← Product Code [DRF](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DRF) · K161769

# Reprocessed Supreme Diagnostic Electrophysiology Catheters (K161769)

_Innovative Health, LLC · DRF · Dec 6, 2016 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DRF/K161769

## Device Facts

- **Applicant:** Innovative Health, LLC
- **Product Code:** [DRF](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DRF.md)
- **Decision Date:** Dec 6, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1220
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The Reprocessed Supreme Diagnostic Electrophysiology Catheters can be used in the evaluation of a variety of cardiac arrhythmias from endocardial and intravascular sites.

## Device Story

Reprocessed diagnostic electrophysiology (EP) catheters used for mapping cardiac arrhythmias from endocardial and intravascular sites. Device consists of a catheter body with electrodes; inserted by a physician into the heart to record electrical signals. Signals are transmitted to external electrophysiology recording equipment to assist in arrhythmia diagnosis. Reprocessing involves cleaning, decontamination, inspection, and functional testing of previously used catheters. Each device is limited to one reprocessing cycle. Benefits include providing diagnostic data for arrhythmia management while extending the utility of the original device.

## Clinical Evidence

Bench testing only. Testing included biocompatibility, cleaning and sterilization validation, functional testing (visual inspection, dimensional verification, electrical continuity/resistance, simulated use, mechanical characteristics), and electrical safety testing (dielectric and current leakage).

## Technological Characteristics

Catheters feature various fixed curves, electrode counts, and spacing. Materials and design are identical to predicate devices. Reprocessing includes cleaning, decontamination, and inspection. Devices are limited to one reprocessing cycle. No software or active electronic components integral to the catheter itself; it functions as a passive electrode probe for external recording systems.

## Regulatory Identification

An electrode recording catheter or an electrode recording probe is a device used to detect an intracardiac electrocardiogram, or to detect cardiac output or left-to-right heart shunts. The device may be unipolar or multipolar for electrocardiogram detection, or may be a platinum-tipped catheter which senses the presence of a special indicator for cardiac output or left-to-right heart shunt determinations.

## Predicate Devices

- Response Electrophysiology Catheter, Supreme Electrophysiology Catheter ([K002976](/device/K002976.md))
- Daig Electrophysiology Catheter ([K894500](/device/K894500.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a symbol of three stylized human profiles facing to the right, with three wave-like lines below them.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

December 6, 2016

Innovative Health, LLC Rafal Chudzik VP, R&D and Operations 1435 North Hayden Road Suite 100 Scottsdale, Arizona 85257

Re: K161769

Trade/Device Name: Reprocessed Supreme Diagnostic Electrophysiology Catheters (see attached list of models) Regulation Number: 21 CFR 870.1220 Regulation Name: Electrode Recording Catheter or Electrode Recording Probe Regulatory Class: Class II Product Code: NLH Dated: November 4, 2016 Received: November 7, 2016

Dear Rafal Chudzik:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely.

