← Product Code [DXN](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DXN) · K030825

# MPM-NET (K030825)

_Aerotel Medical Systems (1998) , Ltd. · DXN · Mar 28, 2003 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DXN/K030825

## Device Facts

- **Applicant:** Aerotel Medical Systems (1998) , Ltd.
- **Product Code:** [DXN](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DXN.md)
- **Decision Date:** Mar 28, 2003
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 870.1130
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Software as a Medical Device

## Indications for Use

The Aerotel Medical Systems (1998) Ltd. MPM-Net system is intended to be used by patients to measure their blood pressure at home and transmit it to a central station by telephone. The system includes both the patient unit and a central computer program (to be used in a Personal Computer) which receives the blood pressure data, stores it in a patient's record, and prepares reports and charts showing the history of the systolic, diastolic pressures and the heart rate. The system is not intended for patients with defibrillators.

## Device Story

MPM-Net is a modem/dialer system extension for cleared home-use blood pressure monitors; enables trans-telephonic transmission of blood pressure data to a central station. System includes a central computer program for PCs; receives data; stores patient records; generates reports/charts of systolic/diastolic pressure and heart rate. Allows report retrieval via internet. Used by patients at home to measure BP; used by physicians at central station to monitor patient progress. Benefits include remote tracking of BP history and heart rate trends.

## Clinical Evidence

No clinical tests were performed. Modifications were validated via non-clinical system validation testing.

## Technological Characteristics

System consists of a modem/dialer interface for home blood pressure monitors and a central station PC software application. Connectivity includes telephone transmission and internet-based report retrieval. Operates as an accessory to cleared home-use blood pressure measurement devices.

## Regulatory Identification

A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.

## Predicate Devices

- BP-Tel ([K983717](/device/K983717.md))
- TeleCliniQ Clinical Data Transmission System ([K021447](/device/K021447.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K030825
page 1 of 2
Page A 2

## Aerotel Medical Systems (1998) Ltd. Special 510(k) MPM-Net

MAR 2 8 2003

## 510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements Regulation par. 807.92, effective March 14, 1995.

#### 1. Submitter

Name: Aerotel Medical Systems (1998) Ltd Address: 5 Hazoref St., Holon 58856, Israel Telephone Number: 972-3-5596111 Contact person: Dr. George Myers, 210-787- 1703 Date prepared: April 1, 2002

### 2. Device

Proprietary name: MPM-Net Common Name: Home blood-pressure measurement device and trans-telephonic transmission system. Monitor, Blood-Pressure, Amplifier and Associated Classification Name: Electronics

### 3. Predicate Devices

BP-Tel K983717, manufactured by Aerotel TeleCliniQ Clinical Data Transmission System, K021447, manufactured by Aerotel

#### 4. Description

. The MPM-Net is an option for a modem/dialer system adapted for the purpose of following patients' blood pressures. It operates in conjunction with a cleared home-use blood pressure device. The system includes the computer program at a central station used for receiving data and preparing reports. The option allows reports to be retrieved over the internet.

#### 5. Intended Use

The intended use of this device in conjunction with the devices previously cleared is to:

{1}------------------------------------------------

- 1. Provide means so that a patient can measure his/her blood pressure at home
- Send the measurement in to central computer 2.
- Provide means so that a physician can monitor progress on the central computer. 3.

# 6. Comparison

The MPM-Net is a modified version of the predicate device. It has basically the same electrical and mechanical characteristics, and the use of the two devices by the patient is equivalent. It permits the reports to be retrieved over the internet.

## 7. Performance Data

(1) Non-clinical tests

The modifications were validated by a validation test for the system.

(2) Clinical Tests

No clinical tests were performed.

## 8. Conclusion

The conclusion drawn from these tests is that the MPM-Net is equivalent in safety and efficacy to its predicate device.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three lines forming its body and head. The eagle is facing to the right. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.

MAR 2 8 2003

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Aerotel Medical Systems (1998) Ltd. c/o Dr. George Myers Medsys Inc. Official Correspondent 377 Rt. 17 Hasbrouck Heights, NJ 07604

Re: K030825 Trade Name: MPM-Net Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive blood pressure measurement system Regulatory Class: Class II (two) Product Code: DXN Dated: March 13, 2003 Received: March 14, 2003

## Dear Dr. Myers:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 -- Dr. George Myers

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Bram D. Zuckerman, M.D.

Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Page _________________________________________________________________________________________________________________________________________________________________________

510(k) Number (if known): K030825

Indications for Use Form

Device Name: MPM-Net

Indications for Use:

The Aerotel Medical Systems (1998) Ltd. MPM-Net system is intended to be used by patients to measure their blood pressure at home and transmit it to a central station by telephone. The system includes both the patient unit and a central computer program (to be used in a Personal Computer) which receives the blood pressure data, stores it in a patient's record, and prepares reports and charts showing the history of the systolic, diastolic pressures and the heart rate. The system is not intended for patients with defibrillators.

# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Ndelotti

510(k) Number K030825

Prescription Use Only

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DXN/K030825](https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DXN/K030825)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
