BLOOD PRESSURE MONITOR, MODELS ZSBP-OXX AND ZSBP-1XX
K070473 · Zen Strong Medical Technology Co., Ltd. · DXN · May 23, 2007 · Cardiovascular
Device Facts
Record ID
K070473
Device Name
BLOOD PRESSURE MONITOR, MODELS ZSBP-OXX AND ZSBP-1XX
Applicant
Zen Strong Medical Technology Co., Ltd.
Product Code
DXN · Cardiovascular
Decision Date
May 23, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1130
Device Class
Class 2
Indications for Use
ZSBP-001 and ZSBP-002 blood pressure monitor measures automatically human being systolic, diastolic blood pressure and heart beat rate from wrist by using the oscillometric method for the patient of the age over 18 years old. ZSBP-101 series of blood pressure monitor measures automatically human being systolic, diastolic blood pressure and heat beat rate from upper arm by using the Oscillometric method for the patient of the age over 18 years old. All measurement values can be read out and keep memory on the LCD panel for home care use.
Device Story
Non-invasive electronic blood pressure monitors (wrist and upper arm models) for home use. Device consists of pneumatic cuff, inflation/deflation pump, pressure sensor, and processing unit. Operation: cuff inflates to set pressure, then automatically deflates; pressure changes over time recorded as database; software calculates systolic/diastolic blood pressure and heart rate using oscillometric method. Results displayed on integrated LCD panel; device includes memory function for result storage. Operated by patient without physician involvement.
Clinical Evidence
Bench testing only. Compliance with ANSI/AAMI SP 10-1992, EN 60601-1, and EN 60601-1-2 standards. Software verification performed per FDA guidance.
Technological Characteristics
Non-invasive electronic oscillometric blood pressure monitor. Components: pneumatic pump, pressure sensor, cuff, LCD, and measuring software. Standards: ANSI/AAMI SP 10-1992, EN 60601-1, EN 60601-1-2. Standalone device for home use.
Indications for Use
Indicated for patients over 18 years old for the automatic measurement of systolic and diastolic blood pressure and heart rate via oscillometric method at the wrist (ZSBP-001, ZSBP-002) or upper arm (ZSBP-101). Intended for home care use.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
{0}------------------------------------------------
K070473/51
p1/2
## 510(K) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510(k) number is:
1. Submitter's Identifications: Zen Strong Medical Technology Co., Ltd. 6F No. 88, Ning Jin St., Keelung, Taiwan, R.O.C. Contact: Cheng Roei-Sheng Date of Summary Preparation: February 15, 2007 MAY 2 3 2007
- 2. Name of the Device: Blood Pressure Monitor, models ZSBP-001 and ZSBP-002 for wrist type and ZSBP-101 for upper arm type.
- 3. Classification information:
| Regulation Number | 870.1130 |
|-------------------|-----------|
| Medical Specialty | Neurology |
| Product Code | DXN |
| Device Class | II |
| Tier | II |
## 4. Device Description:
Basically two different type of blood pressure monitors are to be included in this 510(K) submission, the wrist type and upper arm type. The main intended use for these two tvpe of blood pressure monitor is as the following description:
1> ZSBP-001 and ZSBP-002 blood pressure monitor. This series of device measures automatically the human being systolic, diastolic blood pressure and heart beat rate from wrist by using the oscillometric method for the patient of ages above 18 years old.
- 2> ZSBP-101 blood pressure monitor measures automatically human being systolic, diastolic blood pressure and heart beat rate from upper arm by using the Oscillometric method for the patient of the age above 18 tears old.
For all of these three blood pressure monitors, the measurement values can be read out and keep memory on the LCD panel for home care use( without the involvement of professional physician).
From the construction point of view, the non-invasive electronic blood pressure monitor was composite of blood pressure measuring circuit via Oscillometric method, pressure sensor, measuring cuff at upper arm or wrist, pneumatic pumping inflation and deflation system, housing, display LCD, and measuring software, and memories ....... etc.
