← Product Code [DXQ](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DXQ) · K152468

# Reusable Blood Pressure Cuff (K152468)

_Shenzhen Vistar Medical Supplies Co., Ltd. · DXQ · Oct 29, 2015 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DXQ/K152468

## Device Facts

- **Applicant:** Shenzhen Vistar Medical Supplies Co., Ltd.
- **Product Code:** [DXQ](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DXQ.md)
- **Decision Date:** Oct 29, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1120
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Pediatric

## Indications for Use

The reusable blood pressure cuff is an accessory used in conjunction with noninvasive blood pressure measurement systems. The cuff is non-sterile and may be reused. It is available in infant, pediatric and adult sizes.

## Device Story

Reusable blood pressure cuff; accessory for noninvasive blood pressure measurement systems. Consists of inflatable bladder, hook-and-loop closure, and single or double air hose. Wrapped around patient limb; air hose connects to measurement system. Used in clinical settings to facilitate blood pressure monitoring. Benefits patient by enabling noninvasive, reusable blood pressure measurement.

## Clinical Evidence

No clinical data.

## Technological Characteristics

Materials: 210D Nylon TPU (cuff), PVC (air hose). Pressure range: 0-300 mmHg. Max pressure: 400 mmHg. Leakage: < 4 mmHg/min. Standards: ISO 10993-5 (cytotoxicity), ISO 10993-10 (irritation/sensitization), ISO 81060-1 (non-invasive sphygmomanometers). Non-sterile, reusable.

## Regulatory Identification

A blood pressure cuff is a device that has an inflatable bladder in an inelastic sleeve (cuff) with a mechanism for inflating and deflating the bladder. The cuff is used in conjunction with another device to determine a subject's blood pressure.

## Predicate Devices

- Reusable/Reusable Blood Pressure Cuff ([K151290](/device/K151290.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus, which is a symbol often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus. The logo is simple and recognizable, representing the department's role in promoting the health and well-being of Americans.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

October 29, 2015

Shenzhen Vistar Medical Supplies Co., Ltd. Mr. Ray Wang Official Correspondent 1-202, Build 3, Beijing New World, No.5 Chaoyang Rd., Chaoyang District Beijing, 100024 CN

Re: K152468

> Trade/Device Name: Reusable Blood Pressure Cuff Regulation Number: 21 CFR 870.1120 Regulation Name: Blood Pressure Cuff Regulatory Class: Class II Product Code: DXQ Dated: August 25, 2015 Received: August 31, 2015

Dear Mr. Ray Wang,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply 

{1}------------------------------------------------

Page 2 - Mr. Ray Wang

with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Mitchell Stein

forBram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K152468

Device Name Reusable Blood Pressure Cuff

### Indications for Use (Describe)

The reusable blood pressure cuff is an accessory used in conjunction with noninvasive blood pressure measurement systems. The cuff is non-sterile and may be reused. It is available in infant, pediatric and adult sizes.

Type of Use (Select one or both, as applicable)

| <span style="font-size:100%;">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size:100%;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|-------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|
|-------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# K152468

# Page 1 of 5

# 510(k) Summary

This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92.

The assigned 510(k) Number: __________________________________________________________________________________________________________________________________________________

- 1. Date of Preparation:2015/8/26
- 2. Sponsor Identification

## Shenzhen Vistar Medical Supplies Co., Ltd. 808, Hanhaida Bldg., No. 7 Songgang Blvd. Songgang Town, Shenzhen, China 518105

Establishment Registration Number: Pending

Contact Person: Zhenxun Chen Position: General Manager Tel: +86-0755-85298331 Fax:+86-0755-85298331 Email: chainy.chen@163.com

- 3. Designated Submission Correspondent
Mr. Ray Wang

### Beijing Believe Tech. Service Co., Ltd

Tel: +86-21-50313932, Fax: +86-21-68093116 Email: Ray.Wang@believe-med.com

{4}------------------------------------------------

K152468 Page 2 of 5

#### Identification of Proposed Device 4.

Trade Name: Reusable Blood Pressure Cuff Common Name: Blood Pressure Cuff Model(s): V0111C, V0112C, V0113C, V0114C, V0114C-L, V0115C, V0115C-L, V0116C; V0121C,V0122C,V0123C,V0124C,V0124C,V0125C,V0125C,V0125C-L,V0126C;

### Regulatory Information

Classification Name: Blood Pressure Cuff Classification: 2 Product Code: DXQ Regulation Number: 870.1120 Review Panel: Cardiovascular

### Indication For Use Statement:

The reusable blood pressure cuff is an accessory used in conjunction with noninvasive blood pressure measurement systems. The cuff is non-sterile and may be reused. It is available in infant, pediatric and adult sizes.

### Device Description

The proposed device is an accessory used in conjunction with noninvasive blood pressure measurement systems. It is available in neonatal, infant, pediatric and adult sizes.

The proposed device is Reusable blood pressure cuff, which contains Cuff with bladder and Air Hose. Air hose has single tube and double tube.

During the operation process, the integrated inflatable bladder will wrapped around the patient's limb and secured by hook and loop closure, and the air hose will connected to the noninvasive blood pressure measurement systems.

The reusable blood pressure cuff is reusable device, and which is made of 210D Nylon TPU (Cuff) and PVC (Air Hose), thereinto, the 210D Nylon TPU (Cuff) is the material used to contacting with the patient.

The reusable blood pressure cuff has 16 models with different size for different population with different arm size, and the 8 models of 16 are use single tube of air hose and other 8 models are use double tube of air hose, which is used for different noninvasive blood pressure measurement system.

