← Product Code [DXN](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DXN) · K182127

# Wrist Blood Pressure Monitor Model BP6100 (K182127)

_Omron Healthcare, Inc. · DXN · Nov 2, 2018 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DXN/K182127

## Device Facts

- **Applicant:** Omron Healthcare, Inc.
- **Product Code:** [DXN](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DXN.md)
- **Decision Date:** Nov 2, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1130
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The device is a digital monitor intended for use in measuring blood pressure and pulse rate in adult patient population with wrist circumference ranging from 5.3 inches to 8.5 inches (13.5 cm to 21.5 cm). The device detects the appearance of irregular heartbeats during measurement and gives a warning signal with readings.

## Device Story

Wrist-worn, automatic, non-invasive blood pressure monitor for OTC home use. Input: pulse waveform data extracted from cuff pressure via electric pump inflation. Processing: microprocessor software algorithms analyze waveform to determine systolic/diastolic pressure and pulse rate using oscillometric method. Output: LCD display of blood pressure, pulse rate, irregular heartbeat warning, and stored/averaged readings. Operated by patient; no external power or connectivity. Benefits: enables self-monitoring of blood pressure and pulse rate in home environment.

## Clinical Evidence

Clinical investigation conducted per ANSI/AAMI/ISO 81060-2:2013. Accuracy of BP6100 oscillometric measurements validated against auscultation method using calibrated sphygmomanometer by trained staff. Results demonstrated equivalence to auscultation method and conformance with standard.

## Technological Characteristics

Battery-operated (2x AAA) wrist-worn monitor. Piezo-resistive pressure sensor. Oscillometric measurement method. Pressure range 0-299 mmHg; pulse range 40-180 bpm. LCD display. Biocompatible materials per ISO 10993-1. Electrical safety and EMC testing performed. Software-based algorithm for oscillometric analysis.

## Regulatory Identification

A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.

## Predicate Devices

- Omron Healthcare, Inc. HEM-6131 ([K131742](/device/K131742.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 2, 2018

Omron Healthcare, Inc. % Ronald Warren Senior Director, Regulatory Affairs Experien Group, LLC 224 Airport Pkwy.. Ste. 250 San Jose, California 95110

Re: K182127

Trade/Device Name: Wrist Blood Pressure Monitor Model BP6100 Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: August 3, 2018 Received: August 6, 2018

Dear Ronald Warren:

We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Shawn W. Forrest -A

Digitally signed by Shawn W. Forrest -A N: c=US, o=U.S. Government, ou=HHS Qu=FDA ou=People. 0.9.2342.19200300.100.1.1=1300403341 cn=Shawn W. Forrest -A Date: 2018.11.02 10:07:35 -04'00'

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known)

K182127

Device Name

Wrist Blood Pressure Monitor Model BP6100

Indications for Use (Describe)

The device is a digital monitor intended for use in measuring blood pressure and pulse rate in adult patient population with wrist circumference ranging from 5.3 inches to 8.5 inches (13.5 cm to 21.5 cm).

The device detects the appearance of irregular heartbeats during measurement and gives a warning signal with readings.

| Type of Use (Select one or both, as applicable)                       |                                                                                 |
|-----------------------------------------------------------------------|---------------------------------------------------------------------------------|
| <input type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) | <input checked="" type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Notification K

## GENERAL INFORMATION [807.92(a)(1)]

#### Applicant:

Omron Healthcare, Inc. 1925 West Field Court Lake Forest, IL 60045 USA Phone: 847-247-5626 FAX: 847-680-6269

#### Contact Person:

Ronald S. Warren Senior Director, Regulatory Affairs Experien Group, LLC 224 Airport Parkway, Suite 250 San Jose, CA 95110 USA Phone: 408-505-3926 FAX: 408-400-0856

Date Prepared: August 3, 2018

#### DEVICE INFORMATION [807.92(a)(2)]

#### Trade Name:

Wrist Blood Pressure Monitor Model BP6100

## Generic/Common Name:

Noninvasive blood pressure measurement system

#### Classification:

Class II per 21 CFR §870.1130

Product Code:

DXN

## PREDICATE DEVICE(S) [807.92(a)(3)]

The Wrist Blood Pressure Monitor Model BP6100 is substantially equivalent to the Omron Healthcare, Inc. HEM-6131 non-invasive blood pressure measurement system (K131742).

