← Product Code [DXN](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DXN) · K181931

# GMC Upper Arm Automatic Blood Pressure Monitor (K181931)

_Gmc, Inc. · DXN · Apr 4, 2019 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DXN/K181931

## Device Facts

- **Applicant:** Gmc, Inc.
- **Product Code:** [DXN](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DXN.md)
- **Decision Date:** Apr 4, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1130
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The GMC Upper Arm Automatic Blood Pressure Monitor, Model X3, is a non-invasive blood pressure measurement device that is used for measuring systolic and diastolic blood pressure and pulse rate using the oscillometric method for adults at home. The device has an irregular heartbeat (IHB) indicator. The device detects the appearance of an irregular heartbeat during measurement, and displays an IHB symbol on the LCD with the reading once the irregular heartbeat is detected.

## Device Story

Device measures systolic/diastolic blood pressure and pulse rate via oscillometric method; uses electronic pressure sensor to convert cuff pressure variations into electrical signals; signals analyzed to determine BP and pulse. Features include irregular heartbeat (IHB) detection (triggered by >25% deviation from average pulse), body movement detection, and hypertension classification indicator. Intended for adult home use; operated by patient. Output displayed on LCD; includes BP, pulse, memory results, date/time, and status icons (IHB, movement, battery). Data used by patients for personal health monitoring.

## Clinical Evidence

Clinical validation performed per ISO 81060-2. Study included 85 patients (39 male, 46 female). Accuracy of the device was compared against the standard auscultatory method using alternating measurements in the same arm. Results confirmed clinical accuracy.

## Technological Characteristics

Oscillometric measurement; semi-conductor pressure sensor; pump-driven inflation; automatic air release valve. Powered by batteries or AC adapter. LCD display. Complies with ANSI/AAMI ES60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 80601-2-30, ISO 81060-2, and ISO 10993-1. Materials are biocompatible per ISO 10993-1.

## Regulatory Identification

A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.

## Predicate Devices

- Omron HEM-7311 ([K133379](/device/K133379.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font below.

April 4, 2019

GMC Inc. Yolanda Lin Director of Quality Management Div. No. 686, Su Chu Rd. Chuzhou, 239000 Cn

Re: K181931

Trade/Device Name: GMC Upper Arm Automatic Blood Pressure Monitor Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: January 11, 2019 Received: January 11, 2019

Dear Yolanda Lin:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Please wait...

If this message is not eventually replaced by the proper contents of the document, your PDF viewer may not be able to display this type of document.

You can upgrade to the latest version of Adobe Reader for Windows®, Mac, or Linux® by visiting http://www.adobe.com/go/reader_download.

For more assistance with Adobe Reader visit http://www.adobe.com/go/acrreader.

Windows is either a registered trademark of Microsoft Corporation in the United States and/or other countries. Mac is a trademark of Apple Inc., registered in the United States and other countries. Linux is the registered trademark of Linus Torvalds in the U.S. and other countries.

{3}------------------------------------------------

Section 5- 510(k) Summary

Image /page/3/Picture/3 description: The image shows the logo for GMC 尚钧医疗. The logo consists of the letters "GMC" in a dark gray sans-serif font, followed by a red circle with a white plus sign inside. To the right of the circle are the Chinese characters 尚钧医疗, also in a dark gray color. The characters are written in a traditional Chinese calligraphy style.

GMC Inc. No. 686, Su Chu Rd., Chuzhou, Anhui, China

X3

## 510(k) Summary

| 510(k) owner's information: | GMC Inc.<br>No. 686, Su Chu Rd., Chuzhou,<br>Anhui, China<br>Tel: +86-550-2172888<br>Fax: +86-550-2172889 |
|-----------------------------|-----------------------------------------------------------------------------------------------------------|
| Official Contact:           | Yolanda Lin<br>Tel: +86-550-2172888<br>Fax: +86-550-2172889<br>E-mail: linyolanda23@gmail.com             |
| Proprietary or Trade Name:  | GMC Upper Arm Automatic Blood<br>Pressure Monitor, Model X3                                               |
| Common/Usual Name:          | Non-Invasive Blood Pressure Monitor                                                                       |
| Classification Name:        | System, Measurement, Blood-Pressure,<br>Non-Invasive                                                      |
| Regulation                  | CFR 870.1130                                                                                              |
| Product Code                | DXN                                                                                                       |
| Device Class:               | II                                                                                                        |
| Predicate Device:           | Omron HEM-7311 (K133379)                                                                                  |
| Date Summary Prepared:      | 07/16/2018                                                                                                |

{4}------------------------------------------------

Section 5- 510(k) Summary

Image /page/4/Picture/3 description: The image shows the logo for GMC 尚钧医疗. The logo consists of the letters "GMC" in a dark gray sans-serif font, followed by a red circle with a white plus sign in the center. To the right of the circle are four Chinese characters in a dark gray, traditional font. 

