Disposable NIBP Cuff

K193629 · Shenzhen Changke Connect Electronics Co., Ltd. · DXQ · Mar 25, 2020 · Cardiovascular

Device Facts

Record IDK193629
Device NameDisposable NIBP Cuff
ApplicantShenzhen Changke Connect Electronics Co., Ltd.
Product CodeDXQ · Cardiovascular
Decision DateMar 25, 2020
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1120
Device ClassClass 2
AttributesPediatric

Indications for Use

The disposable blood pressure cuff is an accessory used in conjunction with noninvasive blood pressure measurement systems. The cuff is non-sterile and for single-patient use. It is available in neonate, infant, child and adult sizes. The cuff is not designed, sold, or intended for use except as indicated.

Device Story

Disposable NIBP cuff; accessory for noninvasive blood pressure measurement systems. Device consists of cuff and single air tube; cuff wraps around patient limb; secured by hook and loop closure. Air hose connects to blood pressure monitor. Used in clinical settings; operated by healthcare professionals. Provides pneumatic interface for blood pressure measurement; enables monitoring of patient blood pressure. Benefits include single-patient use to reduce cross-contamination risk.

Clinical Evidence

Bench testing only. Biocompatibility testing (cytotoxicity, sensitization, irritation) performed per ISO 10993-1. Performance testing conducted per ISO 81060-1 (Non-Invasive Sphygmomanometers) to verify design specifications and leakage requirements (< 4 mmHg/min).

Technological Characteristics

Materials: non-woven fabrics (patient-contacting), PVC, ABS. Principle: pneumatic bladder with hook and loop closure. Connectivity: single tube interface to NIBP monitor. Form factor: 6 models covering neonate to adult thigh sizes. Sterilization: non-sterile. Standards: ISO 81060-1, ISO 10993-1.

Indications for Use

Indicated for use as an accessory with noninvasive blood pressure measurement systems for neonate, infant, child, and adult patient populations.

