HDA is a clinical decision support medical device intended to assist healthcare professionals in managing blood pressure and the cardiovascular system in patients undergoing anesthesia during surgical procedures.
Device Story
HDA-OR2 is a clinical decision support SaMD installed on a medical-grade tablet computer. It connects to external multi-parameter patient monitors via wired or wireless networks to acquire continuous arterial blood pressure waveforms and cardiovascular numeric data. The device processes these inputs to derive and display parameters including MAP, heart rate, systolic/diastolic blood pressure, cardiac output, and systemic vascular resistance. It provides graphical and numeric trends, allows users to set MAP targets, and enables manual marking of vasopressor administration and volume challenges. Used in surgical settings by anesthesia professionals, the device acts as a visual aid to assist in hemodynamic management. It supports secure cloud log uploads and remote software updates. By providing real-time cardiovascular status and trend visualization, it assists clinicians in making timely hemodynamic interventions, potentially improving patient stability during anesthesia.
Clinical Evidence
Bench testing only. Verification included measurement accuracy for physiologic parameters per IEC 60601-2-34, artifact detection, power interruption testing, and software verification/validation for a moderate level of concern device. Electrical safety and EMC testing were performed per ES60601-1 and IEC60601-1-2. Testing confirmed the device maintains accuracy and performance over network connections and in electromagnetically noisy hospital environments.
Technological Characteristics
SaMD on medical-grade tablet with internal battery. Connectivity: wired/wireless network (TCP/IP). Standards: IEC 60601-2-34 (measurement accuracy/power), ES60601-1 (safety), IEC60601-1-2 (EMC). Software: Moderate level of concern. Features: Cloud-based log storage and remote update capability via secure API.
Indications for Use
Indicated for adult patients (≥18 years) undergoing surgery with continuous arterial pressure monitoring. Used to assist anesthesia professionals in managing blood pressure, hemodynamic stability, and the cardiovascular system.
Regulatory Classification
Identification
A single-function, preprogrammed diagnostic computer is a hard-wired computer that calculates a specific physiological or blood-flow parameter based on information obtained from one or more electrodes, transducers, or measuring devices.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG ADMINISTRATION" in blue as well. The FDA is a federal agency responsible for regulating and supervising the safety of food, drugs, and other products.
November 23, 2021
Directed Systems Ltd Mark Leaning Chief Executive Officer 47-51 Norfolk Street Cambridge, Cambridgeshire CB1 2LD United Kingdom
Re: K212529
Trade/Device Name: Hypotension Decision Assist Model HDA-OR2 Regulation Number: 21 CFR 870.1435 Regulation Name: Single-function, preprogrammed diagnostic computer Regulatory Class: Class II Product Code: DXG Dated: October 27, 2021 Received: October 29, 2021
Dear Mark Leaning:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for
LCDR Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics, and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
510(k) Number (if known) K212529
#### Device Name
Hypotension Decision Assist HDA-OR2
#### Indications for Use (Describe)
Hypotension Decision Assist is indicated to acquire, process and display arterial pressure and other key cardiovascular characteristics of adult patients who are at least eighteen years of age that are undergoing surgery where their arterial pressure is being continuously monitored by a vital-signs monitor. It is indicated for use to assist anesthesia healthcare professionals manage the blood pressure, hemodynamic stability and the cardiovascular system during such surgery.
| Type of Use (Select one or both, as applicable) | <span> <span style="padding-right: 20px;"> <span style="vertical-align: middle;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> <span> <span style="vertical-align: middle;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </span> |
|-------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."*
FORM FDA 3881 (6/20)
{4}------------------------------------------------
# 510(k) Summary
This 510(k) summary for Hypotension Decision Assist HDA-OR2 is supplied as required by 21 CFR 807.92(c)
# Applicant:
Directed Systems Ltd
47-51 Norfolk Street,
Cambridge
Cambridgeshire CB1 2LD
United Kingdom
Telephone: 011 44 7552 747389
# Contact:
Mark S. Leaning
Chief Executive Officer
Telephone: 011 44 7552 747389
Email: markleaning@directedsystems.com
# Preparation Date:
27 October 2021
# Trade / Proprietary Name:
Hypotension Decision Assist
Model #:
HDA-OR2
# Common Name:
Clinical decision support software
# Classification Name:
Single-function, preprogrammed diagnostic computer
### Regulation:
21 CFR 870.1435
### Product Code:
DXG
# Predicate Device:
Hypotension Decision Assist HDA-OR1 (K190955)
{5}------------------------------------------------
# Device Description:
Hypotension Decision Assist (HDA) is a clinical decision support Software as a Medical Device (SaMD) that is installed upon a medically-rated touch-screen computer. HDA connects to a multi-parameter patient monitor supplied by other manufacturers, from which it acquires vital signs data continuously including the arterial blood pressure waveform and cardiovascular-related numeric parameters.
