← Product Code [DXQ](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DXQ) · K171889

# Reusable Blood Pressure Cuff (K171889)

_Suzhou Minhua Medical Apparatus Supplies Co., Ltd. · DXQ · Dec 19, 2017 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DXQ/K171889

## Device Facts

- **Applicant:** Suzhou Minhua Medical Apparatus Supplies Co., Ltd.
- **Product Code:** [DXQ](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DXQ.md)
- **Decision Date:** Dec 19, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1120
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Pediatric

## Indications for Use

The Reusable Blood Pressure Cuff is an accessory used in conjunction with noninvasive blood pressure measurement systems. The cuff is non-sterile and reusable. It is available in infant, pediatric but not neonatal, and adult sizes.

## Device Story

Reusable blood pressure cuff; wraps around patient limb; secured by hook and loop closure. Air hose connects to noninvasive blood pressure measurement system. Device functions as pneumatic interface for blood pressure monitoring. Used in clinical settings; operated by healthcare professionals. Output is pressure transmission to monitor; enables blood pressure measurement. Benefits include non-invasive, reusable patient monitoring.

## Clinical Evidence

No clinical data; bench testing only. Device verified against IEC 80601-2-30:2013 for safety and performance, and ISO 10993-5 and ISO 10993-10 for biocompatibility.

## Technological Characteristics

Reusable cuff with internal or integrated bladder; materials include polyester or TPU leather with nylon. Available in one or two-tube configurations. Pressure range 0-300mmHg. Biocompatibility per ISO 10993-5 and ISO 10993-10. Mechanical device; no software or electronics.

## Regulatory Identification

A blood pressure cuff is a device that has an inflatable bladder in an inelastic sleeve (cuff) with a mechanism for inflating and deflating the bladder. The cuff is used in conjunction with another device to determine a subject's blood pressure.

## Predicate Devices

- Andon Blood Pressure Cuff ([K133117](/device/K133117.md))
- Reusable Blood Pressure Cuff ([K152468](/device/K152468.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

December 19, 2017

Suzhou Minhua Medical Apparatus Supplies Co., LTD % Diana Hong General Manager Mid-Link Consulting Co., Ltd P.O. Box 120-119 Shanghai, 200120 CN

Re: K171889

Trade/Device Name: Reusable Blood Pressure Cuff, with list of 164 models attached Regulation Number: 21 CFR 870.1120 Regulation Name: Blood Pressure Cuff Regulatory Class: Class II Product Code: DXO Dated: November 10, 2017 Received: November 15, 2017

Dear Diana Hong:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);

{1}------------------------------------------------

## Page 2 - Diana Hong

and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

M.A. Wilhelm
for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## List of 164 models

| No. | Product code | Name                | Tube | Arm size |
|-----|--------------|---------------------|------|----------|
| 1   | MH002RCP10-1 | Small adult         | One  | 17-22cm  |
| 2   | MH002RCP10-2 | Small adult         | two  | 17-22cm  |
| 3   | MH002RCP11-1 | Small adult<br>long | One  | 22-30 cm |
| 4   | MH002RCP11-2 | Small adult<br>long | two  | 22-30 cm |
| 5   | MH002RCP12-1 | Adult               | One  | 22-30 cm |
| 6   | MH002RCP12-2 | Adult               | two  | 22-30 cm |
| 7   | MH002RCP14-1 | Large Adult         | One  | 30-42 cm |
| 8   | MH002RCP14-2 | Large Adult         | two  | 30-42 cm |
| 9   | MH002RCP15-1 | Large Adult<br>long | One  | 42-48 cm |
| 10  | MH002RCP15-2 | Large Adult<br>long | two  | 42-48 cm |

