← Product Code [DXN](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DXN) · K190893

# Non-invasive Hemodynamic Blood Pressure Monitor (K190893)

_Vita-Course Technologies Co., Ltd. · DXN · Aug 7, 2019 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DXN/K190893

## Device Facts

- **Applicant:** Vita-Course Technologies Co., Ltd.
- **Product Code:** [DXN](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DXN.md)
- **Decision Date:** Aug 7, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1130
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Intended Use

The Discovery 1 - Non-invasive Hemodynamic Blood Pressure Monitor device is a small, lightweight, handheld, device intended for measuring and display of Blood Pressure trending (systolic and diastolic) and spot-check of Peripheral pulse rate (PPR) and Peripheral pulse wave (PPW). Measurement is performing on capillary finger tip tissue (other than the thumb). The ring finger is the recommended site. The results of each measurement are stored in the system memory. The device is intended for use in the home environment. It is intended to be used by any person aged above 18 years old.

## Device Story

Portable, handheld, non-invasive hemodynamic blood pressure monitor; powered by internal rechargeable lithium polymer battery. Device uses pulse wave probe with two LEDs to collect PPG signals from capillary fingertip tissue. Principle of operation involves calculating blood pressure via pulse wave conduction time and pulse wave characteristics; systolic/diastolic pressures derived through fitting operation; pulse rate calculated from pulse wave signal. Intended for home use by adults 18+; results stored in memory. Provides users with blood pressure trends and spot-check pulse data to assist in monitoring hemodynamic status.

## Clinical Evidence

Bench testing only. Compliance with IEC 60601-1 (safety), IEC 60601-1-2 (EMC), and ISO 81060-2:2013 (clinical validation of automated measurement) demonstrated.

## Technological Characteristics

Handheld device; internal 3.7V 2200mAh Li-Pol battery; DC 5V adapter. Sensing via two-LED pulse wave probe. Compliant with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, ISO 81060-2, and IEC 62133. IP22 ingress protection.

## Regulatory Identification

A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.

## Predicate Devices

- Cnoga Medical ACAL, TensorTip - Non-invasive Hemodynamic Blood Pressure Monitor ([K093981](/device/K093981.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

August 7, 2019

Vita-Course Technologies Co., Ltd. % Kevin Wang Consultant Chonconn Medical Device Consulting Co., Ltd. No. A415, Block A. NanShan Medical Devices Industrial Park, Nanshan District, Shenzhen, 518067 Cn

Re: K190893

Trade/Device Name: Discovery 1 - Non-invasive Hemodynamic Blood Pressure Monitor Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: May 9, 2019 Received: May 9, 2019

Dear Kevin Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K190893

### Device Name

Discovery 1 - Non-invasive Hemodynamic Blood Pressure Monitor

### Indications for Use (Describe)

The Discovery 1 - Non-invasive Hemodynamic Blood Pressure Monitor device is a small, lightweight, handheld, device intended for measuring and display of Blood Pressure trending (systolic and diastolic) and spot-check of Peripheral pulse rate (PPR) and Peripheral pulse wave (PPW). Measurement is performing on capillary finger tip tissue (other than the thumb). The ring finger is the recommended site. The results of each measurement are stored in the system memory. The device is intended for use in the home environment. It is intended to be used by any person aged above 18 years old.

Type of Use (Select one or both, as applicable)

| <span>☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> |
|-------------------------------------------------------------|
| <span>☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span>  |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(K) Summary

## Prepared in accordance with the requirements of 21 CFR Part 807.92

## Prepared Date: 2019/04/01

#### 1. Submission sponsor

Name: Vita-Course Technologies Co., Ltd.

Address: 3F, Block B, Dajiahao Plaza, No.362, Yu'an 2nd Road, Bao'an District, Shenzhen, Guangdong, China

Contact person: Kezheng Ma

Title: Management Representative

E-mail: makezheng@vita-course.com

Tel: +86-755-23188226

#### 2. Submission correspondent

Name: Chonconn Medical Device Consulting Co., Ltd.

