Polaris X™ Catheter Cable, Blazer™ Dx-20 Catheter Cable

K230503 · Boston Scientific · DRF · Apr 25, 2023 · Cardiovascular

Device Facts

Record IDK230503
Device NamePolaris X™ Catheter Cable, Blazer™ Dx-20 Catheter Cable
ApplicantBoston Scientific
Product CodeDRF · Cardiovascular
Decision DateApr 25, 2023
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1220
Device ClassClass 2

Indications for Use

The BSC cables are intended to be used with the BSC cardiac diagnostic catheters during an electrophysiology procedure for intracardiac stimulation (pacing) and/or recording of electrical potentials.

Device Story

Blazer Dx-20 and Polaris X Catheter Cables serve as electrical interfaces between BSC cardiac diagnostic catheters and electrophysiology (EP) recording systems. The Blazer Dx-20 cable features a 20-conductor trunk; the Polaris X cable features a 10-conductor trunk. Both cables utilize an over-molded splitter yoke to transition to individual leads terminating in universal 2 mm shrouded male pins. Used in fully-equipped electrophysiology labs, these cables are operated by physicians trained in invasive cardiac mapping and ablation. The cables facilitate the transmission of electrical signals from the heart (via the catheter) to the recording system for diagnostic monitoring or pacing. Both devices are provided sterile and are reusable, supporting up to 10 reprocessing cycles via EO or STERRAD methods, enabling repeated clinical use in mapping and ablation procedures.

Clinical Evidence

No clinical data provided. Substantial equivalence is supported by bench testing, including design verification, bioburden testing per ISO 11737-1, and qualification of 10x manual cleaning and resterilization cycles using EO or STERRAD methods.

Technological Characteristics

Cables consist of jacketed multi-conductor trunks (20-conductor for Blazer Dx-20; 10-conductor for Polaris X) with over-molded splitter yokes. Proximal ends feature 2 mm universal shrouded male pins. Sterilization via EO; reusable with up to 10 reprocessing cycles (EO or STERRAD).

Indications for Use

Indicated for use with BSC cardiac diagnostic catheters during electrophysiology procedures for intracardiac stimulation (pacing) and/or recording of electrical potentials in patients undergoing invasive cardiac mapping and ablation.

