← Product Code [DXN](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DXN) · K201302

# LiveOne (K201302)

_Livemetric (Medical) S.A. · DXN · May 4, 2022 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DXN/K201302

## Device Facts

- **Applicant:** Livemetric (Medical) S.A.
- **Product Code:** [DXN](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DXN.md)
- **Decision Date:** May 4, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1130
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

LiveOne is a non-invasive blood pressure monitor, intended to capture the arterial waveform on the wrist and use it to as the basis of measurement of the systolic and diastolic blood pressure and pulse rate in the adult patient population. The blood pressure is thereby measured on the patient's wrist. The LiveOne provides intermittent blood pressure values to a medical professional. The LiveOne displays continuous radial arterial pressure waveform prior measurement to a medical professional. The Interpretation of the numerical values will be made by medical professionals only. The device does not report any diagnosis but provides numerical values. It is the physician's responsibility to make proper judgments based on these numbers and the tracing. The LiveOne is intended to be used for patients who have a need for a noninvasive blood pressure monitor and is intended for use in clinical settings only. It does not provide any alarms and is not intended to clinically monitor patients in operating rooms, intensive care units (ICU) and electrophysiology laboratories.

## Device Story

LiveOne LM1P is a non-invasive blood pressure monitor using applanation tonometry. A wristband with a high-resolution MEMS-based sensor array is placed over the radial artery to capture arterial pressure waveforms. An electronic interface module, connected to an off-the-shelf tablet, processes the signal. The device uses patient-specific data (weight, height, age, sex) in its algorithms to calculate systolic/diastolic blood pressure and heart rate. Used in clinical settings by healthcare professionals; provides intermittent numerical values and continuous waveform display. No alarms; not for use in OR, ICU, or EP labs. Healthcare providers interpret the displayed data to inform clinical decision-making. Benefits include non-invasive, cuffless, beat-to-beat monitoring of arterial pressure.

## Clinical Evidence

Two clinical studies conducted per ISO 81060-2:2018. 1) Multi-center ICU study (n=26): systolic bias 0.7±6.6 mmHg, diastolic bias 2.0±5.5 mmHg, HR relative distance 2.9. 2) General population study (n=8): systolic bias -1.3±7.2 mmHg, diastolic bias -0.4±5.7 mmHg, HR relative distance 1.36. Results met acceptance criteria for accuracy.

## Technological Characteristics

Applanation tonometry; high-resolution MEMS-based sensor array; 100Hz sampling frequency. Wrist-worn bracelet with off-the-shelf tablet companion. Auto-calibrating (no cuff). Complies with IEC 60601-1 (safety), IEC 60601-1-2 (EMC), and ISO 10993-1 (biocompatibility).

## Regulatory Identification

A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.

## Predicate Devices

- TL300 Tensymeter Noninvasive Blood Pressure Monitor ([K123446](/device/K123446.md))
- TL10 Tensymeter Non-invasive Blood Pressure Monitor ([K020537](/device/K020537.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

May 4, 2022

LiveMetric (Medical) S.A. Chen Botvin Senior Regulatory and Clinical Manager 40 Rue Glesener Luxembourg, L-1630 Luxembourg

Re: K201302

Trade/Device Name: LiveOne LM1P Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: April 22, 2022 Received: April 22, 2022

Dear Chen Botvin:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

LCDR Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics, and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K201302

Device Name LiveOne LM1P

Indications for Use (Describe)

The LiveOne is intended to be used in patients who have a need for a noninvasive blood pressure and hemodynamic monitor.