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

| Description                          | Item<br>Number | French<br>Size | Electrode Spacing<br>(mm)                          | Curve<br>Type | Usable<br>Length<br>(cm) | Number of<br>Electrodes |  |        |   |       |     |    |    |  |        |   |       |     |    |    |  |        |   |       |     |     |    |  |        |   |       |     |    |    |  |        |   |       |     |     |    |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |               |       |     |   |  |        |   |               |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |       |     |   |  |        |   |    |     |     |   |  |        |   |    |       |     |   |  |        |   |    |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |       |     |   |  |        |   |   |       |     |   |  |        |   |    |     |     |   |  |        |   |    |     |     |   |  |        |   |    |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |   |     |     |   |  |        |   |   |     |     |   |  |        |   |    |     |     |   |
|--------------------------------------|----------------|----------------|----------------------------------------------------|---------------|--------------------------|-------------------------|--|--------|---|-------|-----|----|----|--|--------|---|-------|-----|----|----|--|--------|---|-------|-----|-----|----|--|--------|---|-------|-----|----|----|--|--------|---|-------|-----|-----|----|--|--------|---|-------|-------|-----|---|--|--------|---|-------|-------|-----|---|--|--------|---|-------|-------|-----|---|--|--------|---|-------|-------|-----|---|--|--------|---|-------|-------|-----|---|--|--------|---|-------|-------|-----|---|--|--------|---|-------|-------|-----|---|--|--------|---|-------|-------|-----|---|--|--------|---|---------------|-------|-----|---|--|--------|---|---------------|-----|-----|---|--|--------|---|-------|-----|-----|---|--|--------|---|-------|-----|-----|---|--|--------|---|-------|-----|-----|---|--|--------|---|-------|-----|-----|---|--|--------|---|-------|-----|-----|---|--|--------|---|-------|-----|-----|---|--|--------|---|-------|-------|-----|---|--|--------|---|----|-----|-----|---|--|--------|---|----|-------|-----|---|--|--------|---|----|-----|-----|---|--|--------|---|---|-----|-----|---|--|--------|---|---|-------|-----|---|--|--------|---|---|-----|-----|---|--|--------|---|---|-----|-----|---|--|--------|---|---|-------|-----|---|--|--------|---|-------|-----|-----|---|--|--------|---|-------|-----|-----|---|--|--------|---|-------|-------|-----|---|--|--------|---|---|-------|-----|---|--|--------|---|----|-----|-----|---|--|--------|---|----|-----|-----|---|--|--------|---|----|-----|-----|---|--|--------|---|---|-----|-----|---|--|--------|---|---|-------|-----|---|--|--------|---|---|-------|-----|---|--|--------|---|---|-----|-----|---|--|--------|---|---|-------|-----|---|--|--------|---|---|-----|-----|---|--|--------|---|-------|-----|-----|---|--|--------|---|-------|-----|-----|---|--|--------|---|-------|-----|-----|---|--|--------|---|---|-----|-----|---|--|--------|---|---|-----|-----|---|--|--------|---|----|-----|-----|---|
| Supreme<br>Diagnostic EP<br>Catheter | 401957         | 5              | 10                                                 | CRD           | 120                      | 2                       |  |        |   |       |     |    |    |  |        |   |       |     |    |    |  |        |   |       |     |     |    |  |        |   |       |     |    |    |  |        |   |       |     |     |    |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |               |       |     |   |  |        |   |               |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |       |     |   |  |        |   |    |     |     |   |  |        |   |    |       |     |   |  |        |   |    |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |       |     |   |  |        |   |   |       |     |   |  |        |   |    |     |     |   |  |        |   |    |     |     |   |  |        |   |    |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |   |     |     |   |  |        |   |   |     |     |   |  |        |   |    |     |     |   |
|                                      | 401978         | 6              | 2-2-2                                              | CSL           | 65                       | 20                      |  |        |   |       |     |    |    |  |        |   |       |     |    |    |  |        |   |       |     |     |    |  |        |   |       |     |    |    |  |        |   |       |     |     |    |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |               |       |     |   |  |        |   |               |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |       |     |   |  |        |   |    |     |     |   |  |        |   |    |       |     |   |  |        |   |    |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |       |     |   |  |        |   |   |       |     |   |  |        |   |    |     |     |   |  |        |   |    |     |     |   |  |        |   |    |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |   |     |     |   |  |        |   |   |     |     |   |  |        |   |    |     |     |   |
