BP5S Wireless Blood Pressure Monitor

K171062 · Andon Health Co, Ltd. · DXN · Aug 1, 2017 · Cardiovascular

Device Facts

Record IDK171062
Device NameBP5S Wireless Blood Pressure Monitor
ApplicantAndon Health Co, Ltd.
Product CodeDXN · Cardiovascular
Decision DateAug 1, 2017
DecisionSESE
Submission TypeAbbreviated
Regulation21 CFR 870.1130
Device ClassClass 2

Indications for Use

BP5S Wireless Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. The cuff circumference is limited to 22cm-48cm.

Device Story

BP5S Wireless Blood Pressure Monitor is a non-invasive system for measuring systolic/diastolic blood pressure and pulse rate. Device utilizes an inflatable cuff (22-48cm) wrapped around the upper arm. Operational principle relies on oscillometric measurement and silicon-integrated pressure sensor technology. Device calculates blood pressure values, detects irregular heartbeats, and displays results on an integrated LED screen. Used by medical professionals or patients in home settings. Output provides blood pressure readings and irregular heartbeat indicators, assisting users in monitoring cardiovascular health. Device design conforms to IEC 80601-2-30 standards.

Clinical Evidence

No clinical data. Substantial equivalence supported by non-clinical bench testing, including electromagnetic compatibility (IEC 60601-1-2), electrical safety (IEC 60601-1), and performance characteristics for automated noninvasive sphygmomanometers (IEC 80601-2-30).

Technological Characteristics

Oscillometric measurement principle using silicon-integrated pressure sensors. Designed per IEC 80601-2-30. Features integrated LED display for result output. Electrical safety per IEC 60601-1; EMC per IEC 60601-1-2. Wireless connectivity implied by trade name.

Indications for Use

Indicated for adult individuals requiring non-invasive measurement of diastolic and systolic blood pressure and pulse rate via upper arm cuff (22cm-48cm circumference). Suitable for use by medical professionals or patients at home.

Regulatory Classification

Identification

A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it. The caduceus is a symbol of medicine and healing. The logo also includes the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern around the caduceus symbol. Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 August 3, 2017 Andon Health Co., Ltd. Ms. Liu Yi President No.3 Jin Ping Street Ya An Road, Nankai District Tianjin, 300190 CN Re: K171062 Trade/Device Name: BP5S Wireless Blood Pressure Monitor Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: June 8, 2017 Received: July 3, 2017 Dear Ms. Liu Yi: This letter corrects our substantially equivalent letter of August 1, 2017. We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply {1}------------------------------------------------ Page 2 - Ms. Liu Yi with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely, M.A. Hilleman for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ### Indications for Use 510(k) Number (if known) K171062 Device Name BP5S Wireless Blood Pressure Monitor Indications for Use (Describe) BP5S Wireless Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. The cuff circumference is limited to 22cm-48cm. Type of Use (Select one or both, as applicable) | Prescription Use (Part 21 CFR 801 Subpart D) | <input type="checkbox"/> | |----------------------------------------------|-------------------------------------| | Over-The-Counter Use (21 CFR 801 Subpart C) | <input checked="" type="checkbox"/> | ne-Counter Use (21 CFR 801 Subbart Ci #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # K171062 ## Page 1 of 3 # 510(k) Summary This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirement of SMDA and 21 CFR 807.92. #### 1.0 submitter's information | Name: | Andon Health Co., Ltd. | |----------------------|--------------------------------------------------------------------------| | Address: | No 3, Jinping Street Ya An Road, Nankai District, Tianjin,<br>P.R. China | | Phone number: | 86-22-6052 6161 | | Fax number: | 86-22-6052 6162 | | Contact: | Liu Yi | | Date of Preparation: | 3/31/2017 | #### 2.0 Device information Trade name: BP5S Wireless Blood Pressure Monitor Common name: Noninvasive blood pressure measurement system Classification name: Noninvasive blood pressure measurement system #### 3.0 Classification Production code: DXN- Noninvasive blood pressure measurement system. Regulation number: 870.1130 Classification: II Panel: Cardiovascular #### 4.0 Predicate device information Manufacturer: Andon Health Co., Ltd. Device: KD-936 Fully Automatic Electronic Blood Pressure Monitor 510(k) number: K120672 {4}------------------------------------------------ K171062 Page 2 of 3 ### 5.0 Intended use BP5S Wireless Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. The cuff circumference is limited to 22cm-48cm. The intended use and the indication for use of BP5S, as described in its labeling are the same as the predicate device KD-936 (K120672). ### 6.0 Device description BP5S Wireless Blood Pressure Monitor is designed and manufactured according to IEC 80601-2-30. The operational principle is based on oscillometric and silicon integrates pressure sensor technology. It can calculate the systolic and diastolic blood pressure, and display the result on the LED. The measurements results can also be classified by the function of blood pressure classification indicator. If any irregular heartbeat is detected, it can be shown to the user. ### 7.0 Summary comparing technological characteristics with predicate device | Technological Characteristics | Comparison result | |-------------------------------|-------------------| | Design principle | Identical | | Appearance | Similar | | Patients contact Materials | Similar | | Performance | Similar | | Biocompatibility | Identical | | Mechanical safety | Identical | | Standards met | Identical | | Electrical safety | Identical | | EMC | Identical | | Function | Similar | {5}------------------------------------------------ #### K171062 Page 3 of 3 #### 8.0 Discussion of non-clinical and clinical test performed #### Non-clinical Tests have been done as follows: a. Electromagnetic compatibility test according to IEC 60601-1-2; b. Electrical safety according test to IEC 60601-1; c. Safety and performance characteristics of the test according to IEC 80601-2-30 None of the test demonstrates that BP5S Blood Pressure Monitor bring new questions of safety and effectiveness. #### 9.0 Performance summary BP5S Wireless Blood Pressure Monitor conforms to the following standards: - 60601-1:2005/(R)2012 And A1:2012,C1:2009/(R)2012 And ● IEC A2:2010/(R)2012 (Consolidated Text) Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance (IEC 60601-1:2005, MOD) - · IEC 60601-1-2:2014,Medical Electrical Equipment -- Part 1-2: General Requirements For Basic Safety And Essential Performance -- Collateral Standard: Electromagnetic Disturbances -- Requirements And Tests - IEC 80601-2-30:2009 & A1:2013,Medical Electrical Equipment Part 2-30: Particular Requirements For The Basic Safety And Essential Performance Of Automated Noninvasive Sphygmomanometers #### 10.0 Comparison to the predicate device and the conclusion Our device BP5S is substantially equivalent to the Fully Automatic Electronic Blood Pressure Monitor KD-936 whose 510(k) number is K120672. BP5S is very similar with its predicate device in the intended use, the design principle, the memory capacity, the material, the performance and the applicable standards. Only their appearance is changed, and BP5S can display test results itself. However, the test in this submission provides demonstrates that these small differences do not raise any new questions of safety and effectiveness.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...