Digital Automatic Blood Pressure Monitor BPM41 Series, Digital Automatic Blood Pressure Monitor BPM45 Series

K163648 · Grandway Technology (Shenzhen) Limited · DXN · Jul 7, 2017 · Cardiovascular

Device Facts

Record IDK163648
Device NameDigital Automatic Blood Pressure Monitor BPM41 Series, Digital Automatic Blood Pressure Monitor BPM45 Series
ApplicantGrandway Technology (Shenzhen) Limited
Product CodeDXN · Cardiovascular
Decision DateJul 7, 2017
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1130
Device ClassClass 2

Indications for Use

This device is for use by medical professional or home users. It is intended to measure the systolic and diastolic blood pressure of an adult individual by using a non-invasive technique, in which an inflatable cuff is wrapped around the upper arm.

Device Story

Digital Automatic Blood Pressure Monitor (BPM41 and BPM45 series) measures systolic/diastolic blood pressure and pulse rate via oscillometric method; inflatable cuff wrapped around upper arm detects brachial artery blood movement; converts signals to digital readings displayed on LCD panel. Used in clinical or home settings by medical professionals or patients. Features include IHB (irregular heartbeat) detection, memory storage for multiple users, voice function, alarm clock, and backlight (model dependent). Provides objective BP/pulse data to assist healthcare providers in monitoring cardiovascular status and clinical decision-making.

Clinical Evidence

Clinical validation performed per ISO 81060-2. BPM41 Series study: 100 patients (50 male, 50 female). BPM45 Series study: 100 patients (51 male, 49 female). Reference method: standard auscultation. Measurements taken alternatively in the same arm. Results demonstrate clinical accuracy equivalent to predicate device.

Technological Characteristics

Oscillometric non-invasive blood pressure measurement. Components: pressure sensor, CPU, PCB, amplifiers, transistors, resistances, capacitance. Materials: ABS cabinet/buttons. Power: 1.5V AA or AAA batteries, AC adaptor (6V/600mA). Connectivity: None. Standards: ISO 81060-1:2007, EN 1060-3:1997+A2:2009, IEC 60601-1:2012, IEC 60601-1-2:2014, IEC 60601-1-11:2015, IEC 81060-2:2013, IEC 80601-2-30:2009+A1:2013.

Indications for Use

Indicated for adult individuals to measure systolic and diastolic blood pressure and pulse rate using a non-invasive oscillometric technique with an upper arm cuff. For use by medical professionals or home users.

