← Product Code [DXN](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DXN) · K093452

# FULLY AUTOMATIC ELECTRONIC BLOOD PRESSURE MONITOR MODEL KD-571, KD-7971 (K093452)

_Andon Health Co, Ltd. · DXN · Dec 23, 2009 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DXN/K093452

## Device Facts

- **Applicant:** Andon Health Co, Ltd.
- **Product Code:** [DXN](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DXN.md)
- **Decision Date:** Dec 23, 2009
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 870.1130
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

KD-5971 Fully Automatic Electronic Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. The cuff circumference is limited to 22cm-48cm. KD-7971 Fully Automatic Electronic Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the wrist. The cuff circumference is limited to 14cm-25cm.

## Device Story

Non-invasive blood pressure monitors (KD-5971 upper arm; KD-7971 wrist) measure diastolic/systolic pressure and pulse rate. Operation uses oscillometric method with silicon integrated pressure sensor. Device inflates cuff, detects pressure oscillations, and processes signals via internal MCU. Outputs include blood pressure values, pulse rate, irregular heartbeat detection, and blood pressure classification displayed on LCD. Includes voice function, memory storage (2x30), and histogram display. Used by clinicians or patients at home. Provides objective data for monitoring cardiovascular status.

## Clinical Evidence

Bench testing only. Conformance to IEC 60601-1 (safety), IEC 60601-1-2 (EMC), and AAMI SP10:2002 (sphygmomanometers) standards.

## Technological Characteristics

Oscillometric measurement using silicon integrated pressure sensor. Powered by electronic interface module. LCD display. Memory capacity 2x30. Features: irregular heartbeat detection, voice function, histogram, touch button (KD-5971). Standards: IEC 60601-1, IEC 60601-1-2, AAMI SP10:2002.

## Regulatory Identification

A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.

## Predicate Devices

- KD-5915 Fully Automatic Electronic Blood Pressure Monitor ([K091737](/device/K091737.md))
- KD-7962 Fully Automatic Electronic Blood Pressure Monitor ([K091997](/device/K091997.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# 510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirement of SMDA and 21 CFR 807.92 ..

UEL Z 3 2009

# 1.0 submitter's information

| Name:                | Andon Health Co., Ltd.                                          |
|----------------------|-----------------------------------------------------------------|
| Address:             | No 31, Changjiang Road, Nankai District, Tianjin,<br>P.R. China |
| Phone number:        | 86-22-6052 6161                                                 |
| Fax number:          | 86-22-6052 6162                                                 |
| Contact:             | Liu Yi                                                          |
| Date of Application: | 10/30/2009                                                      |

## 2.0 Device information

Trade name: Fully Automatic Electronic Blood Pressure Monitor Common name: Noninvasive blood pressure measurement system Classification name: Noninvasive blood pressure measurement system

### 3.0 Classification

Production code: DXN- Noninvasive blood pressure measurement system. Regulation number: 870.1130 Classification: II Panel: Cardiovascular

# 4.0 Predict device information

Manufacturer: Andon Health Co., Ltd. Device: KD-5915 Fully Automatic Electronic Blood Pressure Monitor KD-7962 Fully Automatic Electronic Blood Pressure Monitor 510(k) number: K091737, K091997

## 5.0 Device description

KD-5971 Fully Automatic Electronic Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systelic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. The cuff circumference is limited to 22cm-48cm.

ર- 1

{1}------------------------------------------------

KD-7971 Fully Automatic Electronic Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the wrist. The cuff circumference is limited to 14cm-25cm.

It is designed and manufactured according to ANSI/AAMI SP10--manual, electronic or automated sphygmanometers.

The operational principle is based on oscillometric and silicon integrate pressure sensor technology, the result will be shown on a LCD with an electronic interface module, the result can also be classified and displayed by the function of blood pressure classification indicator the memory capability of KD-5971 and KD-7971 are both 2*30 times. If any irregular heartbeat is detected, it can be shown on the LCD, KD-5971 and KD-7971 also have the voice function.

#### 6.0 Intended use

KD-5971 Fully Automatic Electronic Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. The cuff circumference is limited to 22cm-48cm.

KD-7971 Fully Automatic Electronic Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the wrist. Theavyt circumference is limited to 14cm-25cm.

