Statcorp Medical UltraCheck® Spacelabs ABP Cuffs are used with identified devices intended for the non-invasive measurement of pediatric and adult human blood pressure. Statcorp Medical UltraCheck® Spacelabs ABP Cuffs are intended for use by or under the supervision of qualified medical personnel.
Device Story
UltraCheck Spacelabs ABP Cuffs are non-invasive blood pressure cuffs; applied to patient limb; pneumatically connected to oscillometric blood pressure monitor. Device consists of flexible polymeric material with integrated inflatable bladder; secured via oblong ring loop mechanism for easier application. Features fringe on weld side for patient comfort. Used in clinical settings by medical personnel. Output is pneumatic pressure signal processed by monitor to determine blood pressure. Benefits include reliable, non-invasive blood pressure monitoring.
Clinical Evidence
Clinical accuracy testing performed per AAMI ANSI ISO 81060-2 (2013) requirements. Mechanical testing included pressure cycle testing (10,000 cycles), application testing (1,000 cycles), ring durability, cleaning validation, and label integrity testing.
Technological Characteristics
Flexible polymeric material; integrated inflatable bladder; pneumatic connection via one-piece tube to Spacelabs ABP monitor; ring-loop securement mechanism; fringe added to weld side for comfort.
Indications for Use
Indicated for non-invasive blood pressure measurement in pediatric and adult patients. Intended for use by or under the supervision of qualified medical personnel.
Regulatory Classification
Identification
A blood pressure cuff is a device that has an inflatable bladder in an inelastic sleeve (cuff) with a mechanism for inflating and deflating the bladder. The cuff is used in conjunction with another device to determine a subject's blood pressure.
K143728 — Minhua Disposable Blood Pressure Cuff · Suzhou Minhua Medical Apparatus · May 15, 2015
Submission Summary (Full Text)
{0}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 27, 2015
Statcorp Medical Wayne Emmert Director of Operations 14476 Duval Place West Suite 303 Jacksonville, Florida 32218
Re: K152801
Trade/Device Name: Ultracheck Spacelabs ABP Cuff-child, Ultracheck Spacelabs ABP Cuff-small Adult, Ultracheck Spacelabs ABP Cuff-adult, Ultracheck Spacelabs ABP Cuff-large Adult
Regulation Number: 21 CFR 870.1120 Regulation Name: Blood Pressure Cuff Regulatory Class: Class II Product Code: DXO Dated: September 25, 2015 Received: September 28, 2015
Dear Wayne Emmert:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{1}------------------------------------------------
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Bram D. Zuckerman -S
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K152801
Device Name
Statcorp Medical UltraCheck Spacelabs ABP Cuffs
Indications for Use (Describe)
Statcorp Medical UltraCheck® Spacelabs ABP Cuffs are used with identified devices intended for the non-invasive measurement of pediatric and adult human blood pressure.
Statcorp Medical UltraCheck® Spacelabs ABP Cuffs are intended for use by or under the supervision of qualified medical personnel.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------------------|---------------------------------------------|
| <span style="text-decoration: overline;">x</span> Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K152801
Image /page/3/Picture/1 description: The image shows the logo for Statcorp Medical. The logo features the word "Statcorp" in bold, blue letters at the top, followed by a graphic of horizontal blue lines that resemble a sound wave or a stylized medical chart. Below the graphic, the word "Medical" is written in bold, blue letters, mirroring the style of "Statcorp" above. The overall design is clean and professional, suggesting a company in the healthcare or medical technology sector.
# UltraCheck® Spacelabs ABP Blood Pressure Cuffs 510(k) Summary of Safety and Effectiveness
#### 1. Applicant Information
A. Applicant Information: Statcorp Medical 14476 Duval Place West, Suite 303 Jacksonville, FL 32218 1-904-786-5113
| B. Official Contact: | Wayne Emmert<br>Director of Operations |
|----------------------|----------------------------------------|
|----------------------|----------------------------------------|
- C. Date of Summary: 9/25/15
## II. Device Information
- A. Proprietary Name : UltraCheck® Spacelabs ABP Cuffs
- B. Common Names: Blood Pressure Cuff
- C. Classification Device Name: Blood Pressure Cuff
- D. Classification Regulatory Description: Blood Pressure Cuff
- E. Product Code: DXQ
- F. Regulatory Class: ==
- G. Panel: Cardiovascular
#### III. Predicate Device
The predicate devices for these cuffs are the Ultracuff Blood Pressure Cuffs (for Spacelabs ABP), 510(k) # K954282.
Statcorp Special 510(k) Section 8
{4}------------------------------------------------
#### IV. General Description
The Statcorp UltraCheck® Spacelabs ABP cuffs in this 510(k), are applied to a patient limb and connected pneumatically to an oscillometric blood pressure monitor to enable noninvasive blood pressure measurements. The cuff is made of flexible polymeric material, a section of which forms an integrated inflatable bladder. The cuffs in this 510(k) are secured by looping the cuff through an attached oblong ring back onto itself to enable easier selfapplication. These cuffs are connected pneumatically through a one piece tube to the Spacelabs ABP Monitor.
#### V. Indications
The Statcorp UltraCheck® Spacelabs ABP Cuffs are used with identified devices intended for the non-invasive measurement of pediatric and adult human blood pressure.
UltraCheck® Blood Pressure Cuffs are intended for use by or under the supervision of qualified medical personnel.
### VI. Comparison to Predicate
The UltraCheck® Spacelabs ABP cuffs in this 510(k) are similar to the Statcorp Ultracuff described in K954282 in that use the same cuff materials, bladder sizes and methods of construction. It differs from the fact that these cuffs are designed to be applied by passing the extended portion through a ring and attaching the cuff back onto itself. The predicate device is merely wrapped around the arm in the same direction. There is also an added fringe on the weld side of the cuff which eases the transition from the stiffer portion of the weld to the edge of the cuff which is added to provide for patient comfort.
#### VII. Test Summary
Mechanical tests included
- . Pressure cycle testing to 10,000 cycles and application testing to 1,000 cycles to validate life.
- Mechanical stress testing to validate ring durability .
- Cleaning validation tests to validate cleaning recommendations .
- Label integrity testing to validate label chemical resistance and legibility .
Clinical Accuracy testing performed per the methodology and requirements of AAMI ANSI ISO 81060-2 (2013)
#### VIII. Conclusions
Based upon the above test results the Statcorp Medical UltraCheck® Spacelabs ABP Cuffs are substantially equivalent to the predicate devices cited and safe and effective for their stated intended use.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.