SIEMENS SC/9000/SC9015 SERIES SURGICAL DISPLAY CONTROLLER

K970348 · Siemens Medical Solutions USA, Inc. · DSI · Aug 15, 1997 · Cardiovascular

Device Facts

Record IDK970348
Device NameSIEMENS SC/9000/SC9015 SERIES SURGICAL DISPLAY CONTROLLER
ApplicantSiemens Medical Solutions USA, Inc.
Product CodeDSI · Cardiovascular
Decision DateAug 15, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1025
Device ClassClass 2

Indications for Use

The intended use of the Siemens SC9000/9015 Series Surgical Display Controller is to display SC9000/9015 Bedside Monitor functions on a Siemens Remote Display or on a standard video display.

Device Story

The Siemens SC9000/SC9015 Series Surgical Display Controller acts as a remote interface for the SC9000/9015 Bedside Monitor. It receives physiological data and monitoring parameters generated by the bedside monitor and transmits them for display on a Siemens Remote Display or a standard video monitor. Used in clinical settings like the operating room, it allows surgeons and other healthcare personnel to view patient monitoring information at a distance from the primary bedside unit. The device does not perform independent measurements or data processing; it functions solely as a display controller to facilitate clinical visualization of patient status, aiding in real-time clinical decision-making during procedures.

Clinical Evidence

No clinical data; device is substantially equivalent to the predicate device.

Technological Characteristics

The device is a display controller for medical monitoring systems. It functions as a signal interface to route data from a bedside monitor to a remote or standard video display. It does not perform physiological sensing or measurement. It is not MRI compatible.

Indications for Use

Indicated for use in environments where patient care is provided by healthcare professionals, such as surgeons and operating room personnel, whenever the professional determines that parameters and functions generated by the SC 9000/9015 Bedside Monitor should be displayed on a remote terminal. Not intended for use in an MRI environment.

Regulatory Classification

Identification

The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.

Special Controls

*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.

Predicate Devices

Submission Summary (Full Text)

