CORONARY SINUS DEFLECTABLE MAPPING CATHETER, MODEL D-1263-01, WEBSTER CS CATHETER WITH EZ STEER
Applicant
Biosense Webster, Inc.
Product Code
DRF · Cardiovascular
Decision Date
Sep 4, 2009
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1220
Device Class
Class 2
Indications for Use
The Webster CS Catheter is indicated for electrophysiological mapping of cardiac structures; i.e., stimulation and recording only. The catheter is designed for use in the coronary sinus.
Device Story
Diagnostic 7Fr deflectable electrophysiology mapping catheters; used for mapping electrical activity in coronary sinus. Input: cardiac electrical signals via distal Platinum/Iridium electrodes. Operation: physician-controlled deflection (uni-directional or bi-directional) via handle (thumb knob or rocker lever) to position catheter tip; high-torque shaft allows rotation. Output: electrical signals transmitted to external recording system or Carto EP Navigation System. Auto ID models (D-1263-06-S/07-S) include EEPROM for automatic catheter recognition by Carto systems. Used in clinical EP labs; assists physicians in diagnosing cardiac arrhythmias by visualizing electrical activity; facilitates accurate catheter placement.
Clinical Evidence
Bench testing only. No clinical data provided.
Technological Characteristics
7Fr deflectable EP catheters; Platinum/Iridium electrodes; braided, high-torque shaft; uni-directional or bi-directional deflection mechanisms; handle-based control; EEPROM for Auto ID (specific models); interface via 10-pin connector or dedicated interface cable; sterilization method not specified.
Indications for Use
Indicated for electrophysiological mapping, stimulation, and recording of cardiac structures within the coronary sinus.
Regulatory Classification
Identification
An electrode recording catheter or an electrode recording probe is a device used to detect an intracardiac electrocardiogram, or to detect cardiac output or left-to-right heart shunts. The device may be unipolar or multipolar for electrocardiogram detection, or may be a platinum-tipped catheter which senses the presence of a special indicator for cardiac output or left-to-right heart shunt determinations.
{0}------------------------------------------------
# Ko90898
**SEP 04 2009**
# 510(K) SUMMARY
# Applicant:
Biosense Webster, Inc. 3333 Diamond Canyon Rd. Diamond Bar, CA 91765, USA Phone: 800-839-8599 Fax: 909-839-8804
#### Date:
September 2, 2009
#### Contact Person:
Natalie Bennington, Regulatory Affairs Manager
# Proprietary Device Name / Manufacturing Part Number:
- · Coronary Sinus Deflectable Mapping Catheters (D-1263-01-S & D-1263-02-S)
- · Webster CS Catheter with EZ Steer Technology (D-1263-04-S & D-1263-05-S)
- · Webster CS Catheter with EZ Steer Technology and Auto ID (D-1263-06-S & D-1263-07-S)
#### Common Device Name:
Electrophysiology Mapping Catheter
#### Classification Name:
Electrode Recording Catheter (per 21 CFR 870.1220, Product code DRF)
# Predicate Device:
Biosense Webster Coronary Sinus Deflectable Mapping Catheters (D-1263-01-S & D-1263-02-S), originally cleared via 510(k) K050877 on June 24, 2005.
{1}------------------------------------------------
## Manufacturing Facilities:
1
· Biosense Webster, Inc. Circuito Interior Norte #1820 Parque Industrial Salvacar Juarez, Chihuahua Mexico, Mexico 32599
- . Biosense Webster, Inc. 15715 Arrow Highway Irwindale, CA 91706
#### Device Description
# The Webster Coronary Sinus Catheter (D-1263-01-S & D-1263-02-S)
The Coronary Sinus Deflectable Mapping Catheters (D-1263-01-S & D-1263-02-S) are diagnostic, 7Fr, deflectable, mapping electrophysiology (EP) catheters with the ability to map electrical activity within the Coronary Sinus through distal Platinum/Iridium electrodes located along the catheter's pre-shaped tip. The catheters incorporate a 2 mm tip electrode, 10 total electrodes, 2-5-2 mm or 2-8-2 mm electrode spacing, have uni-directional deflection with an MZ curve, and are 115 cm long. The catheters include a handle with a thumb knob, which allows deflection of the distal end of the catheter. The handle also has a 10-pin connector for interfacing to a recording system though a standard catheter cable. The braided, bendable/conformable shaft allows torque and contact force to be transmitted from the handle to the distal end. The distal tip is unbraided and soft. These two predicate models are identical except that the electrode spacing is 2-5-2 mm for D-1263-01-S and 2-8-2 mm for D-1263-02-S. The following cables are used to provide a means for interface of the catheters with the appropriate equipment:
- · D-1221-21-S
- · D-1221-26-S
- . D-1221-25-S
### The Webster Coronary Sinus Catheter with EZ Steer Technology (D-1263-04-S & D-1263-05-S)
The Webster CS Catheters with EZ Steer Technology (D-1263-04-S & D-1263-05-S) are diagnostic, 7Fr. deflectable, mapping electrophysiology (EP) catheters with the ability to map electrical activity within the Coronary Sinus through distal Platinum/Iridium electrodes located along the catheter's pre-shaped tip. The catheters incorporate a 2 mm tip electrode, 10 total electrodes, 2-8-2 mm electrode spacing, have bi-directional deflection and are 115 cm long. These