Table 1 Specifications of Reusable Blood Pressure Cuff – with Internal Bladder

| No. | Product code  | Name                | Tube | Arm size    |
|-----|---------------|---------------------|------|-------------|
| 1   | MH002RCT06HP1 | Infant              | One  | 9-14.8cm    |
| 2   | MH002RCT06HP2 | Infant              | two  | 9-14.8cm    |
| 3   | MH002RCT07HP1 | Infant long         | One  | 9-14.8cm    |
| 4   | MH002RCT07HP2 | Infant long         | two  | 9-14.8cm    |
| 5   | MH002RCT08HP1 | Child               | One  | 13.8-21.5cm |
| 6   | MH002RCT08HP2 | Child               | two  | 13.8-21.5cm |
| 7   | MH002RCT09HP1 | Child long          | One  | 13.8-21.5cm |
| 8   | MH002RCT09HP2 | Child long          | two  | 13.8-21.5cm |
| 9   | MH002RCT10HP1 | Small adult         | One  | 20.5-28.5cm |
| 10  | MH002RCT10HP2 | Small adult         | two  | 20.5-28.5cm |
| 11  | MH002RCT11HP1 | Small adult<br>long | One  | 20.5-28.5cm |
| 12  | MH002RCT11HP2 | Small adult<br>long | two  | 20.5-28.5cm |
| 13  | MH002RCT12HP1 | Adult               | One  | 27.5-36.5cm |
| 14  | MH002RCT12HP2 | Adult               | two  | 27.5-36.5cm |
| 15  | MH002RCT13HP1 | Adult long          | One  | 27.5-36.5cm |
| 16  | MH002RCT13HP2 | Adult long          | two  | 27.5-36.5cm |

{3}------------------------------------------------

| 17 | MH002RCT14HP1 | Big adult         | One | 35.5-46cm |
|----|---------------|-------------------|-----|-----------|
| 18 | MH002RCT14HP2 | Big adult         | two | 35.5-46cm |
| 19 | MH002RCT15HP1 | Big adult<br>long | One | 35.5-46cm |
| 20 | MH002RCT15HP2 | Big adult<br>long | two | 35.5-46cm |
| 21 | MH002RCT16HP1 | Thigh             | One | 45-56.5cm |
| 22 | MH002RCT16HP2 | Thigh             | two | 45-56.5cm |

Table 3 Specifications of Reusable Blood Pressure Cuff – with Integrated Bladder (SL)

| No. | Product code  | Name                | Tube | Arm size    |
|-----|---------------|---------------------|------|-------------|
| 1   | MH002RCT06SL1 | Infant              | One  | 9-14.8cm    |
| 2   | MH002RCT06SL2 | Infant              | two  | 9-14.8cm    |
| 3   | MH002RCT07SL1 | Infant long         | One  | 9-14.8cm    |
| 4   | MH002RCT07SL2 | Infant long         | two  | 9-14.8cm    |
| 5   | MH002RCT08SL1 | Child               | One  | 13.8-21.5cm |
| 6   | MH002RCT08SL2 | Child               | two  | 13.8-21.5cm |
| 7   | MH002RCT09SL1 | Child long          | One  | 13.8-21.5cm |
| 8   | MH002RCT09SL2 | Child long          | two  | 13.8-21.5cm |
| 9   | MH002RCT10SL1 | Small adult         | One  | 20.5-28.5cm |
| 10  | MH002RCT10SL2 | Small adult         | two  | 20.5-28.5cm |
| 11  | MH002RCT11SL1 | Small adult<br>long | One  | 20.5-28.5cm |
| 12  | MH002RCT11SL2 | Small adult<br>long | two  | 20.5-28.5cm |
| 13  | MH002RCT12SL1 | Adult               | One  | 27.5-36.5cm |
| 14  | MH002RCT12SL2 | Adult               | two  | 27.5-36.5cm |
| 15  | MH002RCT13SL1 | Adult long          | One  | 27.5-36.5cm |
| 16  | MH002RCT13SL2 | Adult long          | two  | 27.5-36.5cm |
| 17  | MH002RCT14SL1 | Big adult           | One  | 35.5-46cm   |
| 18  | MH002RCT14SL2 | Big adult           | two  | 35.5-46cm   |
| 19  | MH002RCT15SL1 | Big adult<br>long   | One  | 35.5-46cm   |
| 20  | MH002RCT15SL2 | Big adult<br>long   | two  | 35.5-46cm   |
| 21  | MH002RCT16SL1 | Thigh               | One  | 45-56.5cm   |
| 22  | MH002RCT16SL2 | Thigh               | two  | 45-56.5cm   |

