FULLY AUTOMATIC ELECTRONC BLOOD PRESSURE MONITOR, MODEL KD-5905

K090770 · Andon Health Co, Ltd. · DXN · Jun 4, 2009 · Cardiovascular

Device Facts

Record IDK090770
Device NameFULLY AUTOMATIC ELECTRONC BLOOD PRESSURE MONITOR, MODEL KD-5905
ApplicantAndon Health Co, Ltd.
Product CodeDXN · Cardiovascular
Decision DateJun 4, 2009
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.1130
Device ClassClass 2

Indications for Use

KD-5905 Fully Automatic Electronic Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. The cuff circumference is limited to 22cm-48cm.

Device Story

KD-5905 is a non-invasive, oscillometric blood pressure monitor for adult use in clinical or home settings. Device uses an inflatable upper arm cuff (22-48cm) and silicon integrated pressure sensor to detect arterial oscillations. System processes pressure signals to calculate systolic/diastolic blood pressure and pulse rate; displays results on an LCD interface. Includes memory for 2x60 readings, irregular heartbeat detection, and a radio-controlled clock (RCC) for time synchronization. Operated by patients or clinicians; provides immediate visual feedback on LCD. Output assists in monitoring cardiovascular status. Differences from predicate include updated MCU, modified pulse rate range (40-180 bpm), expanded humidity operational range (<90%), and additional features like RCC.

Clinical Evidence

Bench testing only. Device performance validated against AAMI/ANSI SP10:2002 (including amendments A1:2003 and A2:2006) for manual, electronic, or automated sphygmomanometers. Compliance with IEC 60601-1 (safety) and IEC 60601-1-2 (EMC) confirmed.

Technological Characteristics

Oscillometric measurement principle using silicon integrated pressure sensor. Powered by electronic interface module. Features LCD display, 2x60 memory, irregular heartbeat detection, and radio-controlled clock. Cuff circumference 22-48cm. Complies with ANSI/AAMI SP10, IEC 60601-1, and IEC 60601-1-2 standards.

Indications for Use

Indicated for adult individuals to measure diastolic and systolic blood pressure and pulse rate via non-invasive upper arm cuff (22cm-48cm circumference). Suitable for use by medical professionals or at home.

