← Product Code [DXQ](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DXQ) · K173197

# Reusable NIBP Cuff (K173197)

_Orantech, Inc. · DXQ · Aug 21, 2018 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DXQ/K173197

## Device Facts

- **Applicant:** Orantech, Inc.
- **Product Code:** [DXQ](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DXQ.md)
- **Decision Date:** Aug 21, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1120
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Pediatric

## Indications for Use

The reusable blood pressure cuff is an accessory used in conjunction with noninvasive blood pressure measurement systems. The cuff is non-sterile and may be reused. It is available in neonate, infant, child and adult sizes.

## Device Story

Reusable NIBP cuff; inelastic sleeve with integrated inflatable bladder; secures to patient limb via hook and loop closure; connects via tubing to non-invasive blood pressure measurement system. Used in clinical settings; operated by healthcare professionals. Device functions as pneumatic interface for blood pressure monitoring systems; cuff inflation/deflation cycles facilitate pressure measurement. Benefits include reliable, reusable blood pressure monitoring across diverse patient sizes.

## Clinical Evidence

No clinical data. Bench testing only. Performance verified via ISO 81060-1 (non-invasive sphygmomanometer requirements) and biocompatibility testing per ISO 10993-5 (cytotoxicity) and ISO 10993-10 (irritation and skin sensitization).

## Technological Characteristics

Materials: PU synthetic leather cuff, transparent TPU film bladder, PVC tubing, molded nylon hook/loop fasteners. Pressure limit: 0-300 mmHg. Durability: 10,000 inflation cycles, 3,000 hook/loop closures. Non-sterile. Complies with ISO 81060-1, ISO 10993-5, and ISO 10993-10.

## Regulatory Identification

A blood pressure cuff is a device that has an inflatable bladder in an inelastic sleeve (cuff) with a mechanism for inflating and deflating the bladder. The cuff is used in conjunction with another device to determine a subject's blood pressure.

## Predicate Devices

- Unimed Medical Supplies Inc. Blood Pressure Cuff ([K112544](/device/K112544.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

August 21, 2018

Orantech Inc. % Kevin Wang Consultant Chonconn Medical Device Consulting Co., Ltd. 22A, Haijing Square, No. 18, Taizi Road Nanshan District, Shenzhen, P.R. China, 518067

## Re: K173197

Trade/Device Name: Reusable NIBP Cuff Regulation Number: 21 CFR 870.1120 Regulation Name: Blood Pressure Cuff Regulatory Class: Class II Product Code: DXO Dated: July 12, 2018 Received: July 18, 2018

## Dear Kevin Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Stephen C. Browning -S5

for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K173197

Device Name Reusable NIBP Cuff

Indications for Use (Describe)

The reusable blood pressure cuff is an accessory used in conjunction with noninvasive blood pressure measurement systems. The cuff is non-sterile and may be reused. It is available in neonate, infant, child and adult sizes.

| Type of Use (Select one or both, as applicable)                                                                                                                                                                                            |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span> <span style="padding-right: 5px;"> </span>    Prescription Use (Part 21 CFR 801 Subpart D)   </span> <span> <span style="padding-left: 20px; padding-right: 5px;"> </span>    Over-The-Counter Use (21 CFR 801 Subpart C)   </span> |

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

 Department of Health and Human Services  

 Food and Drug Administration  

 Office of Chief Information Officer  

 Paperwork Reduction Act (PRA) Staff  

 PRAStaff@fda.hhs.gov

*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."*

{3}------------------------------------------------

## 510(K) Summary

Prepared in accordance with the requirements of 21 CFR Part 807.92

Prepared Date: 2017/09/15

### Submission sponsor 1.

Name: Orantech Inc.

Address: Zone#A,4F,1st Bld,7th Industrial Zone, Yulv Community, GongMing, Guangming New District, Shenzhen, P.R. China 518106 Contact person: Hsin Xiong Title: General manager E-mail: Yunxi.xiong@orantech.cn Tel: +86-755 23243594

### Submission correspondent 2.

Name: Chonconn Medical Device Consulting Co., Ltd. Address: 22A, Haijing Square, No. 18, Taizi Road, Nanshan District, Shenzhen, P.R. China 518067 Contact person: Kevin Wang E-mail: kevin@chonconn.com Tel: +86-755 33941160

| Trade/Device Name | Reusable NIBP Cuff                                                                 |
|-------------------|------------------------------------------------------------------------------------|
| Model             | BP-10BS, BP-20BS, BP-30BS, BP-40BS, BP-50BS, BP-60BS,<br>BP-70BS, BP-80BS, BP-90BS |
| Common Name       | Non-invasive Blood pressure cuff                                                   |
| Regulatory Class  | Class II                                                                           |
| Classification    | 21CFR 870.1120 / Blood pressure cuff / DXQ                                         |
| Submission type   | Traditional 510(K)                                                                 |

### Subject Device Information 3.

### 4. Predicate Device

By submission of the Traditional 510(k), Orantech Inc. is requesting clearance for Reusable NIBP Cuff. It is comparable to the following legally marketed system:

- 1. Unimed Medical Supplies Inc. Blood Pressure Cuff under K112544.
The subject device has same intended use, same target patient population, same performance effectiveness, performance safety as the predicate devices and no question is raised regarding to effectiveness and safety. So, the conclusion is that the subject device is substantial equivalent to the predicate.

### Device Description 5.

The device comprises tubing attached to an inelastic sleeve with an integrated inflatable bladder that is

{4}------------------------------------------------

# Orantech

wrapped around the patient's limb and secured by hook and loop closure. The device tubing is connected to a non-invasive blood pressure measurement system. The reusable NIBP cuff has 9 models with different size for different population with different arm size.

### Intended use & Indication for use 6.

The reusable blood pressure cuff is an accessory used in conjunction with noninvasive blood pressure measurement systems. The cuff is non-sterile and may be reused. It is available in neonate, infant, child and adult sizes.

| Features     | Subject Device                                                                                                                                                                                                                                 | Predicate Device K112544                                                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                             |                                                                                                                                                                                                                                            |
|--------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|              | Orantech Reusable NIBP Cuff                                                                                                                                                                                                                    | Unimed Blood Pressure Cuff                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                             |                                                                                                                                                                                                                                            |
| Intended use | The reusable blood pressure cuff is an<br>accessory used in conjunction with<br>noninvasive blood pressure measurement<br>systems. The cuff is non-sterile and may be<br>reused. It is available in neonate, infant,<br>child and adult sizes. | The Unimed Blood Pressure Cuff is<br>an accessory used in conjunction<br>with noninvasive blood pressure<br>measurement systems. The cuff is<br>non-sterile and may be reused. It is<br>available in pediatric and adult<br>sizes. The cuff is not designed, sold,<br>or intended for use except as<br>indicated. |                                                                                                                                                                                                                                             |                                                                                                                                                                                                                                            |
|              | Patient<br>Populations                                                                                                                                                                                                                         | Adults/Pediatrics/Infants/Neonates                                                                                                                                                                                                                                                                                | Adults/Pediatrics/Infants/Neonates                                                                                                                                                                                                          |                                                                                                                                                                                                                                            |
|              |                                                                                                                                                                                                                                                | Material                                                                                                                                                                                                                                                                                                          | Cuff: PU Synthetic Leather<br>Bladder: Transparent Polyurethane (TPU<br>Film)<br>Tubing: PVC<br>Hook: Molded Nylon<br>Loop: Nylon                                                                                                           | Cuff: PU Synthetic Leather<br>Bladder: Transparent Polyurethane<br>(TPU Film)<br>Tubing: PVC<br>Hook: Molded Nylon<br>Loop: Nylon                                                                                                          |
|              | Tube Number                                                                                                                                                                                                                                    |                                                                                                                                                                                                                                                                                                                   | One                                                                                                                                                                                                                                         | One or two                                                                                                                                                                                                                                 |
|              | Limb<br>Circumference<br>(Range in cm)                                                                                                                                                                                                         |                                                                                                                                                                                                                                                                                                                   | Conform to AHA bladder sizes<br>recommendations<br>Neonatal (6-11cm)<br>Infant (8-13cm)<br>Child (12-19cm)<br>Small Adult (17-25cm)<br>Adult (23-33cm)<br>Adult Extra Long (23-33cm)<br>Large Adult (31-40cm)<br>Large Adult Long (31-40cm) | Conform to AHA bladder sizes<br>recommendations<br>Neonatal (6-11cm)<br>Infant (10-19cm)<br>Pediatric (18-26cm)<br>Small Adult (20-28cm)<br>Adult (25-35cm)<br>Adult Long (25-35cm)<br>Large Adult (33-47cm)<br>Large Adult Long (33-47cm) |

### Comparison to the Predicate Device 7.

{5}------------------------------------------------

|                       | Adult Thigh (38-50cm)                             | Adult Thigh (46-66cm)                             |
|-----------------------|---------------------------------------------------|---------------------------------------------------|
| Repeated<br>inflation | 10,000 inflations<br>3,000 hook and loop closures | 10,000 inflations<br>3,000 hook and loop closures |
| Pressure limits       | 0-300mmHg                                         | 0-300mmHg                                         |
| Sterility             | Non-sterile                                       | Non-sterile                                       |
| Biocompatibility      | Comply with ISO 10993 biocompatibility evaluation | Comply with ISO 10993 biocompatibility evaluation |

### Performance Data 8.

## Clinical test:

Clinical testing is not required.

## Non-clinical data

The Orantech Reusable NIBP Cuff has been tested according to the following standards:

(1) ISO 81060-1, Non-Invasive Sphygmomanometers - Part 1: Requirements and Test Methods for Non-Automated Measurement Type, First Edition 2007.

(2) ISO 10993-5, Biological evaluation of medical devices -- Part 5: Tests for In Vitro cytotoxicity, 2009 (3) ISO 10993-10, Biological evaluation of medical devices -- Part 10: Tests for irritation and skin sensitization, 2010

The tests were selected to show substantial equivalence between the subject device and the predicate.

### 9. Conclusion

Based on the safety and performance testing and compliance with acceptable voluntary standards, we believe that subject device is substantially equivalent to the predicate device and don't raise any new safety and/or effectiveness issues.

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DXQ/K173197](https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DXQ/K173197)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
