K060898 · Pulsion Medical Systems AG · DXG · Jul 25, 2006 · Cardiovascular
Device Facts
Record ID
K060898
Device Name
PULSION PICCO PLUS, MODEL 8100
Applicant
Pulsion Medical Systems AG
Product Code
DXG · Cardiovascular
Decision Date
Jul 25, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1435
Device Class
Class 2
Indications for Use
The PULSION PiCCO Plus is intended for determination and monitoring of cardiopulmonary and circulatory variables. Cardiac output is determined both continuously through pulse contour analysis and intermittently through thermodilution technique. In addition, the PiCCO Plus measures heart rate, systolic, and diastolic and derives mean arterial pressure. Analysis of thermodilution curve in terms of mean transit time and downslope time is used for determination of intravascular and extravascular fluid volumes. If a patient's weight and height are entered. the PiCCO Plus presents the derived parameters indexed to body surface area.
Device Story
PiCCO Plus monitors hemodynamic status via pulse contour analysis and thermodilution. Inputs: arterial pressure waveform (via arterial catheter) and thermodilution data (via central venous and arterial thermistor catheters). Device processes signals to calculate cardiac output, heart rate, systolic/diastolic/mean arterial pressure, and fluid volumes (intravascular/extravascular). Used in clinical settings (e.g., ICU) by trained healthcare professionals. Output displayed on monitor for real-time hemodynamic assessment. Assists clinicians in fluid management and cardiovascular therapy decisions; benefits patient through precise monitoring of circulatory function.
Clinical Evidence
No clinical data provided; substantial equivalence based on technological characteristics and performance specifications.
Technological Characteristics
Hemodynamic monitoring system utilizing pulse contour analysis and thermodilution. Interfaces with arterial and central venous catheters. Provides continuous and intermittent measurements. Standalone diagnostic computer. Software-based signal processing for hemodynamic parameter derivation.
Indications for Use
Indicated for determination and monitoring of cardiopulmonary and circulatory variables in patients requiring hemodynamic assessment. No specific age or gender contraindications provided.
Regulatory Classification
Identification
A single-function, preprogrammed diagnostic computer is a hard-wired computer that calculates a specific physiological or blood-flow parameter based on information obtained from one or more electrodes, transducers, or measuring devices.
Related Devices
K072735 — PULSION PICCO-2, MODEL 8500 · Pulsion Medical Systems AG · Oct 18, 2007
K132528 — E-PICCO CONTINUOUS CARDIAC OUTPUT MODULE AND ACCESSORIES · GE Healthcare Finland Oy · Jan 16, 2014
K091786 — PULSION PICCO2 (8500), CEVOX OPTICAL MODULE (PC3015), CEVOX DISPOSABLE PROBES (PV2022-30 THRU -38, PV2022-46 THRU -48) · Pulsion Medical Systems AG · Jun 18, 2010
K122121 — PULSIOFLEX MONITORING SYSTEM WITH PICCO MODULE · Pulsion Medical Systems SE · Aug 2, 2012
K171620 — PiCCO Catheter · Pulsion Medical Systems SE · Feb 21, 2018
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized depiction of a human figure with three heads, representing health, services, and people. The figure is positioned to the right of the department's name, which is arranged in a circular pattern around the logo.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUL 25 2006
Pulsion Medical Systems c/o Dr. Jamie Sulley Triangulum Consulting Services, Inc. PO Box 99033 Raleigh, NC 27624
Re: K060898
> Trade Name: Pulsion Continuous Pulse Contour Cardiac Output (PiCCO Plus) System Regulation Number: 21 CFR 870.1435 Regulation Name: Single-function, Preprogrammed Diagnostic Computer Regulatory Class: Class II (two) Product Code: DXG Dated: June 29, 2006 Reccived: June 30, 2006
Dear Dr. Sulley:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements not the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation enfitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html .
Sincerely yours,
Bhimima for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## INDICATIONS FOR USE STATEMENT 4.0
## Indications for Use
510(k) Number (if known): _ K 06 Od 98
Device Name: PiCCO Plus
Indications for Use:
The PULSION PiCCO Plus is intended for determination and monitoring of cardiopulmonary and circulatory variables. Cardiac output is determined both continuously through pulse contour analysis and intermittently through thermodilution technique. In addition, the PiCCO Plus measures heart rate, systolic, and diastolic and derives mean arterial pressure. Analysis of thermodilution curve in terms of mean transit time and downslope time is used for determination of intravascular and extravascular fluid volumes. If a patient's weight and height are entered. the PiCCO Plus presents the derived parameters indexed to body surface area.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Plan 2/3)
(Division Sign-Off) Division of Cardiovascular Devices 510(k) Number
Page *of*
(Posted November 13, 2003)
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches