← Product Code [DXN](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DXN) · K083248

# FULLY AUTOMATED ELECTRONIC BLOOD PRESSURE MONITOR, MODEL KD-7902 (K083248)

_Andon Health Co, Ltd. · DXN · Jan 29, 2009 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DXN/K083248

## Device Facts

- **Applicant:** Andon Health Co, Ltd.
- **Product Code:** [DXN](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DXN.md)
- **Decision Date:** Jan 29, 2009
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 870.1130
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

KD-7902 Fully Automatic Electronic Blood Pressure Monitor is intended for use by medical professionals or at home to monitor and display diastolic, systolic blood pressure and pulse rate of adult each time, with the cuff around the wrist according to the instruction in the user's guide manual, which is same as predicated device.

## Device Story

KD-7902 is a non-invasive, fully automatic electronic blood pressure monitor. Input: oscillometric signals from a wrist-worn inflatable cuff. Processing: silicon integrated pressure sensor converts cuff pressure oscillations into digital signals; microprocessor analyzes signals to determine systolic/diastolic blood pressure and pulse rate. Output: results displayed on an integrated LCD. Features: 120-reading memory (2 users), average of last 3 readings. Used in clinical or home settings by professionals or patients. Benefits: provides convenient, non-invasive blood pressure monitoring for adults.

## Clinical Evidence

Bench testing only. The device was validated for performance against the ANSI/AAMI SP-10 standard for non-invasive blood pressure measurement systems. Electrical safety and electromagnetic compatibility were verified against IEC 60601-1 and IEC 60601-1-2, respectively.

## Technological Characteristics

Oscillometric measurement method; silicon integrated pressure sensor; LCD interface; microprocessor-based control. Powered by internal electronics. Wrist cuff circumference: 6.1023 to 9.8425 inches. Standards met: ANSI/AAMI SP-10, IEC 60601-1, IEC 60601-1-2.

## Regulatory Identification

A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.

## Predicate Devices

- KD-795 ([K070826](/device/K070826.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

## 510(k) Summary

### Identification of the submitter:

| Submitter:           | Andon Health Co., Ltd.<br>No 31, Changjiang Road, Nankai District, Tianjin,<br>P.R. China, 300193 |             |
|----------------------|---------------------------------------------------------------------------------------------------|-------------|
| Telephone number:    | 86-22-6052 6161                                                                                   |             |
| Fax number:          | 86-22-6052 6162                                                                                   |             |
| Contact:             | Liu Yi                                                                                            | JAN 29 2009 |
| Date of Application: | 10/29/08                                                                                          |             |

#### ldentification of the product:

|                      | Device proprietary Name: Fully Automatic Electronic Blood Pressure Monitor |
|----------------------|----------------------------------------------------------------------------|
| Common name:         | Noninvasive blood pressure measurement systems                             |
| Classification name: | Noninvasive blood pressure measurement system                              |
|                      | Class II per 21 CFR 870.1130                                               |

### Marketed Devices to which equivalence is claimed:

Device KD-795

manufacture Andon Health Co., Ltd 510(k) number K070826

#### Device description:

KD-7902 Fully Automatic Electronic Blood Pressure Monitor is Non-invasive blood pressure measurement system for only one person each time. Based on oscillometric and silicon integrate pressure sensor technology, the device is used to monitor systolic, diastolic blood pressure and pulse rate which will be shown on a LCD with an electronic interface module. Buckling a cuff around the wrist, which cuff circumference is limited to 6.1023 inches to 9.8425 inches, the device can analyze the signals promptly and display the results. It has some other functions, such as 120 memory recall / 2 users and calculating average of last 3 readings.

#### Intended use:

KD-7902 Fully Automatic Electronic Blood Pressure Monitor is intended for use by medical professionals or at home to monitor and display diastolic, systolic blood pressure and pulse rate of adult each time, with the cuff around the wrist according to the instruction in the user's guide manual, which is same as predicated device.

{1}------------------------------------------------

KD-7902 Fully Automatic Electronic Blood Pressure Monitor has the same principle with predicated device, which utilizes Oscilliometric measurement method to monitor the blood pressure and the result can be shown on the LCD.

The modifications that were made are:

#### 1. Appearance

- 2. No voice
- 3. Change into 120 memory recall / 2 users (2x60) from 1x60
- 4. Memory average function: calculating the average of last 3 readings
- 5. Change microprocessor

Please find the following tabulated comparison supporting that the proposed device is substantially equivalent to the predicated device.

| FDA file reference number     | 510K# K070826     |
|-------------------------------|-------------------|
| Technological Characteristics | Comparison result |
| Indications for use           | Identical         |
| Target population             | Identical         |
| Design                        | Similar           |
| Materials                     | Identical         |
| Performance                   | Identical         |
| Sterility                     | Not Applicable    |
| Biocompatibility              | Identical         |
| Mechanical safety             | Identical         |
| Chemical safety               | Not Applicable    |
| Energy used and/or delivered  | Identical         |
| Where used                    | Identical         |
| Standards met                 | Identical         |
| Electrical safety             | Identical         |

### Device testing:

KD-7902 Fully Automatic Electronic Blood Pressure Monitor meet the following standards:

- · ANSI/AAMI SP-10 standard
- · IEC 60601-1 Medical electrical equipment Part 1: General requirements for safety
- · IEC 60601-1-2 Electromagnetic Compatibility

Refer to the submission for more information.

{2}------------------------------------------------

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of flowing lines. The overall design is simple and conveys a sense of government authority and public service.

#### Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Andon Health Co., Ltd. c/o Ms. Liu Yi No 31, Changjiang Road, Nankai District, Tianjin, P.R. China, 300193

JAN 2 9 2009

Re: K083248

Trade/Device Name: KD-7902 Fully Automatic Electronic Blood Pressure Monitor Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II (two) Product Code: DXN Dated: December 19, 2008 Received: December 29, 2008

Dear Ms. Yi:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act

{3}------------------------------------------------

#### Page 2 - Ms. Liu Yi

or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

510K submission for KD-7902 Fully Automatic Electronic Blood Pressure Monitor

Ko83248

# Statement of Indications for Use

510(k) Number :

Applicant: Andon Health Co., Ltd

Device name:

KD-7902 Fully Automatic Electronic Blood Pressure Monitor

### Indications for use:

KD-7902 Fully Automatic Electronic Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the wrist. The cuff circumference is limited to 6.1023 inches to 9.8425 inches.

Prescription use Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

YES

(Please Do Not WRITE BELOW THIS LINE-COUNTINUE ON ANOTHER PAGE IF NEEDED)

| Concurrence of CDRH, Office of Device Evaluation (ODE) |         |
|--------------------------------------------------------|---------|
| (Division Sign-Off)                                    |         |
| Division of Cardiovascular Devices                     |         |
| 510(k) Number                                          | K083248 |
| Page 1 of                                              | 1       |

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DXN/K083248](https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DXN/K083248)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