The main operation for the non-invasive electronic blood pressure monitor is carried out in such a way that the measuring cuff at wrist or upper arm is inflated to the set pressure level, then deflated to zero automatically. During the inflation and deflation, the pressure change with respective to time were recorded as the database of measurement. Then the following measuring results will be calculated against the measurement database, and displayed on the LCD of device:
{1}------------------------------------------------
-
- Blood pressure information including systolic and diastolic pressure (calculated via -Oscillometric method)
- Heart beat rate. ।
In addition to the main blood pressure and heart beat rate measuring function, the ZSBP series of blood pressure monitor provides also the memory function for user to store the result of measurement.
- 5. Intended Use:
ZSBP-001 and ZSBP-002 blood pressure monitor measures automatically human being systolic, diastolic blood pressure and heart beat rate from wrist by using the oscillometric method for the patient of the age over 18 years old.
ZSBP-101 series of blood pressure monitor measures automatically human being systolic, diastolic blood pressure and heat beat rate from upper arm by using the Oscillometric method for the patient of the age over 18 years old.
All measurement values can be read out and keep memory on the LCD panel for home care use.
- 6. Comparison to the 510(k) Cleared Device (Predicate Device):
- 1> Health & Life model HL-168 (K050249).
- 2> Micro life model BP-3BT0-AP (K041411).
- 7. Discussion of Non-Clinical Tests Performed Determination of Substantial Equivalence are as follows:
Compliance to applicable voluntary standards includes ANSI/AAMI, SP 10-1992, as well as EN 60601-1, and EN 60601-1-2 requirement. For the conformity of standards, the representative models of ZSBP-001 and ZSBP-101 were chosen as testing samples.
In addition to the compliance of voluntary standards, the software verification has been carried out according to the FDA software guidance.
- 8. Conclusions
The Zen Strong mode! ZSBP-001 and ZSBP-002 blood pressure monitor (measurement at wrist) has the same intended use and technical characteristics as the cleared model HL-168 (K050249), and ZSBP-101 series of blood pressure monitor (measured at upper arm) has the same intended use and technical characteristics as the cleared model BP-3BT0-AP (K041411).
Moreover, verification and validation tests contained in this submission demonstrate that the difference in the submitted demonstrate that the difference in the submitted models could maintain the same safety and effectiveness as that of cleared devices.
In the other words, those engineering difference do not: (1) affect the intended use or (2) alter the fundamental scientific technology of the device, and the new models as mentioned on this submission are considered substantial equivalent to the predicate devices.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the Department of Health and Human Services (HHS). The logo features a stylized caduceus, a symbol often associated with medicine and healthcare, with three parallel lines that curve and converge. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES . USA" is arranged in a circular fashion around the caduceus.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAY 2 3 2007
Zen Strong Medical Technologies Co. c/o Cheng- Roei-Sheng 6F, No. 88 Ning Jing St. Keelung China (Taiwan)
Re: K070473
Trade/Device Name: Blood Pressure Monitor, models ZSBP-001 and ZSBP-002 for wrist type and ZSBP-101 for upper arm type. Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive blood pressure measurement system Regulatory Class: Class II Product Code: DXN Dated: April 16, 2007 Received: April 16, 2007
Dear Mr. Roei-Sheng:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Mr. Roei-Sheng
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Blyminuma for
Bram B. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
070473/s1
## Indications For Use
510(k) Number (if known):
Device Name: Blood Pressure Monitor, models ZSBP-001and ZSBP-002 for wrist type, and ZSBP-101 for upper arm type.
Indications For Use:
ZSBP-001 and ZSBP-002 blood pressure monitor measures automatically human being systolic, diastolic blood pressure and heart beat rate from wrist by using the Oscillometric method for the patient of the age over 18 years old.
ZSBP-101 series of blood pressure monitor measures automatically human being systolic, diastolic blood pressure and heart beat rate from upper arm by using the Oscillometric method for the patient of the age over 18 years old.
All measurement values can be read out and keep memory on the LCD panel for home care use.
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use V (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
OR
Concurrence of CDRH, Office of Device Evaluation (ODE)
Bllminson
Page 1 of 1
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.