← Product Code [DXN](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DXN) · K112690

# BLOOD PRESSURE MONITOR (K112690)

_Nihon Seimitsu Sokki Co., Ltd. · DXN · Feb 24, 2012 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DXN/K112690

## Device Facts

- **Applicant:** Nihon Seimitsu Sokki Co., Ltd.
- **Product Code:** [DXN](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DXN.md)
- **Decision Date:** Feb 24, 2012
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1130
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

WSK-1011 system is intended for noninvasive measurement of systolic and diastolic blood pressure, determination of pulse rate and calculation of pulse pressure in adults in a homecare environment. The device features include display of irregular pulse rhythm detection, classification display of measured blood pressure values against WHO (World Health Organization) guidelines, display of body movement detection and two memory account to save measurement results.

## Device Story

Automatic wrist-worn sphygmomanometer for home use; utilizes oscillometric method to measure systolic/diastolic blood pressure and pulse rate. Device consists of main unit and preformed nylon cuff (wrist circumference 125-225 mm); powered by two AAA batteries. Input: pressure oscillations against cuff. Processing: automated analysis of pulse waves to determine blood pressure, pulse rate, pulse pressure, irregular rhythm, and body movement. Output: digital display of readings, WHO-based blood pressure classification, and memory storage (2 banks). Operated by patient; provides data for self-monitoring of cardiovascular health.

## Clinical Evidence

No clinical data submitted. Substantial equivalence supported by non-clinical bench testing in accordance with IEC and SP-10 standards.

## Technological Characteristics

Oscillometric measurement; automatic air pump/electric control valve; 15-digit LCD; ABS and PMMA housing; preformed nylon cuff; 2 AAA battery power; wrist-worn form factor. No connectivity. Software performs signal processing for oscillation analysis, rhythm regularity, and movement detection.

## Regulatory Identification

A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.

## Predicate Devices

- Wrist blood Pressure Monitor, Model WS-1100/WS-1100PV ([K080177](/device/K080177.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K112690

#### 5. 510(K) Summary

This document was prepared in accordance with 21 CFR 807.92.

Section (a)

- (1) Name of the submitter: Nihon Seimitsu Sokki Co., Ltd. Address: 2508-13 Nakago, Shibukawa, Gunma 377-0293, Japan Telephone number: 81-279-20-2311 Contact person: Mitsuo Kanai Date of documentation: September 14, 2011
- (2) Trade name of the device: Blood Pressure Monitor WSK-1011 Common name: Blood pressure monitor Classification: Noninvasive blood pressure measurement system, DXN, 870.1130, Class II
- (3) Predicate devices: Wrist blood Pressure Monitor, Model WS-1100/WS-1100PV, K080177, Nihon Seimitsu Sokki, Co., Ltd.

#### (4) Description of the device:

Blood Pressure Monitor WSK-1011 is an automatic sphygmomanometer to be used in a homecare environment. Blood pressure, systolic and diastolic, and pulse rate are taken at wrist non-invasively using oscillometric method, which is one of the most common methods with recent automatic sphygmomanometer that determines blood pressure and pulse rate with oscillations against cuff applied to measurement site. The device consists of the main unit and the cuff that is applicable to wrist circumference between 4.9 and 8.9 inches (between 125 and 225 mm), singly mounted. The device is powered by two AAA alkaline batteries. The device not only determines blood pressure and pulse rate from oscillations but also analyses pulse wave and determines whether measurement was made with or without body movement and regularity of pulse rhythm. Besides these auxiliaries, user can get pulse pressure value and blood pressure level according to WHO guideline also on the display. User can chose to activate clock function of the device to review measured readings with measurement date and time.

{1}------------------------------------------------

- (5) Intended use of the device:
WSK-1011 system is intended for noninvasive measurement of systolic and diastolic blood pressure, determination of pulse rate and calculation of pulse pressure in adults in a homecare environment.

The device features include display of irregular pulse rhythm detection, classification display of measured blood pressure values against WHO (World Health Organization) guidelines, display of body movement detection and two memory account to save measurement results.

The indication for use of WSK-1011 system is not exactly same with but similar to the predicate device. The fundamental intended use, which is to measure adults' blood pressure non-invasively in home care environment, remains the same. The difference between the indications for use of the subject device and the predicate device lies in supplemental product features of the predicate device are not provided with the new device and some new features are introduced with the subject device. As demonstrated in relevant sections of this submission, these features are concluded not to affect the device safety and effectiveness.

- (6) Technological characteristics of the subject device and the predicate device:
The subject device holds the same technological fundamentals with the predicate device. Both devices consist of the single-mounted main unit and cuff for wrist and powered with two AAA alkaline batteries. The patient contacting materials used for the subject device had been used with the predicate device and these materials do not go through any different process from the predicate devices. The list of patient contacting materials and components is included in the relevant section of this application.