|                                      | 402033         | 6              | 2-2-2-2-2-2-2-2-2-<br>45-2-2-2-2-2-2-2-2-2         | CSL           | 120                      | 20                      |  |        |   |       |     |    |    |  |        |   |       |     |    |    |  |        |   |       |     |     |    |  |        |   |       |     |    |    |  |        |   |       |     |     |    |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |               |       |     |   |  |        |   |               |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |       |     |   |  |        |   |    |     |     |   |  |        |   |    |       |     |   |  |        |   |    |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |       |     |   |  |        |   |   |       |     |   |  |        |   |    |     |     |   |  |        |   |    |     |     |   |  |        |   |    |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |   |     |     |   |  |        |   |   |     |     |   |  |        |   |    |     |     |   |
|                                      | 402020         | 6              | 2-5-2-5-2-5-2-5-2-<br>60-2-10-2-10-2-10-<br>2-10-2 | CRD           | 120                      | 20                      |  |        |   |       |     |    |    |  |        |   |       |     |    |    |  |        |   |       |     |     |    |  |        |   |       |     |    |    |  |        |   |       |     |     |    |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |               |       |     |   |  |        |   |               |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |       |     |   |  |        |   |    |     |     |   |  |        |   |    |       |     |   |  |        |   |    |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |       |     |   |  |        |   |   |       |     |   |  |        |   |    |     |     |   |  |        |   |    |     |     |   |  |        |   |    |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |   |     |     |   |  |        |   |   |     |     |   |  |        |   |    |     |     |   |
|                                      | 401956         | 6              | 10                                                 | CRD           | 120                      | 2                       |  | 401893 | 4 | 2-8-2 | CSL | 65 | 10 |  | 401979 | 4 | 2-5-2 | CSL | 65 | 10 |  | 401863 | 5 | 2-8-2 | CSL | 120 | 10 |  | 401864 | 5 | 2-8-2 | CSL | 65 | 10 |  | 401865 | 5 | 5-5-5 | CSL | 120 | 10 |  | 401966 | 4 | 5-5-5 | CRD-2 | 120 | 6 |  | 401967 | 4 | 2-2-2 | CRD-2 | 120 | 6 |  | 401968 | 4 | 2-5-2 | CRD-2 | 120 | 6 |  | 401872 | 5 | 2-2-2 | CRD-1 | 120 | 6 |  | 402008 | 5 | 2-5-2 | CRD-2 | 120 | 6 |  | 402009 | 5 | 5-5-5 | CRD-2 | 120 | 6 |  | 402010 | 6 | 2-5-2 | CRD-2 | 120 | 6 |  | 402011 | 6 | 5-5-5 | CRD-2 | 120 | 6 |  | 401876 | 6 | 5-5-5-175-175 | CRD-1 | 120 | 6 |  | 401877 | 6 | 5-5-5-175-175 | JSN | 120 | 6 |  | 401891 | 4 | 5-5-5 | CRD | 120 | 4 |  | 401890 | 4 | 5-5-5 | JSN | 120 | 4 |  | 401892 | 4 | 5-5-5 | DAO | 120 | 4 |  | 401993 | 4 | 2-2-2 | JSN | 120 | 4 |  | 401994 | 4 | 2-5-2 | JSN | 120 | 4 |  | 401996 | 4 | 2-5-2 | CRD | 120 | 4 |  | 402012 | 4 | 5-5-5 | CRD-2 | 120 | 4 |  | 401433 | 5 | 10 | CRD | 120 | 4 |  | 401859 | 5 | 10 | CRD-2 | 120 | 4 |  | 401435 | 5 | 10 | JSN | 120 | 4 |  | 401441 | 5 | 5 | CRD | 120 | 4 |  | 401860 | 5 | 5 | CRD-2 | 120 | 4 |  | 401443 | 5 | 5 | JSN | 120 | 4 |  | 401444 | 5 | 5 | DAO | 120 | 4 |  | 401466 | 5 | 5 | DAO-1 | 120 | 4 |  | 401448 | 5 | 2-5-2 | CRD | 120 | 4 |  | 401450 | 5 | 2-5-2 | JSN | 120 | 4 |  | 401468 | 5 | 2-5-2 | DAO-1 | 120 | 4 |  | 401871 | 5 | 2 | CRD-1 | 120 | 4 |  | 401434 | 6 | 10 | CRD | 120 | 4 |  | 401436 | 6 | 10 | JSN | 120 | 4 |  | 401438 | 6 | 10 | DAO | 120 | 4 |  | 401442 | 6 | 5 | CRD | 120 | 4 |  | 401474 | 6 | 5 | CRD-1 | 120 | 4 |  | 402004 | 6 | 5 | CRD-2 | 120 | 4 |  | 401430 | 6 | 5 | JSN | 120 | 4 |  | 401475 | 6 | 5 | JSN-1 | 120 | 4 |  | 401445 | 6 | 5 | DAO | 120 | 4 |  | 401449 | 6 | 2-5-2 | CRD | 120 | 4 |  | 401451 | 6 | 2-5-2 | JSN | 120 | 4 |  | 401453 | 6 | 2-5-2 | DAO | 120 | 4 |  | 401950 | 6 | 5 | CRD | 120 | 4 |  | 401951 | 6 | 5 | JSN | 120 | 4 |  | 401952 | 6 | 10 | CRD | 120 | 4 |
|                                      | 401956         | 6              | 10                                                 | CRD           | 120                      | 2                       |  |        |   |       |     |    |    |  |        |   |       |     |    |    |  |        |   |       |     |     |    |  |        |   |       |     |    |    |  |        |   |       |     |     |    |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |               |       |     |   |  |        |   |               |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |       |     |   |  |        |   |    |     |     |   |  |        |   |    |       |     |   |  |        |   |    |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |       |     |   |  |        |   |   |       |     |   |  |        |   |    |     |     |   |  |        |   |    |     |     |   |  |        |   |    |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |   |     |     |   |  |        |   |   |     |     |   |  |        |   |    |     |     |   |
|                                      | 401893         | 4              | 2-8-2                                              | CSL           | 65                       | 10                      |  |        |   |       |     |    |    |  |        |   |       |     |    |    |  |        |   |       |     |     |    |  |        |   |       |     |    |    |  |        |   |       |     |     |    |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |               |       |     |   |  |        |   |               |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |       |     |   |  |        |   |    |     |     |   |  |        |   |    |       |     |   |  |        |   |    |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |       |     |   |  |        |   |   |       |     |   |  |        |   |    |     |     |   |  |        |   |    |     |     |   |  |        |   |    |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |   |     |     |   |  |        |   |   |     |     |   |  |        |   |    |     |     |   |
