← Product Code [DXN](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DXN) · K231984

# Clinical Automatic Blood Pressure Monitor (DBP-01P,DBP-01HP) (K231984)

_Shenzhen Hingmed Medical Instrument Co., Ltd. · DXN · Mar 26, 2024 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DXN/K231984

## Device Facts

- **Applicant:** Shenzhen Hingmed Medical Instrument Co., Ltd.
- **Product Code:** [DXN](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DXN.md)
- **Decision Date:** Mar 26, 2024
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1130
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

This device is a digital monitor intended for use in measuring blood pressure(SYS and DIA) and pulse rate ,and the physician reference the result to diagnose.

## Device Story

Clinical Automatic Blood Pressure Monitor (models DBP-01P, DBP-01HP) measures blood pressure and pulse rate using oscillometric method; device consists of main unit, power wire, and cuff (17-42cm). Operated by clinicians in hospitals/medical establishments; provides SYS/DIA blood pressure and pulse rate readings via LCD (DBP-01HP) or LED (DBP-01P) screen; includes printing capability. Device automates inflation via internal pump and rapid deflation via solenoid valve. Clinicians use displayed results to assist in patient diagnosis. Benefits include standardized, automated non-invasive blood pressure monitoring for adult populations.

## Clinical Evidence

Clinical validation study per ISO 81060-2:2018 conducted on 85 subjects (mean age 69.9, 43 men). Mean device-observer difference: 1.32±3.18 mmHg (SBP) and 0.78±3.15 mmHg (DBP). Criterion 1 (≤5±8 mmHg) and Criterion 2 (SDs 6.80 mmHg SBP, 6.89 mmHg DBP) were satisfied, confirming accuracy within acceptable limits.

## Technological Characteristics

Oscillometric measurement; automatic internal pump inflation; solenoid valve deflation. Power: mains. Display: LCD (DBP-01HP) or LED (DBP-01P). Patient contact materials: skin-contacting, limited duration (<24h). Standards: IEC 60601-1, IEC 80601-2-30 (electrical safety); IEC 60601-1-2 (EMC); ISO 10993-5, ISO 10993-10 (biocompatibility).

## Regulatory Identification

A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.

## Predicate Devices

- UA-1200BLE Ultraconnect Digital Blood Pressure ([K163683](/device/K163683.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath. The logo is simple and professional, and it is easily recognizable.

March 26, 2024

Shenzhen Hingmed Medical Instrument Co., Ltd. Huang Yongban, RA Manager 4th Floor, Zhonghangfeixiang Building, NO. 371, Guangshen Road, Baoan District Shenzhen. GuangDong 518102 China

Re: K231984

Trade/Device Name: Clinical Automatic Blood Pressure Monitor (DBP-01P,DBP-01HP) Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: February 20, 2024 Received: February 20, 2024

Dear Huang Yongban:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

{2}------------------------------------------------

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Stephen C. Browning -S

LCDR Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

#### Indications for Use

510(k) Number (if known) K231984

Device Name

Clinical Automatic Blood Pressure Monitor (DBP-01P, DBP-01HP)

Indications for Use (Describe)

This device is a digital monitor intended for use in measuring blood pressure(SYS and DIA) and pulse rate ,and the physician reference the result to diagnose.

Environments of use: Hospital and other medical establishment.

Patient population: Adult (exclude pregnant women ).

| Type of Use (Select one or both, as applicable)            |                                                           |
|------------------------------------------------------------|-----------------------------------------------------------|
| <span></span> Prescription Use (Part 21 CFR 801 Subpart D) | <span></span> Over-The-Counter Use (21 CFR 801 Subpart C) |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

# 510(k) Summary

This summary of 510(K) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR 807.92.

