The Globalcare GUS610 automatic Blood Pressure Monitor is indicated for home use for the non-invasive measurement of diastolic and systolic blood pressures and pulse rate of adults by means of an inflatable cuff which is wrapped around the upper arm. The cuff circumference is limited to 22 to 52 cm.
Device Story
Automatic upper arm blood pressure monitor; intended for home use by adults. Device uses inflatable cuff to measure arterial blood pressure and pulse rate. Inputs: oscillometric pressure signals from cuff. Outputs: systolic/diastolic pressure (mmHg), pulse rate (1/min), cardiac arrhythmia (IHB) symbol, risk level, and Hemodynamic Rest Condition (HSD) indication. Powered by batteries or mains adapter. User wraps cuff around upper arm; device performs automated measurement; results displayed on screen. Provides objective data for patient monitoring; assists in tracking blood pressure trends. New optional cuff (42-52 cm) added to existing GUS610 platform.
Clinical Evidence
Clinical investigation per ISO 81060-2:2018 conducted on 87 subjects with arm circumferences 42-52 cm (259 datasets). Results met criteria for mean difference and standard deviation for SBP and DBP. No adverse incidents or failed measurements reported. Accuracy for BP and pulse rate met manufacturer specifications.
Technological Characteristics
Non-invasive oscillometric blood pressure monitor. Cuff material: Polyester 200D. Power: Battery or 100-240 VAC adapter. Connectivity: None. Compliance: IEC 80601-2-30, ISO 81060-2:2018. Form factor: Electronic main unit with LCD display and inflatable upper arm cuff.
Indications for Use
Indicated for home use for non-invasive measurement of diastolic and systolic blood pressure and pulse rate in adults with upper arm circumference 22 to 52 cm.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
April 8, 2025
Globalcare Medical Technology Co., Ltd
Valentina Sassi
Quality and Regulatory Affairs Manager
No.7 Factory, European Industrial Park,
No. 39 Mid Industrial Main Road Xiaolan Town, Zhongshan City
Guangdong, 528415
China
Re: K250136
Trade/Device Name: Globalcare Blood Pressure Monitor (GUS610)
Regulation Number: 21 CFR 870.1130
Regulation Name: Noninvasive blood pressure measurement system
Regulatory Class: Class II
Product Code: DXN
Dated: January 17, 2025
Received: March 11, 2025
Dear Valentina Sassi:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K250136 - Valentina Sassi
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K250136 - Valentina Sassi
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Robert T. Kazmierski -S
for
LCDR Stephen Browning
Assistant Director
DHT2A: Division of Cardiac
Electrophysiology, Diagnostics, and
Monitoring Devices
OHT2: Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K250136
Device Name
Globalcare Blood Pressure Monitor (GUS610)
Indications for Use (Describe)
The Globalcare GUS610 automatic Blood Pressure Monitor is indicated for home use for the non-invasive measurement of diastolic and systolic blood pressures and pulse rate of adults by means of an inflatable cuff which is wrapped around the upper arm. The cuff circumference is limited to 22 to 52 cm.
Type of Use (Select one or both, as applicable)
☐ Prescription Use (Part 21 CFR 801 Subpart D)
☑ Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}
| 510(k) #: K250136 | 510(k) Summary | Prepared on: 2025-03-11 |
| --- | --- | --- |
| Contact Details | | 21 CFR 807.92(a)(1) |
| Applicant Name | Globalcare Medical Technology Co., Ltd | |
| Applicant Address | No.7 Factory, European Industrial Park, No. 39 Mid Industrial Main Road Xiaolan Town, Zhongshan City Guangdong 528415 China | |
| Applicant Contact Telephone | +39 02 83451194 | |
| Applicant Contact | Ms. Valentina Sassi | |
| Applicant Contact Email | valentina.sassi@globalcare.com.hk | |
| Device Name | | 21 CFR 807.92(a)(2) |
| Device Trade Name | Globalcare Blood Pressure Monitor (GUS610) | |
| Common Name | Noninvasive blood pressure measurement system | |
| Classification Name | System, Measurement, Blood-Pressure, Non-Invasive | |
| Regulation Number | 870.1130 | |
| Product Code(s) | DXN, DXQ (CLASS 2) - BLOOD PRESSURE CUFF | |
| Legally Marketed Predicate Devices | | 21 CFR 807.92(a)(3) |
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
| K222160 | Globacare GUS610 Blood Pressure Monitor | DXN |
| Device Description Summary | | 21 CFR 807.92(a)(4) |
{5}
The Globalcare GUS610 Blood Pressure Monitor is an automatic upper arm blood pressure monitor intended for non-invasive measurement or monitoring of adults' arterial blood pressure and pulse rate.
