Non-invasive blood pressure measurement systems

K190886 · Joytech Healthcare Co. , Ltd. · DXN · Jul 29, 2019 · Cardiovascular

Device Facts

Record IDK190886
Device NameNon-invasive blood pressure measurement systems
ApplicantJoytech Healthcare Co. , Ltd.
Product CodeDXN · Cardiovascular
Decision DateJul 29, 2019
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.1130
Device ClassClass 2

Indications for Use

The Fully Automatic Blood Pressure Monitors are intended to measure blood pressure (systolic and diastolic) and pulse rate of adults and adolescents age 12 through 21 years of age.

Device Story

Arm-type digital blood pressure monitor; uses inflatable cuff (22-48cm circumference) wrapped around upper arm; internal pump inflates cuff; oscillometric method with silicon integrated pressure sensor determines systolic/diastolic pressure and pulse rate; mechanical valve controls deflation; device displays results on LCD for three minutes; maximum pressure safety limit 300 mmHg; intended for home or clinical use; provides objective measurements to assist healthcare providers in monitoring cardiovascular health; benefits patient through non-invasive, automated blood pressure tracking.

Clinical Evidence

Clinical validation performed per ISO 81060-2:2013 for cuff sizes 22-42cm and 32-48cm. The 33-43cm cuff was validated by equivalence to the 32-48cm cuff due to identical internal bladder construction. Results confirmed compliance with accuracy requirements for automated non-invasive sphygmomanometers.

Technological Characteristics

Oscillometric measurement; silicon integrated pressure sensor; inflatable cuff (22-48cm); powered by 4 AA batteries or AC adapter; LCD display; maximum pressure 300 mmHg; safety/performance per AAMI/ANSI/IEC 80601-2-30; biocompatibility per ISO 10993-5 and ISO 10993-10.

Indications for Use

Indicated for measuring systolic/diastolic blood pressure and pulse rate in adults and adolescents aged 12-21 years.

