← Product Code [DXT](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DXT) · K151960

# Sterile High-pressure Angiographic Syringes for Single-use (K151960)

_Shenzhen Baoan Medical Supplies Co., Ltd. · DXT · Jan 7, 2016 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DXT/K151960

## Device Facts

- **Applicant:** Shenzhen Baoan Medical Supplies Co., Ltd.
- **Product Code:** [DXT](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DXT.md)
- **Decision Date:** Jan 7, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1650
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

Sterile High-pressure Angiographic Syringes for Single-use are intended for the injection of contrast media or saline; they shall be used with an US legally marketed angiographic injectors.

## Device Story

Sterile, single-use plastic syringes designed for high-pressure delivery of contrast media or saline; used in conjunction with US-marketed angiographic injectors. Device includes syringe barrel, connecting tubes (straight, Y, or T configurations), and J-shape draw tubes. Operated by clinicians in clinical settings; device attaches to injector systems to facilitate fluid delivery into patient vasculature. Output is controlled fluid injection; assists in diagnostic imaging procedures. Benefits include reliable, sterile, single-use fluid delivery for angiography.

## Clinical Evidence

No clinical data provided. Substantial equivalence is supported by bench testing, including biocompatibility (ISO 10993-4, 5, 10, 11), sterility (ISO 11737-2, ISO 10993-7), seal integrity (ASTM F88, F1929), internal pressure resistance (ASTM F1140), and mechanical performance (ISO 7886-1/2, ISO 594-1/2).

## Technological Characteristics

Plastic, single-use, high-pressure angiographic syringes. Features Luer taper fittings (ISO 594-1/2). Sterilized via Ethylene Oxide (EO). Compatible with various power-driven angiographic injectors. Materials meet ISO 10993 biocompatibility requirements. Includes connecting tubes and J-shape draw tubes.

## Special Controls

*Classification.* Class II (special controls). The device, when it is a non-patient contacting balloon inflation syringe intended only to inflate/deflate balloon catheters and monitor pressure within the balloon, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.

## Predicate Devices

- ANT Angiographic Syringes ([K072696](/device/K072696.md))

## Reference Devices

- MCT & MCT plus CT ([K924116](/device/K924116.md))
- Visitron CT ([K991557](/device/K991557.md))
- En Vision CT ([K993782](/device/K993782.md))
- Stellant CT ([K023183](/device/K023183.md))
- Solaris ([K033247](/device/K033247.md))
- Mark V ([K903390](/device/K903390.md))
- Mark III & Mark IV ([K903390](/device/K903390.md))
- CT 9000 & CT9000 ADV ([K912944](/device/K912944.md))
- Optistar LE, Elite ([K073592](/device/K073592.md))
- ILLUMENA ([K963071](/device/K963071.md))
- Empower CT ([K073178](/device/K073178.md))
- CT ([K052633](/device/K052633.md))
- Sonic Shot ([K091734](/device/K091734.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract image of an eagle or bird-like figure, composed of three stylized human profiles facing to the right.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

January 7, 2016

Shenzhen Baoan Medical Supplies Co., Ltd % Diana Hong General Manager Mid-link Consulting Co., Ltd P.O. Box 120-119 Shanghai, 200120 China

Re: K151960

Trade/Device Name: Sterile High-pressure Angiographic Syringes For Single-use Regulation Number: 21 CFR 870.1650 Regulation Name: Angiographic Injector And Syringe Regulatory Class: Class II Product Code: DXT Dated: November 26, 2015 Received: December 1, 2015

Dear Diana Hong:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Kenneth J. Cavanaugh -S

for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K151960

### Device Name

Sterile High-pressure Angiographic Syringes for Single-use

Indications for Use (Describe)

Sterile High-pressure Angiographic Syringes for Single-use are intended for the injection of contrast media or saline; they shall be used with an US legally marketed angiographic injectors.

Type of Use (Select one or both, as applicable)

| <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) |
|--------------------------------------------------------------------------------------|
| <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)                 |

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# Exhibit #2 510(k) Summary

This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92.

The assigned 510(k) Number: K151960

- 1. Date of Preparation: 11/26/2015
- 2. Sponsor Identification

Shenzhen BaoAn Medical Supplies Co., Ltd Industrial Zone, Shenkeng, Liuyue Community, Henggang Street, Longgang District, Shenzhen, Guangdong, 518173, China

Establishment Registration Number: Not yet registered

Contact Person: Mr. Mingan Mu Position: Deputy General Manager Tel: +86-755-28638515 Fax: +86-755-28638033 Email: boon@szboon.com

- 3. Designated Submission Correspondent
Ms. Diana Hong (Primary Contact Person) Mr. Lee Fu (Alternative Contact Person)

## Mid-Link Consulting Co., Ltd P.O. Box 120-119, Shanghai, 200120, China

Tel: +86-21-22815850, Fax: 240-238-7587 Email: info@mid-link.net

{4}------------------------------------------------

#### 4. Identification of Proposed Device

Trade Name: Sterile High-pressure Angiographic Syringes for Single-use Common Name: Disposable angiographic syringe Models: 100101, 100103, 100104, 100113, 100114, 100121, 100124, 200101, 200102, 200103, 200104, 300101, 300103 and 300105

Regulatory Information

Classification Name: Angiographic injector and syringe Classification: II Product Code: DXT Regulation Number: CFR 870.1650 Review Panel: Cardiovascular

Indication for Use:

Sterile High-pressure Angiographic Syringes for Single-use are intended for the injection of contrast media or saline; they shall be used with an US legally marketed angiographic injectors.

Device Description

The proposed devices are available in packs, which may include different configurations of syringes and accessories. The syringes are plastic, disposable syringes, which are available in various models and configurations. They are intended to be used with an U.S. legally marketed angiography injector, compatibilities are shown in Table 1.

| Model (Syringe)                                                                        | Volume (ml) | Type        | Compatible Angiographic Injector                                            |
|----------------------------------------------------------------------------------------|-------------|-------------|-----------------------------------------------------------------------------|
| The Sterile High-pressure Angiographic Syringes for Single-use models listed below are |             |             |                                                                             |
| compatible with the following Medrad Injector models                                   |             |             |                                                                             |
| 100101                                                                                 | 200ml       | Single Shot | MCT & MCT plus CT, K924116<br>Visitron CT, K991557<br>En Vision CT, K993782 |
| 100103                                                                                 | 200ml       | Single Shot | Stellant CT, K023183                                                        |
| 100113                                                                                 | 200/200ml   | Dual Shots  | Stellant CT, K023183                                                        |
| 200101                                                                                 | 65/65ml     | Dual Shots  | Solaris, K033247                                                            |
| 200102                                                                                 | 65/115ml    | Dual Shots  | Solaris, K033247                                                            |
| 300101                                                                                 | 150ml       | Single Shot | Mark V, K903390                                                             |
| 300105                                                                                 | 130ml       | Single Shot | Mark III & Mark IV, K903390                                                 |
| The Sterile High-pressure Angiographic Syringes for Single-use models listed below are |             |             |                                                                             |
| compatible with the following LIEBEL – FLARSHEIM Injector models                       |             |             |                                                                             |
| 100104                                                                                 | 200ml       | Single Shot | CT 9000 & CT9000 ADV, K912944                                               |
| 100114                                                                                 | 200/200ml   | Dual Shots  | CT 9000 & CT9000 ADV, K912944                                               |

Table 1 Compatibility between Syringes and Injectors

{5}------------------------------------------------

| 200103                                                                                 | 60/60ml   | Dual Shots  | Optistar LE, Elite, K073592 |
|----------------------------------------------------------------------------------------|-----------|-------------|-----------------------------|
| 300103                                                                                 | 150ml     | Single Shot | ILLUMENA, K963071           |
| The Sterile High-pressure Angiographic Syringes for Single-use models listed below are |           |             |                             |
| compatible with the following E-Z-EM Injector models                                   |           |             |                             |
| 100111                                                                                 | 200ml     | Single Shot | Empower CT, K073178         |
| 100121                                                                                 | 200/200ml | Dual Shots  | Empower CT, K073178         |
| The Sterile High-pressure Angiographic Syringes for Single-use models listed below are |           |             |                             |
| compatible with the following Nemoto Injector models                                   |           |             |                             |
| 100124                                                                                 | 60/100ml  | Dual Shots  | CT, K052633                 |
| 200104                                                                                 | 60/60ml   | Dual Shots  | Sonic Shot, K091734         |

The connecting tubes, which is used to connect the syringe and the catheter. The tubes are also available in three configurations, which are straight tube (used with single shot syringe), type Y or type T tube (used with dual shot syringe) and high pressure tube, used with syringes which can withstanding maximum pressure of 1200 psi.

J shape tube, which is used to draw contrast media/saline into the syringe barrel.

- ડ. Identification of Predicate Device
510(k) Number: K072696 Product Name: ANT Angiographic Syringes Manufacturer: Shenzhen Ant Hi-Tech Industrial Co., Ltd.