## DEVICE DESCRIPTION [807.92(a)(4)]

The Wrist Blood Pressure Monitor (BPM) Model BP6100 ("BP6100") is an automatic, non-invasive, wrist-worn blood pressure measurement system for over-the-counter (OTC) home use. During measurements, an electric pump within the main unit slowly inflates the wrist cuff, generating cuff pressure which is monitored and from which pulse waveform data is extracted. This waveform data is analyzed by software algorithms

{4}------------------------------------------------

## 510(k) SUMMARY

within the microprocessor to determine pulse rate, systolic pressure, and diastolic pressure. The systolic and diastolic pressures are determined via the oscillometric cuff method. The cuff pressure range is 0 to 299mmHg and the pulse rate range is 40 to 180 beats/min.

The BP6100 is intended for use in the adult patient population, in wrist circumferences ranging between 5.3 - 8.5 inches (13.5 - 21.5 cm). During measurement, the device also detects the appearance of irregular heartbeat rhythms. BP6100 further includes administrative functions for automatic storage of the 60 most recent readings taken by the device. The device also provides an "average" function, which can calculate an average reading based on the latest two (2) or three (3) readings, when taken within a 10-minute span.

The operations of BP6100 are primarily through use of buttons. Functions and other features that are controlled by the end user include applying the wrist cuff to the wrist, powering on/off the system, starting or stopping measurements, and installing and changing batteries as needed. It has no potential connection to external power. There are no additional parts or accessories to the BP6100. The BP6100 does not connect to any other devices.

## INDICATIONS FOR USE [807.92(a)(5)]

The device is a digital monitor intended for use in measuring blood pressure and pulse rate in adult patient population with wrist circumference ranging from 5.3 inches to 8.5 inches (13.5 cm to 21.5 cm).

The device detects the appearance of irregular heartbeats during measurement and gives a warning signal with readings.

## COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICES [807.92(a)(6)]

The proposed BP6100 device and the HEM-6131 predicate device have similar technological characteristics. Both devices are battery-operated wrist blood pressure monitors intended for home use and employ the oscillometric method for measuring blood pressure and pulse rate. Both devices have the same cuff pressure range of 0 to 299mmHg and pulse rate range of 40 to 180 beats/min. The devices are intended for the same wrist circumference range of 13.5 cm to 21.5 cm. The devices share the same accuracy range of pressure reading of ±3 mmHg and the same accuracy range for pulse rate reading of ± 5%.

There are minor differences in technical specifications and features between the BP6100 proposed device and HEM-6131 predicate device. These differences relate primarily to user convenience and simplicity of operation. The BP6100 is slightly smaller in dimension and weighs slightly less than the HEM-6131 predicate device. The BP6100 has a different user interface configuration from the HEM-6131; however, the displayed information is the same between the two devices. The BP6100 software also differs slightly from the HEM-6131 software in processing workflow; however, the underlying principle of operation for the algorithms are the same.

{5}------------------------------------------------

# SUBSTANTIAL EQUIVALENCE

The proposed Indications for Use for the Wrist Blood Pressure Monitor Model BP6100 is substantially equivalent to the Indications for Use for the predicate HEM-6131 device. Comparison testing demonstrated that the proposed device is equivalent to the predicate device with regard to measurement of blood pressure in a pulse wave generator test. Minor differences in the technological features relate mostly to convenience considerations or routine performance updates and do not impact the safety or performance of blood pressure or pulse rate measurements. Any differences in the technological characteristics between the devices do not raise any different questions of safety or effectiveness. Thus, the BP6100 is substantially equivalent to the predicate HEM-6131 device.