Regulatory Classification

Identification

A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # 510(k) Summary This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirement of SMDA and 21 CFR FDA CDRH DMC 807.92. #### 1.0 submitter's information MAR 2-3-2009 JUN - 4 2009 | Name: | Andon Health Co., Ltd. | |----------------------|---------------------------------------------------| | Address: | No 31, Changjiang Road, Nankai District, Tianjin, | | | P.R. China | | Phone number: | 86-22-6052 6161 | | Fax number: | 86-22-6052 6162 | | Contact: | Liu Yi | | Date of Application: | 03/16/2009 | #### 2.0 Device information | Trade name: | Fully Automatic Electronic Blood Pressure Monitor | |----------------------|---------------------------------------------------| | Common name: | Noninvasive blood pressure measurement system | | Classification name: | Noninvasive blood pressure measurement system | #### 3.0 Classification Production code: DXN- Noninvasive blood pressure measurement system. Regulation number: 870.1130 Classification: II Panel: Cardiovascular #### 4.0 Predict device information Manufacturer: Andon Health Co., Ltd. Device: KD-5902 Fully Automatic Electronic Blood Pressure Monitor 510(k) number: K083317 ### 5.0 Device description KD-5905 Fully Automatic Electronic Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. The cuff circumference is limited to 22cm-48cm. {1}------------------------------------------------ It is designed and manufactured according to ANSI/AAMI SP10-manual, electronic or automated sphygmanometers. The operational principle is based on oscillometric and silicon integrate pressure sensor technology, the result will be shown on a LCD with an electronic interface module, the result can also be classified and displayed by the function of blood pressure classification indicator, the memory capability is 2×60 times. If any irreqular heartbeat is detected, it can be shown on the LCD. The RCC (radio controlled clock) function can automatically receive the radio clock signal to adjust the clock accurately. # 6.0 Intended use KD-5905 Fully Automatic Electronic Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. The cuff circumference is limited to 22cm-48cm. The intended use and the indication for use of KD-5905, as described in its labeling are the same as the predict device KD-5902. | Technological Characteristics | Comparison result | |-------------------------------|-------------------| | Design principle | Identical | | Appearance | Similar | | Patients contact Materials | Identical | | Performance | Similar | | Biocompatibility | Identical | | Mechanical safety | Identical | | Energy source | Identical | | Standards met | Identical | | Electrical safety | Identical | | EMC | Identical | | Function | Similar | # 7.0 Summary comparing technological characteristics with predicate device {2}------------------------------------------------ #### 8.0 Performance summary KD-5905 Fully Automatic Electronic Blood Pressure Monitor conforms to the following standards: - · IEC 60601-1. Medical Electrical Equipment Part 1: General Requirements for Safety, 1988; Amendment 1, 1991-11, Amendment 2, 1995. - · IEC 60601-1-2, Medical Electrical Equipment Part 1-2: General Requirements for Safety - Collateral standard: Electromagnetic Compatibility - Requirements and Tests (Edition 2:2001 with Amendment 1:2004; Edition 2.1 (Edition 2:2001 consolidated with Amendment 1:2004)). - · AAMI SP10:2002, Manual, electronic or automated sphygmomanometers. - · AAMI / ANSI SP10:2002/A1:2003 -- Amendment 1 to ANSI/AAMI SP10:2002 Manual, electronic, or automated sphygmomanometers. - · AAMI / ANSI SP10:2002/A2:2006 --, Amendment 2 to ANSI/AAMI SP10:2002 Manual, electronic, or automated sphyqmomanometers. #### 9.0 Comparison to the predict device and the conclusion Our device KD-5905 Fully Automatic Electronic Blood Pressure Monitor is substantially equivalent to the Fully Automatic Electronic Blood Pressure Monitor KD-5902 whose 510(k) number is K083317. The two devices are very similar in the intended use, the design principle, the material, the energy source and the applicable standards. Only their appearance and some functions such as memory mode and the RCC (radio controlled clock) function are different, moreover, the KD-5905 use a different MCU from KD-5902, the performance parameter of cuff pressure range and overpressure limit are different from the predicted device KD-5902. The operational range for humidity (<90%) are changed from the predict device whose operational range for humidity is <80%. The pulse rate range is changed from 30-180 times/min to 40-180 times/min. However, the test in this submission provides demonstration that these small differences do not raise any new questions of safety and effectiveness. : . . . {3}------------------------------------------------ #### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an image of an eagle. #### Public Health Service od and Drug Administration 200 Corporate Boulevard Rockville MD 20850 JUN - 4 2009 Andon Health Co., Ltd. c/o Mr. Liu Yi No 31 Changjiang Road Nankai District, Tianjin China 300193 Re: K090770 > Trade/Device Name: KD-5905 Fully Automatic Electronic Blood Pressure Monitor Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II (two) Product Code: DXN Dated: undated Received: May 7, 2009 Dear Mr. Yi: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ ## Page 2 - Mr. Liu Yi Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please contact the CDRH/Office of Surveillance and Biometrics/Division of Postmarket Surveillance at 240-276-3464. For more information regarding the reporting of adverse events, please go to http://www.fda.gov/cdrh/mdr/. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Steth 1861 Bram D. Zuckerman, M.D. Director · Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # Statement of Indications for Use 1090 770 510(k) Number : KD-5905 Fully Automatic Electronic Blood Pressure Device name: Monitor #### Indications for use: KD-5905 Fully Automatic Electronic Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. The cuff circumference is limited to 22cm-48cm. Prescription use Part 21 CFR 801 Subpart D) Over-The-Counter Use YES AND/OR (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-COUNTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) / Division of Cardiovascular Devices **510(k) Number** K096770 Page 1 of 1
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...