The Cardio Perfect ST is an ECG recording and analysis system designed for exercise stress testing. The system consists of a recorder unit and a CPCOM interface card for a PC. The recorder captures ECG signals, which are transmitted via an optical connection to the CPCOM card. The system is intended for use by clinicians to perform symptom-limited exercise tests. The software processes the incoming ECG data, allowing for real-time monitoring and comparison against reference ECGs on a color monitor. By providing diagnostic ECG data during exercise, the device assists healthcare providers in assessing cardiac function, identifying arrhythmias, and evaluating prognosis in patients with suspected or known coronary artery disease. The system facilitates clinical decision-making by providing objective measurements of exercise tolerance and cardiac electrical activity.
Clinical Evidence
No clinical data provided. Evidence consists of bench testing, including unit testing for compliance with IEC-601 standards, recorder burn-in and calibration verification using simulators, and CPCOM card data integrity testing using bit-pattern verification at high baud rates.
Technological Characteristics
System comprises a recorder unit and a PC-based CPCOM interface card. Connectivity utilizes an optical transmission link between the recorder and the PC. The system is designed to meet IEC-601 safety standards. Software includes a TSR (Terminate and Stay Resident) program for monitoring CPCOM card performance via DMA channels. Parameters such as defibrillation overload recovery and gain drift are managed by design.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
Related Devices
K022618 — X-SCRIBE II · Mortara Instrument, Inc. · Sep 6, 2002
K024283 — CARDIODIRECT, MODELS 12M, 12I, 12S · Reynolds Medical , Ltd. · Aug 1, 2003
K121638 — ST80I STRESS TEST SYSTEM · Philips Health Care · Jul 18, 2012
K052898 — IQMARK EZ STRESS · Brentwood Medical Technology Corp. · Mar 28, 2006
K082173 — Q-STRESS AND HEARTSTRIDE, MODELS QSTRESS/Q40 AND H4/H4S · Cardiac Science Corporation · Aug 26, 2008
Submission Summary (Full Text)
{0}
K935732
CARDIO PERFECT ST 510k application K935732.
Summary & certification
SIOUKS Summary
# 2. Summary and Certification.
## 2.1. Safety and efficacy of exercise testing in general.
### 2.1.1. Safety of the exercise test.
Rochmis and Blackburn reported a survey of the safety experience of 170000 exercise stress tests performed in 73 medical centers [9]. Sixteen documented deaths were attributed to the exercise, a mortality of less than 1 in 10000. Nonfatal complications that required hospitalization within 1 week after the test occurred at the rate of about 3 per 10000. A more recent survey reported 17 deaths among 712285 patients tested (1 in 40000) [6], with nonfatal ventricular fibrillation occurring at a rate of one per 7000 tests, and nonfatal myocardial infarction at a rate of one per 70000 tests.
In a report of 1377 symptom-limited exercise tests performed in 263 patients with histories of ventricular tachycardia or fibrillation, serious but nonfatal arrhythmia complications occurred during 2.3% of the tests, involving 9.1% of the patients [12]. The incidence of such complications in 8221 maximal exercise tests involving 3444 patients without histories of sustained ventricular arrhythmias performed in the same institution was 0.05%.
### 2.1.2. Efficacy of the exercise test.
Sensitivity and specificity of exercise testing are dependent on a number of variables. Among these are the lead system employed, the (ST) criteria used, and the incidence of CAD in the test population. In a study reported by McNeer and associates in 1978 [8] of 1472 patients, the sensitivity was 57% and the specificity 90%. Deckers [3] reports a sensitivity and specificity, respectively, of 77% and 90% in females and 74% and 90% in males, in 1988 at the Thorax Centrum in Rotterdam, the Netherlands. Simoons at the same institute attributes reports of high incidence of both false positives and false negatives in subgroups of patients largely on patient selection.
Both coronary arteriography with angiographic evaluation of left ventricular function and exercise testing can provide prognostic information in patients with CAD [2]. Gohlke et al. [5] analyzed the additional prognostic value of exercise testing in 1034 patients with known CAD and normal or mildly impaired left ventricular function. In various subgroups, five-year survival was highest in patients with a better exercise tolerance. For example, in patients with double-vessel disease, five-year survival was 95±2% in those with exercise tolerance greater than 110W and only 81±5% in those who performed less than 90W.
0107
{1}
CARDIO PERFECT ST 510k application K93
^{}[]ification
## 2.2. Safety and efficacy of Cardio Perfect ST.
### Unit testing.
A total of about 10 Cardio Perfect MD units were tested for compliance with all system specifications. Some of these were tested in-house as part of the design verification and testing phases of development, and some at institutions like the German TÜV for testing of compliance to local and international standards like the IEC-601.
In the volume production, tests are performed to verify correct operation of each produced device.
### Recorder testing
After confirming that the optical transmission section of the recorder operates properly, the devices are burnt in for 24 hours. After that, the lead connections, gain, time base etc. are measured by recording an ECG with a simulator. Measurements of these recordings are stored and compared, and the ECG is superimposed on a reference ECG on a color monitor for visual inspection.
Before the unit is sent to the client (which may be after a time of storage), four more ECGs are recorded with standard simulator settings, and compared against reference recordings.
Parameters like defibrillation overload recovery and gain drift are met by design and are not tested for each produced device.
About 450 Cardio Perfect recorders have been produced to date.
### CPCOM testing
The first CPCOM tests are performed after production at the same time as testing of the recorders. This is a simple check of proper operation which lasts between one and two minutes (not exhaustive), by recording a resting ECG.
The CPCOM cards are then tested by inserting them into a PC on which special test software is running. A small device generates a pattern of bytes which are sent through the same optical connection as is utilized with the Cardio Perfect recorder. Only, the pattern is known (actually a simple counter), so the software can check every bit of incoming information. With a standard recorder, only 5 bits in every 2-byte sample are predictable. Also, the baudrate used for testing is the maximum baudrate (460800) which the CPCOM can accommodate, which is eight times faster than needed for a standard Cardio Perfect recorder.
Through a DMA channel on the card, the testing software (a TSR) can monitor two cards in one PC at the same time while allowing the PC to be used for other purposes as well. When errors are encountered, a window pops up to report this fact. The cards are tested in this fashion over a period of at least an hour, usually longer.
To date, one CPCOM card was returned which was reported defective, but the jumper settings turned out not to match the settings the client entered in the program.
0108
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.