|                                      | 401979         | 4              | 2-5-2                                              | CSL           | 65                       | 10                      |  |        |   |       |     |    |    |  |        |   |       |     |    |    |  |        |   |       |     |     |    |  |        |   |       |     |    |    |  |        |   |       |     |     |    |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |               |       |     |   |  |        |   |               |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |       |     |   |  |        |   |    |     |     |   |  |        |   |    |       |     |   |  |        |   |    |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |       |     |   |  |        |   |   |       |     |   |  |        |   |    |     |     |   |  |        |   |    |     |     |   |  |        |   |    |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |   |     |     |   |  |        |   |   |     |     |   |  |        |   |    |     |     |   |
|                                      | 401863         | 5              | 2-8-2                                              | CSL           | 120                      | 10                      |  |        |   |       |     |    |    |  |        |   |       |     |    |    |  |        |   |       |     |     |    |  |        |   |       |     |    |    |  |        |   |       |     |     |    |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |               |       |     |   |  |        |   |               |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |       |     |   |  |        |   |    |     |     |   |  |        |   |    |       |     |   |  |        |   |    |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |       |     |   |  |        |   |   |       |     |   |  |        |   |    |     |     |   |  |        |   |    |     |     |   |  |        |   |    |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |   |     |     |   |  |        |   |   |     |     |   |  |        |   |    |     |     |   |
|                                      | 401864         | 5              | 2-8-2                                              | CSL           | 65                       | 10                      |  |        |   |       |     |    |    |  |        |   |       |     |    |    |  |        |   |       |     |     |    |  |        |   |       |     |    |    |  |        |   |       |     |     |    |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |               |       |     |   |  |        |   |               |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |       |     |   |  |        |   |    |     |     |   |  |        |   |    |       |     |   |  |        |   |    |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |       |     |   |  |        |   |   |       |     |   |  |        |   |    |     |     |   |  |        |   |    |     |     |   |  |        |   |    |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |   |     |     |   |  |        |   |   |     |     |   |  |        |   |    |     |     |   |
|                                      | 401865         | 5              | 5-5-5                                              | CSL           | 120                      | 10                      |  |        |   |       |     |    |    |  |        |   |       |     |    |    |  |        |   |       |     |     |    |  |        |   |       |     |    |    |  |        |   |       |     |     |    |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |               |       |     |   |  |        |   |               |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |       |     |   |  |        |   |    |     |     |   |  |        |   |    |       |     |   |  |        |   |    |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |       |     |   |  |        |   |   |       |     |   |  |        |   |    |     |     |   |  |        |   |    |     |     |   |  |        |   |    |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |   |     |     |   |  |        |   |   |     |     |   |  |        |   |    |     |     |   |
|                                      | 401966         | 4              | 5-5-5                                              | CRD-2         | 120                      | 6                       |  |        |   |       |     |    |    |  |        |   |       |     |    |    |  |        |   |       |     |     |    |  |        |   |       |     |    |    |  |        |   |       |     |     |    |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |               |       |     |   |  |        |   |               |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |       |     |   |  |        |   |    |     |     |   |  |        |   |    |       |     |   |  |        |   |    |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |       |     |   |  |        |   |   |       |     |   |  |        |   |    |     |     |   |  |        |   |    |     |     |   |  |        |   |    |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |   |     |     |   |  |        |   |   |     |     |   |  |        |   |    |     |     |   |