{4}------------------------------------------------

| No. | Product code  | Name             | Tube | Arm size    |
|-----|---------------|------------------|------|-------------|
| 1   | MH002RCT06MQ1 | Infant           | One  | 9-14.8cm    |
| 2   | MH002RCT06MQ2 | Infant           | two  | 9-14.8cm    |
| 3   | MH002RCT07MQ1 | Infant long      | One  | 9-14.8cm    |
| 4   | MH002RCT07MQ2 | Infant long      | two  | 9-14.8cm    |
| 5   | MH002RCT08MQ1 | Child            | One  | 13.8-21.5cm |
| 6   | MH002RCT08MQ2 | Child            | two  | 13.8-21.5cm |
| 7   | MH002RCT09MQ1 | Child long       | One  | 13.8-21.5cm |
| 8   | MH002RCT09MQ2 | Child long       | two  | 13.8-21.5cm |
| 9   | MH002RCT10MQ1 | Small adult      | One  | 20.5-28.5cm |
| 10  | MH002RCT10MQ2 | Small adult      | two  | 20.5-28.5cm |
| 11  | MH002RCT11MQ1 | Small adult long | One  | 20.5-28.5cm |
| 12  | MH002RCT11MQ2 | Small adult long | two  | 20.5-28.5cm |
| 13  | MH002RCT12MQ1 | Adult            | One  | 27.5-36.5cm |
| 14  | MH002RCT12MQ2 | Adult            | two  | 27.5-36.5cm |
| 15  | MH002RCT13MQ1 | Adult long       | One  | 27.5-36.5cm |
| 16  | MH002RCT13MQ2 | Adult long       | two  | 27.5-36.5cm |
| 17  | MH002RCT14MQ1 | Big adult        | One  | 35.5-46cm   |
| 18  | MH002RCT14MQ2 | Big adult        | two  | 35.5-46cm   |
| 19  | MH002RCT15MQ1 | Big adult long   | One  | 35.5-46cm   |
| 20  | MH002RCT15MQ2 | Big adult long   | two  | 35.5-46cm   |
| 21  | MH002RCT16MQ1 | Thigh            | One  | 45-56.5cm   |
| 22  | MH002RCT16MQ2 | Thigh            | two  | 45-56.5cm   |

Table 4 Specifications of Reusable Blood Pressure Cuff – with Integrated Bladder (MQ)

Table 5 Specifications of Reusable Blood Pressure Cuff – with Integrated Bladder (DM)

| No. | Product code  | Name        | Tube | Arm size    |
|-----|---------------|-------------|------|-------------|
| 1   | MH002RCT06DM1 | Infant      | One  | 9-14.8cm    |
| 2   | MH002RCT06DM2 | Infant      | two  | 9-14.8cm    |
| 3   | MH002RCT07DM1 | Infant long | One  | 9-14.8cm    |
| 4   | MH002RCT07DM2 | Infant long | two  | 9-14.8cm    |
| 5   | MH002RCT08DM1 | Child       | One  | 13.8-21.5cm |
| 6   | MH002RCT08DM2 | Child       | two  | 13.8-21.5cm |
| 7   | MH002RCT09DM1 | Child long  | One  | 13.8-21.5cm |
| 8   | MH002RCT09DM2 | Child long  | two  | 13.8-21.5cm |

{5}------------------------------------------------

| 9   | MH002RCT10DM1 | Small adult | One | 20.5-28.5cm |
|-----|---------------|-------------|-----|-------------|
| 10  | MH002RCT10DM2 | Small adult | two | 20.5-28.5cm |
| 11  |               | Small adult | One | 20.5-28.5cm |
|     | MH002RCT11DM1 | long        |     |             |
| 12  | MH002RCT11DM2 | Small adult | two | 20.5-28.5cm |
|     |               | long        |     |             |
| 13  | MH002RCT12DM1 | Adult       | One | 27.5-36.5cm |
| 14  | MH002RCT12DM2 | Adult       | two | 27.5-36.5cm |
| ા ર | MH002RCT13DM1 | Adult long  | One | 27.5-36.5cm |
| 16  | MH002RCT13DM2 | Adult long  | two | 27.5-36.5cm |
| 17  | MH002RCT14DM1 | Big adult   | One | 35.5-46cm   |
| 18  | MH002RCT14DM2 | Big adult   | two | 35.5-46cm   |
| 19  | MH002RCT15DM1 | Big adult   | One | 35.5-46cm   |
|     |               | long        |     |             |
| 20  |               | Big adult   | two | 35.5-46cm   |
|     | MH002RCT15DM2 | long        |     |             |
| 21  | MH002RCT16DM1 | Thigh       | One | 45-56.5cm   |
| 22  | MH002RCT16DM2 | Thigh       | two | 45-56.5cm   |

| Table 6 Specifications of Reusable Blood Pressure Cuff - with Integrated Bladder (NA) |
|---------------------------------------------------------------------------------------|
|---------------------------------------------------------------------------------------|