The device is powered by batteries or by a mains-connected power adapter and is designed for home use.
The Globalcare GUS610 Blood Pressure Monitor is supplied complete with the following components:
- Blood pressure monitor main device - 1
- Upper arm cuff 22-42 cm - 1
- 1.5 V LR03 AAA batteries - 4
- Storage bag - 1
- Instructions for use - 1
- 100-240 VAC 50/60 Hz power supply (optional) - 1
None of the parts included with the device are supplied in a sterile condition.
The Globalcare GUS610 Blood Pressure Monitor main unit is an electronic unit incorporating a screen that displays results and other information relevant to device operation. The device is powered by batteries or by a mains-connected power adapter and is designed for home ('OTC') use.
The clinical outputs from the GUS610 Blood Pressure Monitor are:
- Systolic pressure (mm Hg)
- Diastolic pressure (mm Hg)
- Pulse rate (1/min)
- Cardiac arrhythmia (Irregular heartbeat (IHB)) symbol
- Level of risk
- Hemodynamic Rest Condition (HSD) indication
The GUS610 Upper Arm Cuff, with a circumference range of 22 cm to 42 cm, is manufactured by Globalcare China. The cuff is manufactured from gray color Polyester 200D (CAS n. 80595-68-2), supplied by Anhui Wuyixing Fabric and Weaving. This is identical to the cuff provided with the predicate device cited in K222160.
The purpose of this Special 510(k) is to add to the availability of an accessory for the GUS610, this being an optional Upper Arm Cuff with an increased circumference range of 42 to 52 cm. The testing undertaken for the predicate device in accordance with ISO 81060-2:2018, 'Non-invasive sphygmomanometers - Part 2: Clinical investigation of intermittent automated measurement type', has been repeated with the new larger cuff, with 88 subjects having upper arm circumferences between 42 and 52 cm screened, from which 259 effective analysis datasets were obtained from 87 subjects. Gender distribution met the requirements of the standard (M: 55.2% / F 44.8%), arm size distribution met the requirements of the standard, and blood pressure measurements met the requirements of the standard clause 5.2.4.1.2 for Criterion 1 'mean difference' and 'standard deviation' for SBP and DBP, and also for Criterion 2 for SBP and DBP 'standard. deviation'. The accuracy of the BP and pulse rate measurements also met the manufacturer's specifications for accuracy.
There were no adverse incidents during the investigation and no failed measurements. There were no protocol deviations or violations.
A third party test house (TUV SUD China) has assessed the continued compliance of the GUS610 subject device when used together with the new optional XL cuff in accordance with IEC 80601-2-30 and found that the combination continues to meet the requirements of this standard.
Intended Use/Indications for Use
21 CFR 807.92(a)(5)
{6}
The Globalcare GUS610 automatic Blood Pressure Monitor is indicated for home use for the non-invasive measurement of diastolic and systolic blood pressures and pulse rate of adults by means of an inflatable cuff which is wrapped around the upper arm. The cuff circumference is limited to 22 to 52 cm.
| Indications for Use Comparison | 21 CFR 807.92(a)(5) |
| --- | --- |
| The indications statement for the subject device differs from the predicate device only with regard to the labeled upper arm cuff maximum circumference. The intended use is unchanged. | |
| Technological Comparison | 21 CFR 807.92(a)(6) |
| The only difference between the subject device and the predicate device is the availability of an optional upper arm cuff with a 52 cm maximum circumference, which is 10 cm larger than that cleared with the predicate device.
The larger circumference upper arm cuff is manufactured from the same material as that provided with the predicate device. | |
| Non-Clinical and/or Clinical Tests Summary & Conclusions | 21 CFR 807.92(b) |
| The testing undertaken for the predicate device in accordance with ISO 81060-2:2018, 'Non-invasive sphygmomanometers - Part 2: Clinical investigation of intermittent automated measurement type', has been repeated with the new larger cuff, with 88 subjects having upper arm circumferences between 42 and 52 cm screened, from which 259 effective analysis datasets were obtained from 87 subjects. Gender distribution met the requirements of the standard (M: 55.2 % / F 44.8 %), arm size distribution met the requirements of the standard, and blood pressure measurements met the requirements of the standard clause 5.2.4.1.2 for Criterion 1 'mean difference' and 'standard deviation' for SBP and DBP, and also for Criterion 2 for SBP and DBP 'standard. deviation'. The accuracy of the BP and pulse rate measurements also met the manufacturer's specifications for accuracy. There were no adverse incidents during the investigation and no failed measurements. There were no protocol deviations or violations. | |
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.