The intended use and the indication for use of KD-5971 and KD-7971, as described in its labeling are the same as the predict device KD-5915 and KD-7962.

# 7.0 Summary comparing technological characteristics with predicate device

{2}------------------------------------------------

ત્ત્વ

.

| Technological Characteristics | Comparison result |
|-------------------------------|-------------------|
| Design principle              | Identical         |
| Appearance                    | Similar           |
| Patients contact Materials    | Identical         |
| Performance                   | Identical         |
| Biocompatibility              | Identical         |
| Mechanical safety             | Identical         |
| Energy source                 | Identical         |
| Standards met                 | Identical         |
| Electrical safety             | Identical         |
| EMC                           | Identical         |
| Function                      | Similar           |

{3}------------------------------------------------

#### 8.0 Performance summary

KD-5971 and KD-7971 Fully Automatic Electronic Blood Pressure Monitor conform to the following standards:

- · IEC 60601-1, Medical Electrical Equipment Part 1: General Requirements for Safety, 1988; Amendment 1, 1991-11, Amendment 2, 1995.
- · IEC 60601-1-2, Medical Electrical Equipment Part 1-2: General Requirements for Safety - Collateral standard: Electromagnetic Compatibility - Requirements and Tests (Edition 2:2001 with Amendment 1:2004, Edition 2.1 (Edition 2:2001 consolidated with Amendment 1:2004)).
- · AAMI SP10:2002, Manual, electronic or automated sphygmomanometers.
- · AAMI / ANSI SP10:2002/A1:2003 --, Amendment 1 to ANSI/AAMI SP10:2002 Manual, electronic, or automated sphygmomanometers,
- · AAMI / ANSI SP10:2002/A2:2006 --, Amendment 2 to ANSI/AAMI SP10:2002 Manual, electronic, or automated sphygmomanometers.

# 9.0 Comparison to the predict device and the conclusion

Our device KD-5971 and KD-7971 Fully Automatic Electronic Blood Pressure Monitor is substantially equivalent to the Fully Automatic Electronic Blood Pressure Monitor KD-5915 and KD-7962 whose 510(k) number is K091737 and K091997.

KD-7971 and KD-7962 is very similar in the intended use, the design principle, the material, the energy source and the applicable standards. Their appearance are different, they have different memory times. KD-7971 has a new function of histogram.The cuff pressure range of KD-7971 is different from KD-7962. KD-7971 also has a different MCU.

KD-5971 and KD-5915 is very similar in the intended use, the design principle, the material, the energy source and the applicable standards. Their appearance are different, they have different memory times. KD-7971 has new functions of histogram and touch button. The cuff pressure range of KD-7971 is different from KD-7962. KD-7971 also has a different MCU.

However, the test in this submission provides demonstration that these small differences do not raise any new questions of safety and effectiveness.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it, topped by a single wing. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-066-0609 Silver Spring, MD 20993-0002

DEC 2 3 2009

Andon Health Co., Ltd. c/o Mr. Liu Yi President No. 31, Changjiang Road, Nankai District Tianjin, P.R. China, 300193

Re: K093452

Trade/Device Name: KD-5971, KD-7971 Fully Automatic Electronic Blood Pressure Monitors Regulation Number: 21 CFR 870.1130 Regulation Name: Non-invasive blood pressure measurement systems Regulatory Class: Class II (two) Product Code: DXN Dated: November 2, 2009 Received: November 9, 2009

Dear Mr. Yi:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

Page 2 - Ms. Liu Yi

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

# Statement of Indications for Use

510(k) Number :

6093457

Device name:

KD-5971 and KD-7971 Fully Automatic Electronic Blood Pressure Monitor

#### Indications for use:

KD-5971 Fully Automatic Electronic Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and syst olic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. The cuff circumference is limited to 22cm-48cm.

KD-7971 Fully Automatic Electronic Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the wrist. The cuff circumference is limited to 14cm-25cm.

Prescription use AND/OR Over-The-Counter Use YES Part 21 CFR 801 Subpart D) (21 CFR-807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-COUNTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH evice Evaluation (ODE)

(Division Sign-Off)
Division of Cardiovascular Devices

510(k) Number K69345

4-1

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DXN/K093452](https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DXN/K093452)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