- 6. Non-Clinical Test Conclusion
Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:

- A ISO 10993-7:2008 Biological evaluation of medical devices- Part 7: Ethylene oxide sterilization residuals:
- > ASTM F 88/F88M-09 Standard test method for seal strength of flexible barrier materials;
- > ASTM F1140/F1140M-13 Standard test methods for internal pressurization failure resistance of unrestrained packages:
- > ASTM F1929-12 Standard Test Method for Detecting Seal Leaks in Porous Medical Package by Dye Penetration;
- > USP37-NF32 <85> Bacterial Endotoxins Limit;
- > ISO 11737-2:2009 Sterilization of medical devices- Microbiological methods- Part 2: Test of sterility performed in the definition, validation and maintenance of a sterilization process;
- > ISO 7886-1:1993 Sterile hypodermic syringes for single use - Part 1: Svringes for manual use;

{6}------------------------------------------------

- > ISO 7886-2:1996 Sterile hypodermic syringes for single use - Part 2: Syringes for use with power-driven Syringe Pumps;
- > ISO594:1986 Conical fittings with a 6% (Luer) taper for syringes, needles and certain other medical equipment - Part 1: General Requirements;
- > ISO594:1998 Conical fittings with a 6% (Luer) taper for syringes, needles and certain other medical equipment - Part 2: Lock Fitting;
- > ISO 10993-4:2002 AMD 2006 Biological evaluation of medical devices -- Part 4: Selection of tests for interactions with blood:
- > ISO 10993-5:2009 Biological evaluation of medical devices -- Part 5: Tests for In Vitro cytotoxicity;
- > ISO 10993-10:2010 Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization;
- > ISO 10993-11:2006 Biological evaluation of medical devices -- Part 11: Tests for systemic toxicity;
- > ASTM F 756-08, Standard practice for assessment of hemolytic properties of material
- > USP37-NF32 <151> Pyrogen test
- 7. Clinical Test Conclusion

No clinical study is included in this submission.

- Substantially Equivalent (SE) Comparison 8.

| Item              | Proposed Device                                                                                                                                                                                           | Predicate Device                                                                                                                                                                  |
|-------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Code      | DXT                                                                                                                                                                                                       | DXT                                                                                                                                                                               |
| Regulation Number | CFR 870.1650                                                                                                                                                                                              | CFR 870.1650                                                                                                                                                                      |
| Intended Use      | Sterile High-pressure Angiographic<br>Syringes for Single-use are intended for<br>the injection of contrast media or saline;<br>they shall be used with an US legally<br>marketed angiographic injectors. | ANT Angiographic Syringes are<br>syringes for the injection of contrast<br>media or saline. This syringe is for<br>single use with US legally marketed<br>angiographic injectors. |
| Mode of operation | Power-driven operation, single use                                                                                                                                                                        | Power-driven operation, single use                                                                                                                                                |
| Configuration     | Angiographic Syringe<br>Connecting tube<br>J shape tube/Spike                                                                                                                                             | Angiographic Syringe<br>Connecting tube<br>J shape tube/Spike                                                                                                                     |
| Sterility         | EO Sterilized                                                                                                                                                                                             | EO Sterilized                                                                                                                                                                     |
| Single Use        | Yes                                                                                                                                                                                                       | Yes                                                                                                                                                                               |
| Performance       | Comply with:<br>ISO 7886-1                                                                                                                                                                                | Comply with:<br>ISO 7886-1                                                                                                                                                        |

## Table 2 Comparison of Technology Characteristics

{7}------------------------------------------------

|                  | ISO 594-1                           | ISO 594-1                           |
|------------------|-------------------------------------|-------------------------------------|
|                  | ISO 594-2                           | ISO 594-2                           |
| Biocompatibility | Conforms to the requirements of ISO | Conforms to the requirements of ISO |
|                  | 10993 series Standards              | 10993 series Standards              |

{8}------------------------------------------------

#### Substantially Equivalent (SE) Conclusion 9.

Based on the comparison, the intended use, mode of operation, and sterility of proposed device is determined to be Substantially Equivalent (SE) to the predicate device.

In addition, the results of performance tests performed on the proposed device can also demonstrate the proposed devices are complied with FDA recognized standards, which the predicate device was also complied with. The results of biocompatibility studies performed on the proposed device demonstrate that the patient materials used in proposed device are biocompatible.

Based on the comparison above, the proposed devices are determined to be Substantially Equivalent (SE) to the predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DXT/K151960](https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DXT/K151960)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
