K212784 · Spectranetics, Inc. · DRE · Aug 24, 2022 · Cardiovascular
Device Facts
Record ID
K212784
Device Name
TightRail Guardian Motorized Dilator Sheath
Applicant
Spectranetics, Inc.
Product Code
DRE · Cardiovascular
Decision Date
Aug 24, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1310
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The TightRail Guardian Motorized Dilator Sheath is intended for use in patients requiring the percutaneous dilation of tissue to facilitate removal of cardiac leads.
Device Story
Sterile, single-use motorized dilator sheath for cardiac lead removal. Device components: drive assembly, dilation extension selection (DES) assembly, motor drive handle, outer sheath accessory, fish tape accessory. Operation: device advanced/retracted along target lead; trigger activation initiates bidirectional rotation of inner shaft and cam blade. Two modes: shielded or extended blade dilation. Outer sheath provides support/conduit and maintains venous access. Fish tape feeds target lead/locking stylet through inner lumen. Used in clinical settings by physicians. Motorized mechanism replaces manual operation of predicate devices; facilitates tissue dilation and fibrous attachment removal to assist lead extraction.
Clinical Evidence
No clinical data. Substantial equivalence supported by design verification and validation testing, including simulated use, human factors, dimensional, mechanical (axial load, tensile, flexibility, compression), EMC/EMI, electrical safety, radiopacity, and comprehensive biocompatibility testing (cytotoxicity, sensitization, irritation, pyrogenicity, systemic toxicity, hemolysis).
Technological Characteristics
Battery-powered motorized dilation mechanism; two modes (Shielded, Extended). Laser-cut hypotube shaft design. Sizes: 9F to 13F. Nominal effective length: 47.5 cm. Sterilization via product adoption equivalency per AAMI TIR28:2016. Biocompatibility per ISO 10993 standards.
Indications for Use
Indicated for patients requiring percutaneous dilation of tissue to facilitate removal of cardiac leads.
Regulatory Classification
Identification
A vessel dilator for percutaneous catheterization is a device which is placed over the guide wire to enlarge the opening in the vessel, and which is then removed before sliding the catheter over the guide wire.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a stacked format.
August 24, 2022
Spectranetics Corporation Jordan Baum Regulatory Affairs Specialist 9965 Federal Drive Colorado Springs, Colorado 80921
Re: K212784
Trade/Device Name: TightRail Guardian Motorized Dilator Sheath Regulation Number: 21 CFR 870.1310 Regulation Name: Vessel Dilator For Percutaneous Catheterization Regulatory Class: Class II Product Code: DRE Dated: August 16, 2022 Received: August 22, 2022
#### Dear Jordan Baum:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Hetal Odobasic Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K212784
Device Name
TightRail Guardian Motorized Dilator Sheath
Indications for Use (Describe)
The TightRail Guardian Motorized Dilator Sheath is intended for use in patients requiring the percutaneous dilation of tissue to facilitate removal of cardiac leads.
Type of Use (Select one or both, as applicable)
| <div> <span> </span> Prescription Use (Part 21 CFR 801 Subpart D) </div> |
|------------------------------------------------------------------------------------------------------|
| <div> <span> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Spectranetics. The logo features a blue geometric shape resembling a stylized atom or starburst to the left of the company name, "Spectranetics," which is written in a bold, sans-serif blue font. Below the company name, there is a tagline that reads "Always Reaching Farther" in a smaller, italicized font.
# 510(k) Summary
This 510(k) summary was prepared in accordance with 21 CFR 807.92(c) Prepared on 24 August 2022
| 510(k) Submitter / Holder: | Spectranetics<br>9965 Federal Drive<br>Colorado Springs, CO 80921.3617<br>Establishment Registration No: 3007284006 |
|----------------------------|-------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Mr. Jordan Baum<br>Regulatory Affairs Specialist<br>Mobile: (719) 247-0183<br>Fax: (719) 447-2070<br>Email: Jordan.Baum@philips.com |
#### Subject Device
| Device Trade Name: | TightRail Guardian Motorized Dilator Sheath |
|----------------------------|-------------------------------------------------|
| Device Common Name: | Sheath |
| Device Class: | II |
| Classification Regulation: | 21 CFR 870.1310 |
| Regulation Description: | Vessel dilator for percutaneous catheterization |
| Product Code: | DRE |
| 510(k) Type: | Traditional |
| Model Numbers: | 575-009, 575-011, 575-013 |
#### Predicate Devices
The TightRail Guardian Motorized Dilator Sheath is being compared to the following legally marketed predicate devices:
| Primary Predicate Device: TightRail | |
|---------------------------------------------|------------------------------------------|
| 510(k) Number: | K150360 (cleared 4 March 2015) |
| Manufacturer: | The Spectranetics Corporation |
| Trade Name: | TightRail Rotating Dilator Sheaths |
| Device Common Name: | Sheath |
| Reference Predicate Device: TightRail Sub-C | |
| 510(k) Number: | K161333 (Cleared 8 July 2016) |
| Manufacturer: | The Spectranetics Corporation |
| Trade Name: | TightRail Sub-C Rotating Dilator Sheaths |
| Device Common Name: | Sheath |
#### Intended and Indications for Use
The TightRail Guardian Motorized Dilator Sheath is intended for use in patients requiring the percutaneous dilation of tissue to facilitate removal of cardiac leads.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the Spectranetics logo. The logo consists of a blue diamond shape with a white circle inside, followed by the word "Spectranetics" in blue, bold font. Below the word "Spectranetics" is the tagline "Always Reaching Farther" in a smaller, regular font.
