← Product Code [DXN](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DXN) · K051375

# VITAL SIGNS MONITOR (K051375)

_Mediana Co., Ltd. · DXN · Jul 28, 2005 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DXN/K051375

## Device Facts

- **Applicant:** Mediana Co., Ltd.
- **Product Code:** [DXN](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DXN.md)
- **Decision Date:** Jul 28, 2005
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1130
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Pediatric

## Indications for Use

The purpose and function of the Mediana ARGUS VCM vital signs monitor is to monitor noninvasive blood pressure (systolic, diastolic, and mean arterial pressures), functional arterial oxygen saturation, pulse rate for adult, pediatric and neonate patients and temperature for adult and pediatric patients in all hospital areas and hospital-type facilities. It may be used during hospital transport and in mobile, land-based environments, such as ambulances, within the specification of the environmental characteristics. Note: Hospital use typically covers such areas as general care floors, operating rooms, special procedure areas, intensive and critical care areas, within the hospital plus hospital-type facilities. Hospital-type facilities include physician office based facilities, sleep labs, skilled nursing facilities, surgicenters, and sub-acutecenters. Note: Intra-hospital transport includes transport of a patient within the hospital or hospital-type facility.

## Device Story

Compact, lightweight vital signs monitor (2.7 kg) for hospital and mobile environments; operated by medically trained personnel. Inputs: NIBP (via cuff), SpO2 (via pulse oximetry sensor), pulse rate (derived from NIBP or SpO2), and temperature (via thermometry probe). Device integrates third-party modules: Welch Allyn POEM for NIBP, Nellcor MP506 for SpO2, and Welch Allyn SureTemp for temperature. Data processed and displayed on 7-segment and LED displays. Provides real-time patient monitoring; assists clinicians in assessing patient status during care or transport. Benefits include continuous, non-invasive tracking of vital parameters to support clinical decision-making.

## Clinical Evidence

Bench testing only. System verification and validation performed per plan #PA300-09000. Risk analysis and mitigation documented per design control procedures. Quality system certified to 21CFR820, ISO 9001, and ISO 13485.

## Technological Characteristics

Dimensions: 180x278x130 mm; Weight: 2.7 kg. Power: AC mains (100-240VAC) or internal battery. Modules: Welch Allyn POEM (NIBP), Nellcor MP506 (SpO2), Welch Allyn SureTemp (thermometry). Display: 7-segment and LED. Connectivity: Standalone monitor.

## Regulatory Identification

A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.

## Predicate Devices

- Welch Allyn Protocol Inc. Vital Signs Monitors, Model 53000 Series ([K031740](/device/K031740.md))
- Welch Allyn Protocol Inc. Vital signs Monitors, Model Propaq LT 802 Series ([K033378](/device/K033378.md))
- Nellcor Puritan Bennett Pulse Oximeter, Model N-550 ([K021090](/device/K021090.md))
- Welch Allyn Inc. Vital Signs Monitor (VSM) ([K024005](/device/K024005.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# JUL 2 8 2005

# 510(k) Summary of Safety and Effectiveness

#### Submitter

Mediana Co.,Ltd. 231-6 Nonhyun-Dong, Kangnam-Ku, Seoul, Korea 135-010 Tel) +822-542-3375 Fax) +822-542-1621 Company Contact: Gina Y.J.LEE, Specialist of Regulatory Affairs Date Summary Prepared: March 2nd, 2005

#### Device Name

| Trade Name:          | ARGUS VCM series, Vital Signs Monitor                                                                                                                              |
|----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name:         | Vital signs measurement devices                                                                                                                                    |
| Classification Name: | NIBP measurement system (21CFR870.1130), also contains<br>non-invasive pulse oximetry, SpO2 (21CFR870.2700) and<br>clinical electronic thermometer (21CFR870.2910) |
| Classification:      | Class II                                                                                                                                                           |

#### Predicate Devices (Legally Marketed Devices)

The predicate devices for Vital signs monitors, Model ARGUS VCM series are:

- Welch Allyn Protocol Inc. Vital Signs Monitors, Model 53000 Series cleared by FDA through 510(k) No. K031740, and
- Welch Allyn Protocol Inc. Vital signs Monitors, Model Propaq LT 802 Series cleared by FDA through 510(k) No. K033378, and
- Nellcor Puritan Bennett (division of Tyco Healthcare Inc.) Pulse Oximeter, Model N-550 cleared by FDA through 510(k) No. K021090, and
- Welch Allyn Inc. Vital Signs Monitors, Welch Allyn Vital Signs Monitor (VSM), cleared by FDA through 510(k) No. K024005.

#### Device Description

The ARGUS VCM vital signs monitor is to monitor non-invasive blood pressure, pulse rate, non-invasive functional oxygen saturation of arterial hemoglobin (SpO2) for adult, pediatric and neonate patients and body temperature for adult and pediatric patients in general hospital and alternate care facilities by medically trained personnel. This monitor is available for sale only upon the order of a physician or licensed health care professional.

The Mediana ARGUS VCM vital signs monitor is a lightweight and compact device (180×278×130 mm and 2.7 kg) powered by AC mains (100-240VAC, 50-60Hz) and also powered by internal battery. The monitor provides patient data and monitoring status on 7segment and LED displays.

