← Product Code [DXN](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DXN) · K092061

# FORECARE ADVANCED SPORT WRIST BLOOD PRESSURE MONITOR, MODEL FP-0217 (K092061)

_Genexel Medical Instrument, Inc. · DXN · Mar 26, 2010 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DXN/K092061

## Device Facts

- **Applicant:** Genexel Medical Instrument, Inc.
- **Product Code:** [DXN](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DXN.md)
- **Decision Date:** Mar 26, 2010
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 870.1130
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

This device is an over the counter device, and its intended use is to measure systolic and diastolic pressure and pulse rate of adults, using wrist cuff and oscillometric method of measurement.

## Device Story

Forecare Advanced Sport Wrist Blood Pressure Monitor (Model FP-0217) is a digital, non-invasive blood pressure monitor for home use. Device utilizes an automatic inflating wrist cuff and oscillometric measurement method to determine systolic/diastolic pressure and pulse rate. Patient operates device by applying cuff and pressing power button; device automatically governs pressurization and deflation. If initial inflation (180mmHg) is insufficient, pump automatically re-pressurizes. Cuff deflates at a stepwise rate of 5-6 mmHg per 2 pulses. Results (systolic, diastolic, pulse, date) are displayed on an LCD panel. Device includes memory capacity for 58 measurements. System functions are preprogrammed; user cannot modify bleed-down rate. Device provides patients with a means to monitor blood pressure at home, facilitating self-management of cardiovascular health.

## Clinical Evidence

No clinical data provided. Substantial equivalence is based on technological characteristics and design similarities to the predicate device.

## Technological Characteristics

Non-invasive oscillometric blood pressure measurement system. Features automatic cuff inflation/deflation, LCD display, and 58-measurement memory. Operates via battery/power button interface. Preprogrammed system functions; no user-adjustable parameters for measurement logic.

## Regulatory Identification

A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.

## Predicate Devices

- Full Auto Wrist Digital Blood Pressure Monitor, Model SE-312 ([K012054](/device/K012054.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K09 2061

# 510(k) Summary As Required by 21 section 807.92 ( c )

Genexel Medical Instrument Inc

1- Submitter Name: FDA Registration number 8043684 111 Yangjae-dong, Seocho-gu 2- Address: MAR 2 6 2010 Seoul, Republic of Korea (82) 2-575-1141 3- Phone: (82) 2-575-1688 4- Fax: Mr Jonathan Seo 5- Contact Person: 6- Date summary prepared: July 3, 2009 7 - Official Correspondent: Mansour Consulting LLC 845 Aronson Lake Ct. Roswell, GA 30075 USA 8- Address: 9- Phone: 678-908-8180 678-623-3765 10- Fax: Jav Mansour, President 11- Contact Person: 12- Device Trade or Proprietary Name: FORECARE ADVANCED SPORT WRIST BLOOD PRESSURE MONITOR, MODEL FP-0217 13- Device Common or usual name: Digital Blood Pressure Monitor

14- Device Classification Name: Non Invasive blood pressure measuring system

15- Substantial Equivalency is claimed against the following device:

510K #K012054 Full Auto Wrist Digital Blood Pressure Monitor, Model SE-312. manufactured by Genexel Medical Instrument Inc. formerly Sein Electronics

#### 16- Description of the Device:

FORECARE ADVANCED SPORT WRIST BLOOD PRESSURE MONITOR, MODEL FP-0217 is intended to measure systolic and diastolic pressure and pulse rate of adults in a home care environment using wrist cuff and oscillometric method of measurement.

There are no contraindications; the subject device may be employed in the care of normotensive, hypertensive, or hypotensive patents.

The user interface panel of FP-0217 has power button, mode button, memory button and liquid crystal display ("LCD") for displaying systolic pressure, diastolic pressure, pulse rate and date. FP-0217 has memory capacity to store the 58 most recent measurement results.

The device measures blood pressure through the use of an automatic inflating cuff. Pressurization is automatically governed. If the initial inflation pressure (180mmHg) is inadequate for measurement, i.e. lower that the patient's systolic pressure, the pump will automatically re-pressurize to a preset level (30 mmHg) above the initial level. Symbols in the LCD indicate pressurization status at all times. The cuff automatically deflates with stepwise pressure drop of 5~6 mmHg per 2 pulses during blood pressure measurement.

\$\rho\$

1/2

{1}------------------------------------------------

The patient is responsible for applying the cuff, for initiating the measurements sequence by pressuring the "Power" button, and for recording results. The patient cannot alter bleed-down rate.

All system functions are preprogrammed. The user is cautioned in the instruction manual against attempting any programming or other modification.

No special training beyond basic ability to follow instruction is required. Since the products are designed for home use, detailed instructions on avoidance of practices that adversely affect the accuracy of measurements are included in the instruction manual.

### 17- Intended use of the device: (refer to FDA form attached)

This device is an over the counter device, and its intended use is to measure systolic and diastolic pressure and pulse rate of adults, using wrist cuff and oscillometric method of measurement.

### 18- Safety and Effectiveness of the device:

This device is safe and effective as the predicate device cited above.

## 19- Summary comparing technological characteristics with predicate device:

Comparison supporting that this device is substantially equivalent to other medical devices in commercial distribution is included within this submission. Also. Equivalency overview chart path is attached. Indeed, this device is SUBSTANTIALLY EQUIVALENT to the predicate device.

2 //

{2}------------------------------------------------

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling a stylized caduceus or a human figure with outstretched arms.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-066-0609 Silver Spring, MD 20993-0002

MAR 2 6 2010

Genexel Medical Instrument Inc. c/o Mr. Jay Mansour Mansour Consulting LLC 845 Aronson Lake Court Roswell, GA 30075

Re: K092061

Trade Name: Forecare Advanced Sport Wrist Blood Pressure Monitor, Model FP-0217 Regulatory Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II (Two) Product Code: DXN Dated: February 24, 2010 Received: February 26, 2010

Dear Mr. Mansour:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Mr. Jay Mansour

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# Indications for Use

510(k) Number (if known):

Device Name: Forecare Advanced Sport Wrist Blood Pressure Monitor, Model FP-0217

K092061

Indications For Use:

This device is an over the counter device, and its intended use is to measure systolic and diastolic pressure and pulse rate of adults, using wrist cuff and oscillometric method of measurement.

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use __ (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

nuaccii

510(k) Number

Page 1 of __ ﺪ ﺍﻟﻤﺪﻳﻨﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤ

p.29

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DXN/K092061](https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DXN/K092061)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
