PORTAPRES AMBULATORY CONTINUOUS NON INVASIVE BLOOD PRESSURE MONITOR,MODEL-2
K023338 · Fms, Finapres Medical Systems BV · DXN · May 9, 2003 · Cardiovascular
Device Facts
Record ID
K023338
Device Name
PORTAPRES AMBULATORY CONTINUOUS NON INVASIVE BLOOD PRESSURE MONITOR,MODEL-2
Applicant
Fms, Finapres Medical Systems BV
Product Code
DXN · Cardiovascular
Decision Date
May 9, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1130
Device Class
Class 2
Indications for Use
The Portapres Ambulatory Continuous Non-Invasive Blood Pressure Monitor is intended to provide the user with continuous, noninvasive blood pressure and pulse rate monitoring. The Portapres enables 24-hour continuous ambulatory measurements. The PC-based BeatScope software used with the data from the device provides Modelflow-based computation of real-time and beat-to-beat blood pressure as well as hemodynamic parameters from the pressure waveform, including, cardiac output and total peripheral resistance.
Device Story
Portable, battery-powered ambulatory monitor; records continuous non-invasive arterial blood pressure. System components: waist-worn main unit (electronics/memory), pump unit, battery pack, wrist unit, and finger cuff. Finger cuff uses infrared photoplethysmograph to measure arterial volume; wrist unit controls cuff pressure to maintain constant arterial volume (setpoint). Pressure transducer measures cuff pressure as indirect blood pressure proxy. Hydrostatic height correction unit compensates for finger position relative to heart. Data stored on internal flash memory; transferred via RS232 to PC. BeatScope software performs Modelflow-based computation of beat-to-beat blood pressure, cardiac output, and total peripheral resistance. Used by clinicians to monitor hemodynamic status; provides continuous data for clinical decision-making; enables ambulatory monitoring outside stationary clinical settings.
Clinical Evidence
No clinical data provided; substantial equivalence based on design, methodology, and technological characteristics compared to predicate devices.
Technological Characteristics
Battery-powered, portable monitor. Components: Neoprene waist belt, air pump, NiCd/Li battery, wrist unit, finger cuff with infrared photoplethysmograph. Sensing: photoplethysmography and pressure transduction. Connectivity: RS232 serial interface for PC data transfer. Software: BeatScope (PC-based) for hemodynamic computation.
Indications for Use
Indicated for continuous, noninvasive blood pressure and pulse rate monitoring in patients requiring 24-hour ambulatory measurements.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
{0}------------------------------------------------
# 510(k) SUMMARY FMS PORTAPRES AMBULATORY NONINVASIVE BLOOD PRESSURE MONITOR
This 510(k) summary is being submitted in accordance with the requirements of SMDA and 21CFR § 807.92
# I. NAME OF SUBMITTER
FMS, Finapres Medical Systems BV Simon Stevinweg 48 NL-6827 BT Arnhem The Netherlands
Phone Number: +31 26 384 9080
Contact Person: Gerard J. Langewouters, PhD. General Manager FMS
U. S Agent for FMS: Richard Meissel Propel Biomedical Consulting 15500 Rockfield Boulevard, Suite A-1 Irvine, CA 92618 Phone Number: 949-768-0444 Fax Number: 949-768-8400
# Date Prepared:
9/27/02
# II. DEVICE NAME AND CLASSIFICATION
| Proprietary Name: | Portapres Ambulatory Continuous Non Invasive Blood<br>Pressure Monitor |
|-----------------------|------------------------------------------------------------------------|
| Common or Usual Name: | Continuous Non-Invasive Blood Pressure Monitor |
| Classification: | Class II; CFR 870.1130, Product Code DXN |
{1}------------------------------------------------
### III. PREDICATE DEVICES
The Portapres is substantially equivalent in design (methodology) and indications for use to the following devices that have been or still are in commercial distribution:
- Ohmeda 2300 Finapres® Blood Pressure Monitor, Ohmeda Medical, Englewood, . CO 80112-5810, K822055
- Ohmeda 2350 Finapres® Blood Pressure Monitor, Ohmeda Medical, Englewood, ● CO 80112-5810, K880572
- Task Force Monitor 3040, CNSystems Medizintechnik GMBH, Europe, AU, ● K014063
- (BeatScope software): RODA Monitoring System, Metracor Technologies, Inc., San Diego, CA 921221, K011238
Portapres and Finapres are substantially equivalent in design, methodology, software, manufacture, materials, intended use and principle of operation. The Portapres, the Finapres, and the Task Force Monitor devices are intended for use for continuous monitoring of finger arterial blood pressure, the Portapres in an ambulatory model and the Finapres and Task Force Monitor in stationary models. FMS considers the use of the Portapres to be substantially equivalent to its predicate devices, the Ohmeda 2300 and 2350 Finapres and the CNSystems Task Force Monitor, and considers the BeatScope software cardiac output technology to be equivalent to the RODA Monitoring System cardiac output technology.