| No. | Product code  | Name                | Tube | Arm size    |
|-----|---------------|---------------------|------|-------------|
| 1   | MH002RCT06NA1 | Infant              | One  | 9-14.8cm    |
| 2   | MH002RCT06NA2 | Infant              | two  | 9-14.8cm    |
| 3   | MH002RCT07NA1 | Infant long         | One  | 9-14.8cm    |
| 4   | MH002RCT07NA2 | Infant long         | two  | 9-14.8cm    |
| 5   | MH002RCT08NA1 | Child               | One  | 13.8-21.5cm |
| 6   | MH002RCT08NA2 | Child               | two  | 13.8-21.5cm |
| 7   | MH002RCT09NA1 | Child long          | One  | 13.8-21.5cm |
| 8   | MH002RCT09NA2 | Child long          | two  | 13.8-21.5cm |
| 9   | MH002RCT10NA1 | Small adult         | One  | 20.5-28.5cm |
| 10  | MH002RCT10NA2 | Small adult         | two  | 20.5-28.5cm |
| 11  | MH002RCT11NA1 | Small adult<br>long | One  | 20.5-28.5cm |
| 12  | MH002RCT11NA2 | Small adult<br>long | two  | 20.5-28.5cm |
| 13  | MH002RCT12NA1 | Adult               | One  | 27.5-36.5cm |
| 14  | MH002RCT12NA2 | Adult               | two  | 27.5-36.5cm |
| 15  | MH002RCT13NA1 | Adult long          | One  | 27.5-36.5cm |
| 16  | MH002RCT13NA2 | Adult long          | two  | 27.5-36.5cm |

{6}------------------------------------------------

| 17 | MH002RCT14NA1 | Big adult | One | 35.5-46cm |
|----|---------------|-----------|-----|-----------|
| 18 | MH002RCT14NA2 | Big adult | two | 35.5-46cm |
| 19 | MH002RCT15NA1 | Big adult | One | 35.5-46cm |
|    |               | long      |     |           |
| 20 |               | Big adult | two | 35.5-46cm |
|    | MH002RCT15NA2 | long      |     |           |
| 21 | MH002RCT16NA1 | Thigh     | One | 45-56.5cm |
| 22 | MH002RCT16NA2 | Thigh     | two | 45-56.5cm |

Table 7 Specifications of Reusable Blood Pressure Cuff – with Integrated Bladder (FL)

| No. | Product code  | Name             | Tube | Arm size    |
|-----|---------------|------------------|------|-------------|
| 1   | MH002RCT06FL1 | Infant           | One  | 9-14.8cm    |
| 2   | MH002RCT06FL2 | Infant           | two  | 9-14.8cm    |
| 3   | MH002RCT07FL1 | Infant long      | One  | 9-14.8cm    |
| 4   | MH002RCT07FL2 | Infant long      | two  | 9-14.8cm    |
| 5   | MH002RCT08FL1 | Child            | One  | 13.8-21.5cm |
| 6   | MH002RCT08FL2 | Child            | two  | 13.8-21.5cm |
| 7   | MH002RCT09FL1 | Child long       | One  | 13.8-21.5cm |
| 8   | MH002RCT09FL2 | Child long       | two  | 13.8-21.5cm |
| 9   | MH002RCT10FL1 | Small adult      | One  | 20.5-28.5cm |
| 10  | MH002RCT10FL2 | Small adult      | two  | 20.5-28.5cm |
| 11  | MH002RCT11FL1 | Small adult long | One  | 20.5-28.5cm |
| 12  | MH002RCT11FL2 | Small adult long | two  | 20.5-28.5cm |
| 13  | MH002RCT12FL1 | Adult            | One  | 27.5-36.5cm |
| 14  | MH002RCT12FL2 | Adult            | two  | 27.5-36.5cm |
| 15  | MH002RCT13FL1 | Adult long       | One  | 27.5-36.5cm |
| 16  | MH002RCT13FL2 | Adult long       | two  | 27.5-36.5cm |
| 17  | MH002RCT14FL1 | Big adult        | One  | 35.5-46cm   |
| 18  | MH002RCT14FL2 | Big adult        | two  | 35.5-46cm   |
| 19  | MH002RCT15FL1 | Big adult long   | One  | 35.5-46cm   |
| 20  | MH002RCT15FL2 | Big adult long   | two  | 35.5-46cm   |
| 21  | MH002RCT16FL1 | Thigh            | One  | 45-56.5cm   |
| 22  | MH002RCT16FL2 | Thigh            | two  | 45-56.5cm   |