#### Device Description
The TightRail Guardian Motorized Dilator Sheath is a sterile, single use, prescription only device used for cardiac leads. The device is comprised of:
- 1. Drive Assembly
- 2. Dilation Extension Selection (DES) Assembly
- 3. Motor Drive Handle
- 4. Outer Sheath Accessory
- 5. Fish Tape Accessory
The motorized dilator sheath is advanced or retracted along the target lead to be removed. Pulling the trigger on the handle of the TightRail Guardian device results in activation of the motor and subsequent rotation of the inner shaft and cam blade. Triager activation results in bidirectional rotation mechanism. Actuation of the distal dilation mechanism causes dilation of tissue and fibrous attachments surrounding the lead tarqeted for removal, thereby facilitating removal of said lead. The device has two modes of operation, providing dilation with the blades shielded or extended. The diameter sizes range from 9 French (F) to 13 F. The nominal effective length of the TightRail Guardian is 47.5 cm. The outer sheath accessory may be used for additional support by creating a conduit for the device's shaft assembly. The outer sheath may also be used to maintain venous access for guidewire placement, prior to implantation of a new lead. The fish tape accessory may be used to feed a target lead and locking stylet through the inner lumen of the device.
#### Technological Characteristics
The TightRail Guardian Motorized Dilator Sheath features the same performance characteristics as the predicate devices (K150360 and K161333). There are no significant changes to the principle of operation and no changes to the intended use. The TightRail Guardian device features a battery powered dilation mechanism and two dilation modes (Shielded, Extended). The predicate devices feature a hand-operated dilation mechanism and a single dilation mode (equivalent to Extended Mode). TightRail Guardian features a similar shaft design to TightRail Sub-C, utilizing laser-cut hypotubes, compared to the tri-coil design of TightRail.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Spectranetics. The logo consists of a blue diamond shape on the left, followed by the word "Spectranetics" in a bold, blue font. Below the word "Spectranetics" is the tagline "Always Reaching Farther" in a smaller, lighter font. The logo is clean and professional, and the colors are bright and eye-catching.
#### Performance Data
The following testing was conducted to validate and verify that the subject device met all specifications:
#### Design Verification and Validation Testing
- Simulated Use Testing ●
- Simulated Environment Testing ●
- Human Factors Evaluation
- Linear and Radial Dimensional Testing ●
- Device Weight Measurement ●
- Blade Extension Testing ●
- Atraumatic Surface Testing
- Axial Load Testing
- Tensile Testing ●
- Flexibility Testing ●
- . Outer Sheath Compression Testing
- EMC/EMI and Electrical Medical Safety Testing
- Radiopacity Evaluation .
### Sterilization
- Product adoption equivalency per AAMI TIR28:2016 ●
## Biocompatibility
- Cytotoxicity ●
- Sensitization
- Irriation/Intracutaneous Reactivity ●
- Material Mediated Pyrogenicity
- Acute Systemic Toxicity ●
- Indirect and Direct Hemolysis ●
- Hemocompatibility ●
#### Pre-clinical and Clinical Data:
Preclinical and clinical data was not required to demonstrate substantial equivalence. The design characteristics of the subject device are similar to the predicate. The design verification and validation test results demonstrated that the subject device is as safe and clinically effective as the predicate device.
#### Substantial Equivalence:
Based on the similarities in design between the subject and predicate devices currently in use, and the performance data, the use of the TightRail Guardian device for the proposed indication does not raise new questions related to safety and effectiveness compared with the predicates. Therefore, TightRail Guardian is substantially equivalent to TightRail and TightRail Sub-C.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.