{2}------------------------------------------------

(7) The following Device comparison table,shows the details of differences between the subject device and the claimed predicats.

|                                      | WSK-1011, the subject device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | WS-1100/WS-1100PV<br>(K080177)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Note |
|--------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------|
| Intended use                         | WSK-1011 system is intended for<br>noninvasive measurement of systolic and<br>diastolic blood pressure, determination of<br>pulse rate and calculation of pulse<br>pressure in adults in a homecare<br>environment.<br>The device features include display of<br>irregular pulse rhythm detection,<br>classification display of measured blood<br>pressure values against WHO (World<br>Health Organization) guidelines, display<br>of body movement detection and two<br>memory account to save measurement<br>results. | WS-1100/WS-1100PV system is<br>intended for noninvasive<br>measurement of systolic and<br>diastolic blood pressure and<br>determination of pulse rate in adults<br>in a homecare environment.<br>The device features include the<br>display of irregular pulse rhythm<br>detection, the classification display of<br>measured blood pressure values<br>against the guideline by World Health<br>Organization or equivalent guideline,<br>the personal setting for individual<br>blood pressure target values, the two<br>memory banks to save the<br>measurement results with date and<br>time of measurement and the<br>transferring the saved results to<br>personal computers. |      |
| Operation<br>principle               | Oscillometric method                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Oscillometric method                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Same |
| Pressure<br>indication<br>range      | 3 to 300 mmHg                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | 0 to 300 mmHg                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |      |
| Measurement<br>site and cuff<br>size | Wrist<br>Regular size (Approximately 4.9 to 8.9<br>inches; 125 to 225 mm)                                                                                                                                                                                                                                                                                                                                                                                                                                                | Wrist<br>Regular size (Approximately 4.9 to<br>8.0; 125 to 205 mm)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Same |
| Power source                         | 2 AAA alkaline batteries                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | 2AAA alkaline batteries                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Same |
| Inflation                            | Automatic air pump                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Automatic air pump                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Same |
| Deflation                            | Automatic electric control valve                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Automatic electric control valve                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Same |
| Exhaust                              | Automatic quick exhaust valve                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Automatic quick exhaust valve                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Same |
| Display                              | 15 digits liquid crystal display<br>Date and time display                                                                                                                                                                                                                                                                                                                                                                                                                                                                | 15 digits liquid crystal display<br>Date and time display                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Same |
| Device setting<br>by the user        | Date and time                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Date and time<br>Personal target limits of blood<br>pressure values                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |      |
| Memory<br>features                   | 2 memory banks to save 60<br>easurement results with date and time,<br>when the clock is activated; the saved<br>readings can be intentionally deleted by<br>the user.                                                                                                                                                                                                                                                                                                                                                   | 2 memory banks to save 60 blood<br>pressure and pulse rate readings<br>each with date and time; saved<br>reading(s) can be intentionally<br>deleted by the user.<br>Mode to view AM and PM readings<br>separately<br>Exporting saved readings to personal<br>computers using the designated USB<br>cable                                                                                                                                                                                                                                                                                                                                                                          |      |
| Main unit                            | Size: approximately W;2.76 x D;1.06 x<br>H;2.76 inches<br>Weight: approximately 3.9 oz.<br>Material: ABS and PMMA                                                                                                                                                                                                                                                                                                                                                                                                        | Approximate size: W; 2.79, D; 2.56,<br>H; 1.18 inches<br>Approximate weight: 4.09 oz.<br>Material: ABS and PMMA                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |      |
| Cuff type                            | Preformed nylon cuff                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Preformed nylon cuff                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Same |

# Device comparison table

{3}------------------------------------------------

4/4

## Section (b)

(1) Non-clinical tests

The subject device was evaluated in accordance with IEC and SP-10 not only to demonstrate the substantial equivalence but also to establish the better safety. This is why some reference standards were added to test the subject device when compared to the predicated devices. The detailed information of reference standards is provided in the relevant sections of this submission.

# (2) Clinical tests

No clinical test report is submitted because differences between the subject device and the predicate devices do not affect clinical performance.

# (3) Conclusions drawn from non-clinical tests

It is concluded from the non-clinical tests conducted that the subject device is not only as safe and as effective as the predicate devices but is also safer and more effective than the predicate devices.

{4}------------------------------------------------

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. The logo is black and white.

#### Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

FEB 2 4 2012

Nihon Seimitsu Sokki Co., Ltd. c/o Mr. Koji Kubo Manager Cosmos Corporation 6-5-3 Beaunce Honkomagome 2F, Honkomagome Bunkyo-ku, Tokyo

JAPAN 113-0021

Re: K112690

Trade/Device Name: Blood Pressure Monitor, Model WSK-1011 Regulatory Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II (Two) Product Code: DXN Dated: February 14, 2012 Received: February 16, 2012

Dear Mr. Kubo:

We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your becamined the device is substantially equivalent (for the indications ferenced above and nave determined by marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device American son to cominerce prov to May 20, 1970, the enaocance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The You may, therefore, market the act include requirements for annual registration, listing of general controls provisions of the 110t liveling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability adulteration. Frease note: CDNT does not evaluate in the truthful and not misleading.

{5}------------------------------------------------

#### Page 2 - Mr. Koji Kubo

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

-fs

Sincerely yours,

Brom D. Zuckerman, M.D.

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

510(K) Number: K112690

Blood Pressure Monitor WSK-1011 Device Name:

### Indications for Use:

Wrist Blood Pressure Monitor WSK-1011 is intended for noninvasive measurement of systolic and diastolic blood pressure, determination of pulse rate and calculation of pulse pressure in adults in a homecare environment.

The device features include display of irregular pulse rhythm detection, classification display of measured blood pressure values against WHO (World Health Organization) guidelines, display of body movement detection and two memory account to save measurement results.

Prescription Use (Per 21 CFR 801.109 Subpart D)

X AND / OR Over-The Counter Use _ (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

7

(Division Sign-Off) Division of Čardiovascular Device

510(k) Number K112690

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DXN/K112690](https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DXN/K112690)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