Address: No. A415, Block A, NanShan Medical devices Industrial Park Nanshan District, Shenzhen,

Guangdong, P.R. China 518067

Contact person: Kevin Wang

E-mail: kevin(@chonconn.com

Tel: +86-755-33941160

| Trade/Device Name | Non-invasive Hemodynamic Blood Pressure Monitor                           |
|-------------------|---------------------------------------------------------------------------|
| Model             | Discovery 1                                                               |
| Common Name       | Non-invasive Blood Pressure Monitor                                       |
| Regulatory Class  | Class II                                                                  |
| Classification    | 21CFR §870.1130 / System, measurement, blood-pressure, non-invasive / DXN |
| Submission type   | Traditional 510(K)                                                        |

#### Subject Device Information 3.

#### 4. Predicate Device

By submission of the Traditional 510(k), Vita-Course Technologies Co., Ltd. is requesting clearance for Discovery 1 Non-invasive Hemodynamic Blood Pressure Monitor. It is comparable to the following legally marketed system:

- Cnoga Medical ACAL, TensorTip ""- Non-invasive Hemodynamic Blood Pressure Monitor 1. under K093981.
#### ട്. Intended use & Indication for use

The Discovery 1 Non-invasive Hemodynamic Blood Pressure Monitor device is a small, lightweight, handheld, device intended for measuring and display of Blood Pressure trending (systolic and diastolic) and spot-check of Peripheral pulse rate (PPR) and Peripheral pulse wave (PPW). Measurement is performing on capillary

{4}------------------------------------------------

finger tip tissue (other than the thumb). The ring finger is the recommended site. The results of each measurement are stored in the system memory. The device is intended for use in the home environment. It is intended to be used by any person aged above 18 years old.

#### 6. Device Description

Discovery 1 is a portable non-invasive Hemodynamic Blood Pressure Monitor powered by an internal rechargeable lithium polymer battery. It contains a pulse wave probe incorporating two LEDs, which can collect the pulse wave to the device.

#### 7. Principle of Operation

The product is designed to calculate the blood pressure in terms of conduction time of pulse and wave parameters of pulse. This method is proposed based on interdependent characteristics between conduction time of the pulse wave along the arterial as well as characteristic parameters of the pulse wave and arterial pressure. Through the determination of the PPG signal, conduction time of the pulse wave along thee arterial and other correlation variables would be obtained and accordingly the systolic and diastolic pressures of the body are gotten by fitting operation. The pulse rate can be obtained by calculating the pulse wave signal.

#### Comparison to the Predicate Device 8.

The Discovery 1 non-invasive Hemodynamic Blood Pressure Monitor and TensorTip™ - Non-invasive Hemodynamic Blood Pressure Monitor (K093981), are identical in purpose, function, core technology and method of operation. Only minor differences exist between the Discovery 1 and predicate, which do not affect the safety or effectiveness of the subject device. Table 1 provides a comparison of the subject and predicate devices.