Regulatory Classification

Identification

A blood pressure cuff is a device that has an inflatable bladder in an inelastic sleeve (cuff) with a mechanism for inflating and deflating the bladder. The cuff is used in conjunction with another device to determine a subject's blood pressure.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath. The logo is simple and professional, and it is easily recognizable. March 25, 2019 Shenzhen Changke Connect Electronics Co., Ltd. % Kevin Wang, Consultant Chonconn Medical Device Consulting Co., Ltd. No. A415, Block A. NanShan Medical devices Industrial Park Nanshan District, Shenzhen, 518067 CHINA Re: K193629 Trade/Device Name: Disposable NIBP Cuff Regulation Number: 21 CFR 870.1120 Regulation Name: Blood Pressure Cuff Regulatory Class: Class II Product Code: DXO Dated: December 24, 2019 Received: December 26, 2019 Dear Kevin Wang: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {1}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, LT Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration ## Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below. 510(k) Number (if known) K193629 Device Name Disposable NIBP Cuff Indications for Use (Describe) The disposable blood pressure cuff is an accessory used in conjunction with noninvasive blood pressure measurement systems. The cuff is non-sterile and for single-patient use. It is available in neonate, infant, child and adult sizes. The cuff is not designed, sold, or intended for use except as indicated. | Type of Use (Select one or both, as applicable) | <span> <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> | |-------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| |-------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*** The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov *"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."* | FORM FDA 3881 (7/17) | Page 1 of 1 | |----------------------|-------------| |----------------------|-------------| FORM FDA 3881 (7/17) {3}------------------------------------------------ # 510(K) Summary ### Prepared in accordance with the requirements of 21 CFR Part 807.92 ### Prepared Date: 2019/12/24 #### Submission sponsor 1. Name: Shenzhen Changke Connect Electronics Co., Ltd. Address: A2-4th floor of Xiang dali Technology Park, No.87 of HengPing Road, Henggang, Longgang District, Shenzhen, P.R. China Contact person: Yahui Zhou Title: General manager E-mail: zhouyahui@szcklt.com Tel: +86 136 1301 2560 #### 2. Submission correspondent Name: Chonconn Medical Device Consulting Co., Ltd. Address: No. A415, Block A, NanShan Medical devices Industrial Park Nanshan District, Shenzhen, Guangdong, P.R. China 518067 Contact person: Kevin Wang E-mail: kevin(@chonconn.com Tel: +86-755 33941160 | Trade/Device Name | Disposable NIBP Cuff | |-------------------|----------------------------------------------------------------------------------------------------------| | Model | CK-XT-88062-001, CK-XT-88062-003, CK-XT-88062-005, CK-XT-88062-<br>007, CK-XT-88062-008, CK-XT-88062-010 | | Common Name | Non-invasive Blood pressure cuff | | Regulatory Class | Class II | | Classification | 21CFR 870.1120 / Blood pressure cuff / DXQ | | Submission type | Traditional 510(K) | #### 3. Subject Device Information #### 4. Predicate Device Manufacturer: Shenzhen Caremed Medical Technology Co., Ltd. Device: Caremed Disposable Blood Pressure Cuff 510(k) No.: K182433. #### Device Description 5. The subject device is an accessory used in conjunction with noninvasive blood pressure measurement systems. It is available in neonatal, infant, child and adult sizes. The subject device includes disposable blood pressure cuff. All the disposable blood pressure cuff has same structure, which contains Cuff and single tube. {4}------------------------------------------------ The disposable blood pressure cuff is single use device, and which is made of non-woven fabrics (Cuff) and PVC (Air tube), thereinto, the non-woven fabrics (Cuff) is the material used to contacting with the patient. The disposable blood pressure cuff has 6 models with different population with different circumference size. #### 6. Intended use & Indication for use The disposable blood pressure cuff is an accessory used in conjunction with noninvasive blood pressure measurement systems. The cuff is non-sterile and for single-patient use. It is available in neonate, infant, child and adult sizes. The cuff is not designed, sold, or intended for use except as indicated. | Features | Subject Device<br>Disposable NIBP Cuff | Predicate Device K182433<br>Disposable Blood Pressure Cuff | Remark | |------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------| | Applicant | Shenzhen Changke Connect<br>Electronics Co., Ltd. | Shenzhen Caremed Medical<br>Technology Co., Ltd. | / | | Classification<br>Regulation | 21CRF 870.1120 | 21CRF 870.1120 | Same | | Classification<br>and Code | Class II,<br>DXQ | Class II,<br>DXQ | Same | | Common<br>name | Non-invasive Blood Pressure<br>Cuff | Non-invasive Blood Pressure<br>Cuff | Same | | Intended use | The disposable blood pressure cuff<br>is an accessory used in conjunction<br>with noninvasive blood pressure<br>measurement systems. The cuff is<br>non-sterile and for single-patient<br>use. It is available in neonate,<br>infant, child and adult sizes. The<br>cuff is not designed, sold, or<br>intended for use except as<br>indicated. | The disposable blood pressure cuff<br>is an accessory used in conjunction<br>with noninvasive blood pressure<br>measurement systems. The cuff is<br>non-sterile and for single-patient<br>use. It is available in neonate,<br>infant, child and adult sizes. The<br>cuff is not designed, sold, or<br>intended for use except as<br>indicated. | Same | | Patient<br>Populations | Adults/Pediatrics | Adults/Pediatrics | Same | | Tube Number | One | One | Same | | Principles of<br>Operation | Bladder is wrapped around the<br>patient's limb and secured by hook<br>and loop closure Air hose is<br>connected to the noninvasive blood<br>pressure measurement systems | Bladder is wrapped around the<br>patient's limb and secured by hook<br>and loop closure Air hose is<br>connected to the noninvasive blood<br>pressure measurement systems | Same | | Limb | Conform to AHA bladder sizes | Conform to AHA bladder sizes | Different | #### Comparison to the Predicate Device 7. {5}------------------------------------------------ | Circumference | recommendations | recommendations | (1) | |-----------------|---------------------------------|---------------------------------|-----------| | (Range in cm) | Neonatal (7-13 cm) | Neonatal 1 (3-6 cm) | | | | Infant (9-14.8 cm) | Neonatal 2 (4-8 cm) | | | | Child (13.8-21.5 cm) | Neonatal 3 (6-11 cm) | | | | Adult (27.5-36.5 cm) | Neonatal 4 (7-13 cm) | | | | Large Adult (35.5-46 cm) | Neonatal 5 (8-15 cm) | | | | Adult Thigh (45-56.5 cm) | Infant (9-14.8 cm) | | | | | Child (13.8-21.5 cm) | | | | | Small Adult (20.5-28.5 cm) | | | | | Adult (27.5-36.5 cm) | | | | | Adult Long (27.5-36.5/46.5 cm) | | | | | Large Adult (35.5-46 cm) | | | | | Large Adult Long (35.5-46 cm) | | | Pressure Range | 0-300 mmHg | 0-300 mmHg | Same | | Sterility | Non-sterile | Non-sterile | Same | | Max. Leakage | < 4mm Hg/ min. | < 4mm Hg/ min. | Same | | Material | Non-woven (Patient contacted); | Non-woven (Patient contacted); | Different | | | PVC; | nylon; | (2) | | | ABS. | PVC. | | | Biocompatibilit | No potential cytotoxicity; | No potential cytotoxicity; | Same | | y | No sensitization observed; | No sensitization observed; | | | | Negligible (no observed primary | Negligible (no observed primary | | | | irritation) | irritation) | | ### Justification of differences: Justifications for differences between subject device and the predicate device are shown as below: Different (1): The difference in the limb circumference. Performance testing accordance with ISO 81060-1 has been conducted and provided with this submission. Therefore, this difference does not raise any safety or effectiveness issue. Different (2): The difference in the material. Biocompatibility testing accordance with ISO 10993-1 has been conducted and provided with this submission. Therefore, this difference does not raise any safety or effectiveness issue. #### Performance Data 8. The following performance data were provided in support of the substantial equivalence determination. ### Biocompatibility testing The biocompatibility evaluation for the subject device was conducted in accordance with the FDA Biocompatibility guidance, 2016 (Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process") and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing {6}------------------------------------------------ Within a Risk Management Process," as recognized by FDA. The battery of testing included the following tests: - Cytotoxicity ● - . Sensitization - Irritation . The subject devices are considered surface contacting for a duration of exceed 24 hours but not 30 days. ### Non-clinical data Non-clinical tests were conducted to verify that the subject device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the subject device complies with the following standards: - . ISO 81060-1, Non-Invasive Sphygmomanometers - Part 1: Requirements and Test Methods for Non-Automated Measurement Type, First Edition 2007. The test was selected to show substantial equivalence between the subject device and the predicate. #### 9. Conclusion Based on the indications for use, technological characteristics, and comparison to predicate devices, the subject Disposable NIBP Cuff has been shown to be substantially equivalent to legally marketed predicate device.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...