The LiveOne is intended to be used on subjects ≥ 27 years old and who have a palpable radial pulse. Software is used to provide data to qualified medical professionals for the purpose of assessing the patient's cardiac health via blood pressure readings in clinic setting by health care professional. LiveOne is intended only for measurement and display; it makes no diagnosis.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) SUMMARY LiveMetric's LiveOne LM1P

## Submitter Information:

| Submitter:            | LiveMetric (Medical) S.A.                             |  |
|-----------------------|-------------------------------------------------------|--|
|                       | 40 Rue Glesener,                                      |  |
|                       | Luxembourg L-1630                                     |  |
|                       | Luxembourg                                            |  |
| Contact Person:       | Tomer Bentzion                                        |  |
|                       | Chief Executive Officer                               |  |
| Phone:                | +1 650-285-7222                                       |  |
| eMail:                | tomer.bentzion@livemetric.com                         |  |
| Date Prepared:        | May 13, 2020                                          |  |
| Device Information:   |                                                       |  |
| Trade Name:           | LiveOne LM1P                                          |  |
| Common or Usual Name: | Noninvasive Blood Pressure Monitor                    |  |
| Classification Name:  | Noninvasive Blood Pressure Monitor Measurement System |  |
| Regulatory Class:     | Class II                                              |  |
| Product Code:         | 21CFR870.1130 DXN                                     |  |

## Predicate Devices:

The LiveOne LM1P is substantially equivalent in indications for use to the devices shown in table below, that have been cleared for marketing:

| Device Name                                            | Manufacturer         | 510(K)  |
|--------------------------------------------------------|----------------------|---------|
| TL300 Tensymeter Noninvasive<br>Blood Pressure Monitor | Tensys Medical, Inc. | K123446 |
| TL10 Tensymeter Non-invasive<br>Blood Pressure Monitor | Tensys Medical, Inc. | K020537 |

{4}------------------------------------------------

#### Device Description

The LiveOne is a non-invasive blood pressure monitor that utilizes a single patient use non-invasive pressure sensor placed on the wrist over the radial artery, and an electronic interface module. The device uses proprietary algorithms to analyze the radial artery pressure and display diastolic, systolic, and pulse rate and a pressure waveform. The device allows the healthcare provider to monitor blood pressure (based on visual waveform) in a clinical setting.

#### Intended Use

LiveOne is a non-invasive blood pressure monitor, intended to capture the arterial waveform on the wrist and use it to as the basis of measurement of the systolic and diastolic blood pressure and pulse rate in the adult patient population. The blood pressure is thereby measured on the patient's wrist.

The LiveOne provides intermittent blood pressure values to a medical professional. The LiveOne displays continuous radial arterial pressure waveform prior measurement to a medical professional.

The Interpretation of the numerical values will be made by medical professionals only. The device does not report any diagnosis but provides numerical values. It is the physician's responsibility to make proper judgments based on these numbers and the tracing.

The LiveOne is intended to be used for patients who have a need for a noninvasive blood pressure monitor and is intended for use in clinical settings only. It does not provide any alarms and is not intended to clinically monitor patients in operating rooms, intensive care units (ICU) and electrophysiology laboratories.

## Indications for Use

The LiveOne is intended to be used in patients who have a need for a non-invasive blood pressure and hemodynamic monitor.

The LiveOne is intended to be used on subjects ≥ 27 years old and who have a palpable radial pulse. Software is used to provide data to qualified medical professionals for the purpose of assessing the patient's cardiac health via blood pressure readings taken in a clinical setting by health care professional . LiveOne is intended only for measurement and display; it makes no diagnosis.