{7}------------------------------------------------

| No. | Product code  | Name             | Tube | Arm size    |
|-----|---------------|------------------|------|-------------|
| 1   | MH002RCT06MF1 | Infant           | One  | 9-14.8cm    |
| 2   | MH002RCT06MF2 | Infant           | two  | 9-14.8cm    |
| 3   | MH002RCT07MF1 | Infant long      | One  | 9-14.8cm    |
| 4   | MH002RCT07MF2 | Infant long      | two  | 9-14.8cm    |
| 5   | MH002RCT08MF1 | Child            | One  | 13.8-21.5cm |
| 6   | MH002RCT08MF2 | Child            | two  | 13.8-21.5cm |
| 7   | MH002RCT09MF1 | Child long       | One  | 13.8-21.5cm |
| 8   | MH002RCT09MF2 | Child long       | two  | 13.8-21.5cm |
| 9   | MH002RCT10MF1 | Small adult      | One  | 20.5-28.5cm |
| 10  | MH002RCT10MF2 | Small adult      | two  | 20.5-28.5cm |
| 11  | MH002RCT11MF1 | Small adult long | One  | 20.5-28.5cm |
| 12  | MH002RCT11MF2 | Small adult long | two  | 20.5-28.5cm |
| 13  | MH002RCT12MF1 | Adult            | One  | 27.5-36.5cm |
| 14  | MH002RCT12MF2 | Adult            | two  | 27.5-36.5cm |
| 15  | MH002RCT13MF1 | Adult long       | One  | 27.5-36.5cm |
| 16  | MH002RCT13MF2 | Adult long       | two  | 27.5-36.5cm |
| 17  | MH002RCT14MF1 | Big adult        | One  | 35.5-46cm   |
| 18  | MH002RCT14MF2 | Big adult        | two  | 35.5-46cm   |
| 19  | MH002RCT15MF1 | Big adult long   | One  | 35.5-46cm   |
| 20  | MH002RCT15MF2 | Big adult long   | two  | 35.5-46cm   |
| 21  | MH002RCT16MF1 | Thigh            | One  | 45-56.5cm   |
| 22  | MH002RCT16MF2 | Thigh            | two  | 45-56.5cm   |

Table 8 Specifications of Reusable Blood Pressure Cuff – with Integrated Bladder (MF)

{8}------------------------------------------------

#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

## Indications for Use

510(k) Number (if known) K171889

Device Name Reusable Blood Pressure Cuff

### Indications for Use (Describe)

The Reusable Blood Pressure Cuff is an accessory used in conjunction with noninvasive blood pressure measurement systems. The cuff is non-sterile and reusable in infant, pediatric but not neonatal, and adult sizes.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW! *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

{9}------------------------------------------------

# 510(k) Summary

This 510(k) Summary is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92.

The assigned 510(k) Number: __K171889

- 1. Date of Preparation: 12/18/2017
- 2. Sponsor Identification

Suzhou Minhua Medical Apparatus Supplies Co., LTD You Yi Industrial Park, Songlin Town, Wujiang, 215222, CHINA

Establishment Registration Number: 3011529836

Contact Person: Yang Qidong Position: Vice general manager Tel: +86-21-59840221 Fax: +86-21-59841086 Email: QD@minhuayiliao.com

- 3. Designated Submission Correspondent
Ms. Diana Hong (Primary Contact Person) Mr. Betty Xiao (Alternative Contact Person)

Mid-Link Consulting Co., Ltd P.O. Box 120-119, Shanghai, 200120, China

Tel: +86-21-22815850 Fax: 240-238-7587 Email: info@mid-link.net

{10}------------------------------------------------

#### 4. Identification of Proposed Device

Trade Name: Reusable Blood Pressure Cuff Common Name: Blood Pressure Cuff

Regulatory Information Classification Name: Blood Pressure Cuff Classification: 2 Product Code: DXQ Regulation Number:870.1120 Review Panel: Cardiovascular

Intended Use Statement:

The Reusable Blood Pressure Cuff is an accessory used in conjunction with noninvasive blood pressure measurement systems. The cuff is non-sterile and reusable. It is available in infant, pediatric but not neonatal, and adult sizes.