The Reusable blood pressure cuff is provided as non-sterile.

{5}------------------------------------------------

K152468 Page 3 of 5

- ર. Identification of Predicate Device(s)
Predicate Device K151290 Reusable/Reusable Blood Pressure Cuff Xuzhou Maicuff Technology Co., Ltd.

- 6. Non-Clinical Test Conclusion
Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:

- > ISO 10993-5: 2009 Biological Evaluation Of Medical Devices -- Part 5: Tests For In Vitro Cytotoxicity
- A ISO 10993-10: 2010 Biological Evaluation Of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization.
- ハ ISO 81060-1:2012 Non-Invasive Sphygmomanometers - Part 1: Requirements And Test Methods For Non-Automated Measurement Type
- 7. Clinical Test Conclusion

No clinical study is included in this submission.

{6}------------------------------------------------

K152468 Page 4 of 5

### 8. Substantially Equivalent (SE) Comparison

| ITEM                       | Proposed Device<br>(Reusable<br>Blood Pressure Cuff)                                                                                                                                                                                           | Predicate Device (K151290)                                                                                                                                                                                                                                                 | Remark     |
|----------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Intended Use               | The reusable blood pressure cuff is an<br>accessory used in conjunction with<br>noninvasive blood pressure measurement<br>systems. The cuff is non-sterile and may be<br>reused. It is available in infant, pediatric<br>and adult sizes.      | The reusable blood pressure cuff is an<br>accessory used in conjunction with<br>noninvasive blood pressure measurement<br>systems. The cuff is non-sterile and may be<br>reused. It is available in infant, pediatric<br>and adult sizes.                                  | SE         |
| Patient<br>Populations     | Adults/Pediatrics                                                                                                                                                                                                                              | Adults/Pediatrics                                                                                                                                                                                                                                                          | SE         |
| Tube<br>Configuration      | One or two tube                                                                                                                                                                                                                                | One or two tube                                                                                                                                                                                                                                                            | SE         |
| Principles of<br>Operation | Bladder is wrapped around the patient's<br>limb and secured by hook and loop closure<br>Air hose is connected to the noninvasive<br>blood pressure measurement systems                                                                         | Bladder is wrapped around the patient's<br>limb and secured by hook and loop closure<br>Air hose is connected to the noninvasive<br>blood pressure measurement systems                                                                                                     | SE         |
| Size                       | Conform to AHA bladder sizes<br>recommendations<br>Infant (8-13 cm)<br>Child (12-19 cm)<br>Small Adult (17-25 cm)<br>Adult (23-33 cm)<br>Adult Long (23-33 cm)<br>Large Adult(31-40 cm)<br>Large Adult Long (31 - 40 cm)<br>Thigh (38 - 50 cm) | Conform to AHA bladder sizes<br>recommendations<br>Neonatal (6-11 cm)<br>Infant (10-19 cm)<br>Pediatric (18-26 cm)<br>Small Adult (20-28 cm)<br>Adult (25-35 cm)<br>Adult Long(25-35 cm)<br>Large Adult (33-47 cm)<br>Large Adult Long(33-47 cm)<br>Adult Thigh (46-66 cm) | Analysis 1 |
| Single Use                 | Y                                                                                                                                                                                                                                              | Y                                                                                                                                                                                                                                                                          | SE         |
| Sterile                    | No                                                                                                                                                                                                                                             | No                                                                                                                                                                                                                                                                         | SE         |
| Pressure Range             | 0-300 mmHg                                                                                                                                                                                                                                     | 0-300 mmHg                                                                                                                                                                                                                                                                 | SE         |
| Max. Pressure              | 400 mmHg                                                                                                                                                                                                                                       | >350 mmHg                                                                                                                                                                                                                                                                  | SE         |
| Tubing<br>Size(mm)         | OD = 8.0; ID=4.0; L=200;                                                                                                                                                                                                                       | OD = 8.0; ID=4.0; L=200;                                                                                                                                                                                                                                                   | SE         |
| Max. Leakage               | < 4mm Hg/ min.                                                                                                                                                                                                                                 | < 4mm Hg/ min.                                                                                                                                                                                                                                                             | SE         |
| Material                   | 210D Nylon TPU (Cuff) and PVC (Air<br>Hose)                                                                                                                                                                                                    | 210D Nylon TPU (Cuff) and PVC (Air<br>Hose)                                                                                                                                                                                                                                | SE         |
| ITEM                       | Proposed Device                                                                                                                                                                                                                                | Predicate Device                                                                                                                                                                                                                                                           | Remark     |
| Cytotoxicity               | Under the conditions of the study, not cyteotoxicity effect                                                                                                                                                                                    | Comply with ISO 10993-5                                                                                                                                                                                                                                                    | SE         |
| Irritation                 | Under the conditions of the study, not an irritant                                                                                                                                                                                             | Comply with ISO 10993-10                                                                                                                                                                                                                                                   | SE         |
| Sensitization              | Under conditions of the study, not a sensitizer.                                                                                                                                                                                               |                                                                                                                                                                                                                                                                            | SE         |

# Table 1 General Comparison

{7}------------------------------------------------

Analysis 1:

The proposed device has different size with predicate device for different patient arm circumference, the proposed device are conducted the ISO 81060-1 for the performance and Biocompatibility test as ISO 10993-1 for safety, so we consider that this difference do not affect the SE with the predicate device.

Table 2 Biocompatibility Comparison

#### Substantially Equivalent (SE) Conclusion 9.

Based on the comparison and analysis above, the proposed devices are determined to be Substantially Equivalent (SE) to the predicate devices.

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DXQ/K152468](https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DXQ/K152468)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