{5}------------------------------------------------

#### Summary of Technological Characteristics

The intended use of LiveOne LM1P has the same intended use as its predicate device(s) TL300 (K123446) and its predicate TL10 (K020537). The technological characteristics of LiveOne LM1P are substantially similar to the technological characteristics of the predicate device(s), TL300 (K123446) and TL10 (K020537). Subject device and predicate device(s) are considered NIBP monitors; both subject device and predicate device(s) are based on applanation tonometry; they both apply the physical principle of applanation tonometry on the radial artery with the use of pressure sensor data. The reading from the sensor are transformed to a pressure waveform, from which, the values of systolic and diastolic pressure are calculated. Both subject device and predicate device(s) are auto-calibrating; do not require any external calibration using a cuff. Both devices use the radial artery site as the primary pressure source for the system, whereas, TL300 utilizes a single sensor on its wrist-worn bracelet, LiveOne LM1P also operates under this same underlying principle, but it reads and processes the pressure waveform using a high-resolution, MEMS-based sensor array. The predicate device(s) utilizes an algorithm to define an optimum pressure sensing position over the radial artery. The subject device uses a placement signal score module that ensures high fidelity, pressure waveform signal, sampled at a frequency of 100Hz, and extracts the systole, diastole and heart rate to be displayed on the device's dedicated monitor. Any differences in the technological characteristics between the LiveOne LM1P and the predicate device do not raise any new issues of safety or effectiveness. Therefore, the LiveOne LM1P is substantially equivalent to the predicate device(s).

