DIGITAL AUTOMATIC WRIST BLOOD PRESSURE MONITOR SWBPM22 SERIES
Applicant
Grandway Technology (Shenzhen) Limited
Product Code
DXN · Cardiovascular
Decision Date
Oct 31, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1130
Device Class
Class 2
Indications for Use
This device is for use by medical professional or home users. It is intended to measure the systolic and diastolic blood pressure of an adult individual by using a non-invasive technique, in which an inflatable cuff is wrapped around the wrist.
Device Story
Digital Automatic Wrist Blood Pressure Monitor (SWBPM22 Series) measures systolic/diastolic blood pressure and pulse rate via oscillometric method. Device consists of pressure sensor, electric valve, electronic control module, and electric pump. User wraps inflatable cuff around wrist; device automatically inflates/deflates cuff to detect brachial artery blood movement. Readings displayed on digital LCD panel. Used by medical professionals or patients at home. Output provides blood pressure and pulse rate data, aiding clinical assessment of cardiovascular status. Includes WHO classification and irregular heartbeat detection. Benefits include convenient, non-invasive, self-administered blood pressure monitoring.
Clinical Evidence
Clinical study conducted per ISO 81060-2:2009 with 100 patients (46 male, 54 female). Compared subject device against standard auscultation method. Results demonstrated clinical accuracy equivalent to predicate device.
Indicated for adult individuals for non-invasive measurement of systolic and diastolic blood pressure and pulse rate. For use by medical professionals or home users.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
Predicate Devices
Digital Automatic Wrist Blood Pressure Monitor WBPM22 Series (K133618)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of a bird-like figure with three human profiles forming the body.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 31, 2014
Grandway Technology (Shenzhen) Limited Patrick Chow General Manager Block 6 and 7. Zhu Keng Industrial Zone. Ping Shan, Long Gang District, Shenzhen, Guang Dong, China
Re: K142088
Trade/Device Name: Digital Automatic Wrist Blood Pressure Monitor SWBPM22 Series Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: September 30, 2014 Received: September 30, 2014
Dear Patrick Chow,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Melissa A. Torres -S
Bram D. Zuckerman, M.D. For Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K142088
Device Name
Digital Automatic Wrist Blood Pressure Monitor SWBPM22 Series, MD2200, MD2210
Page 1 of 1
Indications for Use (Describe)
This device is for use by medical professional or home users. It is intended to measure the systolic and diastolic blood pressure of an adult individual by using a non-invasive technique, in which an inflatable cuff is wrapped around the wrist.
Type of Use (Select one or both, as applicable)
- Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| 510(k) Submitter | GRANDWAY TECHNOLOGY (SHENZHEN) LIMITED |
|--------------------|---------------------------------------------------------------|
| Address | Block 6 and 7, Zhu Keng Industrial Zone, Ping Shan, Long Gang |
| | District, Shenzhen, Guang Dong, People's Republic of China |
| Phone Number | (00852)-2851-6789 |
| Fax Number | (00852)-2851-6278 |
| Contact Person | Mr. Patrick Chow |
| Date of Submission | 30th July, 2014 |
#### Submitter Identification 1.
#### Device Identification 2.
| Trade Name | Digital Automatic Wrist Blood Pressure Monitor SWBPM22 Series<br>[Model No.: MD22x0]<br>x --- The first character (0 and 1) is for the identification of outlook. |
|---------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name | Non-invasive Blood Pressure Measurement System |
| Classification Name | Non-invasive Blood Pressure Measurement System<br>(CFR 870.1130, Class II, Product Code DXN) |
#### 3. Predicate Device
| Predicate Device | Digital Automatic Wrist Blood Pressure Monitor WBPM22 Series |
|------------------|--------------------------------------------------------------|
| Manufacturer | GRANDWAY TECHNOLOGY (SHENZHEN) LIMITED |
| 510(k) Number | K133618 |
#### Device Description 4.
Digital Automatic Wrist Blood Pressure Monitor SWBPM22 Series is a non-invasive blood pressure measurement system for use by medical professional or at home. It is designed to measure the systolic and diastolic blood pressure, and pulse rate of an individual in each measurement and then display the readings on a digital panel.
