/7/Picture/1 description: The image is a logo for Stereotaxis. The logo features an orange abstract shape on the left, followed by the word "STEREOTAXIS" in a bold, sans-serif font. Below the word, there is a tagline that reads "Improving Life with Robotic Precision and Safety." Animal Testing: Stereotaxis performed an animal study in a porcine model to evaluate the safety and effectiveness of Vdrive® with V-CAS™ to perform sheath and catheter movements according to product requirements, product usability and extremes of use. This study demonstrated that Vdrive™ with V-CASTM met its performance and user requirements. Clinical Testing: Stereotaxis performed a prospective, randomized multi-center clinical trial of the Vdrive® with V-Loop™ system compared to conventional manual methods of navigation in patients who underwent a pulmonary vein (PV) isolation procedure for treatment of atrial fibrillation (The VERSATILE Study, http://clinicaltrials.gov/show/NCT01656772). Out of the 120 enrolled subjects in this trial, investigators employed Vdrive Duo™ in 33 subjects, providing safety data on use of the combinations V-Loop™ + V-Sono™ (n=18) and V-Loop™ + V-CASTM (n=15). V-Sono™ was used to control the Biosense Webster, Inc. Soundstar™ 3D Ultrasound Catheter; V-Loop™ was used to control the Lasso 2515 NAV Circular Mapping Catheter; and V-CAS™ was used to control compatible magnetic electrophysiology mapping and ablation catheters in conjunction with a Stereotaxis Magnetic Navigation System and the St. Jude Medical® Transseptal Sheath or the Swartz™ Braided Transseptal Sheath. Results of this study supported the safety and effectiveness of the Vdrive® system, including the use of Vdrive Duo™ with combinations of two disposables. Analysis of safety data showed that no adverse events related to the Vdrive® system occurred. Electrical PV isolation was achieved in 57/59 (96.6%) of targeted PVs using the ablation catheters controlled by the V-CASTM disposable in the 15 subjects who underwent the ablation procedure using Vdrive Duo™ with V-CASTM. Based upon the documentation presented in this 510(k) it has been demonstrated that the Vdrive® with V-CAS™ device is safe and effective for its intended use. Date summary prepared: December 16, 2014
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...