#### Intended Use

The purpose and function of the Mediana ARGUS VCM vital signs monitor is to monitor noninvasive blood pressure (systolic, diastolic, and mean arterial pressures), functional arterial oxygen saturation, pulse rate for adult, pediatric and neonate patients and temperature for adult and pediatric patients in all hospital areas and hospital-type facilities. It may be used during hospital transport and in mobile, land-based environments, such as ambulances, within the specification of the environmental characteristics.

{1}------------------------------------------------

- Note: Hospital use typically covers such areas as general care floors, operating rooms, special procedure areas, intensive and critical care areas, within the hospital plus hospital-type facilities. Hospital-type facilities include physician office based facilities, sleep labs, skilled nursing facilities, surgicenters, and sub-acutecenters.
- Note: Intra-hospital transport includes transport of a patient within the hospital or hospitaltype facility.

### Summary of Technical Characteristics of the Device Compared to the Predicate Devices (Legally Marketed Devices)

The Mediana vital signs monitors, Model ARGUS VCM Series are substantially equivalent to the Welch Allyn Protocol vital signs monitors, Model 53000 Series and Model Propaq LT 802 Series, Nellcor Puritan Bennett (division of Tyco Healthcare Inc.) Pulse Oximeter Model N-550, and Welch Allyn Inc. Vital Signs Monitor, Welch Allyn Vital signs Monitor (VSM).

- The Non-Invasive Blood Pressure (NIBP) measurement specifications and performance are equivalent to the Welch Allyn Protocol vital signs monitors, Model 53000 Series and Model Propaq LT 802 Series. Welch Allyn NIBP module POEM is the same NIBP module used in the Mediana vital signs monitors, Model ARGUS VCM Series.
- The Pulse rate specifications and performance derived from either Non-Invasive Blood Pressure (NIBP) or Pulse Oximetry (SpO2) are equivalent to the Welch Allyn Protocol vital signs monitors, Model 53000 Series, and the Nellcor Puritan Bennett Pulse Oximeter Model N-550.
- The Pulse Oximetry (SpO2) specifications and performance are equivalent to the Nellcor Puritan Bennett Pulse Oximeter Model N-550. Nellcor Oximetry module MP506 is the same SpO2 module used in the Mediana vital signs monitors, Model ARGUS VCM Series.
- The Temperature specifications and performance are equivalent to the Welch Allyn Protocol vital signs monitors, Model 53000 Series and the Vital Signs Monitor (VSM). Welch Allyn thermometry module SureTemp® is the same temperature module used in the Mediana vital signs monitors, Model ARGUS VCM Series.

#### Summary of Performance Testing

The Mediana vital signs monitors, Model ARGUS VCM Series substantially have been tested in accordance with the system V & V plan (#PA300-09000) and summary (#PA300-09013) included with the submission using production equivalent units prior to release to market.

A risk analysis identifying potential hazards and documenting mitigation of the hazards has been developed and applied as part of Mediana design control procedure. Mediana's quality system confirms to 21CFR820, ISO 9001, ISO13485 and CMDCAS ISO 13485 certified by DNV (Det Norske Veritas) and NSAI (National Standards Authority of Irelands).

#### Conclusions

As stated above, the Mediana vital signs monitors, Model ARGUS VCM Series are safe and effective and comply with the appropriate medical device standards and are substantially equivalent to the earlier identified predicate devices.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES, U.S.A." arranged around the perimeter. Inside the circle is an abstract symbol resembling three human profiles facing to the right, stacked on top of each other, with flowing lines suggesting movement or connection.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Mediana Co., LTD. c/o Mr.Charlie Mack International Regulatory Consultants 231-6 Nonhyun-Dong Kangnam-Ku, Seoul, Kerea 135-110

#### Re: K051375

Trade Name: Argus VCM series, vital signs monitors Regulation Number: 21 CFR 870.1130, 21 CFR 870.2700, 21 CFR 880.2910 Regulation Name: System, Measurement, Blood Pressure, Non-Invasive Regulatory Class: Class II (two) Product Code: DXN Dated: May 24, 2005 Received: May 26, 2005

Dear Mr. Mack:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act

JUL 2 8 2005

{3}------------------------------------------------

Page 2 - Mr.Charlie Mack

or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0295. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Bfimmermon for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# Indications for Use

510(k) Number (if known): K051375

Device Name: Arqus VCM Series, Vital Signs Monitor

Indications For Use: The purpose and function of the Mediana ARGUS VCM vital signs monitor is to monitor non-invasive blood pressure (systolic, diastolic, and mean arterial pressures), functional arterial oxygen saturation, pulse rate for adult, pediatric and neonate patients and temperature for adult and pediatric patients in all hospital areas and hospital-type facilities. It may be used during hospital transport and in mobile, land-based environments, such as ambulances, within the specification of the environmental characteristics.

- Note: Hospital use typically covers such areas as general care floors, operating rooms, special procedure areas, intensive and critical care areas, within the hospital plus hospital-type facilities. Hospital-type facilities include physician office based facilities, sleep labs, skilled nursing facilities, surgicenters, and sub-acutecenters.
- Note: Intra-hospital transport includes transport of a patient within the hospital or hospital-type facility.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

R. Hammonds

ivision Sign-Off) Division of Cardlovascular Devices KOSIS IS 510(k) Number__

Page 1 of 1

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DXN/K051375](https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DXN/K051375)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