Please refer to the table below for a comparison of Indications for Use and technological characteristics between the BP6100 and the HEM-6131 predicate device.

| Feature                   | BP6100<br>(K____)                                                                                                                                                                                                                                                                                                                                                        | HEM-6131<br>(K131742)                                                                                                                                                                                                                                                                                                                                                | Comparison |
|---------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Classification            | 21 CFR§870.1130,<br>Noninvasive blood<br>pressure measurement system                                                                                                                                                                                                                                                                                                     | 21 CFR§870.1130,<br>Noninvasive blood<br>pressure measurement system                                                                                                                                                                                                                                                                                                 | Same       |
| Product Code              | DXN - Noninvasive blood<br>pressure<br>measurement                                                                                                                                                                                                                                                                                                                       | DXN - Noninvasive blood<br>pressure<br>measurement                                                                                                                                                                                                                                                                                                                   | Same       |
| Indications for<br>Use    | The device is a digital monitor<br>intended for use in measuring<br>blood pressure and pulse rate in<br>adult patient population with<br>wrist circumference ranging<br>from 5.3 inches to 8.5 inches<br>(13.5 cm to 21.5 cm).<br><br>The device detects the<br>appearance of irregular<br>heartbeats during measurement<br>and gives a warning signal with<br>readings. | The device is a digital monitor<br>intended for use in measuring<br>blood pressure and pulse rate in<br>adult patient population with<br>wrist circumference ranging<br>from 5¼ inches to 8½ inches<br>(13.5cm to 21.5cm).<br><br>The device detects the<br>appearance of irregular<br>heartbeats during measurement<br>and gives a warning signal with<br>readings. | Equivalent |
| Environment of<br>Use     | Home Use                                                                                                                                                                                                                                                                                                                                                                 | Home Use                                                                                                                                                                                                                                                                                                                                                             | Same       |
| Patient Population        | Adults                                                                                                                                                                                                                                                                                                                                                                   | Adults                                                                                                                                                                                                                                                                                                                                                               | Same       |
| Over-The-Counter<br>(OTC) | Yes                                                                                                                                                                                                                                                                                                                                                                      | Yes                                                                                                                                                                                                                                                                                                                                                                  | Same       |
| Measurement<br>method     | Cuff oscillometric method                                                                                                                                                                                                                                                                                                                                                | Cuff oscillometric method                                                                                                                                                                                                                                                                                                                                            | Same       |
| Measurement<br>Range      | Cuff pressure range 0 to 299<br>mmHg<br>Pulse Rate: 40 to 180<br>beats/min.                                                                                                                                                                                                                                                                                              | Pressure: 0 to 299 mmHg<br>Pulse Rate: 40 to 180<br>beats/min.                                                                                                                                                                                                                                                                                                       | Same       |
| Accuracy                  | Blood Pressure: Within<br>±3mmHg<br>Pulse Rate: Within ±5 % of<br>reading                                                                                                                                                                                                                                                                                                | Blood Pressure: Within<br>±3mmHg or 2% of reading<br>Pulse Rate: Within ±5 % of<br>reading                                                                                                                                                                                                                                                                           | Equivalent |
| Pressure Sensor           | Piezo resistance sensor                                                                                                                                                                                                                                                                                                                                                  | Piezo resistance sensor                                                                                                                                                                                                                                                                                                                                              | Same       |