| Features                  | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Predicate Device K093981                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Remark       |
|---------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|
|                           | Discovery 1 non-invasive<br>Hemodynamic Blood Pressure<br>Monitor                                                                                                                                                                                                                                                                                                                                                                                                  | TensorTipTM - Non-invasive<br>Hemodynamic Blood Pressure Monitor                                                                                                                                                                                                                                                                                                                                                                                                                                                               |              |
| Indication<br>for use     | The Discovery 1 - Non-invasive<br>Hemodynamic Blood Pressure Monitor<br>device is a small, lightweight,<br>handheld, device intended for<br>measuring and display of Blood<br>Pressure trending (systolic and<br>diastolic) and spot-check of Peripheral<br>pulse rate (PPR) and Peripheral pulse<br>wave (PPW). Measurement is<br>performing on capillary finger tip tissue<br>(other than the thumb). The ring finger<br>is the recommended site. The results of | The TensorTip - Non-invasive blood<br>pressure trending device is a small,<br>lightweight, handheld, device intended for<br>measuring and display of Blood Pressure<br>trending (systolic and diastolic) and spot-<br>check of Peripheral pulse rate (PPR) and<br>Peripheral pulse wave (PPW).<br>Measurement is performing on capillary<br>fingertip tissue (other than the thumb). The<br>ring finger is the recommended site. The<br>results of each measurement are stored in<br>the system memory. The device is intended | Same         |
| Features                  | Subject Device<br>Discovery 1 non-invasive<br>Hemodynamic Blood Pressure<br>Monitor<br>system memory. The device is intended<br>for use in the home environment. It is<br>intended to be used by any person aged<br>above 18 years old.                                                                                                                                                                                                                            | Predicate Device K093981<br>TensorTipTM- Non-invasive<br>Hemodynamic Blood Pressure Monitor<br>for use in the home environment. It is<br>intended to be used by any person aged<br>above 18 years old.                                                                                                                                                                                                                                                                                                                         | Remark       |
| Patient<br>Population     | >18 years old                                                                                                                                                                                                                                                                                                                                                                                                                                                      | >18 years old                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Same         |
| Principle of<br>Operation | Hemodynamics                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Hemodynamics                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Same         |
| Power<br>source           | DC adaptor: 5Vd.c.,<br>Internal lithium polymer (Li-Pol)<br>battery: 3.7Vd.c., 2200 mAh                                                                                                                                                                                                                                                                                                                                                                            | DC adaptor: 5Vd.c.<br>Internal Li-Polymer battery: 3.7Vd.c.,<br>740mAh                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Different 1) |
| Blood<br>pressure         | Systolic pressure range: 50 mmHg ~200 mmHg<br>Diastolic pressure range: 40 mmHg ~160 mmHg<br>Accuracy: ±5 mmHg                                                                                                                                                                                                                                                                                                                                                     | Systolic pressure range: 90 mmHg ~ 240 mmHg<br>Diastolic pressure range: 30 mmHg ~ 130 mmHg<br>Accuracy: <10%                                                                                                                                                                                                                                                                                                                                                                                                                  | Different 2) |
| Pulse rate                | Measurement range: 40 bpm ~ 240 bpm<br>Accuracy: ±2% or ±2 bpm, whichever<br>is greater                                                                                                                                                                                                                                                                                                                                                                            | Measurement range: 30 bpm ~200 bpm<br>Accuracy: Mean Square Error 0.5digits                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Different 3) |
| Operating<br>environment  | Temperature: 5 °C ~ 45 °C<br>Humidity: 10% ~ 95%                                                                                                                                                                                                                                                                                                                                                                                                                   | Temperature: 6 °C ~ 45 °C<br>Humidity: 10 ~ 80%                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Different 4) |
| IP degree                 | IP22                                                                                                                                                                                                                                                                                                                                                                                                                                                               | IP00                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Different 5) |
| Standards                 | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-1-11<br>ISO 81060-2                                                                                                                                                                                                                                                                                                                                                                                                      | IEC 60601-1<br>IEC 60601-1-2<br>ISO 81060-2                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Different 6) |

Table 1: Comparison to Predicate Device

each measurement are stored in the

{5}------------------------------------------------

Justifications for differences between Discovery 1 and the predicate device are shown as below:

Different (1): The capacity of battery is different. However, the battery of subject device is in compliance with IEC 62133: 2012- Secondary cells and batteries containing alkaline or other non-acid electrolytes -Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications. Thus, this difference will not raise any safety or effectiveness issue.

Different (2): The measurement range and the accuracy of blood pressure is different. The subject device is compliant with FDA recognized consensus standard ISO 81060-2: 2013. This difference will not raise any safety or effectiveness issue.

Different (3): The measurement range and the accuracy of pulse rate is different. The subject device has been verified all the full claimed range. This difference will not raise any safety or effectiveness issue.

{6}------------------------------------------------

Different (4): The operating environment is minor different. However, the difference will not raise any safety or effectiveness issue.

Different (5): The degree of protection is different and the subject device is better than predicate device. The subject device is compliant with FDA recognized consensus standard IEC 60601-1-11: 2015. This difference will not raise any safety or effectiveness issue.

Different (6): The compliance standards are different. The subject device is compliant with FDA recognized consensus standard IEC 60601-1-11: 2015. This difference will not raise any safety or effectiveness issue.

#### 9. Performance Data

The following performance data were provided in support of the substantial equivalence determination.

| NO. | Item              | Description                                                                                                                                                              |
|-----|-------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1   | Safety            | Ensures the device meet the requirements of IEC 60601-1: 2012 and<br>IEC 60601-1-11: 2015                                                                                |
| 2   | EMC               | Ensures the temperature probes meet the requirements of IEC<br>60601-1-2: 2014, IEC 60601-1-11: 2015 clause 12 and ISO 80601-<br>2-61: 2017 clause 201.17 and clause 202 |
| 3   | Clinical accuracy | ISO 81060-2: 2013 Non-invasive sphygmomanometers - Part 2:<br>Clinical validation of automated measurement type                                                          |

## 10. Conclusion

It has been shown in this 510(k) submission that the difference between the proposed devices and the predicate devices do not raise any questions regarding safety and effectiveness. Performance testing and compliance with voluntary standards demonstrate that the proposed are substantially equivalent to the relevant aspects of the predicate devices in terms of design, components, materials, principals of operation, performance characteristics, and intended use.

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DXN/K190893](https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DXN/K190893)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