|                                      | 401967         | 4              | 2-2-2                                              | CRD-2         | 120                      | 6                       |  |        |   |       |     |    |    |  |        |   |       |     |    |    |  |        |   |       |     |     |    |  |        |   |       |     |    |    |  |        |   |       |     |     |    |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |               |       |     |   |  |        |   |               |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |       |     |   |  |        |   |    |     |     |   |  |        |   |    |       |     |   |  |        |   |    |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |       |     |   |  |        |   |   |       |     |   |  |        |   |    |     |     |   |  |        |   |    |     |     |   |  |        |   |    |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |   |     |     |   |  |        |   |   |     |     |   |  |        |   |    |     |     |   |
|                                      | 401968         | 4              | 2-5-2                                              | CRD-2         | 120                      | 6                       |  |        |   |       |     |    |    |  |        |   |       |     |    |    |  |        |   |       |     |     |    |  |        |   |       |     |    |    |  |        |   |       |     |     |    |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |               |       |     |   |  |        |   |               |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |       |     |   |  |        |   |    |     |     |   |  |        |   |    |       |     |   |  |        |   |    |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |       |     |   |  |        |   |   |       |     |   |  |        |   |    |     |     |   |  |        |   |    |     |     |   |  |        |   |    |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |   |     |     |   |  |        |   |   |     |     |   |  |        |   |    |     |     |   |
|                                      | 401872         | 5              | 2-2-2                                              | CRD-1         | 120                      | 6                       |  |        |   |       |     |    |    |  |        |   |       |     |    |    |  |        |   |       |     |     |    |  |        |   |       |     |    |    |  |        |   |       |     |     |    |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |               |       |     |   |  |        |   |               |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |       |     |   |  |        |   |    |     |     |   |  |        |   |    |       |     |   |  |        |   |    |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |       |     |   |  |        |   |   |       |     |   |  |        |   |    |     |     |   |  |        |   |    |     |     |   |  |        |   |    |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |   |     |     |   |  |        |   |   |     |     |   |  |        |   |    |     |     |   |
|                                      | 402008         | 5              | 2-5-2                                              | CRD-2         | 120                      | 6                       |  |        |   |       |     |    |    |  |        |   |       |     |    |    |  |        |   |       |     |     |    |  |        |   |       |     |    |    |  |        |   |       |     |     |    |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |               |       |     |   |  |        |   |               |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |       |     |   |  |        |   |    |     |     |   |  |        |   |    |       |     |   |  |        |   |    |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |       |     |   |  |        |   |   |       |     |   |  |        |   |    |     |     |   |  |        |   |    |     |     |   |  |        |   |    |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |   |     |     |   |  |        |   |   |     |     |   |  |        |   |    |     |     |   |
|                                      | 402009         | 5              | 5-5-5                                              | CRD-2         | 120                      | 6                       |  |        |   |       |     |    |    |  |        |   |       |     |    |    |  |        |   |       |     |     |    |  |        |   |       |     |    |    |  |        |   |       |     |     |    |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |               |       |     |   |  |        |   |               |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |       |     |   |  |        |   |    |     |     |   |  |        |   |    |       |     |   |  |        |   |    |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |       |     |   |  |        |   |   |       |     |   |  |        |   |    |     |     |   |  |        |   |    |     |     |   |  |        |   |    |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |   |     |     |   |  |        |   |   |     |     |   |  |        |   |    |     |     |   |
|                                      | 402010         | 6              | 2-5-2                                              | CRD-2         | 120                      | 6                       |  |        |   |       |     |    |    |  |        |   |       |     |    |    |  |        |   |       |     |     |    |  |        |   |       |     |    |    |  |        |   |       |     |     |    |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |               |       |     |   |  |        |   |               |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |       |     |   |  |        |   |    |     |     |   |  |        |   |    |       |     |   |  |        |   |    |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |       |     |   |  |        |   |   |       |     |   |  |        |   |    |     |     |   |  |        |   |    |     |     |   |  |        |   |    |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |   |     |     |   |  |        |   |   |     |     |   |  |        |   |    |     |     |   |