### IV. DESCRIPTION
Portapres is a portable, battery-powered device to record continuous non-invasive arterial blood pressure. Portapres is intended for 24-hour continuous recordings in ambulatory subjects. The device includes a Neoprene belt to be worn around the waist, with compartments for the Main Unit with electronics and memory card, a Pump Unit with an air pump, and a NiCd or Li battery pack. Connected to the waist belt Main Unit and Pump Unit is a Wrist Unit, used to interface with the signals from the electronics and air supply in the waist belt to the finger cuffs. The Wrist Unit then connects to a finger cuff, which consists of an inflatable air bladder and an infrared photoplethysmograph. The plethysmograph measures the finger arterial volume which varies with the patient's blood pressure. The Wrist Unit controls the pressure in the cuff bladder so as to keep the arterial volume constant at a level (the setpoint) determined during startup. The Wrist Unit contains a pressure transducer that measures the cuff pressure as an indirect measure of the patient's blood pressure. Also connected to the Wrist Unit is a hydrostatic height correction unit that compensates for the hydrostatic component of the blood pressure in the finger when this is not at heart level. This allows the patient free hand movement in ambulatory applications.
Pressure and volume signals from the finger cuffs are routed from the Wrist Unit to the Main Unit microprocessor for pump control and data acquisition purposes. The
{2}------------------------------------------------
microprocessor has an internal flash memory card for storage of blood pressure wave form data.
The Main Unit is equipped with an RS232 serial interface to transfer the data stored in the flash memory card via a Control Unit to a PC and to allow remote control of the Portapres.
#### V. INTENDED USE
The Portapres Ambulatory Continuous Non-Invasive Blood Pressure Monitor is intended to provide the user with continuous, noninvasive blood pressure and pulse rate monitoring. The Portapres enables 24-hour continuous ambulatory measurements.
The PC-based BeatScope software used with the data from the device provides Modelflow-based computation of real-time and beat-to-beat blood pressure as well as hemodynamic parameters from the pressure waveform, including, cardiac output and total peripheral resistance.
# VI. TECHNOLOGICAL CHARACTERISTICS
No new technology, materials, or change in efficacy have been introduced by FMS in the manufacture of the FMS Portapres Ambulatory Continuous Noninvasive Blood Pressure Monitor. The design, form, and materials of the monitor are equivalent to its predicate devices.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, with the negative space forming a human profile.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAY - 9 2003
FMS, Finapres Medical Systems BV c/o Ms. Christine Emanuel TECSA Technical Services 1205 De La Vina Street Santa Barbara, CA 93101
Re: K023338
Trade Name: Portapres Ambulatory Continuous Non Invasive Blood Pressure Monitor Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive blood pressure measurement system Regulatory Class: Class II (two) Product Code: DXN Dated: March 5, 2003 Received: March 10, 2003
Dear Ms. Emanuel:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annùal registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{4}------------------------------------------------
Page 2 - Ms. Christine Emanuel
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Bram D. Zuckerman, M.D.
Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Include the following "Indications For Use" page that contains the applicant's name, name of the device and the intended use of the device. The information, data and labeling claims in the entire the 510(k) submission must support and agree with the "indications for use" statement.
*For a new submission, do NOT fill in the 510(k) number blank.
## INDICATIONS FOR USE
Applicant: FMS, Finapres Medical Systems BV
510(k) Number (if known): N/A*
Device Name: Portapres Ambulatory Continuous Non-Invasive Blood Pressure Monitor
Indications For Use:
The Portapres Ambulatory Continuous Non-Invasive Blood Pressure Monitor is intended to provide the user with continuous, noninvasive blood pressure and pulse rate monitoring. The Portapres enables 24-hour continuous ambulatory measurements.
The PC-based BeatScope software used with the data from the device provides Modelflowbased computation of real-time and beat-to-beat blood pressure as well as hemodynamic parameters from the pressure waveform, including, cardiac output and total peripheral resistance.
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Cardiovascular Devices
Prescription Use ✗
Per 21 CFR 801.109
510(k) Number K023338
OR
Over-the-Counter ***_***
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.