← Product Code [DXN](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DXN) · K121355

# WRIST-TYPE FULLY AUTOMATIC DIGITAL BLOOD PRESSURE MONITOR (K121355)

_Sejoy Electronics & Instruments Co., Ltd. · DXN · Jul 18, 2012 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DXN/K121355

## Device Facts

- **Applicant:** Sejoy Electronics & Instruments Co., Ltd.
- **Product Code:** [DXN](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DXN.md)
- **Decision Date:** Jul 18, 2012
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1130
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Pediatric

## Indications for Use

The Wrist-type Fully Automatic Blood Pressure Monitors BP series are intended for used by adults with 12 years and older to measure the systolic blood pressure and pulse rate.

## Device Story

Wrist-type fully automatic digital blood pressure monitor; uses inflatable cuff wrapped around patient's wrist. Internal pump automatically inflates cuff; oscillometric method and silicon integrated pressure sensor technology determine systolic/diastolic blood pressure and pulse rate. Deflation rate controlled by preset mechanical valve. Device features maximum pressure safety setting at 300 mmHg. Measurements displayed on LCD for three minutes; device turns off automatically after three minutes of inactivity. Select models (BP-202H, BP-2208) compare results against WHO blood pressure classifications; select models (BP-202H, BP-2206, BP-2208) detect irregular heartbeats. BP-2208 provides average results for all, AM, and PM measurements; BP-202H and BP-2206 calculate average of last three measurements. Used by adults in home or clinical settings. Output allows users to monitor blood pressure trends and identify potential hypertension or irregular heartbeats, facilitating clinical decision-making by healthcare providers.

## Clinical Evidence

Clinical tests performed in compliance with ANSI/AAMI SP10:2002/(R) 2008 & A1:2003/(R) 2008 & A2:2006/(R) 2008 standards for manual, electronic, or automated sphygmomanometers. Results demonstrated compliance with accuracy requirements.

## Technological Characteristics

Wrist-type oscillometric blood pressure monitor; silicon integrated pressure sensor; internal pump; mechanical deflation valve. Powered by 2 AAA batteries. LCD display. Models vary in memory capacity, averaging functions, and backlight. Complies with ANSI/AAMI SP10:2002, IEC/EN 60601-1, and IEC/EN 60601-1-2. Biocompatibility per ISO 10993-5 and ISO 10993-10.

## Regulatory Identification

A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.

## Predicate Devices

- AM/PM Memory Wrist Blood Pressure Monitor (Model: UB-512) ([K042967](/device/K042967.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K121355

page 1/5

**SEJOY**

丰田子有限公司 Sejoy Electronics & Instruments Co., Ltd.

# 510(k) Summary

JUL 18 2012

The assigned 510(k) number is:

#### 1. Date Prepared:

2011.11.25

#### 2. Submitter's Identification:

Name: Sejoy Electronics & Instruments Co., Ltd.

Add .: Building 2, No.202, Zhenzhong Road, West Lake Economy & Technology Zone, Hangzhou, China 310030 Contact Person: Yunhua Ren Phone: +86-571-81957767 Fax: +86-571-81957750 Email: info@sejoy.com

#### 3. Name of the Device:

Trade Name: Wrist-type Fully Automatic Digital Blood Pressure Monitor Including the following models: - BP-201M, BP-202H, BP-202N, BP-2206, BP-2208 Common Name: Blood Pressure Monitor Classification name: Non-invasive blood pressure measurement System 21 CFR 870-1130, Class II, 74-DXN.

#### Classification Information: 4.

Regulation Number: 870.1130 Product Code: DXN Device Class: II Panel: 74 Cardiovascular

#### న్. Predicate Device Information:

The Wrist-type Fully Automatic Digital Blood Pressure Monitors are substantially equivalent to the following device:

7

{1}------------------------------------------------

page 2/5

Image /page/1/Picture/2 description: The image shows the word "SEJOY" in a bold, sans-serif font. The letters are black against a white background. There is a small, triangle-like mark above the letter 'J'.

世佳电子有限公司 Sejoy Electronics & Instruments Co., Ltd.

AM/PM Memory Wrist Blood Pressure Monitor (Model: UB-512), FDA 510(k) K042967, distributed by A division of A&D Engineering, Inc.