{2}------------------------------------------------
catheters include a braided bi-directional deflectable tip section. The braided bidirectional tip provides the user with two 180° opposed single plane curves. Currently, the available curves for the Webster CS Catheters with EZ Steer Technology include FJ (D-1263-04-S) and DF (D-1263-05-S). These catheters include a handle with a Rocker Lever, which is used to deflect the tip. The hightorque shaft allows the plane of the curved tip to be rotated to facilitate accurate positioning of the catheter tip at the desired site. The following cables are used to provide a means for interface of the catheters with the appropriate equipment:
- · D-1221-21-S
- · D-1221-25-S
- . D-1221-26-S
# The Webster Coronary Sinus Catheter with EZ Steer Technology (D-1263-06-S & D-1263-07-S)
The Webster CS Catheters with EZ Steer Technology (D-1263-06-S & D-1263-07-S) are diagnostic, 7Fr, deflectable, mapping electrophysiology (EP) catheters with the ability to map electrical activity within the Coronary Sinus through distal Platinum/Iridium electrodes located along the catheter's pre-shaped tip. The catheters incorporate a 2 mm tip electrode, 10 total electrodes, 2-8-2 mm electrode spacing, have bi-directional deflection and are 115 cm long. These catheters include a braided bi-directional deflectable tip section. The braided bidirectional tip provides the user with two 180° opposed single plane curves. Currently, the available curves for the modified Webster CS Catheters include FJ (D-1263-06-S) and DF (D-1263-07-S). These catheters include a handle with a Rocker Lever which is used to deflect the tip. The high-torque shaft allows the plane of the curved tip to be rotated to facilitate accurate positioning of the catheter tip at the desired site. The catheters are equipped with Electronically Erasable Programable Read Only Memory (EEPROM) which is used to store unique catheter identification information. Carto EP Navigation Systems equipped with Auto ID Technology can access the stored information and automatically recognize the catheter information.
The catheters interface with Carto EP Navigation Systems via an interface cable (D-1286-16-S) with the appropriate connectors.
#### Indications for Use
The Webster CS Catheter is indicated for electrophysiological mapping of cardiac structures; i.e., stimulation and recording only. The catheter is designed for use in the coronary sinus.
{3}------------------------------------------------
# Summary of Testing
ﺎ
The Webster CS Catheters have been subjected to performance testing to validate the integrity of the catheters and the performance of the catheters with the ancillary equipment.
### Substantial Equivalence
The Webster CS Catheters (D-1263-01-S, D-1263-02-S, D-1263-04-S, D-1263-05-S, D-1263-06-S & D-1263-07-S) are substantially equivalent to the Biosense Webster Coronary Sinus Deflectable Mapping Catheters (D-1263-01 & D-1263-02), originally cleared via 510(k) K050877 on June 24, 2005.
{4}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features a stylized eagle with three stripes representing the department's mission. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
SEP 0 4 2009
Biosense Webster. Inc. c/o Ms. Natalie Bennington Manager, Regulatory Affairs 3333 Diamond Canyon Road Diamond Bar, CA 91765
#### Re: K090898
Trade/Device Name:
- . Webster CS Catheter (D-1263-01-S and D-1263-02-S);
- Webster CS Catheter with EZ Steer Technology (D-1263-04-S and D-1263-05-S); and
- Webster CS Catheter with EZ Steer Technology and Auto ID (D-1263-06-S and D-1263-07-S).
Regulatory Number: 21 CFR 870.1220 Regulation Name: Electrode Recording Catheter or Electrode Recording Probe Regulatory Class: Class II (Two) Product Code: DRF Dated: July 28, 2009 Received: August 5, 2009
Dear Ms. Bennington:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
{5}------------------------------------------------
### Page 2 - Ms. Natalie Bennington
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050,
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHQffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
W.M.J.
or Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# INDICATIONS FOR USE
510(k) No (if known): K090898
#### Device Name:
- · Webster CS Catheter D-1263-01-S D-1263-02-S
- · Webster CS Catheter with EZ Steer Technology D-1263-04-S D-1263-05-S
- · Webster CS Catheter with EZ Steer Technology and Auto ID D-1263-06-S D-1263-07-S
#### Indications for Use:
The Webster CS Catheter is indicated for electrophysiological mapping of cardiac structures; i.e., stimulation and recording only. The catheter is designed for use in the coronary sinus.
Prescription Use V (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| | <img alt="Signature" src="signature.png"/> |
|--|--------------------------------------------|
|--|--------------------------------------------|
| (Division Sign-Off) | |
|------------------------------------|--|
| Division of Cardiovascular Devices | |
F510(k) Number
Page 1 of 1