K191745 · Medtronic Care Management Services · DXN · May 15, 2020 · Cardiovascular
Device Facts
Record ID
K191745
Device Name
Medtronic Commander Flex
Applicant
Medtronic Care Management Services
Product Code
DXN · Cardiovascular
Decision Date
May 15, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1130
Device Class
Class 2
Indications for Use
The Commander Flex is for use by patients to collect and transmit general health questions and patient vital sign data (such as weight, blood pressure, glucose, pulse oximetry) between the patient, typically at home, and a health care professional at a remote site.
Device Story
Commander Flex is a home-use remote patient monitoring device for patients with chronic conditions (CHF, diabetes, COPD, asthma, CAD). It collects vital sign data (weight, blood pressure, glucose, pulse oximetry) via peripheral devices (scales, meters, cuffs) and subjective health survey responses. The device features an integrated non-invasive blood pressure measurement system. Data is transmitted via cellular network or telephone line to a remote healthcare professional. The device does not provide medical interpretations, evaluations, or treatment recommendations; it serves as a data collection and transmission tool for clinical review. Healthcare providers use the transmitted data to monitor patient status and intervene before symptoms worsen. The device is operated by the patient at home. It benefits patients by facilitating remote monitoring and timely clinical oversight.
Clinical Evidence
Clinical testing of the blood pressure 2.0 algorithm was conducted on 99 human subjects (41% male) using small, medium, and large cuffs. Accuracy was validated by comparing device measurements against a trained observer using the stethoscope auscultatory method. Results confirmed the algorithm meets ANSI/AAMI SP10 and ISO 81060-2 requirements for sphygmomanometers.
Technological Characteristics
Electronic device with embedded firmware; integrated non-invasive blood pressure measurement system. Interfaces with peripherals via RS232, USB 2.0, and RF. Connectivity via cellular or POTS. Security: FIPS 140-2 Level 2 encryption. Power: 120V AC to DC external supply. Tested to IEC 60601-1, IEC 60601-1-2:2014 (EMC), and ISO 80601-2-30. User interface: LCD display, physical keypad, voice prompts.
Indications for Use
Indicated for patients, typically at home, with conditions such as congestive heart failure (CHF), diabetes, chronic obstructive pulmonary disease (COPD), asthma, and coronary artery disease (CAD) to collect and transmit vital signs and health survey data to remote healthcare professionals.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
K091821 — COMMANDER 111, MODEL CD300-CD399 · Cardiocom, LLC · Aug 25, 2009
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath.
May 15, 2020
Medtronic Care Management Services % Charmaine Sutton Consultant The Tamarack Group - MPLS, LLC 2584 Upton Ave. So Minneapolis, Minnesota 55405
Re: K191745
Trade/Device Name: Commander Flex Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN, DQA, DRG Dated: April 10, 2020 Received: April 14, 2020
Dear Charmaine Sutton:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
LT Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K191745
Device Name
Commander Flex
Indications for Use (Describe)
The Commander Flex is for use by patients to collect and transmit general health questions and patient vital sign data (such as weight, blood pressure, glucose, pulse oximetry) between the patient, typically at home, and a health care professional at a remote site.
Type of Use (Select one or both, as applicable)
| <div> <span> <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> |
|--------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="text-decoration: overline;">X</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510k Summary:
### Medtronic Commander Flex Series remote patient monitoring device
#### Date Prepared: May 11, 2020
| Submitter | |
|------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Applicant: | Medtronic Care Management Services<br>7980 Century Boulevard<br>Chanhassen, MN 55317<br>888-243-8881 |
| Contact: | Charmaine Dwyer (Sutton)<br>The Tamarack Group – MPLS, LLC<br>charmaine.sutton@tgmpls.com |
| Device Information | |
| Trade Name:<br>Model Numbers:<br>Common Name:<br>Classification: | Medtronic Commander Flex<br>CD300, CD300C, CD310, CD315, CD320<br>Remote Patient Monitoring Device<br>Class II<br>Noninvasive Blood Pressure Measurement System -<br>21 CFR 870.1130<br>Oximeter – 21 CFR §870.2700<br>Product Codes DXN, DQA, DRG |
### Predicate Device(s)
| | Primary Predicate | Secondary Predicate |
|-------------------|--------------------------------------|-------------------------------------------------------|
| Identification | Commander III<br>Model CD3xx<br>(Rx) | Tablet Commander<br>Model TC200/TC210<br>(Rx and OTC) |
| Manufacturer | Cardiocom, LLC<br>(now Medtronic) | Cardiocom LLC<br>(now Medtronic) |
| 510k number | K053303, K091821 | K122285 |
| Regulation number | 21 CFR 870.1130<br>21 CFR 870.2700 | 21 CFR 870.2910 |
| ProCode | DXN, DQA | DRG |
# Device Description
Commander Flex devices are home health monitoring tools typically used by patients with diseases such as congestive heart failure (CHF), diabetes, chronic obstructive pulmonary disease (COPD) asthma and coronary artery disease (CAD). Patients use Commander Flex to collect and send important health information to their health care team to allow a doctor and other health care team members to act before symptoms worsen.