The device utilizes the oscillometric methodology, in which an inflatable cuff is wrapped around the wrist of an individual, for blood pressure measurement. This means the monitor detects your blood's movement through your brachial artery and converts the movement into a digital reading.
The table below illustrate the feature presence in Digital Automatic Wrist Blood Pressure Monitor SWBPM22 Series.
| Model | Blood Pressure<br>Measurement | Pulse Rate<br>Measurement | WHO<br>Classification | Irregular<br>Heartbeat | LCD Type | User ×<br>Memory |
|--------|-------------------------------|---------------------------|-----------------------|------------------------|------------------------|------------------|
| MD2200 | ✓ | ✓ | ✓ | ✓ | Positive<br>Reflective | 2 × 120 |
| MD2210 | ✓ | ✓ | ✓ | ✓ | Positive<br>Reflective | 2 × 120 |
{4}------------------------------------------------
#### 5. Indication for Use
This device is for use by medical professional or home users. It is intended to measure the systolic and diastolic blood pressure of an adult individual by using a non-invasive technique, in which an inflatable cuff is wrapped around the wrist.
### Comparison of Technological Characteristics between New Device and 6. Predicate Devices
Digital Automatic Wrist Blood Pressure Monitor SWBPM22 Series is compared to the predicate device, WBPM22 Series (K133618) in the device comparison table below.
| | Comparison between SWBPM22 Series and Predicate device | | | |
|--------------------|--------------------------------------------------------|--------------------------------------------|------------|--|
| Item | Predicate Device | WBPM22 Series | Comment | |
| Indication for Use | Digital Automatic Wrist Blood Pressure | This device is for use by medical | Equivalent | |
| | Monitor WBPM22 Series is for use by | professional or home users. It is intended | | |
| | medical professional or home user. The | to measure the systolic and diastolic | | |
| | WBPM22 Series is intended to measure | blood pressure of an adult individual by | | |
| | the systolic and diastolic blood pressure, | using a non-invasive technique, in which | | |
| | and pulse rate of an adult individual by | an inflatable cuff is wrapped around the | | |
| | using a non-invasive technique, in which | wrist. | | |
| | an inflatable cuff is wrapped around the | | | |
| | wrist. | | | |
| Measurement | Non-invasive, Oscillometric | Non-invasive, Oscillometric | Identical | |
| Method | | | | |
| IHB Detection | Yes | Yes | Identical | |
| Patient Population | Adult | Adult | Identical | |
| BP Measurement | Cuff Pressure: 0 - 300 mmHg | Cuff Pressure: 0 - 300 mmHg | Identical | |
| Range | Systolic Pressure: 50 - 250 mmHg | Systolic Pressure: 50 - 250 mmHg | | |
| | Diastolic Pressure: 30 - 200 mmHg | Diastolic Pressure: 30 - 200 mmHg | | |
| Number of User | 2 independent users | 2 independent users | Identical | |
| Memory Space | 2 users × 60 memory space; | 2 users × 120 memory space | Equivalent | |
| | or 2 users × 120 memory space | | | |
| Resolution of | Blood Pressure: 1 mmHg or 0.1kPa | Blood Pressure: 1 mmHg or 0.1 kPa | Identical | |
| Measurement | Pulse Rate: 1 beat/ min | Pulse Rate: 1 beat/ min | | |
| Blood Pressure | ± 3 mmHg or 2% of reading | ± 3 mmHg | Equivalent | |
| Measurement | | | | |
| Accuracy | | | | |
| Pulse Rate | 30 - 180 beats/min | 30 - 180 beats/min | Identical | |
| Measurement | | | | |
| Range | | | | |
| Pulse Rate | ± 5 % of the reading | ± 5 % of the reading | Identical | |
| Measurement | | | | |
| Accuracy | | | | |
| Display Type | LCD | LCD | Identical | |
| Power Source | 2 × 1.5 V AAA-batteries | 2 × 1.5 V AAA-batteries | Identical | |
| Pressurization | Automatic Inflation | Automatic Inflation | Identical | |
| Mode | | | | |
| Deflation Mode | Automatic Exhaust/ Deflation | Automatic Exhaust/ Deflation | Identical | |
| Operating | Temperature: +5 to +40 °C | Temperature: +5 to +40 °C | Identical | |
