← Product Code [DXN](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DXN) · K100658

# U-RIGHT TD-3127 BLOOD PRESSURE MONITORING SYSTEM (K100658)

_Taidoc Technology Corporation · DXN · Jul 9, 2010 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DXN/K100658

## Device Facts

- **Applicant:** Taidoc Technology Corporation
- **Product Code:** [DXN](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DXN.md)
- **Decision Date:** Jul 9, 2010
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 870.1130
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The U-RIGHT TD-3127 Blood Pressure Monitoring System is intended to be used to measure the systolic and diastolic blood pressure and pulse rate by using a non-invasive technique in which an inflatable cuff is wrapped on the upper arm. This system should only be used for the testing on people over 18 years of age and over.

## Device Story

System consists of blood pressure monitor and inflatable upper arm cuff; utilizes oscillometric technology to measure systolic/diastolic blood pressure and pulse rate. Device intended for use by adults 18+; operates via non-invasive cuff inflation. Output displayed to user for blood pressure monitoring. Modifications from predicate include removal of speaking and data transmission functions; software updated accordingly. Device provides clinical data for patient blood pressure management.

## Clinical Evidence

No clinical data provided. Substantial equivalence supported by software verification and validation, and design validation confirming performance equivalent to predicate.

## Technological Characteristics

Oscillometric non-invasive blood pressure measurement. System includes monitor and inflatable cuff. Software-controlled. No specific material standards or connectivity protocols (data transmission removed).

## Regulatory Identification

A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.

## Predicate Devices

- FORA P20/U-RIGHT TD-3132 Blood Pressure Monitoring System ([K092106](/device/K092106.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

### Section 11. 510(k) Summary

– 9 2010

# 510(K) SUMMARY

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.

K100658

The Assigned 510(k) number is:

- 1. Submitter's Identification: TaiDoc Technology Corporation 3F, 5F, No.127, Wugong 2nd Rd., Wugu Township, Taipei County, 248, Taiwan
Correspondence: Nicky Pan Regulatory Affairs Specialist Tel: +886-2-6625-8188 #1196 Fax: +886-2-6625-0288 Email: nicky @taidoc.com.tw

Date of submission: 03/05/2010

- 2. Device name:
Proprietary name: U-RIGHT TD-3127 Blood Pressure Monitoring System

Regulatory information:

|  | A. Regulation section: 21 CFR §870.1130, Noninvasive blood pressure |                                                                                                                  |
|--|---------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------|
|  |                                                                     | measurement system                                                                                               |
|  |                                                                     | The support of the state of the state of the state of the state of the states of the states of the states of the |

- Classification: B. Class II (Blood Pressure Measurement System)
- ். Product Code: DXN, System, Measurement, Blood-Pressure, Non-Invasive
- D. Panel: 74, Cardiovascular - Blood Pressure Measurement System

{1}------------------------------------------------

3. Intended Use:

The U-RIGHT TD-3127 Blood Pressure Monitoring System is intended to be used to measure the systolic and diastolic blood pressure and pulse rate by using a non-invasive technique in which an inflatable cuff is wrapped on the upper arm. This system should only be used for the testing on people over 18 years of age and over.

#### 4. Device Description:

The kit of U-RIGHT TD-3127 Blood Pressure Monitoring System consists of two main products: the Blood Pressure Monitor and the pressure cuff. The product has been designed and tested to work together as a system to produce accurate blood pressure measurements.

5. Substantial Equivalence Information:

A. Predicate device name:

FORA P20/U-RIGHT TD-3132 Blood Pressure Monitoring System

Predicate K number: K092106 B.

C. Comparison with predicate:

The modified U-RIGHT TD-3127 Blood Pressure Monitoring System has the following similarities to the predicate device:

- same operating principle.
- . same fundamental scientific technology,
- I incorporate the same basic circuit design,
- . incorporate the same materials,
- . same shelf life

:"

- 1 packaged using the same materials, and
- manufactured by the same process.

The modifications encompass:

- Removed the functions for speaking and data transmission 1
- . Labeling change due to the software modification

11-2 of 3

{2}------------------------------------------------

- 6. Test Principle:
The measurement is by using oscillometric, non-invasive blood pressure (systolic, diastolic blood pressure and pulse rate) measuring technology.

:

12 11:11 11:11 11:11 11:11 11:11

- 7. Performance Characteristics:
The U-RIGHT TD-3127 Blood Pressure Monitoring System has the same performance characteristics as the predicate device.

.

Software verification and validation, and design validation confirmed that the performance, safety and effectiveness of the U-RIGHT TD-3127 Blood Pressure Monitoring System are equivalent to the predicate device.

- 8. Conclusion:
Based on the information provided in this submission, the U-RIGHT TD-3127 Blood Pressure Monitoring System is substantially equivalent to the predicate FORA P20/U-RIGHT TD-3132 Blood Pressure Monitoring System.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for the Department of Health & Human Services USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" written around the perimeter. Inside the circle is an abstract image of an eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

JUL - 9 2010

TaiDoc Technology Corporation c/o Ms. Nicky Pan Regulatory Affairs Specialist 3F, 5F, No. 127, Wugong 2nd Road Wugu Township, Taipei County 248 TAIWAN

Re: K100658 Trade/Device Name: U-Right TD-3127 Blood Pressure Monitoring System Regulatory Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II (Two) Product Code: DXN Dated: Undated Received: June 11, 2010

#### Dear Ms. Pan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Ms. Nicky Pan

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,
W.M.

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

K10065-f

## Section 8. Indications for Use

### Indications for Use

510(k) Number:

Device Name: U-RIGHT TD-3127 Blood Pressure Monitoring System

Indications for Use:

The U-RIGHT TD-3127 Blood Pressure Monitoring System is intended to be used to measure the systolic and diastolic blood pressure and pulse rate by using a non-invasive technique in which an inflatable cuff is wrapped on the upper arm. This system should only be used for the testing on people over 18 years of age and over.

Prescription Use (21 CFR Part 801 Subpart D) And/Or Over the Counter Use X (21 CFR Part 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)

W.M. Wood.

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety 510(k)

000000

Page 1 of 1

8-1 of 1

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DXN/K100658](https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DXN/K100658)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