Regulatory Classification

Identification

A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the symbol of the Department of Health & Human Services on the left and the FDA acronym with the full name of the agency on the right. The FDA part is in blue, with the acronym in a solid blue square and the words "U.S. FOOD & DRUG ADMINISTRATION" next to it. July 29, 2019 Joytech Healthcare Co., Ltd Ren Yunhua General Manager No. 365, Wuzhou Road, Yuhang Economic Development Zone Hangzhou City Hangzhou. 311100 CHINA Re: K190886 Trade/Device Name: Arm-type Fully Automatic Digital Blood Pressure Monitor - DBP-1204, DBP-1307, DBP-1332, DBP-1305, DBP-1334, DBP-1231, DBP-1314, DBP-1209, DBP-1303, DBP-1335 Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: June 20, 2019 Received: July 1, 2019 Dear Ren Yunhua: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of {1}------------------------------------------------ Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Stephen Browning Assistant Director Division of CardiacElectrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ #### Indications for Use 510(k) Number (if known) K190886 Device Name Arm-type Fully Automatic Digital Blood Pressure Monitor - DBP-1307, DBP-1332, DBP-1305, DBP-1334, DBP-1231, DBP-1314, DBP-1209, DBP-1303, DBP-1335 Indications for Use (Describe) Measure blood pressure (systolic and diastolic) and pulse rate of adults and adolescents age 12 through 21 years of age. | Type of Use (Select one or both, as applicable) | |-------------------------------------------------| | Prescription Use (Part 21 CFR 801 Subpart D) | | Over-The-Counter Use (21 CFR 801 Subpart C) | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary The assigned 510(k) number is: ## 2.1 Subiectter's Identification: Name: JOYTECH Healthcare Co., Ltd. Add.: No. 365. Wuzhou Road, Yuhang Economic Development Zone, Hangzhou,311100, China. Contact Person: Yunhua Ren Phone: +86-571-81957767 Fax: +86-571-81957750 Email: renyh@sejoy.com ## 2.2 Name of the Device: Trade Name: Arm-type Fully Automatic Digital Blood Pressure Monitor Including: DBP-1204,DBP-1307,DBP-1332,DBP-1305,DBP-1334,DBP-1231,DBP-1314, DBP-1209,DBP-1303,DBP-1335 Common Name: Blood Pressure Monitor Classification name: Non-invasive blood pressure measurement System 21 CFR 870-1130, Class II, 74-DXN. ### 2.3 Classification Information: Regulation Number: 870.1130 Product Code: DXN Device Class: II Panel: 74 Cardiovascular ### 2.4 Predicate Device Information: The Arm-Type Fully Automatic Digital Blood Pressure Monitor DBP-1204,DBP-1307, DBP-1332,DBP-1305,DBP-1334,DBP-1231,DBP-1314,DBP-1209,DBP-1303,DBP-1335 are corresponding substantially equivalent to the following device: Blood Pressure Monitor DBP-1204,DBP-1307,DBP-1332,DBP-1305,DBP-1334,DBP-1231,DBP-1314,DBP-1209,DBP-1303, DBP-1335, FDA 510(k) number: K173024, manufactured by JOYTECH Healthcare Co., Ltd.. {4}------------------------------------------------ ### 2.5 Device Description: The Arm-Type Fully Automatic Digital Blood Pressure Monitor (Models: DBP-1204,DBP-1307,DBP-1332,DBP-1305,DBP-1334,DBP-1231,DBP-1314,DBP-1209,DBP-1303,DBP-1335) uses an inflatable cuff which is wrapped around the patient's upper arm. The cuff circumference is limited to 22cm-48cm (be comprised of 3 cuffs, 22cm-36cm,22cm-42cm,32cm-48cm,select one. And the model DBP-1335 is comprised of 4 cuffs,22cm-36cm,22cm-42cm,32cm-48cm,33-43cm, select one.) which is inflated automatically by an internal pump in the device. The systolic and diastolic blood pressures are determined by oscillometric method and silicon integrate pressure sensor technology. The deflation rate is controlled by a preset mechanical valve at a constant rate. The pressure of the cuff is completely released automatically at the end of the measurement. At the same time, the measurements are displayed on the LCD display for three minute. There is a maximum pressure safety setting at 300 mmHg. The device will not inflate the cuff higher than 300 mmHg. | Features<br>Models | A | B | C | D | E | F | G | H | I (mm) | J (cm) | K(mm) | L | M | N | |--------------------|---|---|----------------|---|---|---|---|---|-------------|---------|------------|---|---|---| | DBP-1204 | Y | Y | 120Memories×1 | N | N | N | N | N | 150×112×56 | 22-48 | 83×35 | N | N | N | | DBP-1307 | Y | Y | 60 Memories×2 | Y | Y | Y | N | Y | 166×114× 72 | 22-48 | 102.1×68.9 | O | O | N | | DBP-1332 | Y | Y | 60 Memories×2 | Y | N | Y | Y | Y | 148×100× 56 | 22-48 | 84.1×55.1 | O | O | N | | DBP-1305 | Y | Y | 60 Memories×2 | Y | N | Y | Y | Y | 166×114×72 | 22-48 | 84.1×55.1 | O | O | N | | DBP-1334 | Y | O | 30Memories×4 | Y | N | Y | Y | Y | 110×155×70 | 22-48 | 62.7×46.4 | N | N | N | | DBP-1231 | Y | Y | 120 Memories×1 | Y | N | Y | Y | Y | 140 ×98× 48 | 22-48 | 62.3× 46 | N | O | N | | DBP-1314 | Y | Y | 30 Memories×4 | Y | N | Y | Y | Y | 110×155×70 | 22-48 | 84.1×55.1 | N | N | N | | DBP-1209 | Y | Y | 120 Memories×1 | Y | N | N | N | N | 134×99×66 | 22-48 | 62.7×46.4 | N | N | N | | DBP-1303 | Y | Y | 30Memories×4 | Y | N | N | N | N | 110×155×70 | 22-48 | 62.7×46.4 | N | N | N | | DBP-1335 | Y | N | 60 Memories×2 | Y | Y | Y | N | Y | 135×105×53 | 22-48 * | 59.9×50.9 | N | N | N | The detail comparisons among the Arm-Type Fully Automatic Digital Blood Pressure Monitors are list in table below: A = Powered by 4 AA size batteries B= Powered by AC adaptor C = Memory Size D = Time & Date E = Results Average in Three way F = WHO (World Health Organization) Classification Indicator G = Last 3 Results Average H = Irregular Heartbeat Detection I = Outside Demission (L x W x H in unit mm) J = Cuff Size 22cm-48cm (be comprised of 3 cuffs, 22cm-36cm,22cm-42cm,32cm-48cm,select {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the logo for JOYTECH. The logo is in blue and red. To the right of the word JOYTECH are two Chinese characters. one. Note:'*'And the model DBP-1335 is comprised of 4 cuffs,22cm-36cm,22cm-42cm,32cm-48cm,33-43cm, select one.) K = LCD Size (Viewing Area in unit mm) L = LCD Backlight M= Voice N=Music Y= Yes N= No O= Optional function depending on clients' needs The device are all designed and manufactured according to AAMI / ANSI / IEC 80601-2-30:2009/A1:2013, medical electrical equipment - part 2-30: particular requirements for the basic safety and essential performance of automated noninvasive sphygmomanometers. #### 2.6 Intended Use: The Fully Automatic Blood Pressure Monitors are intended to measure blood pressure (systolic and diastolic) and pulse rate of adults and adolescents age 12 through 21 years of age. #### 2.7 Difference of comparison with predicate device: The subject Arm-type fully automatic blood pressure monitor (Model: DBP-1204.DBP-1307.DBP-1332.DBP-1305.DBP-1334.DBP-1231.DBP-1314.DBP-1209.DBP-1303,DBP-1335) has the same functions and principle with predicated device which utilizes Oscilliometric measurement method to determine the blood pressure and the result can be shown on the LCD. The modification that was occurred is only cuff size from 22cm-36cm change to range within 22cm-48cm (be comprised of 3 cuffs, 22cm-36cm,22cm-42cm,32cm-48cm,select one. And the model DBP-1335 is comprised of 4 cuffs,22cm-36cm,22cm-42cm.32cm-48cm.33-43cm.select one.). ## 2.8. Non-clinical Tests: Electromagnetic Compatibility Test according to IEC 60601-1-2:2014, EN 60601-1-2:2015: General Safety Provisions Test according to AAMI/ANSI ES 60601-1:2005/(R) 2012 {6}------------------------------------------------ and A1:2012 and C1:2009/(R)2012 and A2:2010/(R)2012; Performance Test according to IEC 80601-2-30:2009(First Edition) and A1:2013, medical electrical equipment - part 2-30: particular requirements for the basic safety and essential performance of automated noninvasive sphygmomanometers. The test result all meet or exceed the requirement of the standards. Biocompatibility Test according to FDA Bluebook Memorandum G95-1Use of International Standard ISO 10993. ISO 10993-5: 2009 Biological evaluation of medical devices-Part 5: Tests for in vitro cytotoxicity and ISO 10993-10:2010 Biological evaluation of medical devices Part 10: Tests for irritation and skin sensitization. #### 2.10. Discussion of Clinical Tests Performed: Blood pressure monitor with its accessory which the fit arm circumference cuff sizes are 22-42cm and 32-48cm, these clinical tests were performed and comply with the accuracy requirements of ISO 81060-2 Second edition 2013-05-01, non-invasive sphygmomanometers - part 2: clinical validation of automated measurement type. Due to the cuff size of 33-43cm which is falling into the scope of 32-48cm, and both the double size cuffs (33-43cm and 32-48cm) are constructed with the same size internal bladder. So the clinical test result of the cuff size 32-48cm can represent and validate the another cuff which cuff size is 33-43cm meanwhile. ### 2.11. Conclusions: The subject Arm-type Fully Automatic Blood Pressure Monitor (Model: DBP-1204,DBP-1307,DBP-1332,DBP-1305,DBP-1334,DBP-1231,DBP-1314,DBP-1209,DBP-1303,DBP-1335) is itself of the corresponding predicate device manufactured by JOYTECH Healthcare Co., Ltd. (FDA 510(k) number: K173024). The addition of new arm-cuff range 22cm-48cm (be comprised of 3 cuffs, 22cm-36cm,22cm-42cm,32cm-48cm,select one.And the model DBP-1335 is comprised of 4 cuffs,22cm-36cm,22cm-48cm,33-43cm.select one.) will not: (1) affect the intended use or (2) alter the fundamental scientific technology of the device.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...