HDA continually processes this data to display, in graphical charts and numeric format, vital signs data and derived variables including mean arterial pressure (MAP), heart rate, systolic and diastolic blood pressure, cardiac output and systemic vascular resistance. HDA compares MAP to user set targets to indicate when MAP is above or below the target range. It allows the user to mark the administration of vasopressors and volume challenges to the MAP trend.
### Intended Use:
HDA is a clinical decision support medical device intended to assist healthcare professionals in managing blood pressure and the cardiovascular system in patients undergoing anesthesia during surgical procedures.
#### Indications for Use:
HDA is indicated to acquire, process and display arterial pressure and other key cardiovascular characteristics of adult patients who are at least eighteen years of age that are undergoing surgery where their arterial pressure is being continuously monitored by a vital-signs monitor. It is indicated for use to assist anesthesia healthcare professionals manage the blood pressure, hemodynamic stability and the cardiovascular system during such surgery.
### Comparison of Technological Characteristics with the Predicate Device:
HDA-OR2 has the same intended use and indications for use as the HDA-OR1 predicate. They are both intended to assist physicians to manage the blood pressure and cardiovascular stability of patients undergoing anesthesia during surgery.
The subject and predicate device have the following same technological characteristics:
- Clinical decision support software
- Supplied on a medical grade touch screen computer
- . Used with patients whose arterial pressure is being continuously monitored by a vital signs monitor.
- Input data received from a vital signs monitor ●
- Allows serial connection to vital signs monitor ●
- Input data received using same communications protocols ●
- Acquires arterial waveform signal provided by the vital signs monitor ●
{6}------------------------------------------------
- Derives and displays from the input arterial waveform signal; systolic blood ● pressure, diastolic blood pressure, MAP, heart rate, and short term changes in heart rate, cardiac output and systemic vascular resistance
- Provides clinicians with a graphical display of monitoring and support information ● as a visual aid in determining a patient's cardiovascular state.
- Allows marking of the administration of bolus or infused vasopressors and volume . challenges
The following differences in technological characteristics exist between the subject and predicate device:
- . HDA-OR2 is supplied with a tablet computer with an internal battery whereas the predicate HDA-OR1 is supplied with a mains powered all-in-one computer without a battery.
- HDA-OR2 allows connection to Philips patient monitors via a wireless or wired network connection. The predicate HDA-OR1 only allows connections via a serial connection.
- HDA-OR2 allows uploading of its log data to secure cloud storage over the internet. The predicate HDA-OR1 is not internet enabled and only allows log data to be downloaded to connected USB drives.
- . HDA-OR2 can download software updates from a secure API over the internet. The predicate HDA-OR1 is not internet enabled and can only download software updates from connected USB drives.
### Performance Data
#### System verification testing:
The subject device was subjected to the system regression testing for HDA. This system regression testing consolidates the performance testing that was presented in the submission for the predicate HDA-OR1 and included:
#### Measurement accuracy verification:
The measurement accuracy of HDA across the intended use measuring range for each of its measured physiologic parameters verified by bench testing following the approach of the IEC 60601-2-34 Edition 3.0 2011-05 standard that is recognized by FDA. This testing demonstrated the measurement accuracy over the serial and network connections available for the subject device.
#### Artefact Detection Verification
Bench testing performed over a network connection to confirm that HDA is capable of detecting each of the signal artefacts and anomalies that have the potential to impact its performance.
#### Power Interruption testing
Bench testing performed in accordance with the IEC 60601-2-34 Edition 3.0 2011-05 standard that is recognized by FDA to verify that HDA can tolerate a sudden power interruption without loss of either user-input data, or patient data while remaining in the correct operating mode including when the battery is disconnected.
{7}------------------------------------------------
#### Software Verification and Validation Testing
Software verification and validation testing was performed and documentation was provided in accordance with the FDA guidance document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" requirements for a moderate level of concern device. HDA is considered a moderate level of concern device because a malfunction of, or a latent design flaw in the device could lead to an erroneous diagnosis or a delay in delivery of appropriate medical care that would likely lead to Minor Injury.
#### Electrical safety and electromagnetic compatibility (EMC)
Electrical safety and EMC testing was conducted on the touch screen computer supplied with HDA-OR2. The computer complies with the FDA recognized standards ES60601-1 2005/(R) 2012 and A1:2012 for safety and IEC60601-1-2:2014 for EMC.
#### Remote update of software in an electromagnetically noisy environment testing
Testing was performed of HDA's ability to receive remote updates in an electromagnetically noisy environment such as would be found in a hospital installation site. This testing confirmed that HDA can receive remote updates in such an electromagnetically noisy environment, that it will respond as designed and not install updates that are interrupted (interfered or dropped) during wireless communication and that it can detect and reject malware masquerading as legitimate updates.
### Conclusion:
The bench testing and hardware and software verification support the safety of HDA-OR2 and verify that it can accurately measure the physiologic parameters associated with its intended use. Bench testing has demonstrated that the accuracy of its measurements is substantially equivalent to its predicate device and that HDA-OR2 performs as well as its predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.