No. 686, Su Chu Rd., Chuzhou Anhui, China

X3

## Device Description:

The GMC Upper Arm Automatic Blood Pressure Monitor, Model X3, measures systolic and diastolic blood pressure and pulse rate of an adult individual. The method used to define blood pressure is similar to the auscultatory method but uses an electronic pressure sensor rather than a stethoscope and mercury manometer. The device utilizes the oscillometric method whereby the electronic pressure sensor converts variation in cuff pressure to electrical signals. The electrical signals are analyzed and used to define the systolic and diastolic measurements and determine the pulse rate. An irregular heart beat (IHB) indicator is displayed when there is more than a 25% deviation from the averaged pulse rate measurements.

## Intended Use:

The GMC Upper Arm Automatic Blood Pressure Monitor, Model X3, is a non-invasive blood pressure measurement device that is used for measuring systolic and diastolic blood pressure and pulse rate using the oscillometric method for adults at home.

The device has an irregular heartbeat (IHB) indicator. The device detects the appearance of an irregular heartbeat during measurement, and displays an IHB symbol on the LCD with the reading once the irregular heartbeat is detected.

## Substantially Equivalent (SE) Comparison

The GMC Upper Arm Automatic Blood Pressure Monitor, Model X3 is substantially equivalent to the Omron HEM-7311 (K133379).

{5}------------------------------------------------

# GMC0尚钩医疗

GMC Inc. No. 686, Su Chu Rd., Chuzhou, Anhui, China

X3

| Model name                              | Predicate Device                                                                                                                                                                                                                                                  | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Comparison |
|-----------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
|                                         | Omron HEM-7311 (K133379)                                                                                                                                                                                                                                          | GMC X3                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |            |
| 510(k) submitter/holder                 | Omron Healthcare, Inc.                                                                                                                                                                                                                                            | GMC Inc.                                                                                                                                                                                                                                                                                                                                                                                                                                                                |            |
| Intended use and Indications<br>for Use | The device is a digital monitor intended<br>for use in measuring blood pressure and<br>pulse rate in adult patient population.<br>The device detects the appearance of<br>irregular heartbeats during<br>measurement and gives a warning<br>signal with readings. | The device is a non-invasive blood pressure<br>measurement device that is used for<br>measuring systolic and diastolic blood<br>pressure and pulse rate using the<br>oscillometric method for adults at home.<br>The device has an irregular heartbeat (IHB)<br>indicator. The device detects the appearance<br>of an irregular heartbeat during<br>measurement, and displays an IHB symbol<br>on the LCD with the reading once the<br>irregular heartbeat is detected. | Identical  |
| Use Environment                         | Home use                                                                                                                                                                                                                                                          | Home use                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Identical  |
| Patient Population                      | Adult                                                                                                                                                                                                                                                             | Adult                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Identical  |

{6}------------------------------------------------

## GMC0尚钧医疗

GMC Inc. No. 686, Su Chu Rd., Chuzhou, Anhui, China

Specifications / Features

| Model name                  | Predicate Device<br>Omron HEM-7311 (K133379) |                                | Subject Device<br>GMC X3 | Comparison                                                                                  |
|-----------------------------|----------------------------------------------|--------------------------------|--------------------------|---------------------------------------------------------------------------------------------|
| Measurement method          | Oscillometric methods                        |                                | Oscillometric methods    | Identical                                                                                   |
| Cuff fits arm circumference | 17-22cm/ 22-32cm/ 22-42cm                    |                                | 24-40 cm (9.4"~15.7")    | Similar<br>This minor difference does not impact<br>safety and effectiveness of the device. |
| Measurement range           | Pressure                                     | 0~299 mmHg                     | 30~260 mmHg              | Similar<br>This minor difference does not impact<br>safety and effectiveness of the device. |
|                             | Pulse                                        | 40~180 beats/min               | 40~199 beats/min         | Similar<br>This minor difference does not impact<br>safety and effectiveness of the device. |
| Accuracy                    | Pressure                                     | $\pm$ 3 mmHg or 2 % of reading | $\pm$ 3 mmHg             | Similar<br>There is no difference in the<br>accuracy range based on mmHg<br>scale           |
|                             | Pulse                                        | $\pm$ 5% of reading            | $\pm$ 5% of reading      | Identical                                                                                   |