| Condition | Humidity: 15 to 93 % R.H. max | Humidity: 15 to 93 % R.H. max | | |
| | Atmospheric Pressure: 700 - 1060 hPa | Atmospheric Pressure: 700 - 1060 hPa | | |
| Storage and | Temperature: -25 to +70 °C | Temperature: -25 to +70 °C | Identical | |
| Transportation | Humidity: up to 93% R.H. max | Humidity: up to 93% R.H. max | | |
| Condition | Atmospheric Pressure: 700 - 1060 hPa | Atmospheric Pressure: 700 - 1060 hPa | | |
{5}------------------------------------------------
| Comparison between SWBPM22 Series and Predicate device | | | |
|--------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Item | Predicate Device | WBPM22 Series | Comment |
| Material | Resistances, capacitance, transistors, amplifiers, pressure sensor, CPU, PCB, cuff ABS button, ABS cabinet, batteries and packaging | Resistances, capacitance, transistors, amplifiers, pressure sensor, CPU, PCB, cuff ABS button, ABS cabinet, batteries and packaging | Identical |
| Compatibility with Environment and Other Devices | No influence with environment and other device | No influence with environment and other device | Identical |
| Applicable Standard | - EN 1060-1:1995+A2:2009<br>- EN 1060-3:1997+A2:2009<br>- IEC 60601-1:2012<br>- EN 60601-1-2:2007<br>- FCC Part 15 Subpart B<br>- ISO 10993-5:2009<br>- ISO 10993-10:2010<br>- IEC 62304:2006<br>- IEC 81060-2:2009 | - EN 1060-1:1995+A2:2009<br>- EN 1060-3:1997+A2:2009<br>- IEC 60601-1:2012<br>- IEC 60601-1-2:2007<br>- FCC Part 15 Subpart B<br>- ISO 10993-5:2009<br>- ISO 10993-10:2010<br>- IEC 62304:2006<br>- IEC 81060-2:2009 | Equivalent |
Digital Automatic Wrist Blood Pressure Monitor SWBPM22 Series is a non-invasive measuring device and utilizes the oscillometric methodology to measure the blood pressure reading. The key components of device are: a pressure sensor, a electric valve and an electronic control module together with an electric pump. The electric pump inflate (and deflate) the inflatable cuff automatically according to our designed architecture. The predicate device adopts exactly same methodology and key components for measuring blood pressure.
#### Clinical and Non-clinical Tests 7.
## Clinical Test Summary
Testing to insure clinical accuracy of the device in accordance with ISO 81060-2:2009 as documented in Clinical Test report.
One hundred patients (46 males and 54 females) were invited for the study. Standard auscultation method was used as the reference blood pressure monitor measuring in the left wrist. Blood pressure measurements were repeated alternatively with the device and auscultation in the same arm according to the sequence in ISO 81060-2:2009.
## Non-Clinical Test Summary
Digital Automatic Wrist Blood Pressure Monitor SWBPM22 Series has performed several nonclinical tests to show that all requirement specifications and standard requirements are met. The tests includes the follows:
- EN 1060-1:1995+A2:2009 <>
- >> EN 1060-3:1997+A2:2009
- ◆ IEC 60601-1:2012
- や IEC 60601-1-2:2007
- や FCC Part 15 Subpart B
- ◆ ISO 10993-5:2009
- >> ISO 10993-10:2010
- >> IEC 62304:2006
{6}------------------------------------------------
All of the clinical and non-clinical testing performed on Digital Automatic Wrist Blood Pressure Monitor SWBPM22 Series are same as the predicate device.
Also, bench testing, IEC 80601-2-30, is conducted to show the performance of Digital Automatic Wrist Blood Pressure Monitor SWBPM22 Series is equivalent to the predicate device.
#### Conclusion 8.
Digital Automatic Wrist Blood Pressure Monitor SWBPM22 Series has the same intended use and same technological characteristics as the predicate device, Digital Automatic Wrist Blood Pressure Monitor WBPM22 Series (K133618). Moreover both clinical and non- clinical testing has demonstrated that no differences in the technological characteristics and questioning on safety or effectiveness to be raised. Thus, Digital Automatic Wrist Blood Pressure Monitor SWBPM22 Series is substantially equivalent to the predicate device.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.