← Product Code [DRF](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DRF) · K982835

# ELECTRODE SWITCHBOX, MODEL 11-081002 (K982835)

_Cardima, Inc. · DRF · Oct 29, 1998 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DRF/K982835

## Device Facts

- **Applicant:** Cardima, Inc.
- **Product Code:** [DRF](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DRF.md)
- **Decision Date:** Oct 29, 1998
- **Decision:** SESE
- **Submission Type:** Abbreviated
- **Regulation:** 21 CFR 870.1220
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The Cardima EP SELECT Switchbox is an accessory intended to provide selective connection of multi-electrode catheters for electrogram recording and pacing during diagnostic electrophysiology studies.

## Device Story

Passive switching unit for electrophysiology (EP) diagnostic procedures; accepts up to eight unipolar or bipolar intracardiac electrodes via standard pin connectors. Device placed on horizontal surface or at patient side; operated by clinician. Manual switches allow selection between recording or pacing modes for each electrode. Facilitates connection between catheters and EP mapping systems; enables signal acquisition and cardiac pacing. Benefits include convenient, selective control of multiple electrodes during diagnostic studies.

## Clinical Evidence

Bench testing only. Conformance with IEC-60601-1 (Medical Electrical Equipment - Part 1: General Requirements for Safety) established safety and effectiveness.

## Technological Characteristics

Passive electromechanical switching unit; accepts up to eight standard EP lab pin connectors. Form factor designed for horizontal surface or bedside placement. No active electronic components or software described. Sterilization: non-sterile.

## Regulatory Identification

An electrode recording catheter or an electrode recording probe is a device used to detect an intracardiac electrocardiogram, or to detect cardiac output or left-to-right heart shunts. The device may be unipolar or multipolar for electrocardiogram detection, or may be a platinum-tipped catheter which senses the presence of a special indicator for cardiac output or left-to-right heart shunt determinations.

## Predicate Devices

- Webster CES-500 switching system

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# OCT 29 1998

### Summary of Safety and Effectiveness

# Trade Name:

Cardima EP SELECT Switchbox

## Owner:

Cardima, Inc. 47266 Benicia Street Fremont. CA 94538-7330 Contact: Jack P. Douglas, Ph.D. Establishment Registration Number: 9007594

#### Classification Name:

Electrode Recording Catheter (21 CFR 870.1220)

#### Device Classification:

Class II (21 CFR 870.1220) Panel: Circulatory System Devices Panel, DCRND

### Intended Use and Product Description:

The Cardima EP SELECT Switchbox is a passive switching unit used in conventional electrophysiology (EP) diagnostic procedures. It is designed to provide a convenient means by which the user can record or pace from multiple unipolar or bipolar intracardiac electrodes. The device is designed to be placed on a horizontal surface or at the patient's side. It is designed to accept up to eight electrodes using standard pin connectors typically found in the EP lab. Switching between record or pace modes is accomplished by a simple switch for each electrode.

#### Sterilization, Packaging and Pyrogenicity:

The EP SELECT is individually packaged non-sterile in plastic wrap and cardboard box.

#### Substantial Equivalence:

The EP SELECT is a modified version of other conventional accessory switchboxes found in EP mapping systems. Establishment of equivalence was based on similarities of labeling, design, and functionality with the Webster CES-500 switching system.

## Summarv of Safety and Effectiveness:

Safety and effectiveness were established through a "Declaration of Conformity", in part, with recognized consensus standards using International Electrotechnical Commission (IEC) electromechanical standard IEC-60601-1, Medical Electrical Equipment - Part 1: General Requirements for Safety. The results show that the switchbox meets conformance with the standard and is thus safe and effective for use.

2

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image is a black and white seal for the Department of Health & Human Services USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. In the center of the seal is a stylized image of three human profiles facing right, resembling a bird in flight.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

OCT 2 9 1998

Dr. Jack Douglas Manager, Regulatory Affairs CARDIMA, Inc. 47266 Benicia Street Freemont, CA 94539-1372

Re: . K982835 Electrode Switchbox, Model 11-081002 Requlatory Class: II (two) Product Code: 74 DRF Dated: October 14, 1998 Received: October 15, 1998

Dear Dr. Douglas:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions aqainst misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 A substantially equivalent determination assumes compliance to 895. with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Requlation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such Failure to comply with the GMP regulation may result in assumptions. regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any

{2}------------------------------------------------

Page 2 - Dr. Jack Douglas

obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diaqnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, Misbranding by reference to premarket notification (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Thomas J. Callahon

Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

5 1 0(k) Number (if known):

K 982835

Device Name:

Cardima EP SELECT Switchbox

Indications For Use:

The Cardima EP SELECT Switchbox is an accessory intended to provide selective connection of multi-electrode catheters for electrogram recording and pacing during diagnostic electrophysiology studies.

# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF

NEEDED)

# Concurrence of CDRH, Office of Device Evaluation (ODE)

Vola Telle

(Division Sign-Off) Division of Cardiovascular, Respiratory, and Neurological Devices

510(k) Number K982835

**Prescription Use**
**(Per 21 CFR 801. 109)**

OR

す

Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________

(Optional Format 1-2-96)

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DRF/K982835](https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DRF/K982835)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