|                                      | 402011         | 6              | 5-5-5                                              | CRD-2         | 120                      | 6                       |  |        |   |       |     |    |    |  |        |   |       |     |    |    |  |        |   |       |     |     |    |  |        |   |       |     |    |    |  |        |   |       |     |     |    |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |               |       |     |   |  |        |   |               |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |       |     |   |  |        |   |    |     |     |   |  |        |   |    |       |     |   |  |        |   |    |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |       |     |   |  |        |   |   |       |     |   |  |        |   |    |     |     |   |  |        |   |    |     |     |   |  |        |   |    |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |   |     |     |   |  |        |   |   |     |     |   |  |        |   |    |     |     |   |
|                                      | 401876         | 6              | 5-5-5-175-175                                      | CRD-1         | 120                      | 6                       |  |        |   |       |     |    |    |  |        |   |       |     |    |    |  |        |   |       |     |     |    |  |        |   |       |     |    |    |  |        |   |       |     |     |    |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |               |       |     |   |  |        |   |               |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |       |     |   |  |        |   |    |     |     |   |  |        |   |    |       |     |   |  |        |   |    |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |       |     |   |  |        |   |   |       |     |   |  |        |   |    |     |     |   |  |        |   |    |     |     |   |  |        |   |    |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |   |     |     |   |  |        |   |   |     |     |   |  |        |   |    |     |     |   |
|                                      | 401877         | 6              | 5-5-5-175-175                                      | JSN           | 120                      | 6                       |  |        |   |       |     |    |    |  |        |   |       |     |    |    |  |        |   |       |     |     |    |  |        |   |       |     |    |    |  |        |   |       |     |     |    |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |               |       |     |   |  |        |   |               |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |       |     |   |  |        |   |    |     |     |   |  |        |   |    |       |     |   |  |        |   |    |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |       |     |   |  |        |   |   |       |     |   |  |        |   |    |     |     |   |  |        |   |    |     |     |   |  |        |   |    |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |   |     |     |   |  |        |   |   |     |     |   |  |        |   |    |     |     |   |
|                                      | 401891         | 4              | 5-5-5                                              | CRD           | 120                      | 4                       |  |        |   |       |     |    |    |  |        |   |       |     |    |    |  |        |   |       |     |     |    |  |        |   |       |     |    |    |  |        |   |       |     |     |    |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |               |       |     |   |  |        |   |               |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |       |     |   |  |        |   |    |     |     |   |  |        |   |    |       |     |   |  |        |   |    |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |       |     |   |  |        |   |   |       |     |   |  |        |   |    |     |     |   |  |        |   |    |     |     |   |  |        |   |    |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |   |     |     |   |  |        |   |   |     |     |   |  |        |   |    |     |     |   |
|                                      | 401890         | 4              | 5-5-5                                              | JSN           | 120                      | 4                       |  |        |   |       |     |    |    |  |        |   |       |     |    |    |  |        |   |       |     |     |    |  |        |   |       |     |    |    |  |        |   |       |     |     |    |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |               |       |     |   |  |        |   |               |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |       |     |   |  |        |   |    |     |     |   |  |        |   |    |       |     |   |  |        |   |    |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |       |     |   |  |        |   |   |       |     |   |  |        |   |    |     |     |   |  |        |   |    |     |     |   |  |        |   |    |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |   |     |     |   |  |        |   |   |     |     |   |  |        |   |    |     |     |   |