{7}------------------------------------------------

#### X3

# GMC0尚钧医疗

GMC Inc. No. 686, Su Chu Rd., Chuzhou, Anhui, China

| Model name                              | Predicate Device<br>Omron HEM-7311 (K133379)                                                                                                    | Subject Device<br>GMC X3                                                                                                                                                                                                   | Comparison                                                                                        |
|-----------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|
| Reusable                                | Yes                                                                                                                                             | Yes                                                                                                                                                                                                                        | Identical                                                                                         |
| Display Type                            | LCD                                                                                                                                             | LCD                                                                                                                                                                                                                        | Identical                                                                                         |
| Display Parameter                       | Current cuff pressure, Systolic blood<br>pressure, Diastolic blood pressure,<br>Pulse rate, Error message,<br>measurement results in the memory | Systolic pressure, Diastolic pressure,<br>Pulse rate, Memory sequence<br>number, Date/Time, memory zone<br>icons, pulse mark, weak battery mark,<br>AVG, Hypertension Classification<br>Indicator, IHB, Movement indicator | Similar<br>The different display parameters do not impact safety and effectiveness of the device. |
| Inflation                               | Pump driven                                                                                                                                     | Pump driven                                                                                                                                                                                                                | Identical                                                                                         |
| Deflation                               | Automatic Air Release Valve                                                                                                                     | Automatic Air Release Valve                                                                                                                                                                                                | Identical                                                                                         |
| Pressure Sensor                         | Semi-conductor                                                                                                                                  | Semi-conductor                                                                                                                                                                                                             | Identical                                                                                         |
| Irregular Heart Beat detection          | Yes                                                                                                                                             | Yes                                                                                                                                                                                                                        | Identical                                                                                         |
| Body movement detection                 | Yes                                                                                                                                             | Yes                                                                                                                                                                                                                        | Identical                                                                                         |
| Hypertension indicator                  | Yes                                                                                                                                             | Yes                                                                                                                                                                                                                        | Identical                                                                                         |
| Average of latest three<br>measurements | Yes                                                                                                                                             | Yes                                                                                                                                                                                                                        | Identical                                                                                         |

{8}------------------------------------------------

#### X3

# GMC0尚钧医疗

GMC Inc. No. 686, Su Chu Rd., Chuzhou, Anhui, China

| Model name     |                     | Predicate Device<br>Omron HEM-7311 (K133379)        | Subject Device<br>GMC X3        | Comparison                                                                                  |
|----------------|---------------------|-----------------------------------------------------|---------------------------------|---------------------------------------------------------------------------------------------|
| Multiple Users |                     | Yes(2)                                              | Yes(2)                          | Identical                                                                                   |
| Power Source   |                     | Battery as well as AC adaptor                       | Battery as well as AC adaptor   | Identical                                                                                   |
| Dimension      |                     | 183 X 230 X 99mm (Lx W x H)                         | 96 X 139.7 X 63.2mm (L x W x H) | Similar<br>Size is not a factor in function of the<br>device                                |
| Weight         |                     | Approx. 640g (1 lbs6 5/8 oz)<br>(without batteries) | (without batteries)<br>246g     | Similar<br>Weight is not a factor in function of the<br>device                              |
| Operating      | Temperature         | 10~40℃ (50-104°F)                                   | 10~40℃ (50-104°F)               | Identical                                                                                   |
| Environment    | Humidity            | 15%~90% RH                                          | 15%~85% RH                      | Similar<br>This minor difference does not impact<br>safety and effectiveness of the device. |
|                | Ambient<br>pressure | N/A                                                 | 700-1060 hpa                    | Different<br>This feature does not impact safety and<br>effectiveness of the device         |