#### 1. Submitter's Information

510(k) Owner's Name: Shenzhen Hingmed Medical Instrument Co., Ltd Establishment Registration Number: Applying Address: 4th Floor, Zhonghangfeixiang Building, NO. 371, Guangshen Road, Baoan District, Shenzhen,,GuangDong , People's Republic of China Tel: +86-0755-232069446 Contact Person (including title): Yongban Huang (Manger Representative) E-mail: hyb@hingmed.com

## 2. Subject Device Information

Type of 510(k): Traditional

Common Name:Noninvasive blood pressure measurement system

Classification Name: System, Measurement, Blood-Pressure, Non-Invasive

Trade Name: Clinical Automatic Blood Pressure Monitor

Model Name: DBP-01P,DBP-01HP

Review Panel: Cardiovascular

Product Code: DXN

Regulation Number: 21 CFR 870.1130

Regulatory Class: Class II

#### 3. Predicate Device Information

| Sponsor            | A&D Company, Ltd                               |
|--------------------|------------------------------------------------|
| Device Name        | UA-1200BLE Ultraconnect Digital Blood Pressure |
| Model or type Name | UA-1200BLE                                     |
| 510(k) Number      | K163683                                        |
| Product code       | DXN                                            |

{5}------------------------------------------------

| Regulation Number | 21 CFR 870.1130 |
|-------------------|-----------------|
| Regulation class  | Class II        |

#### 4. Indications for Use

This device is a digital monitor intended for use in measuring blood pressure(SYS and DIA) and pulse rate ,and the physician reference the result to diagnose. Environments of use: Hospital and other medical establishment Patient population: Adult (exclude pregnant women )

#### Device Description 5.

Clinical Automatic blood pressure monitor have two models that are DBP-01HP and DBP-01P,The Clinical Automatic blood pressure monitor is used to measure the blood pressure of adult in hospital or other medical establishment. It's contain of main body , power wire.it can show the time and measure result, print the measure result .There is a difference between DBP-01HP with DBP -01P. DBP-01HP show the measure result and time by the LCD screen,DBP-01P show the measure result and time by the LED screen. Photos of DBP-01HP &DBP-01P:

Image /page/5/Picture/6 description: The image shows two different models of the Hingmed blood pressure monitor. The left image shows the DBP-01HP model, which has a screen and buttons on the base. The right image shows the DBP-01P model, which has a smaller screen and fewer buttons. The screen on the DBP-01P model shows a blood pressure reading of 125/84 mmHg and a pulse of 75 bpm.

## 6. Substantial Equivalence table

| Device      | Subject Device                                  | Predicate Device | Remarks |
|-------------|-------------------------------------------------|------------------|---------|
| Manufacture | Shenzhen Hingmed Medical<br>Instrument Co., Ltd | A&D Company, Ltd | NA      |
| Model       | DBP-01P,DBP-01HP                                | UA-1200BLE       | NA      |