#### Device Description: 6.

The wrist-type fully automatic digital blood pressure monitors uses an inflatable cuff which is wrapped around the patient's wrist. The cuff is inflated automatically by an internal pump in the device. The systolic and diastolic blood pressures are determined by oscillometric method and silicon integrate pressure sensor technology. The deflation rate is controlled by a preset mechanical valve at a constant rate. The pressure of the cuff is completely released automatically at the end of the measurement. At the same time, the measurements are displayed on the LCD display for three minute. There is a maximum pressure safety setting at 300 mmHg. The device will not inflate the cuff higher than 300 mmHg. For BP-202H, BP-2208, the blood pressure results are compared with WHO (World Health Organization) Blood Pressure classification, which are severe Hypertension, Moderate Hypertension, Mild Hypertension, High-normal, Normal, and Optimal. The corresponding LCD segment will be turned on along with the systolic, diastolic, and pulse rate information. BP-202H, BP-2206, BP-2208 will display an irregular heartbeat symbol " if an irregular heartbeat was detected during the measurement process. BP-2208 can display average results in three ways: the average of all measurements, the average of all AM (5:00 AM-8:59AM) measurements, and the average of all PM (18:00 PM-19:59 PM) measurements. BP-202H, BP-2206 can calculate the average of the last three measurements. In addition, BP-2208 have LCD backlight. After three minutes without operation, the blood pressure monitors turn off automatically.

| Features<br>Models | A | B                              | C | D | E | F | G | H            | I            | J     | K |
|--------------------|---|--------------------------------|---|---|---|---|---|--------------|--------------|-------|---|
| BP-201M            | ● | 120 Memories                   | ● | ○ | ○ | ○ | ○ | 76×67.5×28.5 | 13.5-21.5 cm | 45×30 | ○ |
| BP-202N            | ● | 120 Memories in<br>four groups | ● | ○ | ○ | ○ | ○ | 76×67.5×28.5 | 13.5-21.5 cm | 45×30 | ○ |
| BP-202H            | ● | 120 Memories in<br>four groups | ● | ○ | ● | ● | ● | 76×67.5×28.5 | 13.5-21.5 cm | 45×30 | ○ |
| BP-2206            | ● | 120 Memories in<br>two groups  | ● | ○ | ● | ● | ● | 77×64×32.5   | 13.5-21.5 cm | 45×30 | ○ |
| BP-2208            | ● | 120 Memories in<br>two groups  | ● | ● | ● | ○ | ● | 77×64×32     | 13.5-21.5 cm | 49×38 | ● |

The detail comparisons among devices are listed below:

A = Powered by 2 AAA batteries

B = Memory Size

C = Time & Date

{2}------------------------------------------------

K121355

**SEJOY**

世佳电子有限公司

# Sejoy Electronics & Instruments Co., Ltd.

- D = Results Average in Three way
E = WHO (World Health Organization) Classification Indicator

- F = Last 3 Results Average
- G = Irregular Heartbeat Detection
- H = Outside Demission (L x W x H in mm)
- I = Cuff Size
- J = LCD Size (Viewing Area in mm)

K = LCD Backlight

• = Yes

o = No

The devices are all designed and manufactured according to ANSI/AAMI SP10:2002/(R) 2008 & ANSI/AAMI SP10:2002/A1:2003/(R) 2008 & ANSI/AAMI SP10:2002/A2:2006/(R) 2008. manual, electronic or automated sphygmanometers.

### 7. Intended Use:

The Wrist-type Fully Automatic Blood Pressure Monitors BP series are intended for used by adults with 12 years and older to measure the systolic blood pressure and pulse rate.

The intended user and the indication for use of the Wrist-type Fully Automatic Blood Pressure Monitors BP series as described in the labeling are nearly the same as their predicated devices, AM/PM Memory Wrist Blood Pressure Monitor (Model: UB-512).

#### 8. Summary comparing technological characteristics with predicate device:

| Technological Characteristics | Comparison result |
|-------------------------------|-------------------|
| Design principle              | Identical         |
| Appearance                    | Similar           |
| Patients contact Materials    | Identical         |
| Performance                   | Similar           |
| Biocompatibility              | Identical         |
| Mechanical Safety             | Identical         |
| Energy Source                 | Similar           |
| Standards Met                 | Identical         |
| Electrical Safety             | Identical         |
| EMC                           | Identical         |
| Function                      | Similar           |

The difference of technological characteristic between the predicate device and the submit

{3}------------------------------------------------

page 4/5

Image /page/3/Picture/2 description: The image shows the word "SEJOY" in bold, black letters. There is a small symbol above the "J" that resembles a triangle or arrow pointing upwards. The letters are closely spaced together, and the overall appearance is clean and simple.

世佳电子有限公 Sejoy Electronics & Instruments Co., Ltd.

wrist-type fully automatic blood pressure monitors is the appearance, the performance and the function.

## 9. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence are as follows:

Compliance to applicable voluntary standards including ANSI/AAMI SP10:2002/(R) 2008 & ANSI/AAMI SP10:2002/A1:2003/(R) 2008 & ANSI/AAMI SP10:2002/A2:2006/(R) 2008 as well as IEC/EN 60601-1: 2005+CORR.1 (2006) +CORR.2 (2007) / EN 60601-1:2006 /AC: 2010, IEC/EN 60601-1-2:2007, ISO 10993-5: 2009 Biological evaluation of medical devices -Part 5:Tests for in vitro cytotoxicity and ISO 10993-10:2010 Biological evaluation of medical devices Part 10: Tests for irritation and skin sensitization.