{4}------------------------------------------------
Commander Flex devices have a built-in non-invasive blood pressure measurement system that measures systolic and diastolic blood pressure and displays the readings on the Commander Flex panel.
In addition to making blood pressure and heart rate measurements, Commander Flex devices store and transmit data from peripheral devices to remote locations. Commander Flex connects to a user's home phone line or cellular network. It has inputs for peripheral devices such as weight scales, glucose meters, pulse oximeters, thermometers and blood pressure cuff. The display asks health-related questions to obtain objective biometric data such as: weight, blood pressure, glucose, and pulse oximetry. Patients may also answer a series of questions about their current symptoms. The patient can respond Yes or No or select from a list. Commander Flex devices collect data from peripherals and transmit the data using a cellular connection or toll-free telephone number and modelar design of the Commander Flex device allows for selection of the appropriate health monitoring devices for cost-effective patient care.
Commander Flex products are for home use. This 510(k) clearance achieves OTC designation for Commander Flex devices.
The peripheral devices are used as supplied by the manufacturer. The Commander Flex devices make no interpretation, evaluation, medical judgements or recommendations for treatment. Commander Flex is not used as an emergency response device. Clinical judgment and experience are required to interpret the information transmitted. Commander Flex devices are not intended as a substitute for medical care.
### Intended Use
The Commander Flex is for use by patients to collect and transmit general health questions and patient vital sign data (such as weight, blood pressure, glucose, pulse oximetry) between the patient, typically at home, and a health care professional at a remote site.
### Comparison to Predicate Devices
| Feature | Subject Device<br>Commander Flex<br>Model CD300 -CD399<br>with OTC designation | Primary Predicate<br>Commander III, Model<br>CD300-CD399<br>K091821 & K053303 | Secondary Predicate<br>Tablet Commander<br>Model TC200 -TC299<br>K122285 | Comparison |
|---------------------------------------|------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------|-----------------|
| 510(k) Number | | K091821 & K053303 | K122285 | -- |
| Product Code &<br>Classification Name | DXN<br>System, Measurement,<br>Blood Pressure, Non-<br>Invasive<br>DQA<br>Oximeter | DXN<br>System, Measurement,<br>Blood Pressure, Non-<br>Invasive<br>DQA<br>Oximeter | DRG<br>Transmitters and<br>Receivers, Physiological<br>Signal, Radiofrequency | Same as primary |
| Class | Class II | Class II | Class II | Same |
| Regulation | 21 CFR 870.1130<br>Non-Invasive Blood<br>Pressure Measurement<br>System | 21 CFR 870.1130<br>Non-Invasive Blood<br>Pressure Measurement<br>System | 21 CFR 870.2910<br>Radiofrequency<br>Physiological Signal<br>Transmitter and Receiver | Same as primary |
Regulatory and Intended Use Comparison
{5}------------------------------------------------
| Feature | Subject Device<br>Commander Flex<br>Model CD300 -CD399<br>with OTC designation | Primary Predicate<br>Commander III, Model<br>CD300-CD399<br>K091821 & K053303 | Secondary Predicate<br>Tablet Commander<br>Model TC200 -TC299<br>K122285 | Comparison |
|-----------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | 21 CFR 870.2700<br>Oximeter | 21 CFR 870.2700<br>Oximeter | | |
| Intended Users | All three devices are used by home users and health care professionals | | | Same |
| Location of Use | All three devices are used in the home | | | Same |
| Rx/OTC | Rx & OTC | Rx | Rx & OTC | Same as secondary;<br>same Rx as Primary |
| Intended Use | The Commander Flex is<br>for use by patients to<br>collect and transmit<br>general health<br>questions and patient<br>vital sign data (such as<br>weight, blood pressure,<br>glucose, pulse<br>oximetry) between the<br>patient, typically at<br>home, and a health<br>care professional at a<br>remote site. | The Commander III is<br>for use by patients to<br>collect and transmit<br>general health<br>questions and patient<br>vital sign data (such as<br>weight, blood pressure,<br>glucose, pulse<br>oximetry, peak flow)<br>between the patient,<br>typically at home, and<br>a health care<br>professional at a<br>remote site. | The Tablet Commander<br>device is for use by<br>patients to collect and<br>transmit general health<br>information, physiological<br>measurements and other<br>data between themselves<br>and a caregiver. | Same; all are for use<br>by patients to collect<br>and transmit general<br>health information<br>between themselves<br>and a healthcare<br>professional at a<br>remote monitoring<br>station |
| | All three devices make no interpretation, evaluation, medical judgments, or<br>recommendations for treatment. Clinical judgment and experience are<br>required to check and interpret the information transmitted, and they are<br>not intended as a substitute for medical care. | | | Same |
