TRANSTEK BLOOD PRESSURE MONITOR MODEL TMB-986, TMB-987, TMB-995
Applicant
Zhongshan Transtek Electronics Co., Ltd.
Product Code
DXN · Cardiovascular
Decision Date
May 27, 2011
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 870.1130
Device Class
Class 2
Indications for Use
Transtek Blood Pressure Monitor, Models TMB 986, TMB987, and TMB995 are digital monitors intended for use in measuring blood pressure and heartbeat rate in adult patient population with arm circumference ranging from 22 cm to 42 cm (about 9 - 17 inches). These devices detect the appearance of irregular heartbeats during measurement and give a warning signal with readings. The Blood Pressure Monitor compares average blood pressure results to pre-established AHA (American Heart Association) hypertension guideline of 135/85 mmHg. Transtek Blood Pressure Monitor, TMB-986, TMB-987, and TMB-995 models are not intended to be a diagnostic device. Contact your physician if hypertensive values are indicated.
Device Story
Digital oscillometric blood pressure monitor; measures systolic/diastolic pressure and heart rate via inflatable arm cuff. Input: cuff pressure oscillations detected by electronic sensor. Processing: microprocessor analyzes pressure signals to derive blood pressure/heart rate; compares pulse wave intervals to detect irregular heartbeats (threshold >25% difference). Output: LCD displays readings and irregular heartbeat warning; compares results to AHA 135/85 mmHg guideline. Used in home/clinical settings by patients or clinicians. Powered by batteries or AC adapter. Provides non-diagnostic monitoring; alerts users to consult physicians for hypertensive readings.
Clinical Evidence
Bench testing only. Performance verified through biocompatibility, electromagnetic compatibility, and electrical safety testing. Accuracy validated against standard oscillometric requirements.
Technological Characteristics
Oscillometric measurement principle; electronic pressure sensor; microprocessor-controlled. ABS plastic housing; nylon cuff with inflatable bladder. Powered by 4x AAA/AA batteries or 6V DC adapter. Digital LCD display. Continuous operation mode. No wireless connectivity.
Indications for Use
Indicated for adult patients with arm circumferences 22-42 cm for non-invasive measurement of systolic/diastolic blood pressure and heart rate. Includes irregular heartbeat detection. Not for diagnostic use.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
Predicate Devices
OMRON, Automatic Blood Pressure Monitor, Model HEM 780N3 (K061822)
Submission Summary (Full Text)
{1}------------------------------------------------
TRANSTEK
# Section 5 - 510(k) Summary
Date of Summary Preparation: 11/09/2010
#### 1. Submitter's Identifications
Submitter's Name: ZHONGSHAN TRANSTEK ELECTRONIC CO., LTD. Address: Jinbiao Road, Minzhong, Zhongshan City, Guangdong, China Contact Person: Guo Zhiping Contact Email Address: gzp@transtek.cn Telephone: 086-760-88282982 ext. 231 Fax: 086-760-85339231
#### 2. Correspondent's Identifications
Correspondent's Name: A03 Lab of BTS Address: No.1 Fanghua Street, Hi-tech Zone, Chengdu City, Sichuan, China Contact Person: Leo Wang Contact Email Address: leo.w(@hibts.com Telephone: 086-28-86083300 Fax: 086-4006070050 ext. 8
#### 3. Name of the Device
Device Classification Name: System, Measurement, Blood-Pressure, Non-invasive Product Name: Blood Pressure Monitor Trade Name: Transtek Models: TMB-986, TMB-987, TMB-995 Classification Panel: Cardiovascular Common/Usual Name: Automatic Blood Pressure Monitor Product Code: DXN Device Classification: Class II Contraindications: None
#### 4. The Predicate Devices
OMRON, Automatic Blood Pressure Monitor, Model HEM 780N3, K061822
#### 5. Device Description
Transtek Blood Pressure Monitor, TMB-986, TMB-987, and TMB-995 are designed to measure the systolic and diastolic blood pressure and heartbeat rate of an individual by using a non-invasive technique in which an inflatable cuff is wrapped around the arm.