{6}------------------------------------------------

#### 510(k) SUMMARY

| Feature                         | BP6100<br>(K____)                                                                                                                             | HEM-6131<br>(K131742)                                                                                                                         | Comparison |
|---------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Inflation Method                | Automatic inflation with<br>electric pump                                                                                                     | Automatic inflation with<br>electric pump                                                                                                     | Same       |
| Deflation Method                | Automatic rapid deflation valve                                                                                                               | Automatic rapid deflation valve                                                                                                               | Same       |
| Display Type                    | LCD digital display                                                                                                                           | LCD digital display                                                                                                                           | Same       |
| Additional<br>Functionalities   | Irregular heartbeat feature<br>Averaging value<br>Body movement detection<br>Hypertension indicator<br>Cuff wrap guide<br>Storage of readings | Irregular heartbeat feature<br>Averaging value<br>Body movement detection<br>Hypertension indicator<br>Cuff wrap guide<br>Storage of readings | Same       |
| Power Source                    | Two "AAA" batteries                                                                                                                           | Two "AAA" batteries                                                                                                                           | Same       |
| Operating<br>Conditions         | 10 to 40 °C (50 to 104 °F)<br>15 to 90 %RH (non-condensing)<br>800 to 1060 hPa                                                                | 10 to 40 °C (50 to 104 °F)<br>15 to 85 %RH                                                                                                    | Similar    |
| Storage/Transport<br>Conditions | -20 to 60 °C (-4 to 140 °F)<br>10 to 90 %RH (non-condensing)                                                                                  | -20 to 60 °C (-4 to 140 °F)<br>10 to 95 %RH                                                                                                   | Similar    |
| Dimensions                      | 93 (W) × 20 (D) × 62 (H) mm                                                                                                                   | 78 (W) × 60 (D) × 21 (H) mm                                                                                                                   | Similar    |
| Weight                          | Approx. 86g (3.0 oz) (not<br>including battery)                                                                                               | Approx. 101g (not including<br>battery)                                                                                                       | Similar    |

## PERFORMANCE DATA [807.92(b)]

All necessary bench and clinical testing was conducted on the Wrist Blood Pressure Monitor Model BP6100 to support a determination of substantial equivalence to the predicate device.

## [807.92(b)(1)] Nonclinical Testing Summary:

The nonclinical, bench testing included:

- Comparative blood pressure and pulse rate testing to the predicate device
- . Performance verification testing of BP6100 to confirm acceptable performance of device features and functions
- Cleaning verification testing to confirm device retains its performance when . cuff is cleaned with household detergents as may be required in home use environment

Other nonclinical safety testing included:

- Biocompatibility of patient-contacting materials per ISO 10993-1 . requirements
- Electrical safety, electromagnetic compatibility, and electrostatic discharge ● testing
- Software verification and validation ●

{7}------------------------------------------------

# 510(k) SUMMARY

The collective results of the nonclinical testing demonstrate that the materials chosen, the manufacturing processes, and design of the BP6100 meet the established specifications necessary for consistent performance during its Intended Use. In addition, the collective bench testing results demonstrate that the BP6100 does not raise different questions of safety or effectiveness for blood pressure and pulse rate measurement in a home use environment when compared to the predicate device.

# [807.92(b)(2)] Clinical Testing Summary:

A clinical investigation was conducted with the objective of validating the accuracy of blood pressure measurements by BP6100 based on an oscillometric method as compared to an auscultation method using a calibrated sphygmomanometer by trained medical staff. This study was conducted in accordance with guideline per ANSI/AAMI/ISO 81060-2:2013 Non-invasive sphygmomanometers — Part 2: Clinical investigation of automated measurement type. The results demonstrated that BP6100 performed equivalently to the auscultation method and is in conformance with ANSVAAMI/ISO 81060-2:2013.

## CONCLUSIONS [807.92(b)(3)]

The conclusions drawn from the nonclinical and clinical tests performed in support of BP6100 demonstrate that the device is safe and effective, and performs at least as safely and effectively as the legally marketed HEM-6131 predicate device.

#### SUMMARY

The Wrist Blood Pressure Monitor Model BP6100 is substantially equivalent to the Omron Model HEM-6131 predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DXN/K182127](https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DXN/K182127)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