Regulatory Classification

Identification

An electrode recording catheter or an electrode recording probe is a device used to detect an intracardiac electrocardiogram, or to detect cardiac output or left-to-right heart shunts. The device may be unipolar or multipolar for electrocardiogram detection, or may be a platinum-tipped catheter which senses the presence of a special indicator for cardiac output or left-to-right heart shunt determinations.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. Food & Drug Administration". April 25, 2023 Boston Scientific Corporation Stephanie Andre Principal Regulatory Affairs Specialist 4100 Hamline Ave North St. Paul, Minnesota 55112 Re: K230503 Trade/Device Name: Blazer™ Dx-20 Catheter Cable, Polaris X™ Catheter Cable Regulation Number: 21 CFR 870.1220 Regulation Name: Electrode Recording Catheter Or Electrode Recording Probe Regulatory Class: Class II Product Code: DRF Dated: February 23, 2023 Received: February 24, 2023 Dear Stephanie Andre: We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {1}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Aneesh S. Deoras -S Aneesh Deoras Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use ## 510(k) Number (if known) K230503 ### Device Name Blazer™ Dx-20 Catheter Cable Polaris XTM Catheter Cable ## Indications for Use (Describe) Blazer Dx-20 Catheter Cable: The BSC Cables are intended to be used with the BSC Cardiac Diagnostic Catheters during an electrophysiology procedure for intracardiac stimulation (pacing) and/or recording of electrical potentials. Polaris X Catheter Cable: The BSC Cables are intended to be used with the BSC Cardiac Diagnostic Catheters during an electrophysiology procedure for intracardiac stimulation (pacing) and/or recording of electrical potentials. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|--| |-------------------------------------------------|--| X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary for Blazer Dx-20 Cable and Polaris X Cable K230503 #### I. Submitter Boston Scientific Corporation Electrophysiology Division 4100 Hamline Ave North St. Paul, MN 55112 Contact: Stephanie Andre Principal Regulatory Affairs Associate Phone: 651-582-5544 E-mail: stephanie.andre@bsci.com Date Prepared: February 23, 2023 ### II. Device | Name of Device: | Blazer™ Dx-20 Catheter Cable | |-----------------------|-------------------------------------------------------------------| | | Polaris X™ Catheter Cable | | Common or Usual Name: | Blazer Dx-20 Cable | | | Polaris X Cable | | Classification Name: | Electrode recording catheter or electrode recording probe (21 CFR | | | 870.1220) | | Regulatory Class: | II | | Product Code: | DRF | ### III. Predicate Device | Subject Device | Predicate Device | Predicate 510(k) | |-------------------------------|-------------------------------|------------------| | BlazerTM Dx-20 Catheter Cable | BlazerTM Dx-20 Catheter Cable | K081576 | | Polaris XTM Catheter Cable | Electrophysiology Cable | K924163 | The predicate cables have not been subject to a design-related recall. No reference devices were used in this submission. ### IV. Device Description ## Blazer™ Dx-20 Cable: The subject Blazer Dx-20 Catheter cable (sold separately) is used with the Blazer Dx-20 Bidirectional Duodecapolar Catheter during an electrophysiology procedure. The Blazer Dx-20 {4}------------------------------------------------ Catheter Cable provides an electrical connection between the Blazer Dx-20 Diagnostic Catheter and the Electrophysiology (EP) Recording System. The cable has one connector that directly connects to the diagnostic catheter and multiple connection points that connect to an EP Recording System. The main trunk consists of a jacketed 20-conductor cable that is split into individual leads by means of an over-molded splitter yoke. The proximal end of the cable has 20 universal 2 mm shrouded male pins that are typical to electrophysiology recording and stimulation systems. The Blazer Dx-20 cable is provided sterile using an EO sterilization process. The cable is reusable and can be reprocessed (i.e. re-sterilized) up to 10 times using either EO or STERRAD™ sterilization methods. The Blazer Dx-20 cable is to be used in a fully-equipped electrophysiology lab by physicians thoroughly trained in invasive cardiac mapping and ablation procedures. # Polaris X™ Cable: The subject Polaris X Catheter Cable (sold separately) is used with the Polaris X Steerable Decapolar Mapping Catheters during an electrophysiology procedure. The Polaris X cable provides an electrical connection between the Polaris X Catheter and the Electrophysiology (EP) Recording System. The cable has one connector that directly connects to the diagnostic catheter and multiple connection points that connect to an EP Recording System. The main trunk consists of a jacketed 10-conductor cable that is split into individual leads by means of an over-molded splitter yoke. The proximal end of the cable has 10 universal 2 mm shrouded male pins that are typical to electrophysiology recording and stimulation systems. The Polaris X cable is provided sterile using an EO sterilization process. The cable is reusable and can be reprocessed (i.e. re-sterilized) up to 10 times using EO sterilization method. The Polaris X cable is to be used in a fully-equipped electrophysiology lab by physicians thoroughly trained in invasive cardiac mapping and ablation procedures. ## UPN and Technical Description Each cable has a single configuration. The UPN and general technical specifications of the Blazer Dx-20 and Polaris X cables are provided in Table 1. | UPN | Product Description | Length | Number of Contacts | Reprocessing Method | Jacket Color | |------------|--------------------------------|--------|--------------------|---------------------|--------------| | M00420S0 | Blazer Dx-20<br>Catheter Cable | 5 ft | 20 | EO<br>STERRAD | Grey | | M0045454S0 | Polaris X<br>Catheter Cable | 5 ft | 10 | EO | Black | Table 1: Electrophysiology Cable Descriptions and UPNs {5}------------------------------------------------ ### V. Intended Use The BSC cables are intended to be used with the BSC cardiac diagnostic catheters during an electrophysiology procedure for intracardiac stimulation (pacing) and/or recording of electrical potentials. ### VI. Comparison of Technological Characteristics with the Predicate Device The Blazer Dx-20 Cable and Polaris X Cable incorporate substantially equivalent design, packaging, fundamental technology, manufacturing processes, sterilization process and intended use as those featured in the predicate devices. The non-patient contacting material changes are the primary technological differences between the subject and predicate devices. ### Performance Data VII. The following performance data were provided in support of the substantial equivalence determination: # Bioburden Testing Bioburden testing, per ISO 11737-1, Microbiological Methods - Part 1: Determination of a Population of Microorganisms on Products, was performed to confirm bioburden levels of the cables were not adversely affected. # Reusable Device Testing The subject Blazer Dx-20 and Polaris X cables underwent 10x manual cleaning and resterilization qualification. The cables were subjected to 2X EO sterilization in qualified EO cycles and then 10X manual cleaning and resterilization cycles using either EO or STERRAD methods, as described in the Blazer Dx-20 and Polaris X cable IFUs, to confirm the cable design meets its predetermined design input specifications. # Design Verification Testing Design verification testing was performed on a subset of product specifications to provide objective evidence that after 2X EO sterilization cycles, distribution challenge, climatic conditioning, and 10X manual cleaning and resterilization cycles, the Blazer Dx-20 and Polaris X cables meet the design input specifications in accordance with the applicable product specifications. {6}------------------------------------------------ ## VIII. Conclusion The subject Blazer Dx-20 Catheter Cable and Polaris X Catheter Cable are substantially equivalent in indications for use, design, product function, materials, and sterility to the predicate Blazer Dx-20 Catheter Cable and predicate Electrophysiology Cable, respectively. The performance data, including sterilization, and design verification testing supports that the subject catheters are substantially equivalent to their respective predicate devices.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...