{9}------------------------------------------------

#### X3

## GMC0尚钧医疗

GMC Inc. No. 686, Su Chu Rd., Chuzhou, Anhui, China

| Model name             | Predicate Device<br>Omron HEM-7311 (K133379) |                                                                                                       | Subject Device<br>GMC X3                                                                              | Comparison                                                                                  |
|------------------------|----------------------------------------------|-------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|
| Storage<br>Environment | Temperature                                  | -20~60°C (-4-140°F)                                                                                   | -10~60 ℃(14-140°F)                                                                                    | Similar<br>This minor difference does not impact<br>safety and effectiveness of the device. |
|                        | Humidity                                     | 10%~95% RH                                                                                            | 10%~90% RH                                                                                            | Similar<br>This minor difference does not impact<br>safety and effectiveness of the device. |
|                        | Ambient<br>pressure                          | 700-1060 hpa                                                                                          | 700-1060 hpa                                                                                          | Identical                                                                                   |
| Compliance Standards   |                                              | ANSI/AAMI ES60601-1<br>IEC 60601-1-2<br>IEC 60601-1-11<br>IEC 80601-2-30<br>ISO81060-2<br>ISO 10993-1 | ANSI/AAMI ES60601-1<br>IEC 60601-1-2<br>IEC 60601-1-11<br>IEC 80601-2-30<br>ISO81060-2<br>ISO 10993-1 | Identical                                                                                   |

{10}------------------------------------------------

 GMC Inc. No. 686, Su Chu Rd., Chuzhou, Anhui, China

The subject device is viewed as substantially equivalent to the predicate devices Omron HEM-7311 because they have same intended use and use the well known oscillometric method within the software algorithm to determine the systolic and diastolic blood pressure and pulse rate. There are no significant diffect the safety or effectiveness of the intended device as compared to the predicate devices

{11}------------------------------------------------

Image /page/11/Picture/3 description: The image shows the logo for GMC+. The logo consists of the letters "GMC" in a dark gray sans-serif font, followed by a red circle with a white plus sign inside. To the right of the circle are two Chinese characters, also in dark gray. The overall design is simple and modern, with a focus on the company's initials and a medical symbol.

No. 6

## Performance Testing:

Bench testing was conducted to demonstrate that the device meets its requirements and specification.

The following performance tests were completed:

- #19-4
ANSI AAMI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012 (Consolidated Text) Medical electrical equipment - Part 1: General requirements for basic safety

and essential performance (IEC 60601-1:2005, MOD)

- #19-8 •
IEC 60601-1-2 Edition 4.0 2014-02

Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests

- #19-14
IEC 60601-1-11 Edition 2.0 2015-01

Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment

- #3-123
IEC 80601-2-30 Edition 1.1 2013-07

Medical electrical equipment - Part 2-30: Particular requirements for the basic safety and essential performance of automated non-invasive sphygmomanometers

The GMC Upper Arm Automatic Blood Pressure Monitor, Model X3 meets all applicable requirements.

{12}------------------------------------------------

Image /page/12/Picture/3 description: The image shows the logo for GMC+. The logo consists of the letters "GMC" in a dark gray sans-serif font, followed by a red circle with a white plus sign inside. To the right of the circle is a Chinese character, also in dark gray.

No. 6

## Biocompatibility

The materials of construction are widely used in the same or similar applications; therefore, demonstrating the material's use in other medical devices and supporting existing biocompatibility.

It is evaluated according to #2-220 ISO 10993-1 Fourth edition 2009-10-15 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process.

## Clinical testing

Testing to insure clinical accuracy of the device in accordance with ISO 81060-2 as documented in Clinical Investigation Report. 85 patients (39 males and 46 females) were invited for the study. Standard auscultation method was used as the reference blood pressure monitor measuring in the left arm. Blood pressure measurements were repeated alternatively with the device and auscultation in the same arm according to the sequence in #3-122 ISO 81060-2 Second edition 2013-05-01 Non-invasive sphygmomanometers - Part 2: Clinical validation of automated measurement type.

## Conclusion

Based on the evaluations completed, there are no new safety or effectiveness issues introduced with this new medical device. Therefore, the GMC Upper Arm Automatic Blood Pressure Monitor, Model X3, is substantially equivalent to the predicate devices Omron HEM-7311 (K133379).

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DXN/K181931](https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DXN/K181931)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