{6}------------------------------------------------

| classification                                             | II                                                                                                                                                            | II                                                             | Same       |
|------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------|------------|
| Product code                                               | DXN                                                                                                                                                           | DXN                                                            | Same       |
| Regulation No.                                             | 870.1130                                                                                                                                                      | 870.1130                                                       | Same       |
| 510(K) number                                              | N/A                                                                                                                                                           | K163683                                                        | NA         |
| Intended use/ Indicate for use                             | This device is a digital monitor intended for use in measuring blood pressure(SYS and DIA) and pulse rate ,and the physician reference the result to diagnose | Measure blood pressure (systolic and diastolic) and pulse rate | Same       |
| Environmental of use                                       | Hospital and other medical establishment                                                                                                                      | Home or clinical use                                           | Difference |
| Patient population                                         | Adult (exclude pregnant women )                                                                                                                               | Adult                                                          | Same       |
| Measurement site                                           | Upper Arm                                                                                                                                                     | Upper Arm                                                      | Same       |
| Measurement method                                         | Oscillometric method                                                                                                                                          | Oscillometric method                                           | Same       |
| Measurement range                                          | Pressure:0 to 289<br>Pulse rate:40 to 200bpm                                                                                                                  | Pressure:0 to 299mmHg<br>Pulse rate:40 to 180beats per minute  | similar    |
| Measuring accuracy                                         | Pressure:within $\pm$ 3mmHg<br>Pulse Rate: whichever is grater( $\pm$ 3bpm or $\pm$ 3%)                                                                       | Pressure:within $\pm$ 3mmHg<br>Pulse Rate: $\pm$ 5%            | similar    |
| Cuff                                                       | 17-42cm                                                                                                                                                       | 22-42cm                                                        | similar    |
| Inflation                                                  | Automatic internal pump                                                                                                                                       | Automatic internal pump                                        | Same       |
| Deflation                                                  | Automatic rapid deflation                                                                                                                                     | Solenoid valve for rapid exhaust                               | Same       |
| Power source                                               | mains                                                                                                                                                         | 3.7V Li-ion batteries or AC adaptor as an option               | Difference |
| Protection against electrical shock                        | Class I, Type B                                                                                                                                               | Class I, Type B                                                | same       |
| Display                                                    | LCD display or LED display                                                                                                                                    | OLED Display                                                   | Difference |
| Operation Environment condition                            | 5°C-40°C; RH: 10%-95%; non-condensing,atmospheric pressure(70KPa-106Kpa)                                                                                      | 50°F (10°C) to 104°F (40°C)<br>15%RH to 85%RH                  | Difference |
| Storage Environment Condition                              | -20°C-55°C, RH≤90% (non-condensing) 70KPa-106Kpa                                                                                                              | -4°F (-20°C) to 140°F (60°C)<br>10%RH to 95%RH                 | Difference |
| Memory                                                     | NA                                                                                                                                                            | Last 60 measurements                                           | Difference |
|                                                            |                                                                                                                                                               | user                                                           |            |
| Weight                                                     | About 6Kg                                                                                                                                                     | 205g without batteries                                         | Difference |
| Dimension(L*D*H)                                           | About W:310×L:478×H:300mm                                                                                                                                     | 40(W)x126(H)x26(D) mm                                          | Difference |
| Patient<br>contact<br>materials                            | Surface contact Skin<br>Limited duration of use<24 hours                                                                                                      | Surface contact Skin<br>Limited duration of use<24 hours       | Same       |
| Blood<br>Pressure<br>Reading<br>Classification<br>Criteria | WHO Classification<br>Criteria                                                                                                                                | USA JNC VII                                                    | Difference |

{7}------------------------------------------------

#### 5. non-clinical Test Summary

Clinical Automatic blood pressure monitor has been evaluated the safety and performance by lab bench testing,as following:

Electrical safety test according to IEC 60601-1 and IEC 80601-2-30 standards Electromagnetic compatibility test according to IEC 60601-1-2 standard Biocompatibility test according to ISO 10993-5 and ISO 10993-10 standards

#### 6.Brief discussions of clinical tests

ISO 81060-2:2018 Non-invasive sphygmomanometers - Part 2: Clinical validation of automated measurement type;

In this clinical investigation, The DBP-01P was tested in 85 subjects from the general population (mean age, 69.9 years; 43 men) using a wide-range cuff for arm circumferences from 17 to 42 cm. The mean device-observer difference was 1.32±3.18 mmHg for SBP and 0.78±3.15 mmHg for DBP. These data were in agreement with criterion 1 of the protocol standard requirements (≤5±8 mmHg).

Also, criterion 2 was satisfied with the SDs of the 85 participants being well below the maximum values required by the protocol (6.80 and 6.89 mmHg for SBP and DBP pressure, respectively). The results showed the accuracy of the blood pressure monitor is within acceptable scope specified in ISO 81060-2

#### 7.Comparison to predicate device and conclusion

The technological characteristics, features, - specifications, materials, mode of and intended use of clinical automatic blood pressure monitor is operation. substantially equivalent to the predicate devices quoted above.

between the subject device and predicate devices do not rise The differences new issues of safety or effectiveness.

{8}------------------------------------------------

## 8.Conclusion

Non-clinical performance was conducted on the subject device and all tests met specified criteria. Based on the information provided in this submission, the Clinical Automatic blood pressure monitor is substantially equivalent to the predicate device, UA-1200BLE UltraConnect Digital Blood Pressure.

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DXN/K231984](https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DXN/K231984)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