{0} 510 (K) Notification Siemens SC9000 / SC9015 Series Surgical Display Controller AUG 15 1997 # 510(k) SUMMARY as required per 807.92(c) K970348 ## 2: Submitter's Name, Address: Siemens Medical Systems, Inc. 16 Electronics Avenue Danvers, MA 01923 Tel: (508) 750-7500 Fax: (508) 777-3398 Official Correspondent: David Simard, Director of Quality Assurance and Regulatory Affairs Contact person for this submission: Jacqueline E. M. Emery Date submission was prepared: December 10, 1996 ## 3: Trade Name, Common Name and Classification Name: Medical Cathode-Ray Tube Display A. Trade Name: Siemens SC9000/ SC9015 Series Surgical Display Controller B. Common Name, Classification Number, Class and Regulation Number: | Common Name | Classification Number | Class | Regulation Number | | --- | --- | --- | --- | | Cardiac monitor | 74DRT | II | 21 CFR 870.2300 | | Arrhythmia detector & Alarm System | 74DSI | III | 21 CFR 870.1025 | | Breathing frequency monitor | 73BZQ | II | 21 CFR 868.2375 | | Pulse rate monitor | 74BWS | II | 21 CFR 870.2300 | | Non-indwelling blood pressure monitor | 74DXN | II | 21 CFR 870.1130 | | Clinical electronic thermometer | 80BWX | II | 21 CFR 880.2910 | | Pulse Oximeter | 74DQA | II | 21 CFR 870.2700 | | Cardiac Output Monitor | 74KFN | II | 21 CFR 870.1435 | | end-tidal Carbon-Dioxide Monitor | 73CCK | II | 21 CFR 868.1400 | | Indwelling Blood Pressure Monitor | 74CAA | II | 21 CFR 870.1110 | | Indwelling blood pressure monitor | 74CAA | II | 21 CFR 870.1110 | | Heart Rate Monitor, Neonatal | 74FLO | II | 21 CFR 870.2300 | | Ventilatory Effort Monitor (Apnea Detector) | 73FLS | II | 21 CFR 868.2375 | | Monitor Blood Pressure, Neonatal, Invasive | 74FLP | II | 21 CFR 870.1110 | | Medical cathode-ray tube display | 74DXJ | II | 21CFR870.2450 | Siemens Medical Systems, Inc. Electromedical Group 16 Electronics Avenue Danvers, MA 01923 USA Tel: (508) 750-7500 Fax: (508) 777-3398 Telex: 511958 (Siemensm SD) 200·39Hd 020010015 01 {1} 10 (1) NtHcA Siemens SC9000 / SC9015 Series Surgical Display Controller 4: Predicate Device Identification: The Siemens SC9000/9015 Series Surgical Display Controller is substantially equivalent to the Siemens SC9000/9015 Series SC9015 Display. 5. Device Description The SC9000/9015 Series Surgical Display Controller enables SC9000/9015 monitoring functions to be displayed on a Siemens 15 inch Remote Display or on a standard video display. 6. Intended Use: The intended use of the Siemens SC9000/9015 Series Surgical Display Controller is to display SC9000/9015 Bedside Monitor functions on a Siemens Remote Display or on a standard video display. 7. Table of device similarities and differences to predicate device | | Substantial Equivalent Device: Siemens Medical Systems SC9000/9015 Series SC9015 Display | Applicant: Siemens Medical Systems SC9000/9015 Series Surgical Display Controller | | --- | --- | --- | | Manufacturer | Siemens Medical Systems Electromedical Group | Siemens Medical Systems Electromedical Group | | 510K Number | K946306 | K970348 | | Intended Population | Same as Siemens Component Monitoring System | Same | | Displayed Parameters | Same as bedside monitor | Same | | Measuring Methods | None: Information and data generated in bedside monitor | Same | | Waveform Display | Dependent on the bedside monitor's primary display | Same | 8. Assessment of non-clinical performance data for equivalence: The Siemens 9000/9015 Series Surgical Display Controller is substantially equivalent to the predicate device: Siemens 9000/9015 Series SC9015 display. 9. Assessment of clinical performance data for equivalence: Not applicable; device is substantially equivalent to predicate device. 10. Biocompatibility: Not applicable 11. Sterilization: Not applicable Siemens Medical Systems, Inc. Electromedical Group 16 Electronics Avenue Danvers, MA 01923 USA Tel: (508) 750-7500 Fax: (508) 777-3398 Telex: 511958 (Siemensm SD) 000.396d 0200810516 01 SINIH7dW0C SRENNN WORH 16. 26. 97 {2} DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20856 AUG 15 1997 Mr. David Simard Siemens Medical Systems, Inc. 16 Electronics Avenue Danvers, Massachusetts 01923 Re: K970348 Siemens SC9000/SC9015 Series Surgical Display Controller Regulatory Class: III (three) Product Code: 74 DSI Dated: May 16, 1997 Received: May 19, 1997 Dear Mr. Simard: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: This response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {3} Page 2 - Mr. David Simard This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for *in vitro* diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, ![img-0.jpeg](img-0.jpeg) Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4} 510 (K) Notification Siemens SC9000/ SC9015 Series Surgical Display Controller ## Indicated Use Statement The Siemens SC9000/SC9015 Series Surgical Display Controller is indicated for use in an environment where patient care is provided by healthcare professionals (e.g. surgeons and other operating room personnel). It will be used whenever the professional determines that parameters and functions generated by the SC 9000/9015 Bedside Monitor should be displayed on a remote terminal. ## MRI Compatibility Statement: The Siemens SC9000/9015 Series Surgical Display Controller is not intended for use in an MRI environment. *Prescription Use √* *Chouety-Soumen for AAC* (Division Sign-Off) Division of Cardiovascular, Respiratory, and Neurological Devices 510(k) Number **K970348** --- ## Company Confidential **Siemens Medical Systems, Inc.** Electromedical Group 16 Electronics Avenue Danvers, MA 01923 USA Tel: (508) 750-7500 Fax: (508) 777-3398 Telex: 511958 (Siemensm SD)
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...