# Technological and performance characteristics comparison
| Feature | Subject Device<br>Commander Flex<br>Model CD300 -CD399<br>Series with OTC<br>designation | Primary Predicate<br>Commander III, Model<br>CD300-CD399 Series<br>K091821 & K053033 | Secondary Predicate<br>Tablet Commander<br>Model TC200 -TC299<br>series<br>K122285 | Comparison |
|---------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------|
| Basic Technology<br>Description | The Commander Flex is an<br>electronic device with built-<br>in proprietary software.<br>The Commander Flex also<br>includes an integrated<br>blood pressure<br>measurement capability. | The Commander III is an<br>electronic device with<br>built-in proprietary<br>software.<br>The Commander III also<br>includes an integrated<br>blood pressure<br>measurement capability. | The Tablet Commander is<br>a software application<br>capable of running on any<br>hardware platform that<br>uses the Android<br>operating system. | Same as primary;<br>similar to<br>secondary |
| | All three devices interface with other electronic devices, including 510(k)-cleared or<br>Class I exempt medical devices, using standard wired and wireless protocols to<br>collect biometric data and communicate that data to a back-end database<br>application using public communications (phone, cellular or Wi-Fi). All three use the<br>same proprietary software to query and collect patient data. | | | Same |
| Feature | Subject Device<br>Commander Flex<br>Model CD300 -CD399<br>Series with OTC<br>designation | Primary Predicate<br>Commander III, Model<br>CD300-CD399 Series<br>K091821 & K053033 | Secondary Predicate<br>Tablet Commander<br>Model TC200 -TC299<br>series<br>K122285 | Comparison |
| Device Software | Embedded firmware and<br>patient database | Embedded firmware and<br>patient database | Embedded software and<br>patient database | Same as primary |
| | The software hazards for all three are: no data from peripheral, no data transmission<br>to remote monitoring site, and corrupt data at remote monitoring site. The related<br>risk is low because the patient data is used for trending and not to make immediate<br>treatment decisions. | Same | | |
| | Transmission /<br>communication<br>method | Cellular (network<br>connectivity: CDMA or<br>GPRS) or telephone modem<br>(POTS) | Telephone modem<br>(POTS) | Cellular (network<br>connectivity: CDMA or<br>GPRS) |
| Security & Privacy | FIPS 140-2 Level 2 and<br>native 4G/LTE encryption<br>for cellular transmissions<br>FIPS 140-2 Level 2<br>encryption for POTS (CD320<br>model) | Plain old telephone<br>system (POTS) | FIPS Level 140-2 Level 2<br>encryption for WIFI<br>transmissions | Subject device<br>adds FIPS 140-2<br>encryption for<br>cellular and POTS<br>transmissions |
| Peripheral device<br>interfaces | RS232 serial ports<br>USB: 2.0<br>Radio frequency<br>Supports battery charging<br>of MCMS-approved<br>peripherals via USB<br>connection | RS232 serial ports | WIFI: 802.11 b/n/g<br>USB: 2.0<br>Bluetooth: 3.0<br>Radio frequency<br>Supports battery charging<br>of MCMS-approved<br>peripherals via USB<br>connection | Same as<br>predicates,<br>collectively |
| | All three devices interface with the peripheral medical devices using standard<br>connections according to protocols established by the manufacturer of the<br>peripheral device. The protocols include checksums with re-transmit and purge<br>loops for data validation. The list of compatible peripherals will change according to<br>market demand, with each new peripheral being validated prior to making claims of<br>compatibility. | Same | | |
| Built-in<br>measurements | Non-invasive blood<br>pressure reading using<br>small, medium and large<br>size cuffs | Non-invasive blood<br>pressure reading using<br>small, medium and large<br>size cuffs | None | Same as primary |
| Power supply | All three use an external power supply that converts standard wall 120 volts AC to<br>DC | | | Same |
| Electrical Safety | All three devices are tested to IEC 60601-1,2. The subject device is tested to the<br>current version (2014). | | | Same |
| Communications<br>compliance | All three devices comply with FCC Part 68. | | | Same |
| User Interface /<br>Controls | All three communicate with patients using an on-screen LCD display in English,<br>Spanish or both, enlarged font size, voice prompts, user input buttons or keys, quick | | | Same |
| Feature | Subject Device<br>Commander Flex<br>Model CD300 -CD399<br>Series with OTC<br>designation | Primary Predicate<br>Commander III, Model<br>CD300-CD399 Series<br>K091821 & K053033 | Secondary Predicate<br>Tablet Commander<br>Model TC200 -TC299<br>series<br>K122285 | Comparison |
| | Start guide, user manuals, voice prompts, and customer support phone line 800 number. | | | |
| | physical keypad | physical keypad | virtual keypad /<br>touchscreen | Same as primary |
| Cleanability | All three are designed to be cleaned using common household cleaning agents. | | | Same |
| Memory | Biometric data collection: all three devices can display past readings. Biometric<br>review includes weight, blood pressure, glucose, oxygen saturation (and peak<br>flow for the predicate device)<br><br>Health check/subjective data: for all three devices only a single health check is<br>stored in memory at a given time. If a new health check is taken the new answers<br>will overwrite the previous answers. Answers are also removed from memory if the<br>device successfully transmits a health check to the remote monitoring system.<br><br>All three devices can read an external memory card. | | | Nearly identical;<br>the subject<br>device no longer<br>supports a peak<br>flow meter |
{6}------------------------------------------------
{7}------------------------------------------------
# Performance Data
Commander Flex devices are already used and labeled for use in the home, therefore no different or increased risks were identified by the OTC designation. Accordingly, no design changes, including no hardware, software or labeling changes were needed to support the OTC designation.
### Electrical safety and electromagnetic compatibility (EMC)
Electrical safety and electromagnetic compatibility (EMC) Electrical safety and EMC testing conducted on the Commander device showed the device complies with IEC 60601-1-2:2014 Ed.4 Electromagnetic Emissions and Immunity (EMC/EMI) and 80601-2-30 Medical Electrical Equipment - Particular requirements for basic safety and essential performance of automated non-invasive sphygmomanometers.
### Software verification and validation testing
The software for this device is considered a 'moderate' level of concern, since a failure or latent flaw in the software that controls the blood pressure circuit could directly result in minor injury to the patient or indirectly result in minor injury to the patient through incorrect or delayed information or through the action of a care provider. Documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices," including software risk assessments, software verification testing. A cybersecurity threat assessment report was also provided.
### Biomaterial safety information
The patient contact materials in the blood pressure cuff are identical to those in the predicate device, therefore no new biomaterial testing was needed to support substantial equivalence.
#### K191745
{8}------------------------------------------------
#### Human factors engineering information
Reports of usability engineering evaluations of Commander Flex devices, including use error analysis, user interface design analysis, and summative validation were provided. Usability test results show the user interface and user manual are sufficient for lay users to operate the device in a safe and effective manner.
A Use-Related Risk Assessment (URRA) was performed for Commander Flex devices to identify known or foreseeable user actions and identify risk-related use scenarios related to OTC use. The URRA included user profiles, task analysis, use error analysis, user interface design, and summative justification in accordance with FDA 2016 Guidance Applying Human Factors and Usability Engineering to Medical Devices. Commander Flex devices have been home-use devices for over a decade (under an Rx designation). The risk assessment revealed no different or increased risks and no new or additional critical tasks associated with the OTC designation. Therefore, additional Human Factors Validation Testing was deemed not necessary to support the Commander Series OTC designation.
#### Clinical data
Clinical testing of the Commander Flex blood pressure 2.0 algorithm in 99 human subjects (41% male; small, medium and large cuffs) was performed to measure the accuracy of the blood pressure measurement. Simultaneous blood pressure measurements were made by the Commander device and a trained observer using the stethoscope auscultatory method. The results demonstrate that the Commander blood pressure algorithm performance meets ANSI/AAMI SP10 and ISO 81060-2 requirements for sphygmomanometers.
#### Summary
Commander devices have always been used in the home. The updated performance evaluations revealed no new information about the safety and effectiveness profile of Commander. Device labeling was updated to reflect current standards. Commander devices with the OTC designation have a safety and effectiveness profile identical to the predicate device.
#### Conclusion
Commander Flex with the OTC designation has the same regulatory classifications, substantively the same intended use, the same environment of use, the same patient population and the same fundamental technology as the previously cleared Commander III devices. Commander Flex devices have no performance differences when compared to the predicate devices collectively and have been shown to meet all applicable industry standards. There are no new questions of safety or performance compared to the predicate devices. The information supplied demonstrates Commander Flex devices with the OTC designation have substantively the same intended use and the same technological characteristics as the previously cleared Commander III and Tablet Commander devices. Commander Flex devices with the OTC designation are, therefore, substantially equivalent to the legally marketed predicate devices.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.