|                                      | 401892         | 4              | 5-5-5                                              | DAO           | 120                      | 4                       |  |        |   |       |     |    |    |  |        |   |       |     |    |    |  |        |   |       |     |     |    |  |        |   |       |     |    |    |  |        |   |       |     |     |    |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |               |       |     |   |  |        |   |               |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |       |     |   |  |        |   |    |     |     |   |  |        |   |    |       |     |   |  |        |   |    |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |       |     |   |  |        |   |   |       |     |   |  |        |   |    |     |     |   |  |        |   |    |     |     |   |  |        |   |    |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |   |     |     |   |  |        |   |   |     |     |   |  |        |   |    |     |     |   |
|                                      | 401993         | 4              | 2-2-2                                              | JSN           | 120                      | 4                       |  |        |   |       |     |    |    |  |        |   |       |     |    |    |  |        |   |       |     |     |    |  |        |   |       |     |    |    |  |        |   |       |     |     |    |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |               |       |     |   |  |        |   |               |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |       |     |   |  |        |   |    |     |     |   |  |        |   |    |       |     |   |  |        |   |    |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |       |     |   |  |        |   |   |       |     |   |  |        |   |    |     |     |   |  |        |   |    |     |     |   |  |        |   |    |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |   |     |     |   |  |        |   |   |     |     |   |  |        |   |    |     |     |   |
|                                      | 401994         | 4              | 2-5-2                                              | JSN           | 120                      | 4                       |  |        |   |       |     |    |    |  |        |   |       |     |    |    |  |        |   |       |     |     |    |  |        |   |       |     |    |    |  |        |   |       |     |     |    |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |               |       |     |   |  |        |   |               |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |       |     |   |  |        |   |    |     |     |   |  |        |   |    |       |     |   |  |        |   |    |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |       |     |   |  |        |   |   |       |     |   |  |        |   |    |     |     |   |  |        |   |    |     |     |   |  |        |   |    |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |   |     |     |   |  |        |   |   |     |     |   |  |        |   |    |     |     |   |
|                                      | 401996         | 4              | 2-5-2                                              | CRD           | 120                      | 4                       |  |        |   |       |     |    |    |  |        |   |       |     |    |    |  |        |   |       |     |     |    |  |        |   |       |     |    |    |  |        |   |       |     |     |    |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |               |       |     |   |  |        |   |               |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |       |     |   |  |        |   |    |     |     |   |  |        |   |    |       |     |   |  |        |   |    |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |       |     |   |  |        |   |   |       |     |   |  |        |   |    |     |     |   |  |        |   |    |     |     |   |  |        |   |    |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |   |     |     |   |  |        |   |   |     |     |   |  |        |   |    |     |     |   |
|                                      | 402012         | 4              | 5-5-5                                              | CRD-2         | 120                      | 4                       |  |        |   |       |     |    |    |  |        |   |       |     |    |    |  |        |   |       |     |     |    |  |        |   |       |     |    |    |  |        |   |       |     |     |    |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |               |       |     |   |  |        |   |               |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |       |     |   |  |        |   |    |     |     |   |  |        |   |    |       |     |   |  |        |   |    |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |       |     |   |  |        |   |   |       |     |   |  |        |   |    |     |     |   |  |        |   |    |     |     |   |  |        |   |    |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |   |     |     |   |  |        |   |   |     |     |   |  |        |   |    |     |     |   |