Our method to define systolic and diastolic pressure is similar to the auscultatory method but uses
### MAY 2 7 2011
{2}------------------------------------------------
# TRANSTE
an electronic pressure sensor rather than a stethoscope and mercury manometer. The sensor converts tiny alterations in cuff pressure to electrical signals, by analyzing those signals to define the systolic and diastolic blood pressure and calculating heartbeat rate, which is a well-known technique in the market called the "Oscillometric method".
Transtek Blood Pressure Monitor is single-mounted devices of the main unit and cuff unit. ABS is used to outer housing of the main unit. The preformed cuff unit, which is applicable to arm circumference approximately between 22 and 42 cm, includes the inflatable bladder and nylon shell. All models of the arm blood pressure monitor can use the same two size of cuff (22 - 32cm, or 22 - 42cm). Product package will contains only one cuff and which size is decide by business requirement. The device consists of the microprocessor, the pressure sensor, the operation keys, the pump, the electromagnetic deflation control valve and the LCD. The subject devices are powered by four AAA or AA alkaline batteries or by a DC 6V 400mA adapter. I
The device also compares the longest and the shortest time intervals of detected pulse waves to mean time interval and displays a warning signal with the reading to indicate the detection of irregular heartbeat when the difference of the time intervals is over 25%.
#### 6. Intended Use of Device
Transtek Blood Pressure Monitor, Models TMB 986, TMB987, and TMB995 are digital monitors intended for use in measuring blood pressure and heartbeat rate in adult patient population with arm circumference ranging from 22 cm to 42 cm (about 9 - 17 inches).
These devices detect the appearance of irregular heartbeats during measurement and give a warning signal with readings.
The Blood Pressure Monitor compares average blood pressure results to pre-established AHA (American Heart Association) hypertension guideline of 135/85 mmHg.
Transtek Blood Pressure Monitor, TMB-986, TMB-987, and TMB-995 models are not intended to be a diagnostic device. Contact your physician if hypertensive values are indicated.
#### 7. Summary of Substantial Equivalence
Table 1: The difference between Transtek Blood Pressure Monitor
| | The Similarities and Difference among TMB-986, TMB-987 and TMB-995 | | | | | |
|----|--------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|--|
| | Feature | TMB-986 | TMB-987 | TMB-995 | Comparison | |
| 1 | Power supply | Battery powered mode: | | | | |
| | | 6VDC 4*AAA<br>alkaline batteries | 6VDC 4*AAA<br>alkaline batteries | 6VDC 4*AA<br>alkaline batteries | similar | |
| | | AC adaptor powered mode: | | | | |
| | | AC 100-240V,<br>50-60Hz, 400mA | AC 100-240V,<br>50-60Hz, 400mA | AC 100-240V,<br>50-60Hz, 400mA | similar | |
| | Feature | TMB-986 | TMB-987 | TMB-995 | Comparison | |
| 2 | Display mode | Digital LCD<br>V.A.128*50mm | Digital LCD<br>V.A.150*34mm | Digital LCD<br>V.A.128*50mm | TMB-986, and<br>TMB-995:<br>128*50mm<br>TMB-987:<br>150*34mm | |
| 3 | Measurement<br>mode | Oscillographic<br>testing mode | Oscillographic<br>testing mode | Oscillographic<br>testing mode | similar | |
| 4 | Accuracy | Pressure:<br>$15℃~25℃$ :<br>within±0.4kpa /<br>3mmHg<br>$10℃~40℃$ (out of<br>$15℃~25℃$ ):<br>Within ±0.65kpa /<br>5mmHg<br>Heartbeat:<br>value:±5% | Pressure:<br>$15℃~25℃$ :<br>within±0.4kpa /<br>3mmHg<br>$10℃~40℃$ (out of<br>$15℃~25℃$ ):<br>Within ±0.65kpa /<br>5mmHg<br>Heartbeat:<br>value:±5% | Pressure:<br>$15℃~25℃$ :<br>within±0.4kpa /<br>3mmHg<br>$10℃~40℃$ (out of<br>$15℃~25℃$ ):<br>Within ±0.65kpa /<br>5mmHg<br>Heartbeat:<br>value:±5% | similar | |
