K990807 · Health & Life Co., Ltd. · DXN · Dec 3, 1999 · Cardiovascular
Device Facts
Record ID
K990807
Device Name
HL148 & HL168
Applicant
Health & Life Co., Ltd.
Product Code
DXN · Cardiovascular
Decision Date
Dec 3, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1130
Device Class
Class 2
Intended Use
Measures automatically human's Systolic, Diastolic blood pressure and heart rate using the oscillometric method. All values can be read out in one LCD Panel. The intended for use of this over-the-counter device is for age 16 and above.
Device Story
Full Automatic (NIBP) Blood Pressure Monitor; utilizes oscillometric method to measure systolic/diastolic blood pressure and heart rate; intended for over-the-counter use by individuals age 16 and above; device displays results on integrated LCD panel; provides automated monitoring of vital signs; supports patient self-management of blood pressure.
Clinical Evidence
No clinical data provided; substantial equivalence based on technological characteristics and intended use.
Indicated for individuals age 16 and above for the automatic measurement of systolic and diastolic blood pressure and heart rate using the oscillometric method.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
Related Devices
K042802 — H & L FULL AUTOMATIC (NIBP) BLOOD PRESSURE MONITOR, MODELS HL168H, HL168J & HL168Q · Health & Life Co., Ltd. · Oct 26, 2004
K020766 — HL168F · Health & Life Co., Ltd. · Apr 5, 2002
K033043 — MODIFICATION TO HL168 · Health & Life Co., Ltd. · Oct 10, 2003
K014122 — HL168E · Health & Life Co., Ltd. · Jan 16, 2002
K013033 — MODIFICATION TO: HL168W TOUCH SCREEN · Health & Life Co., Ltd. · Oct 23, 2001
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, positioned within a circle. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged around the upper portion of the circle. The logo is presented in black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
JUN 2 4 2010
Health & Life CO., Ltd. c/o Mr. Richard J. Stark Medisana, USA 801 East Morehead Street, Suite 305 Charlotte, NC 28202
- Re: K990807 HL148 and HL168 Regulatory Class: II (two) Product Code: DXN Dated: October 19, 1999 Received: October 27, 1999
Dear Mr. Stark:
This letter corrects our substantially equivalent letter of December 3, 1999.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2 - Mr. Richard J. Stark
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Bram D. Zuckerman, M.D.
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
{2}------------------------------------------------
PAGE 3 OF 4
## 510(K) Number (if known): K990807
Device Name: Full Automatic (NIBP) Blood Pressure Monitor
## Intended for Use
Measures automatically human's Systolic, Diastolic blood pressure and heart rate using the oscillometric manal 3 bystolic, Diastone blood pressure and heart
using the oscillometric method. All values can be read out in one LCD Panel.
The intended for use of this over-the-counter device is for age 16 and above.
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use ( -Or (Per 21 CFR 801.109)
Over-the-Counter Use ( √ )
Bore L. Luyers Qe
(Division Sign-Off)
Division of Cardiovascular, Respiratory, and Neurological Devices
510(k) Number K990807
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.