Guidance documents include the "FDA Non-invasive Blood Pressure (NIBP) Monitor Guidance" and "FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" and "FDA Bluebook Memorandum G95-1Use of International Standard ISO 10993."

### Non-clinical Tests:

- . Electromagnetic Compatibility Test according to IEC/EN 60601-1-2:2007
- General Safety Provisions Test according to IEC/EN 60601-1: 2005+CORR.1 (2006) +CORR.2 (2007) / EN 60601-1:2006 /AC: 2010
- . Performance Test according to ANSI/AAMI SP10:2002/(R) 2008 & ANSI/AAMI SP10:2002/A1:2003/(R) 2008 & ANSI/AAMI SP10:2002/A2:2006/(R) 2008. The test result all meet or exceed the requirement of the standards.
- Biocompatibility Test according to FDA Bluebook Memorandum G95-1Use of International Standard ISO 10993, ISO 10993-5: 2009 Biological evaluation of medical devices-Part 5: Tests for in vitro cytotoxicity and ISO 10993-10:2010 Biological evaluation of medical devices Part 10: Tests for irritation and skin sensitization.

### 10. Discussion of Clinical Tests Performed:

Clinical tests were performed and complied the accuracy requirements of ANSI/AAMI SP10:2002/(R) 2008 & ANSI/AAMI SP10:2002/A1:2003/(R) 2008 & ANSI/AAMI SP10:2002/A2:2006/(R) 2008 "National Standard for Manual, Electronic or Automated Sphygmomanometers".

#### 11. Conclusions:

Our Wrist-type Fully Automatic Blood Pressure Monitor BP-201M, BP-202H, BP-202N, BP-2206, BP-2208 have the same intended use and similar technological characteristics as the

1 ()

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the word "SEJOY" in bold, black letters. The letters are large and spaced closely together. The background is white, providing a strong contrast to the text.

K121355 电子有限公司 Sejoy Electronics & Instruments Co., Ltd.

A & D AM/PM Memory Wrist Blood Pressure Monitor (Model: UB-512) whose 510 (k) number is K042967.

Moreover, verification and validation tests contained in this submission demonstrate that the difference in the submitted models could maintain the same safety and effectiveness as that of cleared devices.

In the other words, those engineering difference do not: (1) affect the intended use or (2) alter the fundamental scientific technology of the device, and the new models as mentioned on this submission are considered substantial equivalent to the predicate devices.

{5}------------------------------------------------

### DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle or bird-like symbol with three curved lines forming its body and wings. The logo is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the emblem.

### Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

JUL 18 2012

Sejoy Electronics & Instruments Co., Ltd. c/o Mr. Edmondo Roark 1259 Monten Street #4 Cincinnati, OH 45208

Re: K121355

Trade/Device Name: Wrist-type Fully Automatic Digital Blood Pressure Monitor Wrist-type Funy Automatio Digital (Five Models: BP-201M, BP-202H, BP-202N, BP-2206, BP-2208) Regulatory Number: 21 CFR 870.1130 Regulation Name: Noninvasive blood pressure measurement system Regulatory Class: II (two) Product Code: DXN Dated: April 6, 2012 Received: May 4, 2012

Dear Mr. Roark:

We have reviewed your Section 510(k) premarket notification of intent to market the indication We have reviewed your Section 310(x) premaited is substantially equivalent (for the indications 
referenced above and have determined the device is substanted in interstate referenced above and have ustchillied the arreed predicate devices marketed in interstate for use stated in the enclosure) to regally manced the Medical Device Amendments, or to
commerce prior to May 28, 1976, the enactment date of the Federal Food. Drug commerce prior to May 28, 1970, the chacanene with the provisions of the Federal Food. Drug, devices that have been reciassified in accordance that in proval application (PMA).
and Cosmetic Act (Act) that do not require approval or a provisions of the Act . The and Cosmetic Act (Act) that do not require approvate of the general controls provisions of the Act. The Act. The You may, therefore, market the device, subject to incements for annual registration, listing of general controls provisions of the Act mender reading and prohibitions against misbranding and devices, good manufacturing placity, label information related to contract liability.
adulteration. Please note: CDRH does not evaluate in trutiful and not mislead adulteration. Please note: CDRF does not evaluate information of the many of misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it 
and he If your device is classified (See above) into enner elass if (Spoth
may be subject to additional controls. Existing major regulations affecting your device can be

{6}------------------------------------------------

Page 2 - Mr. Roark

found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Bram D. Zuckerman, M.D.

Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

K121355

SFÍO

世佳电子有限公司 Sejoy Electronics & Instruments Co., Ltd.

# Indications for Use

510(k) Number (if known): K 12 / 355

Device Name:

Blood Pressure Monitors BP series

Indications For Use:

Measure blood pressure (systolic and diastolic) and pulse rate.

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use V (21 CFR 801 Subpart C)

5

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 1 of 1

Ivision Sign-Off) Division of Cardiovascular Devices

510(k) Number

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DXN/K121355](https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DXN/K121355)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