Regulatory Classification

Identification

A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle with three human profiles superimposed on its body. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 July 7, 2017 GRANDWAY TECHNOLOGY (SHENZEN) LIMITED Mr. Patrick Chow General Manager Block 7. Zhu Keng Industrial Zone. Ping Shan, Long Gang District, Shenzhen, 518118 China Re: K163648 Trade/Device Name: Digital Automatic Blood Pressure Monitor BPM41 Series, Digital Automatic Blood Pressure Monitor BPM45 Series Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: June 7, 2017 Received: June 8, 2017 Dear Mr. Patrick Chow: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely, M.A. Hillemann for for Bram D. Zuckerman, M.D. Director Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration # Indications for Use 510(k) Number (if known) K163648 Device Name Digital Automatic Blood Pressure Monitor BPM41 Series; Digital Automatic Blood Pressure Monitor BPM45 Series Indications for Use (Describe) This device is for use by medical professional or home users. It is intended to measure the systolic and diastolic blood pressure of an adult individual by using a non-invasive technique, in which an inflatable cuff is wrapped around the upper arm. Type of Use (Select one or both, as applicable) ] Prescription Use (Part 21 CFR 801 Subpart D) X Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ #### Submitter identification 1. | 510(k) Submitter | GRANDWAY TECHNOLOGY (SHENZHEN) LIMITED | |--------------------|-------------------------------------------------------------------------------------------------------------------| | Address | Block 7, Zhu Keng Industrial Zone, Ping Shan, Long Gang District, 518118,<br>Shenzhen, People's Republic of China | | Phone Number | (00852)-2851-6789 | | Fax Number | (00852)-2851-6278 | | Contact Person | Mr. Patrick Chow | | Date of Submission | 22-Dec-2016 | #### Device identification 2. | Trade Name | Digital Automatic Blood Pressure Monitor BPM41 Series<br>[Model No.: MD41X0]<br>x --- The character (0, 1, 2, 3, 4, 5, 6, 7, 8, 9, A & B) is for the minor change<br>revision of device. The mentioned "minor change" refers to those device<br>changes not to be affecting the conformity test results of EMC & safety as well<br>as device performance, i.e. IEC 60601-1 and IEC 60601-1-2. | |---------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | Digital Automatic Blood Pressure Monitor BPM45 Series<br>[Model No.: MD45X0]<br>x --- The character (0, 1, 2, 3, 4, 5, 6, 7, 8, 9, A & B) is for the minor change<br>revision of device. The mentioned "minor change" refers to those device<br>changes not to be affecting the conformity test results of EMC & safety as well<br>as device performance, i.e. IEC 60601-1 and IEC 60601-1-2. | | Common Name | Non-invasive Blood Pressure Measurement System | | Classification Name | Non-invasive Blood Pressure Measurement System<br>(CFR 870.1130, Class II, Product Code DXN) | #### Predicate device 3. | Predicate Device | Digital Automatic Blood Pressure Monitor BPM25 & BPM 26 Series | |------------------|----------------------------------------------------------------| | Manufacturer | GRANDWAY TECHNOLOGY (SHENZHEN) LIMITED | | 510(k) Number | K150373 | #### Device Description 4. Digital Automatic Blood Pressure Monitor BPM41 Series and Digital Automatic Blood Pressure Monitor BPM45 Series (Subject Device) are non-invasive blood pressure measurement systems for use by medical professional or at home. It is designed to measure the systolic and diastolic blood pressure, and pulse rate of an individual in each measurement and then display the readings on a digital panel. The Subject Device utilize the oscillometric methodology, in which an inflatable cuff is wrapped around the upper arm of an individual, for blood pressure measurement. This means the monitor detects your blood's movement through your brachial artery and converts the movement into a digital reading. The table below illustrate the feature presence in the Subject Device: {4}------------------------------------------------ K163648 ## GRANDWAY TECHNOLOGY (SHENZHEN) LIMITED 510(k) Summary | Series | Model | LCD type | Backlight | Sound | DC jack | |-------------|--------|-----------------------|-----------|--------|---------| | MD41 Series | MD4100 | Positive Reflective | No | Buzzer | No | | | MD4110 | Positive Reflective | No | Voice | Yes | | | MD4120 | Positive Reflective | No | Buzzer | Yes | | | MD4130 | Positive Reflective | No | Voice | No | | | MD4140 | Negative transmissive | Yes | Buzzer | Yes | | | MD4150 | Negative transmissive | Yes | Buzzer | No | | | MD4160 | Negative transmissive | Yes | Voice | Yes | | | MD4170 | Negative transmissive | Yes | Voice | No | | | MD4180 | Positive transmissive | Yes | Buzzer | Yes | | | MD4190 | Positive transmissive | Yes | Buzzer | No | | | MD41A0 | Positive transmissive | Yes | Voice | Yes | | | MD41B0 | Positive transmissive | Yes | Voice | No | | MD45 Series | MD4500 | Positive Reflective | No | Buzzer | No | | | MD4510 | Positive Reflective | No | Voice | Yes | | | MD4520 | Positive Reflective | No | Buzzer | Yes | | | MD4530 | Positive Reflective | No | Voice | No | | | MD4540 | Negative transmissive | Yes | Buzzer | Yes | | | MD4550 | Negative transmissive | Yes | Buzzer | No | | | MD4560 | Negative transmissive | Yes | Voice | Yes | | | MD4570 | Negative transmissive | Yes | Voice | No | | | MD4580 | Positive transmissive | Yes | Buzzer | Yes | | | MD4590 | Positive transmissive | Yes | Buzzer | No | | | MD45A0 | Positive transmissive | Yes | Voice | Yes | | | MD45B0 | Positive transmissive | Yes | Voice | No | #### 5. Indication for use This device is for use by medical professional or home users. It is intended to measure the systolic and diastolic blood pressure on an adult individual by using a non-invasive technique, in which an inflatable cuff is wrapped around the upper arm. ### 6. Comparison of technological characteristics between the Subject Device and the Predicate Devices The Subject Device are compared to the Predicate device comparison table below: {5}------------------------------------------------ ## K163648 | Item | Predicate Device (K150373) | Subject Device (BPM41 Series) | Subject Device (BPM45 Series) | Comment | |--------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------| | Indication for Use | This device is for use by medical<br>professional or home users. It is<br>intended to measure the systolic and<br>diastolic blood pressure of an adult<br>individual by using a non-invasive<br>technique, in which an inflatable cuff is<br>wrapped around the upper arm. | This device is for use by medical<br>professional or home users. It is<br>intended to measure the systolic and<br>diastolic blood pressure on an adult<br>individual by using a non-invasive<br>technique, in which an inflatable cuff is<br>wrapped around the upper arm. | This device is for use by medical<br>professional or home users. It is<br>intended to measure the systolic and<br>diastolic blood pressure on an adult<br>individual by using a non-invasive<br>technique, in which an inflatable cuff is<br>wrapped around the upper arm. | Identical | | Patient Population | Adult | Adult | Adult | Identical | | Measurement Method | Non-invasive, Oscillometric | Non-invasive, Oscillometric | Non-invasive, Oscillometric | Identical | | BP Measurement Range | Cuff Pressure: 0 - 300 mmHg<br>Systolic Pressure: 50 - 250 mmHg<br>Diastolic Pressure: 30 - 200 mmHg | Cuff Pressure: 0 - 300 mmHg<br>Systolic Pressure: 50 - 250 mmHg<br>Diastolic Pressure: 30 - 200 mmHg | Cuff Pressure: 0 - 300 mmHg<br>Systolic Pressure: 50 - 250 mmHg<br>Diastolic Pressure: 30 - 200 mmHg | Identical | | Resolution of<br>Measurement | Blood Pressure: 1 mmHg or 0.1kPa<br>Pulse Rate: 1 beat/ min | Blood Pressure: 1 mmHg or 0.1kPa<br>Pulse Rate: 1 beat/ min | Blood Pressure: 1 mmHg or 0.1kPa<br>Pulse Rate: 1 beat/ min | Identical | | Blood Pressure<br>Measurement Accuracy | $\pm$ 3 mmHg | $\pm$ 3 mmHg | $\pm$ 3 mmHg | Identical | | Pulse Rate Measurement<br>Range | 40 - 180 beats/min | 40 - 180 beats/min | 40 - 180 beats/min | Identical | | Pulse Rate Measurement<br>Accuracy | $\pm$ 5 % of the reading | $\pm$ 5 % of the reading | $\pm$ 5 % of the reading | Identical | | IHB Detection | Yes | Yes | Yes | Identical | | Number of User | 2 independent users | 4 independent users | 2 independent users | Different (BPM41 Series)1<br>Identical (BPM45 Series) | | Memory Space | 2 users × 120 memory space<br>(Total 240 memory space) | 4 users × 120 memory space<br>(Total 480 memory space) | 2 users x 240 memory space<br>(Total 480 memory space) | Different2 | | Voice function | Present (depends on model) | Present (depends on model) | Present (depends on model) | Identical | | Alarm clock function | Absent | Present | Present | Different3 | | Backlight | Present (depends on model) | Present (depends on model) | Present (depends on model) | Identical | | Display Type | LCD | LCD | LCD | Identical | | Power Source | 4 × 1.5 V AA-batteries; and/or<br>AC adaptor (6V/600mA) | 4 x 1.5 V AAA-batteries; and/or<br>AC adaptor (6V/600mA) | 4 x 1.5 V AA-batteries; and/or<br>AC adaptor (6V/600mA) | Different (BPM41 Series)4<br>Identical (BPM45 Series) | | Pressurization Mode | Automatic Inflation | Automatic Inflation | Automatic Inflation | Identical | | Deflation Mode | Automatic Exhaust/ Deflation | Automatic Exhaust/ Deflation | Automatic Exhaust/ Deflation | Identical | | Item | Predicate Device (K150373) | Subject Device (BPM41 Series) | Subject Device (BPM45 Series) | Comment | | Operating Condition | Temperature: +5 to +40 °C<br>Humidity: 15 to 93 % R.H. max<br>Atmospheric Pressure: 700-1060 hPa | Temperature: +5 to +40 °C<br>Humidity: 15 to 93 % R.H. max<br>Atmospheric Pressure: 700-1060 hPa | Temperature: +5 to +40 °C<br>Humidity: 15 to 93 % R.H. max<br>Atmospheric Pressure: 700-1060 hPa | Identical | | Storage and<br>Transportation Condition | Temperature: -25 to +70 °C<br>Humidity: up to 93% R.H. max<br>Atmospheric Pressure: 700-1060 hPa | Temperature: -25 to +70 °C<br>Humidity: up to 93% R.H. max<br>Atmospheric Pressure: 700-1060 hPa | Temperature: -25 to +70 °C<br>Humidity: up to 93% R.H. max<br>Atmospheric Pressure: 700-1060 hPa | Identical | | Material | Resistances, capacitance, transistors,<br>amplifiers, pressure sensor, CPU, PCB,<br>cuff ABS button, ABS cabinet, batteries<br>and packaging | Resistances, capacitance, transistors,<br>amplifiers, pressure sensor, CPU, PCB,<br>cuff ABS button, ABS cabinet, batteries<br>and packaging | Resistances, capacitance, transistors,<br>amplifiers, pressure sensor, CPU, PCB,<br>cuff ABS button, ABS cabinet, batteries<br>and packaging | Identical | | Compatibility with<br>Environment and Other<br>Devices | No influence with environment and<br>other device | No influence with environment and<br>other device | No influence with environment and<br>other device | Identical | | Applicable Standard | - EN 1060-1:1995+A2:2009<br>- EN 1060-3:1997+A2:2009<br>- IEC 60601-1:2012<br>- EN 60601-1-2:2007<br>- FCC Part 15 Subpart B<br>- IEC 81060-2:2009<br>- IEC 80601-2-30:2009+A1:2013 | - ISO 81060-1:2007<br>- EN 1060-3:1997+A2:2009<br>- IEC 60601-1:2012<br>- IEC 60601-1-2:2014<br>- IEC 60601-1-11:2015<br>- FCC Part 15 Subpart B<br>- IEC 81060-2:2013<br>- IEC 80601-2-30:2009+A1:2013 | - ISO 81060-1:2007<br>- EN 1060-3:1997+A2:2009<br>- IEC 60601-1:2012<br>- IEC 60601-1-2:2014<br>- IEC 60601-1-11:2015<br>- FCC Part 15 Subpart B<br>- IEC 81060-2:2013<br>- IEC 80601-2-30:2009+A1:2013 | Equivalent5 | {6}------------------------------------------------ 1 The number of user betwen the Subject Device is different but this will neither raise any safety issues no affect the esserial performance of the subject device. ? The total memory space between the Predicate Device is different but this will neither raise any safety issues nor affect the essential performance of the subject device. ³ The Subject Device has a new alam this new function will neither raise any safety issues nor affect the essential performance of the subject device. * The battery type used in Subject Device is different but this will neither raise any safety issues nor affect the essential performance of the subject device. 5 The standards used in Subject Device are updated and they are equivalent to the standards used in the Predicate Device. {7}------------------------------------------------ #### Clinical and Non-clinical Tests 7. # Clinical Test Summary Testing to insure clinical accuracy of the device in accordance with ISO 81060-2 as documented in Clinical Test report. For the Subject Device (BPM 41 Series), one hundred patients (50 males and 50 females) were invited for the study. Standard auscultation method was used as the reference blood pressure monitor measuring in the left arm. Blood pressure measurements were repeated alternatively with the device and auscultation in the same arm according to the sequence in ISO 81060-2. For the Subject Device (BPM 45 Series), one hundred patients (51 males and 49 females) were invited for the study. Standard auscultation method was used as the reference blood pressure monitor measuring in the left arm. Blood pressure measurements were repeated alternatively with the device and auscultation in the same arm according to the sequence in ISO 81060-2. ## Non-Clinical Test Summary The Subject Device has performed several non-clinical tests to show that all requirement specifications and standard requirements are met. The tests includes the follows: - ゃ ISO 81060-1:2007 - ゃ EN 1060-3:1997+A2:2009 - ゃ IEC 60601-1:2012 - や IEC 60601-1-2:2007 (for BPM45 Series only) - や IEC 60601-1-2:2014 (for BPM41 Series only) - ゃ IEC 60601-1-11:2015 - や IEC 80601-2-30:2009+A1:2013 All of the clinical and non-clinical testing performed on the Subject Device are same as the Predicate Device. Therefore, no bench test is conducted to show the performance of the Subject Device are equivalent to the Predicate Device. #### Conclusion 8. The Subject Device has the same intended use and same technological characteristics as the Predicate Device. Moreover both clinical and non-clinical testing has demonstrated that no differences in the technological characteristics and questioning on safety or effectiveness to be raised. Thus, the Subject Device are substantially equivalent to the Predicate Device.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...