| 5 | Normal<br>working<br>condition | Temperature: $10℃$<br>$~40℃$<br>Relative humidity<br>≤80% | Temperature: $10℃$<br>$~40℃$<br>Relative humidity<br>≤80%<br>Barometric pressure: 105~80 kPa (790~600 mmHg) | Temperature: $10℃$<br>$~40℃$<br>Relative humidity<br>≤80% | similar | |
| 6 | Storage &<br>transportation<br>condition | Temperature: - $20℃~60℃$ | | | | |
| | | Relative<br>humidity:10~93% | Relative<br>humidity: 10~93% | Relative<br>humidity:10~93% | similar | |
| 7 | Measurement<br>perimeter<br>of upper arm | 22cm~42cm | 22cm~42cm | 22cm~42cm | similar | |
| 8 | Weight | Approx.300g<br>(Excluding the<br>batteries) | Approx.300g<br>(Excluding the<br>batteries) | Approx.300g<br>(Excluding the<br>batteries) | similar | |
| 9 | External<br>dimensions | Approx.<br>180*100*39mm | Approx.<br>200*60*56mm | Approx.<br>180*100*39mm | TMB-986,<br>and<br>TMB-995:<br>Approx.<br>180*100*39mm<br>TMB-987:<br>Approx.<br>200*60*56mm | |
| 10 | Mode of<br>operation | Continuous<br>operation | Continuous<br>operation | Continuous<br>operation | similar | |
ર-2
{3}------------------------------------------------
# TRANSTEK
北
5-3
{4}------------------------------------------------
TRANSTEK
| Feature | TMB-986, TMB-987,<br>TMB-995 | Predicate: HEM 780N3 | Comparison |
|------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------|
| Device name | Blood Pressure Monitor | Automatic Blood Pressure<br>Monitor | Similar |
| Indication for use | Measure the blood pressure<br>and heartbeat rate.<br>Irregular heartbeat detection. | Measure the blood pressure<br>and heartbeat rate. | Add new<br>features |
| Components | Main Unit, Cuff, Adapter,<br>Instruction Manual, 4*AAA<br>/4*AA batteries, Storage<br>Case and Warranty Card | Main Unit, Cuff, Adapter,<br>Instruction Manual,<br>4"AA"batteries, Storage<br>Case and Warranty Card | Similar<br>components<br>and materials |
| Performance Tests | Biocompatibility<br>Electromagnetic<br>Compatibility and Electrical<br>Safety | Biocompatibility<br>Electromagnetic<br>Compatibility and Electrical<br>Safety | Similar |
| Labeling | Company name and address<br>Specification<br>Product description<br>Indication for use<br>Contraindications for use<br>Precautions<br>Warnings<br>Safety terms and conditions<br>Safety alert descriptions<br>Safety and performance<br>standards and so on | Company name and address<br>Specification<br>Product description<br>Indication for use<br>Contraindications for use<br>Precautions<br>Warnings<br>Safety terms and conditions<br>Safety alert descriptions<br>Safety and performance<br>standards and so on | Similar |
| Energy used | Battery (4*AAA/4*AA) or<br>AC adapter (DC 6V Output) | Battery (4*AA) or AC<br>adapter (DC 6V Output) | Similar |
| Display | Liquid crystal digital display | Liquid crystal digital display | Similar |
| Dimensions | TMB-986, TMB-995:<br>180(W)*100(D)*39(H)mm<br>TMB-987:<br>200(W)*60(D)*56(H)mm | 131(W)*155(D)*85(H)mm | Similar |
| Applicable cuff | Wrap around cuff | Wrap around cuff | Similar |
| Accuracy of<br>pressure indicator | Within ±3mmHg(15~25℃)<br>Within ±5mmHg<br>(10°C~40°C(out of<br>15℃~25℃)) | Within ±3mmHg or 2% of<br>reading | Similar |
| Accuracy of<br>heartbeat rate | Within ±5% of reading | Within ±5% of reading | Similar |
| Feature | TMB-986, TMB-987,<br>TMB-995 | Predicate: HEM 780N3 | Comparison |
| Cuff inflation | Automatic inflation with air<br>pump | Automatic inflation with air<br>pump | Similar |