Device Description

The proposed device is an accessory used in conjunction with noninvasive blood pressure measurement systems, and wrap around the patient's limb. The cuff is non-sterile and reusable.

The model difference of the proposed device includes bladder type, tube, color, joint model and size.

#### ર. Identification of Predicate Device(s)

Predicate Device 1 510(k) Number: K133117 Product Name: Andon Blood Pressure Cuff

Predicate Device 2 510(k) Number: K152468 Product Name: Reusable Blood Pressure Cuff

#### 6. Non-Clinical Test Conclusion

Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:

{11}------------------------------------------------

- > IEC 80601-2-30: 2013, Medical Electrical Equipment - Part 2-30: Particular Requirements For The Basic Safety And Essential Performance Of Automated Non-Invasive Sphygmomanometers.
- > ISO 10993-5: 2009, Biological Evaluation Of Medical Devices - Part 5: Tests For In Vitro Cytotoxicity.
- A ISO 10993-10: 2010, Biological Evaluation Of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization.
- 7. Clinical Test Conclusion

No clinical study is included in this submission.

- Substantially Equivalent (SE) Comparison 8.

| Item                       | Proposed Device                                                                                                                                                                                                                                           | Predicate Device 1                                                                                                                                                                                    |
|----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                            | Cuff with Internal Bladder                                                                                                                                                                                                                                | K133117                                                                                                                                                                                               |
| Device class               | 2                                                                                                                                                                                                                                                         | 2                                                                                                                                                                                                     |
| Product code               | DXQ                                                                                                                                                                                                                                                       | DXQ                                                                                                                                                                                                   |
| Regulation Number          | 21 CFR 870.1120                                                                                                                                                                                                                                           | 21 CFR 870.1120                                                                                                                                                                                       |
| Intended use               | The Reusable Blood Pressure Cuff is<br>an accessory used in conjunction with<br>noninvasive blood pressure<br>measurement systems. The cuff is<br>non-sterile and reusable. It is<br>available in infant, pediatric but not<br>neonatal, and adult sizes. | Blood Pressure Cuffs are intended<br>to be wrapped on the upper arm<br>and used with a non-invasive blood<br>pressure monitor system to<br>complete the measurement of<br>blood parameters on adults. |
| Patient populations        | adult                                                                                                                                                                                                                                                     | adult                                                                                                                                                                                                 |
| Tube configuration         | One or two tube                                                                                                                                                                                                                                           | One tube                                                                                                                                                                                              |
| Bladder                    | Internal Bladder                                                                                                                                                                                                                                          | Internal Bladder                                                                                                                                                                                      |
| Principles of<br>Operation | Bladder is wrapped around the<br>patient's limb and secured by hook<br>and loop closure. Air hose is<br>connected to the noninvasive blood<br>pressure measurement systems.                                                                               | Bladder is wrapped around the<br>patient's limb and secured by hook<br>and loop closure. Air hose is<br>connected to the noninvasive blood<br>pressure measurement systems.                           |
| Size                       | Small adult: 17-22cm<br>Small adult long / Adult:<br>22-30 cm<br>Large Adult: 30-42 cm<br>Large Adult long: 42-48 cm                                                                                                                                      | KD-525-P34: 15-24cm<br>KD-525-P31: 20-34 cm<br>KD-525-P32: 30-44 cm<br>KD-525-P33: 40-48 cm                                                                                                           |

## Table 1 Comparison for Cuff with Internal Bladder

{12}------------------------------------------------

| Reusable         | yes                                          | Yes                                          |
|------------------|----------------------------------------------|----------------------------------------------|
| Sterile          | No                                           | No                                           |
| Pressure range   | 0-300mmHg                                    | 0-300mmHg                                    |
| Material         | Polyester                                    | Not available                                |
| Biocompatibility | Comply with ISO 10993-5 and ISO<br>10993-10. | Comply with ISO 10993-5 and<br>ISO 10993-10. |

There are four differences for the proposed cuff with internal bladder and predicate device 1.

#### For intended use.

The proposed devices include two types of bladder. The cuffs with integrated bladder include with sizes of infant, child and adult. But the cuffs with internal bladder only include adult size. The patient population of proposed cuffs with internal bladder include adult only. The intended use of proposed cuffs with internal bladder and predicate device 1 is same actually.