| Functionality                    | LiveOne LM1P                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | TL10 Tensymeter Non-<br>invasive Blood Pressure<br>Monitor<br>K020537                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | TL300 Tensymeter<br>Noninvasive Blood<br>Pressure Monitor<br>K123446                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
|----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Regulatory<br>compliance         | ISO 81060-2:2016 (The<br>successor of AAMI<br>SP10:1992; AAMI<br>SP10:2002; IEC 81060-<br>2:2009 standard; ISO<br>81060-2:2013)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | AAMI Standard SP-10 -<br>2002                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | IEC 81060-2:2009<br>The accuracy of the device<br>will be compared to the<br>standard radial artery<br>catheter, as well as to the<br>non-invasive blood pressure<br>cuff.                                                                                                                                                                                                                                                                                                                                              |
| Product Code<br>Intended Use     | DXN<br>LiveOne is a non-<br>invasive blood pressure<br>monitor. The blood                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | DXN<br>This device is intended<br>for use by medically<br>trained personnel in a                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | DXN<br>TL-300 is a non-invasive<br>hemodynamic monitor.                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|                                  | pressure waveform is<br>measured on the<br>patient's wrist.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | clinical setting to<br>continuously monitor<br>and display diastolic,<br>systolic, and mean blood<br>pressures.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
|                                  | Provides intermittent<br>blood pressure values to<br>a medical professional.<br>The LiveOne displays<br>continuous radial<br>arterial pressure<br>waveform prior<br>measurement to a<br>medical professional.<br>Software is used to<br>provide data to qualified<br>medical professionals<br>for the purpose of<br>assessing the patient's<br>cardiac health via blood<br>pressure readings in a<br>clinical setting.<br>The LiveOne is<br>intended to be used on<br>subjects≥27 years old<br>and who have a<br>palpable radial pulse.                                                                                                                                                                                                                                                                                                                                                                   | This device is intended<br>for use by medically<br>trained personnel in a<br>clinical setting to<br>continuously monitor<br>and display diastolic,<br>systolic, and mean blood<br>pressures.<br>The device is intended<br>for use on adult patients<br>with a palpable pulse.<br>This device is intended<br>for use by medically<br>trained personnel in a<br>clinical setting to<br>continually monitor and<br>display diastolic,<br>systolic, and mean blood                                                                                                                                                          | TL-300 does<br>continuous beat-to-beat<br>waveform and blood<br>pressure monitoring.                                                                                                                                                                                                                                                                                                                                                                                                                                    |
|                                  | The Interpretation of the<br>numerical values will<br>be made by medical<br>professionals only.<br>The device does not<br>report any diagnosis but<br>provides numerical<br>values. It is the<br>physician's<br>responsibility to make<br>proper judgments based<br>on these numbers.<br>LiveOne is intended<br>only for measurement<br>and display; it makes no<br>diagnosis.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | pressures and pulse rate.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Technological<br>Characteristics | LiveOne is placed on<br>the wrist over the area<br>of the radial artery pulse<br>while the sensor<br>interface acts as a<br>pressure transducer.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | The design of the TL10<br>Tensymeter utilizes a<br>semiconductor pressure-<br>sensing element applied<br>to the wrist, over the<br>radial artery, to obtain a                                                                                                                                                                                                                                                                                                                                                                                                                                                           | The T-line device will be<br>placed over the radial artery<br>at the distal wrist.                                                                                                                                                                                                                                                                                                                                                                                                                                      |
|                                  |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
|                                  | LiveOne LM1P also<br>operates under this<br>same underlying<br>principle (Tonometry),<br>but it reads and<br>processes the pressure<br>waveform at the radial<br>artery by the use of a<br>high-resolution,<br>MEMS-based sensor<br>array<br>The LiveOne wristband<br>is similar in size and<br>weight to a sports<br>wristwatch.<br>LiveOne is monitoring<br>at the wrist using<br>arterial tonometry<br>An embedded<br>microcontroller<br>supervises the actions of<br>the monitor<br>Utilizes a<br>semiconductor pressure-<br>sensing element/ array<br>of elements applied to<br>the wrist, over the radial<br>artery, to obtain a<br>pressure waveform<br>The LiveOne LM1P<br>uses weight, height, age<br>and sex in its signal<br>processing and pressure<br>measurement<br>algorithms.<br>LiveOne LM1P is<br>comprised of a<br>wristband and an off-<br>the-shelf tablet as a<br>companion device. | pressure waveform.<br>The TL10 Tensymeter is<br>a non-invasive blood<br>pressure monitor that<br>utilizes a single patient<br>use non-invasive<br>pressure sensor placed on<br>the wrist, over the radial<br>artery, and an electronic<br>interface module.<br><br>The monitor has a size<br>and weight similar to a<br>sports type wrist watch.<br><br>Blood pressure<br>monitoring at the wrist<br>using arterial tonometry.<br>An embedded<br>microcontroller<br>supervises the actions of<br>the monitor.<br>The monitor is applied to<br>the wrist with a pressure<br>transducer placed over<br>the radial artery. | To attach the device, a<br>single-use sterile sensor is<br>placed over the radial artery<br>and the device then gently<br>clamped around the wrist to<br>incorporate the sensor.<br><br><br><br><br><br><br><br><br><br>The T-Line uses Body Mass<br>Index (BMI, computed using<br>height and weight) in its<br>signal processing and<br>pressure measurement<br>algorithms.<br>The TL-300 T-Line system<br>comprises a sensor, bracelet,<br>wrist positioner, and a tablet<br>monitor to form an<br>integrated system. |
| Non-Clinical<br>Tests Submitted  | ANSI/AAMI ES60601-<br>1:2005 / A2:2010<br>Medical electrical<br>equipment - Part 1:<br>General requirements<br>for basic safety and<br>essential performance<br>IEC 60601-1<br>Medical electrical<br>equipment - Part 1:<br>General requirements<br>for basic safety and<br>essential performance<br>IEC 60601-1-2                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | ANSI/AAMI ES60601-<br>1:2005 / A2:2010<br><br>IEC 60601-1<br><br>IEC 60601-1-2                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
|                                  | Medical electrical<br>equipment – Part 1-2:<br>General requirements<br>for safety –<br>Electromagnetic<br>compatibility –<br>Requirements and tests<br>ISO 10993-1<br>Biological evaluation of<br>medical devices – Part<br>1: Evaluation and<br>testing                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | ISO 10993-1                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Target<br>Population             | The LiveOne targets<br>patients, over 27 years<br>of age, who have a need<br>for non-invasive blood<br>pressure monitoring                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | The device is intended<br>for use on adult patients<br>with a palpable pulse.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Anatomical Site                  | LiveOne is placed on<br>the wrist over the area<br>of the radial artery pulse                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | The monitor is placed on<br>the wrist over the area of<br>the radial artery pulse                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| End User                         | The LiveOne is<br>intended to be used for<br>patients who have a<br>need for a noninvasive<br>blood pressure monitor<br>and is intended for use<br>in clinical settings only.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | This device is intended<br>for use by medically<br>trained personnel in a<br>clinical setting to<br>continuously monitor<br>and display diastolic,<br>systolic, and mean blood<br>pressures.                                                                                                                                                                                                                                                                                                                                                                                                                            |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |

{6}------------------------------------------------

{7}------------------------------------------------

{8}------------------------------------------------

#### Performance data

#### Non-clinical testing

#### Biocompatibility testing

The biocompatibility evaluation for the LiveOne device was conducted in accordance with the FDA Blue Book Memorandum #G95-1 "Use of International Standard ISO-10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing, '' May 1, 1995, and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices – Part 1 : Evaluation and Testing Within a Risk Management Process," as recognized by FDA. The battery of testing included the following tests:

- · Cytotoxicity
- Sensitization
- Irritation
- Systemic toxicity

{9}------------------------------------------------

The LiveOne device is considered tissue contacting for a duration of more than 24 hours.

#### Electrical safety and electromagnetic compatibility (EMC)

Electrical safety and EMC testing were conducted on the LiveOne device, consisting of the LiveOne wristband and the Microsoft Surface tablet. The system complies with the IEC 60601-1 standard for safety and the IEC 60601-1-2 standard for EMC.

## Software Verification and Validation Testing

Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "major" level of concern, since a failure or latent flaw in the software could directly result in serious injury or death to the patient or operator.

#### Clinical testing

## Multi-center ICU accuracy study:

This was a multi-center, non-randomized, single arm study of 26 subjects. The study was conducted in accordance with ISO 81060-2:2018 standard. This LiveOne device accuracy was compared to arterial line in ICU patients.

## Primary endpoint

The results of the Subject Device compared to the A-line measurement in the ICU accuracy validation data set were as follows: (1) mean bias of systolic BP: 0.7+6.6, (2) mean diastolic bias: 2.0 ±5.5 mmHg, and (3) HR relative distance: 2.9. Those results are within the acceptance criteria for the accuracy of blood pressure monitors. The study demonstrated that the Subject Device is accurate, and the study population and results comply with the ISO standard requirements.

## General population accuracy study:

{10}------------------------------------------------

This was a single-center, non-randomized, single arm study of 8 subjects. The study was conducted in accordance with ISO 81060-2:2018 standard. This LiveOne device accuracy was compared to arterial line in healthy subjects.

#### Primary endpoint

The results of the Subject Device compared to the A-line measurement in the general population validation data set were as follows: (1) mean bias for systolic BP: -1.3 ± 7.2 mmHg(2) mean bias of diastolic BP: -0.4 ± 5.7 mmHg(3) HR relative distance 1.36. Those results are within the acceptance criteria for the accuracy of blood pressure monitors. The study demonstrated that the Subject Device is accurate, and the study population and results comply with the ISO standard requirements.

#### Conclusion

The LiveOne LM1P is as safe and as effective as the TL300 and its predicate TL10. The LiveOne LM1P has the same intended uses and similar indications, as its predicate device. The minor differences in indications do not alter the intended use. The LiveOne LM1P is as safe and effective as the Tensymeter's TL300 and the TL10. The LiveOne LM1P has the same technological characteristics, and principles of operation as its predicate device(s). In addition, the minor technological differences between the LiveOne LM1P and its predicate device(s) raise no new issues of safety or effectiveness. The nonclinical data support the safety of the device and the hardware and software verification and validation demonstrate that the LiveOne LM1P device should perform as intended in the specified use conditions. The clinical testing demonstrated that the LiveOne LM1P device is accurate according to the applicable ISO 81060-2-2018 standard. Thus, the LiveOne LM1P is substantially equivalent to the predicate.

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DXN/K201302](https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DXN/K201302)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