| Deflation of<br>pressure | Automatic air release | Deflation rate is controlled by<br>an active electronic control<br>valve by 4 to 11 mmHg/s<br>depending on pulse rate | Similar |
| Operating voltage | DC 6V | DC 6V | Similar |
| Measurement<br>perimeter<br>of upper arm | 22cm~42cm | 22cm~42cm | Similar |
| Operation<br>environment | Temperature: 10℃~40℃<br>Relative humidity: ≤80%<br>Barometric pressure:<br>105~80 kPa<br>(790-600 mmHg) | Temperature: 10℃~40℃<br>Relative humidity: 30-80%<br>Barometric pressure:<br>105~80 kPa<br>(790-600 mmHg) | Similar |
| Transport and<br>storage environment | Temperature: -20℃~60℃<br>Relative humidity:10~93% | Temperature: -20℃~60℃<br>Relative humidity: 10~95% | Similar |
Table 2: The difference between Transtek Blood Pressure Monitor and Predicate HEM 780N3
4/5
{5}------------------------------------------------
S/C
#### 8. Conclusions
The subject devices have all features of the predicate device HEM 780N3 except the new features of irregular heartbeat detection, some specification of arm cuff and battery. These differences do not affect the safety and effectiveness of the subject devices.
Irregular heartbeat detection technology is same as what is used in Automatic Blood Pressure Monitor with Com Fit™ Cuff Model HEM-780. Thus, the subject devices are substantially equivalent to the predicate device.
--- End of this section --
{6}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/5/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo features a stylized bird-like symbol with three curved lines forming the body and wings. Below the bird, there are wavy lines that represent water. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Zhongshan Transtek Electronic Co., Ltd. c/o Mr. Leo Wang A03 Lab of BTS No. 1 Fanghua Street, Hi-tech District Chengdu City, Sichuan CHINA
MAY 2 7 2011
Re: K101681
Trade/Device Name: Transtek Blood Pressure Monitor Models TMB-986, TMB-987 and TMB-995
Regulatory Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Monitoring System Regulatory Class: Class II (Two) Product Code: DXN Dated: Undated Received: May 13, 2011
Dear Mr. Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or 10 devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{7}------------------------------------------------
Page 2 – Mr. Leo Wang
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fdagov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
D. Zahariev, M.D.
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
# TRANSTEK
# Section 4 - Indications for Use
510(k) Number (if known): K101681
Device Name: Blood Pressure Monitor
Models: TMB-986, TMB-987, TMB-995
Indications for Use:
This series of devices are digital monitors intended for use in measuring blood pressure and heartbeat rate in adult patient population with arm circumference ranging from 22 cm to 42 cm (about 9 - 17 inches).
These devices detect the appearance of irregular heartbeats during measurement and give a warning signal with readings.
The Blood Pressure Monitor compares average blood pressure results to pre-established AHA (American Heart Association) hypertension guideline of 135/85 mmHg.
Transtek Blood Pressure Monitor, TMB-986, TMB-987, and TMB-995 models are not intended to be a diagnostic device. Contact your physician if hypertensive values are indicated.
Prescription Use
AND/OR
(Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C) メ
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
| Concurrence of CDRH, Office of Device Evaluation (ODE) | |
|--------------------------------------------------------|--|
| (Division Sign-Off) | |
| Division of Cardiovascular Devices | |
| 510(k) Number | |
11/0168
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.