### For tube configuration,

The proposed devices include one two tubes, while the predicate device only includes one tube. More tube configuration can be applicable for more types of noninvasive blood pressure measurement system. And the two tubes configuration is very common for blood pressure cuff. The predicate device 2, Reusable Blood Pressure Cuff, includes two tubes configuration. Therefore, this difference will not result in any safety and effectiveness issue of the proposed device.

#### For size:

The cuff size range of proposed devices is included in the range of the predicate device 1. And the size is indicated in user manual clearly for selection. Therefore, this difference will not result in any safety and effectiveness issue of the proposed device.

For material:

The material of proposed device and predicate device 1 may be different, but both of them comply with ISO 10993-5 and ISO 10993-10. Therefore, this difference will not result in any safety and effectiveness issue of the proposed device.

| ITEM                 | Proposed Device                        | Predicate Device 2                       |
|----------------------|----------------------------------------|------------------------------------------|
|                      | Cuff with Integrated Bladder           | K152468                                  |
| Device class         | 2                                      | 2                                        |
| Product code         | DXQ                                    | DXQ                                      |
| Regulation<br>Number | 21 CFR 870.1120                        | 21 CFR 870.1120                          |
| Intended use         | The Reusable Blood Pressure Cuff       | The reusable blood pressure cuff is an   |
|                      | is an accessory used in conjunction    | accessory used in conjunction with       |
|                      | with noninvasive blood pressure        | blood<br>noninvasive<br>pressure         |
|                      |                                        | measurement systems. The cuff is         |
|                      | measurement systems. The cuff is       |                                          |
|                      | non-sterile and reusable. It<br>is     | non-sterile and may be reused. It is     |
|                      | available in infant, pediatric but not | available in infant, pediatric and adult |
|                      | neonatal, and adult sizes.             | sizes                                    |
| Patient              | Infant / pediatric / adult             | Infant / pediatric / adult               |
| populations          |                                        |                                          |
| Tube                 | One or two tube                        | One or two tube                          |
| configuration        |                                        |                                          |
| Bladder              | Integrated Bladder                     | Integrated Bladder                       |
| Principles<br>of     | Bladder is wrapped around<br>the       | Bladder is wrapped around<br>the         |
| Operation            | patient's limb and secured by hook     | patient's limb and secured by hook       |
|                      | and loop closure. Air hose is          | and loop closure. Air hose is            |
|                      | connected to the noninvasive blood     | connected to the noninvasive blood       |
|                      | pressure measurement systems.          | pressure measurement systems.            |
| Size                 | Infant / infant long: 9-14.8cm         | Infant: 8-13 cm                          |
|                      | Child /child long: 13.8-21.5cm         | Child: 12-19 cm                          |
|                      | Small adult / small adult long:        | Small adult: 17-25cm                     |
|                      | 20.5-28.5cm                            |                                          |
|                      | Small adult long / Adult:              | Adult: 23-33cm                           |
|                      | 27.5-36.5cm                            |                                          |
|                      | Big adult /<br>big adult<br>long:      | Large adult / large<br>adult<br>long:    |
|                      | 35.5-46cm                              | 31-40cm                                  |
|                      | Thigh: 45-56.5cm                       | Thigh: 38-50cm                           |
| Reusable             | yes                                    | Yes                                      |
| Sterile              | No                                     | No                                       |
| Pressure range       | 0-300mmHg                              | 0-300mmHg                                |
| Material             | TPU leather with nylon                 | 210D Nylon TPU and PVC                   |
| Biocompatibility     | Comply with ISO 10993-5 and ISO        | Comply with ISO 10993-5 and ISO          |
|                      | 10993-10.                              | 10993-10.                                |

Table 2 Comparison for Cuff with Integrated Bladder

{13}------------------------------------------------

There are two differences for the proposed cuff with integrated bladder and predicate device 2.

## For size,

 The cuff size range of proposed devices is different to that of predicate device 2. But the different is slight. And the size is indicated in user manual clearly for selection. Therefore, this difference will not result in any safety and effectiveness issue of the proposed device.

{14}------------------------------------------------

For material:

The material of proposed device and predicate device 2 is different, but both of them comply with ISO 10993-5 and ISO 10993-10. Therefore, this difference will not result in any safety and effectiveness issue of the proposed device.

- 9. Substantially Equivalent (SE) Conclusion
Based on the comparison and analysis above, the proposed devices are determined to be Substantially Equivalent (SE) to the predicate devices.

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DXQ/K171889](https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DXQ/K171889)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