|                                      | 401433         | 5              | 10                                                 | CRD           | 120                      | 4                       |  |        |   |       |     |    |    |  |        |   |       |     |    |    |  |        |   |       |     |     |    |  |        |   |       |     |    |    |  |        |   |       |     |     |    |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |               |       |     |   |  |        |   |               |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |       |     |   |  |        |   |    |     |     |   |  |        |   |    |       |     |   |  |        |   |    |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |       |     |   |  |        |   |   |       |     |   |  |        |   |    |     |     |   |  |        |   |    |     |     |   |  |        |   |    |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |   |     |     |   |  |        |   |   |     |     |   |  |        |   |    |     |     |   |
|                                      | 401859         | 5              | 10                                                 | CRD-2         | 120                      | 4                       |  |        |   |       |     |    |    |  |        |   |       |     |    |    |  |        |   |       |     |     |    |  |        |   |       |     |    |    |  |        |   |       |     |     |    |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |               |       |     |   |  |        |   |               |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |       |     |   |  |        |   |    |     |     |   |  |        |   |    |       |     |   |  |        |   |    |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |       |     |   |  |        |   |   |       |     |   |  |        |   |    |     |     |   |  |        |   |    |     |     |   |  |        |   |    |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |   |     |     |   |  |        |   |   |     |     |   |  |        |   |    |     |     |   |
|                                      | 401435         | 5              | 10                                                 | JSN           | 120                      | 4                       |  |        |   |       |     |    |    |  |        |   |       |     |    |    |  |        |   |       |     |     |    |  |        |   |       |     |    |    |  |        |   |       |     |     |    |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |               |       |     |   |  |        |   |               |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |       |     |   |  |        |   |    |     |     |   |  |        |   |    |       |     |   |  |        |   |    |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |       |     |   |  |        |   |   |       |     |   |  |        |   |    |     |     |   |  |        |   |    |     |     |   |  |        |   |    |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |   |     |     |   |  |        |   |   |     |     |   |  |        |   |    |     |     |   |
|                                      | 401441         | 5              | 5                                                  | CRD           | 120                      | 4                       |  |        |   |       |     |    |    |  |        |   |       |     |    |    |  |        |   |       |     |     |    |  |        |   |       |     |    |    |  |        |   |       |     |     |    |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |               |       |     |   |  |        |   |               |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |       |     |   |  |        |   |    |     |     |   |  |        |   |    |       |     |   |  |        |   |    |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |       |     |   |  |        |   |   |       |     |   |  |        |   |    |     |     |   |  |        |   |    |     |     |   |  |        |   |    |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |   |     |     |   |  |        |   |   |     |     |   |  |        |   |    |     |     |   |
|                                      | 401860         | 5              | 5                                                  | CRD-2         | 120                      | 4                       |  |        |   |       |     |    |    |  |        |   |       |     |    |    |  |        |   |       |     |     |    |  |        |   |       |     |    |    |  |        |   |       |     |     |    |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |               |       |     |   |  |        |   |               |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |       |     |   |  |        |   |    |     |     |   |  |        |   |    |       |     |   |  |        |   |    |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |       |     |   |  |        |   |   |       |     |   |  |        |   |    |     |     |   |  |        |   |    |     |     |   |  |        |   |    |     |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |   |       |     |   |  |        |   |   |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |       |     |     |   |  |        |   |   |     |     |   |  |        |   |   |     |     |   |  |        |   |    |     |     |   |
|                                      | 401443         | 5              | 5                                                  | JSN           | 120                      | 4                       |  |        |   |       |     |    |    |  |        |   |       |     |    |    |  |        |   |       |     |     |    |  |        |   |       |     |    |    |  |        |   |       |     |     |    |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |       |       |     |   |  |        |   |               |       |     |   |  |        |   |               |     |     |   |  |        |   |       |     |     |   |  |…

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DRF/K161769](https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DRF/K161769)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
